K173663 · Imediplus, Inc. · DQD · Aug 17, 2018 · Cardiovascular
Device Facts
Record ID
K173663
Device Name
Electronic Stethoscope DS3011A
Applicant
Imediplus, Inc.
Product Code
DQD · Cardiovascular
Decision Date
Aug 17, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1875
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The IMEDIPLUS Electronic Stethoscope DS3011A is intended for the detection, amplification and recording of sounds from the heart, lungs, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. And the stethoscope chest-piece is designed for use with child, adolescent and adult patients. It is used for any subject undergoing a physical examination and intended only for medical diagnostic purposes in clinic or hospital.
Device Story
Handheld electronic stethoscope; captures acoustic signals from heart, lungs, abdomen, neck, limbs, and vasculature via chest-piece. Features digital signal processor for amplification and filtering (Bell, Diaphragm, Wide modes). User interface includes OLED display, keypad, and 1D barcode reader for patient/recording identification. Operates via rechargeable lithium battery. Records audio to internal storage or micro SD card; transmits data via Bluetooth to mobile device running DS3011A_DM software. Software enables data reception, storage, phonogram visualization, and playback. Used by clinicians (physicians, nurses, EMTs, etc.) for auscultation. Provides real-time phonogram display and heart rate monitoring. Benefits include enhanced sound amplification, objective recording for diagnostic review, and digital data management.
Clinical Evidence
Clinical evaluation performed with 4 certified clinical doctors and 15 subjects (4 children, 4 adolescents, 7 adults). Questionnaire-based statistical analysis confirmed device met safety and effectiveness principles for heart, chest, intestinal, and neck auscultation across specified patient groups.
Technological Characteristics
Materials: Silicon diaphragm, brass alloy headset. Sensing: Electronic chest-piece with digital signal processor. Power: NP-120 rechargeable lithium battery. Form factor: Handheld with 1.46" OLED display. Connectivity: Bluetooth, micro SD card. Software: DS3011A_DM data management software. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62133, IEC 62471, ANSI C63.27.
Indications for Use
Indicated for child, adolescent, and adult patients undergoing physical examination for medical diagnostic purposes in clinical or hospital settings.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
3M™ LITTMANN® ELECTRONIC STETHOSCOPE MODEL 3200 (K083903)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 17, 2018
IMEDIPLUS INC. Shirley Lai Regulatory Specialist 2F, 12, Shengvi Rd.Sec.2 Chupei City, Hsinchu County 30261 Taiwan (R.O.C.)
Re: K173663
Trade/Device Name: Electronic Stethoscope DS3011A Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: July 16, 2018 Received: July 18, 2018
Dear Shirley Lai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Shawn W. Forrest -A for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K173663
#### Device Name
IMEDIPLUS Electronic Stethoscope DS3011A
Type of Use (Select one or hoth, as annlicable)
#### Indications for Use (Describe)
The IMEDIPLUS Electronic Stethoscope DS3011A is intended for the detection, amplification and recording of sounds from the heart, lungs, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. And the stethoscope chest-piece is designed for use with child, adolescent and adult patients. It is used for any subject undergoing a physical examination and intended diagnostic purposes in clinic or hospital.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# IMEDIPLUS INC.
## Deficiency Reply Electronic Stethoscope DS3011A
### Section 5
Document Title: 510(k) Summary File Name: 018_510(k)Summary
{4}------------------------------------------------
| Image: iMEDI logo | Document Title:<br>510(k) Summary | | |
|-------------------|-----------------------------------|-------|-----------|
| Document No.: | DS3011A-SDUS01-00501 | Page: | 2 of 20 |
| Product Name: | Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: | 018_510(k)Summary | Date: | 2018/8/17 |
#### 510(k) Summary
The Assigned 510(k) Number: K173663/S002 Date Prepared: August 17, 2018.
