Alere i Strep A 2, Alere i instrument, Alere i Strep A 2 Control Swab Kit
K173653 · Alere Scarborough, Inc. · PGX · May 2, 2018 · Microbiology
Device Facts
Record ID
K173653
Device Name
Alere i Strep A 2, Alere i instrument, Alere i Strep A 2 Control Swab Kit
Applicant
Alere Scarborough, Inc.
Product Code
PGX · Microbiology
Decision Date
May 2, 2018
Decision
SESE
Submission Type
Dual Track
Regulation
21 CFR 866.2680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Alere™ i Strep A 2 is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.
Device Story
Device performs rapid, instrument-based, isothermal nucleic acid amplification for qualitative detection of S. pyogenes DNA in throat swab specimens; used in clinical settings for patients with pharyngitis symptoms. Operator expresses swab in elution buffer; transfers aliquot to Test Base containing lyophilized reagents for enzymatic lysis and amplification. Alere i Instrument automates heating, mixing, and fluorescent detection of target and Internal Control. Instrument displays and archives results; provides rapid diagnosis to guide clinical treatment. Benefits include faster turnaround than traditional culture, aiding timely antibiotic decision-making.
Clinical Evidence
Prospective multi-center clinical study (n=981 evaluable subjects) compared Alere i Strep A 2 to reference bacterial culture. Sensitivity was 98.5% (95% CI: 95.6-99.5%) and specificity was 93.4% (95% CI: 91.4-94.9%). Prevalence was 20.2% by culture. Initial invalid rate was 0.9%, final invalid rate 0.2%.
Technological Characteristics
Isothermal nucleic acid amplification; strand-displacing DNA polymerase; fluorescent Molecular Beacon probes. Single-use disposables: Sample Receiver (elution buffer), Transfer Cartridge, Test Base (lyophilized reagent pellets). Instrument performs automated heating, magnetic mixing, and real-time optical detection. Standalone instrument; no external connectivity. No end-user calibration required.
Indications for Use
Indicated for patients with signs and symptoms of pharyngitis to aid in the rapid diagnosis of Group A Streptococcus bacterial infections. Prescription use only. Follow-up culture required if negative result and symptoms persist or during acute rheumatic fever outbreaks.
Regulatory Classification
Identification
A Streptococcus spp. nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify various Streptococcus spp. nucleic acids extracted directly from clinical specimens. The device detects specific nucleic acid sequences for organism identification. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.
Special Controls
(b) Classification. Class II (special controls). The special controls for this device are: 1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection. 2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination. 3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods. 4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardwarebased devices that incorporate software. 5) Premarket notification submissions must include database implementation methodology, construction parameters and quality assurance protocols, as appropriate. 6) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate. 7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling. 8) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols, as appropriate.
(6) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(8) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.
{1}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K173653
B. Purpose for Submission:
To obtain a Substantial Equivalence determination for the Alere i Strep A 2 performed on the Alere i Analyzer for the detection of Streptococcus pyogenes (Group A Streptococcus).
C. Measurand:
Group A Streptococcus DNA
D. Type of Test:
Isothermal DNA amplification assay for the qualitative detection of Group A Streptococcus in throat swab specimens.
E. Applicant:
Alere Scarborough, Inc.
F. Proprietary and Established Names:
Alere i Strep A 2
Alere i Instrument
Alere i Strep A 2 Control Swab Kit
G. Regulatory Information:
1. Regulation section:
21 CFR 866.2680 Streptococcus spp. Nucleic Acid-Based Assay
2. Classification:
Class II
3. Product codes:
PGX: Groups A, C and G Beta-Hemolytic Streptococcus Nucleic Acid Amplification System
1
{2}
OOI: Real-Time Nucleic Acid Amplification System
4. Panel:
83: Microbiology
### H. Intended Use:
1. Intended use(s):
Alere i Strep A 2 is a rapid, instrument-based, molecular in vitro diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of Streptococcus pyogenes, Group A Streptococcus bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A Streptococcus bacterial infections.
2. Indication(s) for use:
Same as Intended Use.
