K173701 · Implant Logistics, Inc. · DZE · Nov 29, 2018 · Dental
Device Facts
Record ID
K173701
Device Name
Implant-One System
Applicant
Implant Logistics, Inc.
Product Code
DZE · Dental
Decision Date
Nov 29, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Indications for Use
The Implant-One™ System is indicated for surgical placement in partially or completely edentulous upper or lower jaws to provide a means for prosthetic attachment to restore a patient's chewing function. The Implant-One™ system is indicated for immediate loading only when primary stability is achieved and with the appropriate occlusal loading.
Device Story
Implant-One™ System consists of endosseous dental implants and various abutments; used for prosthetic attachment in edentulous jaws. Implants are self-tapping, threaded, root-form or wide-thread designs; manufactured from titanium alloy (ASTM F136/F1472). System includes cover screws and healing caps for protection during healing. Abutments (ball, locator, conical, angled, glueless, custom blank, etc.) facilitate prosthetic connection. Device used by dental professionals in clinical settings. Implants and cover screws provided sterile. Clinical benefit: restoration of patient chewing function. Output is the physical implant/abutment construct; provider uses output to support dental prosthetics. No software or AI components.
Clinical Evidence
No clinical data was used in support of this submission. Evidence consists of non-clinical mechanical testing (ISO 14801) demonstrating performance equivalent to or better than predicates, biocompatibility testing (ISO 10993-5), and sterilization/packaging/shelf-life validations (ISO 11137, 17665, ASTM D4169). Endotoxin testing (LAL) met the <20EU/Device limit.
Technological Characteristics
Endosseous dental implants and abutments made of titanium alloy (ASTM F136) or wrought titanium alloy (ASTM F1472). Features include threaded fixation, hex alignment, and 12° taper. Surface treatment: resorbable media blasted. Sterilization: validated per ISO 11137 and 17665. Dimensions: 3.25mm-6.5mm diameter, 8mm-14mm length. No software or electronic components.
Indications for Use
Indicated for surgical placement in partially or completely edentulous upper or lower jaws for prosthetic attachment to restore chewing function. Indicated for immediate loading only when primary stability is achieved and with appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 29, 2018
Implant Logistics, Inc. % Karen E. Warden, PhD President BackRoads Consulting, Inc. PO Box 566 Chesterland, Ohio 44026
Re: K173701
Trade/Device Name: Implant-One™ System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: October 29, 2018 Received: October 30, 2018
Dear Karen E. Warden, PhD:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known)
Device Name Implant-One™ System
#### Indications for Use (Describe)
The Implant-One™ System is indicated for surgical placement in partially or completely edentulous upper or lower jaws to provide a means for prosthetic attachment to restore a patient's chewing function. The Implant-One™ system is indicated for immediate loading only when primary stability is achieved and with the appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
# 510(k) Summary for K173701
| Date: | 28 November 2018 |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Implant Logistics, Inc.<br>305 3rd Street South<br>Lacrosse, WI 54601<br>Phone 608-498-4855 |
| Sponsor Contact: | Thomas Arendt |
| Trade Name: | Implant-One™ System |
| Common Name: | Dental implant and abutment system |
| Regulatory Class: | Class II |
| Primary Classification<br>Name, Regulation,<br>Product Code: | Endosseous dental implant, 21 CFR 872.3640, DZE |
| Secondary Product<br>Code: | NHA |
| Submission purpose: | To add new components to the system. |