#### l. Submitter Information:
Submitter: IMEDIPLUS INC. Address: 2F, 12, ShengYi Rd. Sec. 2, Chupei City, Hsinchu County 30261, Taiwan (R.O.C.) Phone Number: +886-3-658-7700 Fax Number: +886-3-658-9535
Contact Person: Shirley Lai E-mail: shirley.lai@imediplus.com Phone Number: +886-3-658-7700#326 Fax Number: +886-3-658-9535 Date Prepared: August 17, 2018.
#### II. Predicate Device
#### Predicate Device:
IMEDIPLUS Electronic Stethoscope DS301.
#### Reference Device:
3M™ LITTMANN® ELECTRONIC STETHOSCOPE MODEL 3200
| | Device Name | Manufacturer | Model | 510(k)<br>Number |
|---------------------|-------------------------------------------------------|----------------|---------|------------------|
| Candidate<br>Device | IMEDIPLUS Electronic<br>Stethoscope DS3011A | IMEDIPLUS INC. | DS3011A | K173663 |
| Predicate<br>Device | IMEDIPLUS Electronic<br>Stethoscope DS301 | IMEDIPLUS INC. | DS301 | K160023 |
| Reference<br>Device | 3M™ LITTMANN®<br>ELECTRONIC STETHOSCOPE<br>MODEL 3200 | 3M COMPANY | 3M 3200 | K083903 |
Predicate Device Identification Table 1
{5}------------------------------------------------
| Image: IMEDI logo | Document Title:<br>510(k) Summary |
|----------------------------------------------|-----------------------------------|
| Document No.: DS3011A-SDUS01-00501 | Page: 3 of 20 |
| Product Name: Electronic Stethoscope DS3011A | Ver.: A4 |
| 510(k) File Name: 018_510(k)Summary | Date: 2018/8/17 |
#### III. Device
#### Table 2 Common Name and Classification
| No. | Product Code | Device | Regulation Section | Classification | Panel |
|-----|--------------|----------------------------|--------------------|----------------|----------------|
| 1 | DQD | Stethoscope,<br>Electronic | 870.1875 | II | Cardiovascular |
#### IV. Device Description
When you turn on the IMEDIPLUS Electronic Stethoscope DS3011A for the first time. the IMEDIPLUS Electronic Stethoscope DS3011A in your hand requires your setting password for the security. Your authorization to operate this handheld electronic stethoscope is necessary for the operation.
The IMEDIPLUS Electronic Stethoscope DS3011A picks up the sounds from the heart, lung, anterior/posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs from a patient's body. When picking up the sounds, the phonogram of sounds could simultaneously display after pushing the button 'OK' or 'REC' for recording. When you auscultate with DS3011A, the sounds are conducted simultaneously to the user's ears bilaterally by active speaker embedded at the bottom of the IMEDIPLUS Electronic Stethoscope DS3011A. At the meantime, sound processing is operated with the aid of a digital signal processor. The IMEDIPLUS Electronic Stethoscope DS3011A could identify the recording number by 1-D barcode reader, indicate the sound location by intuitive keypad, and record the sounds from different sites.
The one-hand user interface includes a full-color OLED display, an intuitive keypad at the anterior part, a barcode reader at the posterior part, a chest-piece at the superior part, a tube connector for output of sounds at the inferior part, and a recording button at the left part.
After opening the Electronic Stethoscope DS3011A with your setting password and connection with a wireless device by Bluetooth, the IMEDIPLUS Electronic Stethoscope DS3011A could transmit digital data of recorded sounds under the operation of IMEDIPLUS DS3011A_DM Software (DS3011A Data Management) installed in the connected device. The effective range of transmission by Bluetooth will be influenced when some objects blocking between the IMEDIPLUS Electronic Stethoscope DS3011A and the connected device. (such as wall, human, big objectives as barrier) Reducing the distance or allowing the line of sight between the IMEDIPLUS Electronic Stethoscope DS3011A and the connected device will improve the connection by Bluetooth.
