The TBL-ABI is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD). TBL-ABI is intended for the rapid measurement of ankle-brachial pressure index (ABPI), or ankle-brachial index (ABI), and pulse volume recording (PVR) / volume plethysmography in adults. It is suitable for use in wound care assessment, for assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy. TBL-ABI can be used on patients with unilateral lower limb amputation. The TBL-ABI System is intended to be used to spot-check patients. The TBL-ABI provides information regarding patient risk. The physician has the responsibility of making proper judgments based on this information.
Device Story
TBL-ABI system performs rapid, automated measurement of ankle-brachial index (ABI) and pulse volume recording (PVR) for PAD screening. System consists of three Bluetooth-enabled blood pressure devices with cuffs and host computer software. Operator places cuffs on limbs; software controls automated inflation/deflation via air pump and electromagnetic valve. Device uses pneumo-plethysmography to monitor pressure variations. Data transmitted via Bluetooth 4.0 to computer for calculation and display of ABI, systolic, and diastolic values. Used in clinical environments by technicians/physicians for spot-checking. Results assist clinicians in assessing PAD risk and determining suitability for compression therapy. Benefits include simultaneous limb measurement, eliminating rest periods between readings, and improved convenience via wireless hardware.
Clinical Evidence
Bench testing only. Performance validated through calibration tests, Bluetooth/cybersecurity testing, and laboratory testing against standards including IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30, ISO 10993-5/10, and IEC 62133. Software verification performed via SRS/SDS/STD/STR documentation.
Technological Characteristics
System comprises 3 portable units with internal Li-Po batteries (3.7 VDC). Patient contact materials: lycra/raised fabric cuffs (ISO 10993-5/10 compliant). Sensing: pneumo-plethysmographic method. Connectivity: Bluetooth 4.0 to PC. Software: automated inflation/deflation control. Standards: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30, ISO 14971. IPX21 protection. Dimensions: 60x30x132mm per unit.
Indications for Use
Indicated for adult subjects at risk of or with peripheral arterial disease (PAD), including those with venous/arterial ulcers or unilateral lower limb amputation. Contraindicated for patients undergoing external defibrillation, connected to life support/implanted electronic devices, or those requiring long-term monitoring.
Regulatory Classification
Identification
A hydraulic, pneumatic, or photoelectric plethysmograph is a device used to estimate blood flow in a region of the body using hydraulic, pneumatic, or photoelectric measurement techniques.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
December 19, 2017
LD Technology LLC Albert Maarek Quality Manager 100 N. Biscayne Blvd Suite 502 Miami. Florida 33132
Re: K173696
Trade/Device Name: TBL-ABI System Regulation Number: 21 CFR 870.2780 Regulation Name: Hydraulic, Pneumatic, Or Photoelectric Plethysmographs Regulatory Class: Class II Product Code: JOM Dated: November 3, 2017 Received: December 1, 2017
Dear Albert Maarek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
M.A. Hilleman
for Bram D. Zuckerman, M.D.
Director Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number K173696
Device Name: TBL-ABI System
#### Indications for Use (Describe)
The TBL-ABI is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD). TBL-ABI is intended for rapid measurement of ankle-brachial pressure index (ABPI) or ankle-brachial index (ABI) in adults and pulse volume recording (PVR) / volume plethysmography in adults.
It is suitable for use in wound care assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy.
TBL-ABI can be used on patients with unilateral lower limb amputation.
The TBL-ABI System is intended to be used to spot-check patients. The TBL-ABI provides information regarding patient risk. The physician has the responsibility of making proper judgments based on this information.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# K173696
Image /page/3/Picture/1 description: The image shows a logo for LD Technology. The logo consists of a circular emblem on the left and the text "LD Technology" on the right. The emblem features a green sphere with plant-like structures extending from it, all enclosed within a circle. The text "LD Technology" is written in a light gray color, with the "LD" appearing larger and slightly above the word "Technology".