| Device Description: | Endosseous implants are self-tapping and threaded, and offered having root-<br>form or wide thread forms. Root-form implant diameters range from 3.25mm to<br>5.5mm having lengths from 8mm to 14mm. Wide thread implant diameters are<br>available in 4.1 and 4.5mm (8mm to 14mm lengths), 5.5mm (8mm to 12mm<br>lengths) and 6.5mm (8mm to 10mm lengths). Cover screws and healing caps<br>provide protection to the threads of the abutment connection during endosseous<br>and gingival healing. Cover screws are pre- packaged with each implant.<br>Healing caps are provided as an alternative to the cover screw and are<br>packaged separately. The Implant-One™ dental implants and cover screws are<br>provided sterile.<br>Not all abutments can be used for single-unit restorations. The conical, angled<br>conical, ball, locator and glueless abutments are intended only for multi-unit<br>loaded restorations. The ball, locator and glueless abutments are to be used in<br>fully removable dentures. The conical and angled conical abutments are to be<br>used in screw retained dentures and with the titanium sleeve for screw retention.<br>The final design parameters for the custom blank abutment are as follows:<br>maximum total height, 12.5mm; minimum/maximum gingival height,<br>0.5mm/6mm; minimum post height, 4mm; maximum angulation, 30°; minimum<br>wall thickness, 0.78 (at 1.5mm above the proximal end); minimum diameter,<br>3.75 mm for the 300 Series, 4.25 mm for the 300 Series and 4.75 mm for the<br>500 Series. |
| Indications for Use: | The Implant-One™ System is indicated for surgical placement in partially or<br>completely edentulous upper or lower jaws to provide a means for prosthetic<br>attachment to restore a patient's chewing function. The Implant-One™ system is<br>indicated for immediate loading only when primary stability is achieved and with<br>the appropriate occlusal loading. |
| Materials: | All components (implants, abutments, cover screws, healing caps and abutment<br>screws) are manufactured from titanium alloy (Ti-6Al-4V ELI) as described by<br>ASTM F136 or wrought titanium alloy (Ti-6Al-4V) as described by ASTM F1472. |
| Primary Predicate: | Implant-One™ System (Implant Logistics, Inc. – K102822) |
| Reference Devices: | Ankylos C/X Dental Implant System (Dentsply International – K083805),<br>SynCone Abutment 5° (Dentsply International – K131644), Nobel Active® Multi-<br>unit Abutment (Nobel BioCare® – K072570), Temporary Snap Abutment (Nobel<br>BioCare® - K161435), ArgenIS Titanium Abutment Blanks (The Argen<br>Corporation - K143051), LOCATOR® R-Tx Attachment System (K150295 -<br>Zest Anchors LLC), O-Ring Abutment (K990277 - Biohorizons Implant<br>Systems), MiNi Internal Implant System (K150537 - Megagen Implant Co, Ltd) |
{5}------------------------------------------------
| Performance Data: | Non-clinical mechanical testing of the worst case Implant-One™ System construct was performed according to ISO 14801 and demonstrated that the Implant-One™ system performs as well as or better than the predicate devices. Biocompatibility testing was performed according to ISO 10993-5 and demonstrated substantial equivalence. | | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | For the modified implant surfaces, the following information was submitted per FDA Guidance, "Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments: identification of the blast media, composition of the particles, identification of the treatments and agents used to remove the particles, a chemical analysis of the surface and SEM photomicrographs of the surface. | | |
| | Sterilization validations were performed according to ISO 11137 and 17665 and demonstrated substantial equivalence. Packaging and shelf-life validations were performed according to ASTM D4169 including ASTM F1886, ASTM F88 and ASTM F1929 and demonstrated substantial equivalence. The Limulus amebocyte lysate (LAL) test was performed on sterile devices per AAMI ST72 and met the <20EU/Device endotoxin limit. | | |