The IMEDIPLUS DS3011A_DM Software (DS3011A Data Management) will only operate under the user's authorization with the seething password. The IMEDIPLUS DS3011A-DM Software (IMEDIPLUS DS3011A-Datamanagement Software) installed in mobile device provides the users to connect with the IMEDIPLUS Electronic Stethoscope DS3011A, receive the recorded data of sounds, storage the data, display the phonograms, and replay the recorded
{6}------------------------------------------------
| Image: iMEDI logo | Document Title:<br>510(k) Summary |
|----------------------------------------------|-----------------------------------|
| Document No.: DS3011A-SDUS01-00501 | Page: 4 of 20 |
| Product Name: Electronic Stethoscope DS3011A | Ver.: A4 |
| 510(k) File Name: 018_510(k)Summary | Date: 2018/8/17 |
audio data with the speaker of effective frequency range from 20 to 1000Hz.
The IMEDIPLUS Electronic Stethoscope DS3011A could also exchange audio data with an external personal computing device by micro SD card. Every single audio file stored in the micro SD card was labeled with the user's ID, recording number and indicated position. The IMEDIPLUS Electronic Stethoscope DS3011A does not incorporate any other off-the-shelf (OTS) software. The recorded audio data only can be replayed by the IMEDIPLUS Electronic Stethoscope DS3011A, the IMEDIPLUS Electronic Stethoscope DS301, and the IMEDIPLUS DS3011A_DM Software (DS3011A_Data Management) installed in the mobile device with the speaker of effective range from 20 to 1000Hz. The IMEDIPLUS Electronic Stethoscope DS3011A operates on one (1) NP-120 lithium battery with an included power management system to prolong the battery life.
The associated accessories include:
- · Rechargeable Lithium-ion Battery.
- · Battery charger which includes adaptor and charger.
- Micro SD card for exchanging audio data.
#### V. Indication for Use
The IMEDIPLUS Electronic Stethoscope DS3011A is intended for the detection, amplification and recording of sounds from the heart, lungs, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. And the stethoscope chest-piece is designed for use with child, adolescent and adult patients. It is used for any subject undergoing a physical examination and intended only for medical diagnostic purposes in clinic or hospital.
VI. Comparison of Technology Characteristics between the Predicate Devices (The IMEDIPLUS Electronic Stethoscope DS3011 A reference with 3M™ LITTMANN® ELECTRONIC STETHOSCOPE MODEL 3200 and predicate the IMEDIPLUS Electronic Stethoscope DS301) (A) Same technological elements among these 3 candidates (Table 3)
- Binaural headset send sounds to user's ears. (a)
- Chest-piece pick up sounds from patient's internal organs. (b)
- Sound amplifier amplify the sounds from chest-piece. (c)
- Battery power the device. (d)
- Digital signal processor sound processing. (e)
- Display show the operation information. (f)
- Keypad control and setup the device. (g)
- Detect and display heart rate monitor patient's heart rate. (h)
- Record record a sound track. (i)
- (i) Playback-playback a sound track.
- Filter Selection select certain mode of filter for clinical auscultation, (k) including B/ D/ W mode.
- (I) Sound amplification level adjustment - control the level of sound amplification.