510(k) Premarket Notification Number: 173696 Date of preparation: November 04 .2017
#### 510(k) Summary TBL-ABI device
#### This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92
#### 1. Submitter's Identification: Manufacturer: L.D TECHNOLOGY Address:
100 N. Biscayne Blvd. Suite 502 Miami, FL, 33132, USA Tel: 305-379-9900 E mail: albert.ldteck@gmail.com
### 2. Device Name / Classification
Trade name: Patient monitor Device Name and Model: TBL-ABI Regulation number: 21CFR 870.2780 Product Code: JOM Device Class: Class II Classification Name: Hydraulic, pneumatic, or photoelectric plethysmograph Classification Panel: Cardiology
#### 3. Predicate legally marketed devices
TM-ABI k143152 Applicant: LD TECHNOLOGY LLC. Product Code: JOM.
### 4. Device Description
TBL-ABI System comprises:
- 2 3 Bluetooth blood pressure devices with attached cuffs.
- ∑ Software installed on a computer
It is intended to measure a patient's Ankle Brachial Index (ABI) and provide Pulse Volume Recording (PVR) / volume plethysmography.
This is done through an automated process.
The operator places the three devices with different color-coded electronic boxes on the right or left arm, and on each leg as described in the instructions for use, and the devices connect to the computer via Bluetooth.
Once the devices are connected, the operator clicks start on the software to start the measurement. The devices will then automatically control the inflation of the cuffs and monitor the variations in individual pressures to determine values to be used for the calculation of the ABI
{4}------------------------------------------------
values for both the left and right of the patient.
TBL-ABI uses pneumo-plethysmography in order to obtain physiologic measurements from the patient's limbs.
Measurements are conducted as a single occurrence on the three limbs, thus eliminating any requirement to rest the patient between measurements.
The test period takes approximately 3 minutes.
ABI values, as well as the Pulse Volume recording (PVR), are displayed on the software installed in a computer. The results are saved in a backup file and can also be printed.
## 5. Intended use and indications for use
The TBL-ABI is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD).
TBL-ABI is intended for the rapid measurement of ankle-brachial pressure index (ABPI), or ankle-brachial index (ABI), and pulse volume recording (PVR) / volume plethysmography in adults. It is suitable for use in wound care assessment, for assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy.
TBL-ABI can be used on patients with unilateral lower limb amputation.
The TBL-ABI System is intended to be used to spot-check patients. The TBL-ABI provides information regarding patient risk. The physician has the responsibility of making proper iudgments based on this information.
*Prescription Use: Federal law restricts the use of this device to sale by or on the order of a physician
#### 6. Performances, technical specifications and materials Performances
TBL-ABI comprises 3 portable medical devices, powered by internal Li-Po battery. Each device has a rechargeable battery, and a Bluetooth connection with the software installed in a computer. Each device has a different box color.
The front panel of each device has an on/off button to connect the Bluetooth.,
The main purpose of the TBL-ABI is to measure ABI, SYS and DIA of all limbs. It also allows for checking measurement history.
# Technical specifications i.e. table of comparison with the predicate device Patient contact materials:
The materials in contact with the patient are the cuffs. The Cuff material is lycra fabric, raised fabric.
# 7. Contra-indications
- Σ Patients undergoing external defibrillation.
- > Patients connected to electronic life support devices, or any implanted electronic device.
- 网 Patients moving or undergoing long term monitoring.
- > Do not use this device in the presence of:
- > Magnetic resonance imaging (MR or MRI) equipment. MRI equipment may deliver induced current to the device.
{5}------------------------------------------------
- > Strong electromagnetic sources, such as electro surgery equipment.
- > Computed Tomography (CT) equipment.
- > WIFI router at distance less than 10 feet and in the same room.
- > Cellphones at distance less than 10 feet.
- > Microwaves ovens, cables and connectors associated with satellite dishes, nearby power
- > lines or power stations, cordless telephones, wireless speakers, some monitors and
- > displays, cameras, baby monitors.
The distance between hardware and the computer must be less than 10 feet and in the same room.
- ∑ Arterial catheters, or with an AV fistula or pressure dressing.
- 2 Venous pulsations may cause erroneous reading in blood pressure (e.g. tricuspid valve regurgitation).
- C Take caution with patients that have low perfusion. Using the blood pressure device may cause skin erosion and/or pressure necrosis.