| | No clinical data was used in support of this submission. | | |
| Technological<br>Characteristics: | The fundamental scientific technology of the Implant-One™ System is the same as previously cleared devices as shown below. Minor differences between the subject and predicate devices did not raise new questions of safety or effectiveness. | | |
| | Specifically, the minor differences in dimension, surface or connection features between the subject implants, primary predicate and reference device do not introduce new issues of safety or efficacy as demonstrated by the non-clinical mechanical and biocompatibility testing. | | |
| | Reference devices have been identified to accommodate the dimensional features for the ball, straight, post, wide post, glueless, anatomical angled, solid, custom blank, conical, angled conical, temporary and locator abutments therefore these subject abutments do not introduce new issues of safety or efficacy. | | |
| System: | Implant-One™ 300,<br>400, 500 series<br>(K173701) | Implant-One™ 100,<br>200 series (K102822) | Ankylos C/X Dental<br>Implant (K083805) |
|-----------------------------|-------------------------------------------------------------------------------------------|----------------------------------------------------------|---------------------------------------------------------------|
| Material of<br>manufacture: | Titanium and/or titanium<br>alloy | Titanium and/or titanium<br>alloy | Commercially pure<br>titanium |
| Design: | | | |
| Endosseous implant | Root-form | Root-form | Root-form |
| Method of<br>stabilization | Threaded fixation | Threaded fixation | Threaded fixation |
| Range of<br>Diameters (mm) | 3.5 – 6.5 | 3.25 – 5.5 | 3.5 – 5.5<br>7.0 |
| Range of<br>Lengths (mm) | 8 - 14 | 8 – 14 | 8 - 17<br>8 - 14 |
| Modified surface | Yes, resorbable media<br>blasted | Yes, AIO2 blasted | Yes, grit blasted and<br>etched |
| Connection to<br>abutment | Hex alignment, 12°<br>included taper, screw<br>attachment or integral<br>screw attachment | Hex alignment, 6°<br>included taper, screw<br>attachment | Keyed alignment,<br>friction-lock taper,<br>thread attachment |
{6}------------------------------------------------
| Abutments | | Implant-One™ 300,<br>400, 500 series<br>(K173701) | Implant-One™ 100,<br>200 series<br>(K102822) | Biohorizons O-<br>Ring Abutment<br>(K990277) |
|-----------|-----------------------|---------------------------------------------------|----------------------------------------------|----------------------------------------------|
| Ball | Surface<br>Treatment | Color anodization,<br>full | None | Color anodization,<br>partial |
| | Diameter<br>(mm) | 4.0, 5.0, 5.5 | 3.8 | 3.0, 3.5, 4.5, 5.7 |
| | Collar<br>Height (mm) | 2, 3, 4, 5 | 2, 3, 4, 5 | 1, 3, 5 |
| | Material | ASTM F136 or<br>F1472 | ASTM F1472 | Titanium alloy |
| Abutments | | Implant-One™ 300,<br>400, 500 series<br>(K173701) | Implant-One™ 100,<br>200 series (K102822) | MiNi Implant System<br>"milling"(K150537) |
|-----------|-----------------------|---------------------------------------------------|-------------------------------------------|-------------------------------------------|
| Straight | Surface<br>Treatment | Color anodization | None | Color anodization |
| | Diameter<br>(mm) | 3.1, 4.1, 4.5mm | 4.0mm | 3.0, 3.5mm |
| | Collar<br>Height (mm) | 1mm | 0.5, 1, 2, 3, 4, 5mm | 1.0, 1.5, 2.5, 3.5,<br>4.5mm |
| | Material | ASTM F136 or F1472 | ASTM F1472 | ASTM F136 |
| Abutments | | Implant-One™ 300, 400, 500 series (K173701) | Implant-One™ 100, 200 series (K102822) | |
| Post | Surface Treatment | Color anodization | None | |
| | Diameter (mm) | 4.0, 5.0, 5.5 | 4.0 | |
| | Collar Height (mm) | 0.5, 1, 2, 3, 4, 5 | 0.5, 1, 2, 3, 4, 5 | |
| | Material | ASTM F136 or F1472 | ASTM F1472 | |