{7}------------------------------------------------
| Image: [IMEDI logo] | Document Title:<br>510(k) Summary | |
|----------------------------------------------|-----------------------------------|-----------|
| Document No.: DS3011A-SDUS01-00501 | Page: | 5 of 20 |
| Product Name: Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: 018_510(k)Summary | Date: | 2018/8/17 |
###
| Table 3 Same Technological Elements among These 3 Candidates | | | | |
|--------------------------------------------------------------|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Items | Comparison<br>Items | Candidate Device<br>DS3011A | Reference Device<br>3M 3200 (K083903) | Predicate Device<br>DS301 (K160023) |
| 1 | Regulatory | | | |
| | Regulatory<br>No | 870.1875 | 870.1875 | 870.1875 |
| | Classification | Class II | Class II | Class II |
| 2 | Power Source | | | |
| | Batteries<br>Enclosed | One NP-120 Lithium<br>battery | One AA battery | One NP-120 Lithium<br>battery |
| | Battery Life | 24 Hours | 50-60 Hours | 24 Hours |
| | Special<br>Adaptors | Yes | No | Yes |
| 3 | Features | | | |
| | Binaural<br>headset | Same | Same | Same |
| | Chest-piece<br>Technology | Single Sided | Single Sided | Single Sided |
| | Chest-piece<br>Weight | 69 g | 98 g | 69 g |
| | Clinical Area | Auscultation | Auscultation | Auscultation |
| | Diaphragm<br>Diameter | 4.1cm | 5.1cm | 4.1cm |
| | Diaphragm<br>Material | Silicon | Polyurethane-Coated<br>Silicone | Silicon |
| | Ear-tips Type | Soft Sealing | Soft Sealing | Soft Sealing |
| | Headset<br>Material | Brass alloy to electroplate | Wide diameter<br>aerospace alloy /<br>Anodized aluminum | Brass alloy to<br>electroplate |
| | Length | 77 cm | 69 cm | 77 cm |
| | Net Weight | 304 g | 185 g | 304 g |
| | Tube Color | Black 、Burgundy、Navy<br>Blue | Black 、Burgundy、Navy<br>Blue | Black 、Burgundy、Navy<br>Blue |
| 4 | Intended Use | | | |
| | Intended Use | The IMEDIPLUS<br>Electronic Stethoscope<br>DS3011A is intended for<br>the detection,<br>amplification and<br>recording of sounds from<br>the heart, lungs, anterior<br>and posterior chest,<br>abdomen, neck, limbs,<br>arteries, veins and other<br>internal organs with<br>selective frequency<br>ranges. And the<br>stethoscope chest-piece<br>is designed for use with<br>child, adolescent and<br>adult patients. It is used<br>for any subject<br>undergoing a physical | The 3M™ Littmann®<br>Electronic Stethoscope<br>Model 3200 is intended<br>for medical diagnostic<br>purposes only. It may be<br>used for the detection<br>and amplification of<br>sounds from the heart,<br>lungs, arteries, veins,<br>and other internal<br>organs with the use of<br>selective frequency<br>ranges. It can be used<br>on any person<br>undergoing a physical<br>assessment. | The IMEDIPLUS<br>Electronic Stethoscope<br>DS301 is intended for<br>the detection,<br>amplification and<br>recording of sounds<br>from the heart, lung,<br>anterior/posterior<br>chest, abdomen, neck,<br>limbs, arteries, veins<br>and other internal<br>organs with use of<br>selective frequency<br>ranges. And the<br>stethoscope chest-<br>piece is designed for<br>use with child,<br>adolescent and adult<br>patients. It is used<br>for any subject |
{8}------------------------------------------------
| Image: iMEDI logo | Document Title:<br>510(k) Summary | | |
|-------------------|-----------------------------------|-------|----------------|
| Document No.: | DS3011A-SDUS01-00501 | Page: | <b>6 of 20</b> |
| Product Name: | Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: | 018_510(k)Summary | Date: | 2018/8/17 |
| | | only for medical<br>diagnostic purposes in<br>clinic or hospital. | | undergoing a physical<br>examination and<br>intended only for<br>medical diagnostic<br>purposes in clinic or<br>hospital. |
