The Ureteral Dilator Sets are intended for the dilation of the ureter prior to ureteroscopy or stone manipulation. The Ureteral Dilators are intended for the dilation of the ureter prior to ureteroscopy or stone manipulation.
Device Story
Ureteral dilators and sets facilitate ureteral access for ureteroscopy or stone manipulation. Devices consist of tapered catheters (60 cm length; 6.0-18.0 French diameter) manufactured from radiopaque ethylene vinyl acetate (EVA) or fluorinated ethylene propylene (FEP); some EVA models feature hydrophilic coating. Sets may include a 0.038-inch diameter stainless steel fixed-core wire guide. Used by physicians in clinical settings; dilators are inserted over a wire guide to expand the ureteral lumen. Output is mechanical dilation of the ureter, enabling subsequent endoscopic procedures. Single-use; ethylene oxide sterilized.
Clinical Evidence
Bench testing only. Testing included tensile strength, kink resistance, tip rollback, lubricity, dimensional verification, and radiopacity (ASTM F640-12). Biocompatibility testing (ISO 10993-1) for cytotoxicity, sensitization, and irritation was performed. Sterilization validated per ISO 11135:2014; residuals verified per ANSI/AAMI/ISO 10993-7:2008(R)2012. Packaging integrity tested per ASTM D4169-16 and AAMI/ISO 11607-1:2006/(R)2015.
Technological Characteristics
Materials: Radiopaque ethylene vinyl acetate (EVA) or fluorinated ethylene propylene (FEP). Dimensions: 60 cm length, 6.0-18.0 French diameter. Distal end tapered to 0.040-inch endhole. Wire guide: 0.038-inch diameter stainless steel fixed core. Sterilization: Ethylene oxide. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients requiring dilation of the ureter prior to ureteroscopy or stone manipulation.
Regulatory Classification
Identification
A ureteral dilator is a device that consists of a specially shaped catheter or bougie and is used to dilate the ureter at the place where a stone has become lodged or to dilate a ureteral stricture.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 17, 2018
Cook Incorporated Kotei Aoki Regulatory Affairs Specialist 750 Daniels Way, P.O. Box 489 Bloomington, IN 47404
Re: K173654
> Trade/Device Name: Ureteral Dilator Sets Ureteral Dilators Regulation Number: 21 CFR& 876.5470 Regulation Name: Ureteral Dilator Regulatory Class: II Product Code: EZN Dated: June 27, 2018 Received: June 28, 2018
Dear Kotei Aoki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
#### for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K173654
Device Name Ureteral Dilator Sets Ureteral Dilators
#### Indications for Use (Describe)
The Ureteral Dilator Sets are intended for the dilation of the ureter prior to ureteroscopy or stone manipulation.
The Ureteral Dilators are intended for the dilation of the ureteroscopy or stone manipulation.
Type of Use (Select one or both, as applicable)
∑ Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 ER
{3}------------------------------------------------
#### K173654 Page 1 of 5
Image /page/3/Picture/1 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, bold, sans-serif font on a red background. The red background is shaped like a rectangle with a diagonal cut on the bottom left corner.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
## K173654 Ureteral Dilator Sets Ureteral Dilators 21 CFR 8807.92 Date Prepared: 27 June 2018
#### Submitted By:
Submission: Applicant: Applicant Address:
Traditional 510(k) Premarket Notification Cook Incorporated Cook Incorporated 750 Daniels Way Bloomington, IN 47404 Kotei Aoki regsubmissions(a)cookmedical.com (812) 335-3575 x102630 (812) 332-0281
| Contact: |
|-----------------------|
| Email: |
| Contact Phone Number: |
| Contact Fax Number: |
## Device Information:
| Trade Name: | Ureteral Dilator Sets<br>Ureteral Dilators |
|-----------------------|--------------------------------------------|
| Common Name: | Dilator, Catheter, Ureteral |
| Classification Name: | Ureteral Dilator |
| Regulation: | 21 CFR §876.5470 |
| Product Code: | EZN |
| Device Class: | II |
| Classification Panel: | Gastroenterology/Urology |
## Predicate Device:
The Ureteral Dilator Sets, and each individual Ureteral Dilator, is substantially equivalent to the following device: the AQ Hydrophilic Dilator (K961904, Cook Urological Inc.) cleared on October 18, 1996.
## Device Description:
The Ureteral Dilator Sets, and each individual Ureteral Dilator, are used for dilation of the ureter prior to ureteroscopy and/or stone manipulation. The Ureteral Dilator Sets are available in one of three sets: a set of 6 dilators, a set of 7 dilators and a wire guide, or a set of 9 dilators and a wire guide. The dilators are manufactured from either radiopaque ethylene vinyl acetate (EVA) or fluorinated ethylene propylene (FEP). Dilators manufactured with EVA are also available with hydrophilic coating. Dilators are also
{4}------------------------------------------------
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA ONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
available individually. The wire guide is a stainless steel fixed core wire guide. The Ureteral Dilators and the Ureteral Dilator Sets are sterilized by ethylene oxide and intended for one-time use.
All Ureteral Dilators are manufactured with a length of 60 centimeters and range from 6.0 French to 18.0 French in diameter. The distal end of the dilators is tapered to a 0.040inch diameter endhole with the taper measuring 0.8-2.5 centimeters from the distal end, depending on the device French size. Additionally, the French size of each dilator is stamped at the proximal end.
The stainless steel fixed cored wire guide is manufactured with a diameter of 0.038 inches (0.97 millimeters) and a length of 145 centimeters.