| Wide Post | Surface Treatment | Color anodization | None | |
| | Diameter (mm) | 4.0, 5.0, 5.5 | 4.0 | |
| | Collar Height (mm) | 0.5, 1, 2, 3, 4, 5 | 0.5, 1, 2, 3, 4, 5 | |
| | Material | ASTM F136 or F1472 | ASTM F1472 | |
| Abutments | | Implant-One™ 300, 400,<br>500 series | Dentsply International |
|-----------|--------------------|--------------------------------------|---------------------------|
| Glueless | 510(k) # | K173701 | SynCone 5° (K131644) |
| | Surface Treatment | Color anodization | None |
| | Diameter (mm) | 4 | 4 |
| | Collar Height (mm) | 0.5 | 1.5, 3.0, 4.5 |
| | Angulation | 0° and 15° | 0°, 7.5°, 15°, 22.5°, 30° |
| | Material | ASTM F136 or F1472 | Ti6Al4V ELI |
{7}------------------------------------------------
| Abutments | | Implant-One™ 300, 400,<br>500 series | Dentsply International |
|----------------------|--------------------|------------------------------------------------|--------------------------------------------------|
| Anatomical<br>angled | 510(k) # | K173701 | Ankylos C/X (K083805) |
| | Surface Treatment | Color anodization | None |
| | Diameter (mm) | 4.0, 5.0, 5.5 | 5.7 |
| | Collar Height (mm) | 0.5, 1, 2, 3, 4, 5 | 0.75, 1.5, 3.0, 4.5 |
| | Angulation | 15° and 30° | 7.5°, 15°, 22.5°, 30°, 37.5° |
| | Material | ASTM F136 or F1472 | Titanium |
| Solid | 510(k) # | K173701 | Ankylos C/X (K083805) |
| | Surface Treatment | Color anodization | None |
| | Diameter (mm) | 4.0, 5.0, 5.5 | 3.3, 4.5 |
| | Collar Height (mm) | 0.5, 1, 2, 3, 4, 5 | 1.5, 3.0, 4.5, 6 |
| | Material | ASTM F136 or F1472 | Titanium |
| Abutments | | Implant-One TM, 300,<br>400, 500 series | Nobel BioCare |
| Conical | 510(k) # | K173701 | Nobel Active Multi-unit<br>Abutment (K072570) |
| | Surface Treatment | Color anodization | None |
| | Diameter (mm) | 4.8 | 4.8 |
| | Collar Height (mm) | 0.5, 1, 2, 3, 4, 5 | 1.5, 2.5, 3.5, 4.5 |
| | Material | ASTM F136 or F1472 | Titanium vanadium alloy |
| Angled conical | 510(k) # | K173701 | Nobel Active Multi-unit<br>Abutment (K072570) |
| | Surface Treatment | Color anodization | None |
| | Diameter (mm) | 4.8 | 4.8 |
| | Collar Height (mm) | 1 | 1.5, 2.5, 3.5, 4.5 |
| | Angulation | 15° | 17° and 30° |
| | Material | ASTM F136 or F1472 | Titanium vanadium alloy |
| Temporary | 510(k) # | K173701 | Temporary Snap<br>Abutment (K161435) |
| | Surface Treatment | Color anodization | None |
| | Diameter (mm) | 4.0, 5.0, 5.5 | 4.0, 4.5, 6.0 |
| | Collar Height (mm) | 0.5, 1, 2, 3, 4, 5 | 1.5, 3.0 |
| | Material | ASTM F136 or F1472 | ASTM F136 or F1472 |
| Abutments | | Implant-One™ 300, 400,<br>500 series (K173701) | LOCATOR® R-Tx<br>(K150295) |
| Locator® | Surface Treatment | Color anodization | Titanium nitride (TiN) or<br>CarboNitride (TiCN) |
| | Diameter (mm) | 3.9 | 3.9 |
| | Collar Height (mm) | 0.5, 1, 2, 3, 4, 5 | 1, 2, 3, 4, 5, 6 |
| | Material | ASTM F136 or F1472 | Titanium alloy |
{8}------------------------------------------------
| Abutments | | Implant-One TM, 300, 400,<br>500 series (K173701) | ArgenIS Titanium Abutment<br>Blanks (K143051) |
|--------------|-------------------------|---------------------------------------------------|-----------------------------------------------|
| Custom blank | Surface Treatment | None | None |
| | Total height (mm) | 12.5 (max) | 12.5 (max) |
| | Gingival height<br>(mm) | 0.5 (min) 6 (max) | 0 (min) 6 (max) |
| | Post height (mm) | 4 (min) | 4 (min) |
| | Angulation | 30° (max) | 30° (max) |
| | Wall thickness<br>(mm) | 0.78 (1.5mm above the<br>proximal end) | 0.65 (at abut/implant interface) |
| | Diameter (mm) | 3.75 (min) | 3.5 (min) |
| | Material | ASTM F136 or F1472 | ASTM F136 |
| | Mode of retention | Hex alignment, screw | Screw |
## Conclusion:
The Implant-One™ System is substantially equivalent to the predicate devices referenced above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.