|-----|-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| 5 | Functional<br>Screen | OLED 1.46" Full Color | LCD Monochrome | OLED 1.46" Full Color |
| | Barcode<br>Reader | Yes | No | Yes |
| | Three Filter<br>Modes | Yes | Yes | Yes |
| | Frequency<br>Range of<br>Filter Mode | Bell (20-200 Hz),<br>Diaphragm (100-500 Hz)<br>and Wide (20-1000Hz) | Bell (20-200 Hz),<br>Diaphragm (100-500<br>Hz) and Extend range<br>(20-2000 Hz) | Bell (20-200 Hz),<br>Diaphragm (100-500<br>Hz) and Wide (20-<br>1000Hz) |
| | Human<br>Machine<br>Interface | One-hand user interface<br>for patient's ID scanning,<br>auscultation position<br>selection, recording, and<br>function selection. | Yes | One-hand user<br>interface for patient's<br>ID scanning,<br>auscultation position<br>selection, recording,<br>and function selection. |
| | Recording<br>Number of<br>the sound<br>track | Yes | No | Yes |
| | Recording<br>the Organ<br>and Position<br>of the<br>Sound Track | Yes | No | Yes |
| Dip | Software Comparison | | | |
| | Heart Rate<br>Detection | Yes | Yes | Yes |
| | Heart Rate<br>Display | Yes | Yes | Yes |
| | Recording<br>the Sound<br>Track | Yes | Yes | Yes |
| | Playback the<br>Sound Track | Yes | Yes | Yes |
| | Volume<br>Control | Yes | Yes | Yes |
| | Volume<br>Control Scale | Yes<br>1-10 level | Yes<br>0-9 | Yes |
| | Amplification | Yes,<br>24 times | Yes,<br>24 times | Yes,<br>24 times |
| | Filter mode<br>selection | Yes,<br>Bell, Diaphragm, and<br>Wide modes | Yes,<br>Bell, Diaphragm and<br>Extend Modes | Yes,<br>Bell, Diaphragm, and<br>Wide modes |
| | Automatic<br>turn-off | Yes | Yes | Yes |
| | Monitor the<br>battery level | Yes | Yes | Yes |
| | Password<br>Setting | Yes,<br>The user needs to enter | No | No |
{9}------------------------------------------------
| Image: Imedi logo | Document Title:<br>510(k) Summary | |
|----------------------------------------------|-----------------------------------|-----------|
| Document No.: DS3011A-SDUS01-00501 | Page: | 7 of 20 |
| Product Name: Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: 018_510(k)Summary | Date: | 2018/8/17 |
| | Display the<br>Phonogram | the password for the<br>operation of the device | No | No |
|---|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| | Software of<br>the<br>Connected<br>Device | Yes,<br>IMEDIPLUS DS3011A-<br>DM software (DS3011A<br>DataManagement<br>software)<br>(1) Receive and store<br>the data of sound<br>tracks<br>(2) Display the<br>phonogram of the<br>sound tracks<br>(3) Reply the audio data<br>in the connected<br>device installed with<br>IMEDIPLUS<br>DS3011A-DM<br>software but this<br>software could not<br>operate the<br>IMEDIPLUS<br>Electronic<br>Stethoscope<br>DS3011A to replay<br>the sound track from<br>the connected device. | Yes,<br>3MTM Littmann®<br>Zargis StethAssistTM<br>Heart and Lung Sound<br>Visualization Software | No |
| 7 | Function of Data Transfer and the Interface | | | |
| | Data<br>Transfer of<br>the Recorded<br>Sound<br>Tracks | Yes,<br>Micro SD card and<br>Bluetooth<br>(1) Micro SD card,<br>(Permits the data<br>exchange from the<br>electronic stethoscope to<br>the personal computer)<br><br>(2) Bluetooth, (transmits<br>the audio data from<br>the stethoscope to the<br>connected device) | Yes<br>Bluetooth<br>Permits the data<br>exchange from the<br>electronic stethoscope<br>to the personal<br>computer | Yes<br>Micro SD card<br>Permits the data<br>exchange from the<br>electronic stethoscope<br>to the personal<br>computer |