### 8. Undesirable side effects:
No side effects or adverse reactions are known to date.
### 9. Substantial equivalence
Predicate legally marketed device: TM-ABI, K143152. Applicant: LD TECHNOLOGY LLC. Product Code: JOM.
### Substantial Equivalence Discussion
- Step 1: Does the new device have the same indication statement? Yes
Device Modifications are not affected the intended use of the device.
- V Step 3: The new device has the same intended use and may be "substantially equivalent." Yes
- > Step 5: Does the new device have the same technological characteristics, e.g. design, materials, etc.?
### Similarities:
- ✔ Same intended use
- Scientific Background (oscillometric measurement of the blood pressure and PVR)
- ✔ Placement of the cuff
- V Power supply using a rechargeable battery
- Classification Degree of protection against electric shocks
- V Operating mode
### Device Modifications:
- √ The TBL-ABI and predicate device have different dimension and weight
- The TBL-ABI and predicate device use a battery with different voltage and intensity rate.
{6}------------------------------------------------
- √ The TBL-ABI cuffs are in different material and are directly attached to each device versus cuffs connected with tubes for TM-ABI.
- ✔ The TBL-ABI has a maximal pressure of 270 mmHg versus 299 mmHg for TM-ABI
- V The TM-ABI has small difference in temperature and humidity conditions for storage and working environment
- √ The TBL-ABI uses 3 Bluetooth devices versus one USB connection device for TM-ABI
### Rationale for the Device modifications:
- > Bluetooth communication
- √ Cuffs attached to 3 electronic boxes with pump Allow to remove the tubes and the patient feel more comfortable and it is more convenient for the technician.
#### Step 6: If it has new technological characteristics, could they affect safety or effectiveness? Discussion why any differences between the subject and predicate(s) do not impact safetv and effectiveness [see section 513(i) (1) (A) of the FD&C Act and 21 CFR 807.87(f)] See "
### Does the change in communication to the software affect the safety or effectiveness?
The Bluetooth communication is safe, and we use the same CRC coding test as the previous USB communication. CRC (Cyclic redundancy check) Coding was performing to demonstrate the software performance to accurately capture, store, and analyze the data measured by the hardware. Testing of the Bluetooth has been performed.
### Do the other changes affect the safety or effectiveness?
The fact that the TBL-ABI has:
- √ A different dimension and weight
- V Battery with lower voltage and intensity rate
- > Different cuff material and cuffs directly attached to the device
- > A small lower maximal blood pressure value
- √ A small difference in temperature and humidity environment
do not affect the performances and the effectiveness of the TBL-ABI system compared to the predicate device as shown by the laboratory testing (60601-1-1-1, 60601-1-2, IEC 80601-2-30. ISO 10993 and IEC 62133) and testing performed comparing the results of 2 devices as well as the cybersecurity risk assessment.
{7}------------------------------------------------
# Table of comparison
| Information | TBL-ABI | TM-ABI K143152 |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The TBL-ABI is indicated for use on adult<br>subjects at risk of having or developing<br>peripheral arterial disease (PAD).<br>TBL-ABI is intended for the rapid measurement<br>of ankle-brachial pressure index (ABPI), or<br>ankle-brachial index (ABI), and pulse volume<br>recording (PVR) / volume plethysmography in<br>adults. It is suitable for use in wound care<br>assessment, for assessing symptomatic PAD, and<br>as a screening device for PAD. It may also be<br>used on patients with venous or arterial ulcers<br>prior to the application of compression therapy.<br>TBL-ABI can be used on patients with unilateral<br>lower limb amputation. | The TM-ABI is indicated for use on adult<br>subjects at risk of having or developing<br>peripheral arterial disease (PAD).<br>TM-ABI is intended for the rapid measurement<br>of ankle-brachial pressure index (ABPI), or<br>ankle-brachial index (ABI), and pulse volume<br>recording (PVR) / volume plethysmography in<br>adults. It is suitable for use in wound care<br>assessment, for assessing symptomatic PAD,<br>and as a screening device for PAD. It may also<br>be used on patients with venous or arterial<br>ulcers prior to the application of compression<br>therapy.<br>TM-ABI can be used on patients with unilateral<br>lower limb amputation. |