#### Indications for Use:
The Ureteral Dilator Sets are intended for the dilation of the ureter prior to ureteroscopy or stone manipulation.
The Ureteral Dilators are intended for the dilation of the ureter prior to ureteroscopy or stone manipulation.
## Comparison to Predicate Device:
The Ureteral Dilator Sets, and each individual Ureteral Dilator, and the predicate device, the AQ Hydrophilic Dilator (K961904), are substantially equivalent in that these devices have the similar intended use and are identical in technological characteristics and method of placement. Additionally, the proposed devices have similar indications for use and dimensions as the predicate device. The differences between the proposed device and the predicate device, including the materials, dimensions, and indications for use do not raise any new issues of safety and effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, sans-serif font, with a registered trademark symbol to the right of the word. Below the word "COOK" is the word "MEDICAL" in a smaller, sans-serif font. The background of the logo is a solid red color.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA HONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
#### Table 1: Substantial Equivalence Table
| | PREDICATE DEVICE | SUBJECT DEVICE |
|------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| | AQ Hydrophilic Dilators<br>(K961904) | Ureteral Dilator Sets |
| Regulation | 21 CFR §876.5520<br>21 CFR §876.5470 | 21 CFR §876.5470 |
| Product Code | KOE, EZN | EZN |
| Classification | II | Identical |
| Indications for Use | Intended to dilate the suprapubic and/or<br>nephrostomy fascial tract, urethra, and<br>ureters | Intended for dilation of the ureter prior<br>to ureteroscopy or stone manipulation |
| Device for One-Time<br>Use | Yes | Identical |
| Distal Taper | Yes | Identical |
| Endhole Diameter | 0.040 in | Identical |
| Set Components | Dilator, Wire Guide | Identical |
| Sterilization Method | EtO | Identical |
| Sterility Assurance<br>Level | 10-6 | Identical |
| Dilator O.D. | 5.0 - 36.0 Fr. | 6.0 - 18.0 Fr. |
| Dilator Length | 20 - 60 cm | 60 cm |
| Dilator Material | Polyurethane, Polyethylene, Vinyl | Ethylene Vinyl Acetate<br>Fluorinated Ethylene Propylene |
| Hydrophilic Coating | Yes | Optional (EVA only) |
| Packaging | Tyvek-Poly Pouch | Tray (sets only)<br>Tyvek-Poly Pouch<br>Tyvek Peel Pouch |
#### Technological Characteristics:
The subject devices, Ureteral Dilator Sets and Ureteral Dilators, were subjected to the following applicable testing to assure reliable design and performance under the specified testing parameters:
Bench Testing (including time zero and applicable three-year accelerated aged testing),
- Tensile Testing The peak load preceding failure shall be greater than or equal to 15 . N. Testing performed on the dilator shaft per applicable ISO standards demonstrated that the devices met the acceptance criteria.
- Kink Resistance Testing The dilator shaft shall resist kinking at 80° angle. The . dilator tip shall fit over a 0.038 inch guide wire.
- Dilator Tip Rollback Testing The tip of the dilator shall be successfully introduced ● through the urethane membrane without showing any signs of damage such as rollback, buckle or kink at the tip. Testing performed per applicable ISO standards demonstrated that the dilator will not kink, buckle or rollback.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, with the word "MEDICAL" underneath it in a smaller font. The background is a solid red color. The word "MEDICAL" is on a white angled rectangle.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
- Lubricity Testing The peak load during lubricity testing of the hydrophilic coated ● region of the dilators shall be less than the peak load during lubricity testing of the uncoated dilators.
- Dimensional Verification The length of the dilators shall be 60 cm ± 1 cm. FEP . dilators shall include a handle at the proximal end to facilitate removal of the dilator. Dilator markings shall be legible, defined as being able to read markings on dilators. Testing performed demonstrated that the test specimens' dimensions are within the specified tolerances. Testing performed demonstrated that the dilator is compatible with the wire guide.
- . Radiopacity -Dilators shall be radiopaque in compliance with the methods detailed in ASTM F640-12. Testing performed demonstrated that the devices are visible in the radiographic image.
## Biocompatibility Testing:
- Per ISO 10993-1 and FDA guidance, testing for cytotoxicity, sensitization, and . intracutaneous irritation demonstrated the biocompatibility of the subject device.
# Sterility Testing:
- The established method used to validate the sterilization cycle is consistent with the ● half-cycle method as described in ISO 11135:2014.
- . Residual levels of ethylene oxide (EO) and ethylene chlorohydrin (ECH) have been evaluated and verified to be no greater than the maximum limits as specified in ANSI/AAMI/ISO 10993-7:2008(R)2012.
- The subject device is not an implant, nor labeled non-pyrogenic, nor intended to be in ● contact with intravascular, intralymphatic, or intrathecal fluid systems. Therefore, these devices do not require testing for the presence of bacterial endotoxins according to USP<161>.
# Packaging Integrity and Distribution Testing:
- Packaging integrity testing after simulated distribution of these packaging materials . has been performed on a worst-case representative device. Simulated distribution testing was performed according to ASTM D4169-16. All test methods are as recommended in Annex B of AAMI/ISO 11607-1:2006/(R)2015.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Cook Medical. The word "COOK" is in white, bold letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, bold letters on a red background.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
#### Conclusion:
The results of these tests confirm that the Ureteral Dilator Sets, and each individual Ureteral Dilator, meet the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate device, AQ Hydrophilic Dilator (Cook Urological, Inc., K961904).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.