| | Capable of<br>Storing Data<br>and Latest<br>sound tracks<br>for playback | Save up to 600 10-second<br>sound tracks;<br>Latest 50 sound tracks for<br>playback. | Save up to twelve 30-<br>second sound tracks;<br>Latest 12 sound tracks<br>for playback. | Save up to 999 10-<br>second sound tracks;<br>Latest 15 sound tracks<br>for playback. |
| 8 | Service and Occupation | | | |
{10}------------------------------------------------
| Image: iMEDI logo | Document Title:<br>510(k) Summary | |
|----------------------------------------------|-----------------------------------|-----------|
| Document No.: DS3011A-SDUS01-00501 | Page: | 8 of 20 |
| Product Name: Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: 018_510(k)Summary | Date: | 2018/8/17 |
| Warranty<br>Period | 2 Years | 2 Years | 2 Years |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Occupation | Anesthesiologist,<br>Cardiologist, Emergency<br>Physician, EMT/EMS,<br>Family Practitioner,<br>Internist,<br>Medical Assistant,<br>Medical Student,<br>Nurse,<br>Nursing Student,<br>Pediatrician, Physician,<br>Respiratory Specialist,<br>Teacher/Professor/Instruct<br>or, and Veterinarian. | Anesthesiologist,<br>Cardiologist, Emergency<br>Physician, EMT/EMS",<br>Family Practitioner,<br>Internist,<br>Medical Assistant,<br>Medical Student,<br>Nurse,<br>Nursing Student,<br>Pediatrician, Physician,<br>Respiratory Specialist,<br>Teacher/Professor/Instru<br>ctor, and Veterinarian. | Anesthesiologist,<br>Cardiologist,<br>Emergency Physician,<br>EMT/EMS,<br>Family Practitioner,<br>Internist,<br>Medical Assistant,<br>Medical Student,<br>Nurse,<br>Nursing Student,<br>Pediatrician,<br>Physician,<br>Respiratory Specialist,<br>Teacher/Professor/Inst<br>ructor, and<br>Veterinarian. |
| Teaching<br>Accessories<br>Included | User manual | CD-ROM<br>User manual | User manual |
{11}------------------------------------------------
| Image: IMEDI logo | Document Title:<br>510(k) Summary | | |
|-------------------|-----------------------------------|-------|-----------|
| Document No.: | DS3011A-SDUS01-00501 | Page: | 9 of 20 |
| Product Name: | Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: | 018_510(k)Summary | Date: | 2018/8/17 |
#### (B) Comparison between the Electronic Stethoscope DS3011A and the reference device, the 3M™ LITTMANN® ELECTRONIC STETHOSCOPE MODEL 3200
#### (a) The differences between these two devices: (Table 4)
- (1) 1D Barcode reader identify the user's ID and recording number.
- Micro SD card storage and exchange for sound tracks. (2)
- (3) Organ position indication - indicate the auscultated organ.
- (4) Auscultation position indication used to indicate the auscultation position of organ.
- (5) Intuitive keypad - select the organ position between heart and lung during auscultation.
- (6) A/P CHEST Key switch the organ position between "Anterior Chest" and "Posterior Chest".
- Password the user needs to enter the password to access (7) device.
- Phonogram display for real time monitoring sound wave. (8)
#### (b) The same elements between these two devices: (Table 4)
- Both of these connect with Bluetooth. (1)
- (2) Binaural headset - send sounds to user's ears.
- Chest-piece pick up sounds from patient's internal organs. (3)
- (4) Sound amplifier - amplify the sounds from chest-piece.
- (5) Battery - power the device.
- Digital signal processor sound processing. (6)
- Display show the operation information. (7)
- (8) Keypad - control and setup the device.
- Detect and display heart rate monitor patient's heart rate. (ਰ)
- (10) Record and playback - record and playback a sound track.
- (11) Select filter - select certain mode of filter for clinical auscultation, including B/ D/ W mode.
- Sound amplification level adjustment control the level of sound (12) amplification.
- (13) Monitor battery level - indicate the battery level.