| Dimensions/weight | For each device: Width: 60mm, height: 30mm,<br>depth:132mm Weight: 260g | width: 250 mm, height: 730mm, depth: 200 mm,<br>weight: 0.60kg |
| Power Supply | Output: 3.7 VDC. Battery type: rechargeable<br>lithium polymer<br>Capacity: 530 mAh,<br>number of measurements per charge: 60 | Output: 5VDC. Battery type: rechargeable lithium<br>polymer<br>Capacity: 2,300mAh,<br>number of measurements per charge: 30 |
| Applied parts in contact<br>with the patient | 3 cuffs lycra fabric raised fabric ISO 10993- 10<br>2010. ISO 10993-10: 2012 and<br>ISO 10993-5: 2009 | 3 cuffs, nylon and bladders ISO 10993-10: 2010 |
| Testing Bench | Type of protection against electric shock: Class<br>II. BF<br>Compliant with standards:<br>60601-1,<br>60601-1-2<br>80601-2-30. | Type of protection against electric shock: Class<br>II. BF<br>Compliant with standards:<br>60601-1,<br>60601-1-2<br>80601-2-30. |
| Measurement types | Using the pneumo-plethysmographic method:<br>Right and left Ankle brachial pressure index<br>Systolic blood<br>Diastolic blood pressure | Using the pneumo-plethysmographic method:<br>Right and left Ankle brachial pressure index<br>Systolic blood<br>Diastolic blood pressure |
| Measurement ranges | Pressure: 0 to 270 mmHg | Pressure: 0 to 299 mmHg |
| Limit values of<br>measurement errors | Pressure: ± 3mmhg<br>Ankle brachial pressure index: ± 0.1 | Pressure: ± 3mmhg<br>Ankle brachial pressure index: ± 0.1 |
| Cuffs inflation and deflation | Automatic inflation using an air pump and deflation using an electromagnetic valve. | Automatic inflation using an air pump and deflation using an electromagnetic valve. |
| Temperature and humidity range | working environment: 5 to 40°C, 15 to 80%<br>relative air humidity, IPX21 protection,<br>transport and storage: -20 to 55°C, up to 85%<br>relative air humidity. | working environment: 10 to 40°C, 30 to 85%<br>relative air humidity, IPX0 protection,<br>transport and storage: 0 to 60°C, up to 85%<br>relative air humidity. |
| Target population | Adult | Adult |
| Where used | Clinical environment | Clinical environment |
| PC Data transmission | Bluetooth version 4.0 | USB |
{8}------------------------------------------------
# 10. Non-clinical, Performances and Effectiveness
CRC (Cyclic redundancy check) Coding was performing to demonstrate the software performance to accurately capture, store, and analyze the data measured by the hardware. Testing comprises:
- 1. Calibration tests
- 2. Bluetooth testing and Cybersecurity risk design
- 3. Laboratory tests
- 4. Software verification (SRS/SDS/STD/STR)
# 11. General Safety Concerns
The laboratory test reports of the TBL-ABI System and Bluetooth communication technology testing have demonstrated the general safety of the system compared to the legally marketed predicate device.
### 12. Standards
6 | 7
- V IEC 60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performance. Third Edition December 2006
- > IEC60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests. Third Edition 05/17/2007
- > IEC 80601-2-30: Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. 10/31/2010 Second Edition
- > ISO 10993-10: Tests for irritation and skin sensitization.2010 and 2012
- > ISO 10993-5: Biological evaluation of medical devices. 2009
- > IEC 62133: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. 2nd Edition.2012.
- > Guidance for: Industry; FDA staff; and Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005
- > Wireless Medical device FDA Guidance. 2014.
{9}------------------------------------------------
- ✔ Cybersecurity FDA Guidance. 2013.
- V ISO 14971: Medical devices Application of risk management to medical devices. March 01, 2007
## Conclusion
TBL-ABI is equivalent in performance, technology, safety and efficacy to the legally marketed predicate device. Signature:
**Signature:**
# Albert MAAREK
scribble
Premarket notification [510K] Number: 173696
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.