{12}------------------------------------------------
| Image: [IMEDI logo] Document Title: | 510(k) Summary | | |
|-------------------------------------------------|--------------------------------|-------|-----------|
| Document No.: | DS3011A-SDUS01-00501 | Page: | 10 of 20 |
| Product Name: | Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: | 018_510(k)Summary | Date: | 2018/8/17 |
#### Table 4 Same and Differences Technological Elements between DS3011A and 3M 3200
| Features | allu Jivi JZVV<br>Candidate<br>Device<br>DS3011A | Reference<br>Device<br>3M 3200 | Same/Different |
|--------------------------------------|--------------------------------------------------|--------------------------------|----------------|
| 1D Barcode reader | Yes | No | Different |
| Micro SD card | Yes | No | Different |
| Organ position indication | Yes | No | Different |
| Auscultation position indication | Yes | No | Different |
| Intuitive keypad | Yes | No | Different |
| A/P CHEST Key | Yes | No | Different |
| Password | Yes | No | Different |
| Phonogram display on the screen | Yes | No | Different |
| Wireless connection by Bluetooth | Yes | Yes | Same |
| Binaural headset | Yes | Yes | Same |
| Chest-piece | Yes | Yes | Same |
| Sound amplification | Yes | Yes | Same |
| Battery | Yes | Yes | Same |
| Digital signal processor | Yes | Yes | Same |
| Display | Yes | Yes | Same |
| Keypad | Yes | Yes | Same |
| Detect and display the heart rate | Yes | Yes | Same |
| Record | Yes | Yes | Same |
| Playback | Yes | Yes | Same |
| Filter selection | Yes | Yes | Same |
| Sound amplification level adjustment | Yes | Yes | Same |
| Monitor Battery level | Yes | Yes | Same |
#### (C) Comparison between the Electronic Stethoscope DS3011A and the predicate device, the IMEDIPLUS Electronic Stethoscope DS301 (a) The differences between these two devices (Table 5)
- Bluetooth interface transmit sound data to the connected (1) device installed with DS3011A_DM software (DS3011A Data Management).
- DS3011A DM software (DS3011A Data Management) used to (2) receive sound data.
- (3) Password - the user needs to enter the password to access device.
- Phonogram display on the screen of IMEDIPLUS Electronic (4) Stethoscope DS3011A - to display the phonogram while auscultating and replaying the data of audio sounds.
{13}------------------------------------------------
| Image: Imedi Logo | Document Title: | 510(k) Summary | |
|-------------------|--------------------------------|----------------|-----------|
| Document No.: | DS3011A-SDUS01-00501 | Page: | 11 of 20 |
| Product Name: | Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: | 018_510(k)Summary | Date: | 2018/8/17 |
#### (b) The same elements between these two devices (Table 5)
- (1) 1D Barcode reader - identify the user's ID and recording number.
- Micro SD card exchange for sound tracks. (2)
- Organ position indication indicate the auscultated organ. (3)
- Auscultation position indication used to indicate the auscultation (4) position of organ.
- (5) Intuitive keypad - select the organ position between heart and lung during auscultation.
- A/P CHEST Key switch the organ position between "Anterior Chest" (6) and "Posterior Chest".
- (7) Binaural headset - send sounds to user's ears.
- Chest-piece pick up sounds from patient's internal organs. (8)
- Sound amplification amplify the sounds from chest-piece. (છ)
- (10) Battery power the device.
- (11) Digital signal processor sound processing.
- (12) Display show the operation information.
- (13) Keypad control and setup the device.
- (14) Detect and display heart rate monitor patient's heart rate.
- (15) Record record a sound track.
- (16) Playback-playback a sound track.
- (17) Filter selection select certain mode of filter for clinical auscultation, including B/ D/ W mode.
- (18) Sound amplification level adjustment control the level of sound amplification.
- (19) Monitor Battery level indicate the battery level.
{14}------------------------------------------------
| Image: iMEDI logo | Document Title: 510(k) Summary | |
|----------------------------------------------|--------------------------------|-----------|
| Document No.: DS3011A-SDUS01-00501 | Page: | 12 of 20 |
| Product Name: Electronic Stethoscope DS3011A | Ver.: | A4 |
| 510(k) File Name: 018_510(k)Summary | Date: | 2018/8/17 |
### Table 5 Same and Different Technological Elements between DS3011A and DS301
| Features | Candidate Device<br>DS3011A | Predicate Device<br>DS301 | Same/Differe<br>nt |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--------------------|
| Wireless connection | Yes | No | Different |
| Connected device software | Yes<br>DS3011A DM software<br>(DS3011A Data Management<br>software)<br>1.Receive the data of sounds<br>2.Data storage<br>3.Display the phonogram<br>4.Replay the recorded data<br>with the spectrum of selected<br>frequency mode<br>5. The audio data could be<br>displayed with the connected<br>device but not on the DS3011A<br>screen. | No | Different |
| Password | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.