3. Special conditions for use statement(s):
For prescription use only.
Additional follow-up testing by culture is required if the Alere i Strep A 2 assay result is negative and clinical symptoms persist, or in the event of an outbreak of acute rheumatic fever (ARF).
For optimal performance, use the specimen collection swabs provided in the test kit. Analytical studies have demonstrated that rayon swabs and the BBL CultureSwab Liquid Stuart Medium are not suitable for use with this assay and may produce false negative results.
4. Special instrument requirements:
Alere i Instrument
### I. Device Description:
Alere i Strep A 2 is a rapid, instrument-based isothermal assay for the qualitative detection of Streptococcus pyogenes (Group A Streptococcus) DNA in throat swab specimens from patients suspected of pharyngitis. Alere i Strep A 2 is performed on the Alere i Instrument and is comprised of the following single-use disposables:
- Sample Receiver
○ Containing Elution Buffer for swab expression
2
{3}
• Transfer Cartridge
○ For transfer of the eluted sample to the Test Base
• Test Base
○ Consisting of two sealed reaction tubes, each containing a lyophilized reagent pellet;
○ One tube is for detection the S. pyogenes target, the other for an oligonucleotide Internal Control
Alere i Strep A was originally cleared under K141757. Alere i Strep A 2 includes various modifications to the original Alere i Strep A assay chemistry to improve clinical sensitivity and alleviate the need for culture confirmation of negative test results, as well as to improve manufacturability of the reagents.
The reaction tubes in the Test Base contain the reagents required for enzymatic lysis of S. pyogenes, and the subsequent amplification of the target nucleic acid and an Internal Control. Alere i Strep A 2 uses a pair of templates (similar to primers) and a strand displacing DNA polymerase to amplify the S. pyogenes target DNA and Internal Control. Detection of the amplified products occurs through hybridization of fluorescently labeled Molecular Beacon probes. Alere i Strep A 2 is performed within the confinement of the Test Base, and no other part of the Alere i Instrument is in contact with the sample during the amplification process, thereby reducing the risk of instrument contamination and sample carry-over between measurements.
To perform the assay, the Sample Receiver and Test Base are inserted into the Alere i Instrument. A throat swab is expressed in the Elution Buffer contained within the Sample Receiver and a Transfer Cartridge is used to move the specified volume of swab eluate to the Test Base to dissolve the lyophilized reagent pellets that contain the reagents for bacterial lysis and target amplification. Heating of the reaction tubes, mixing of reagents, fluorescent detection of the amplification products and result interpretation are performed automatically by the instrument.
Inactivated bacterial Positive and Negative External Controls for use with Alere i Strep A 2 are provided within the assay kit and are also available separately. The controls consist of swabs that are coated with enumerated suspensions of inactivated bacteria and dried. The Positive Control is comprised of S. pyogenes while the Negative Control is comprised of Group C Streptococcus.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Alere i Strep A
2. Predicate 510(k) number(s):
K141757
3
{4}
# 3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Alere i Strep A 2 (K173653) | Alere i Strep A (K141757) |
| Intended Use | Alere i Strep A 2 is a rapid, instrument-based, molecular *in vitro* diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of *Streptococcus pyogenes*, Group A *Streptococcus* bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A *Streptococcus* bacterial infections. | Alere i Strep A is a rapid, instrument-based, molecular *in vitro* diagnostic test utilizing isothermal nucleic acid amplification technology for the qualitative detection of *Streptococcus pyogenes*, Group A *Streptococcus* bacterial nucleic acid in throat swab specimens obtained from patients with signs and symptoms of pharyngitis. It is intended to aid in the rapid diagnosis of Group A *Streptococcus* bacterial infections. All negative test results should be confirmed by bacterial culture because negative results do not preclude infection with Group A *Streptococcus* and should not be used as the sole basis for treatment. |
| Regulation | 21 CFR 866.2680 | Same |
| Product Codes | PGX, OOI | Same |
| Sample Type | Throat swab | Same |
| Assay Method | Isothermal nucleic acid amplification and real-time fluorescent detection | Same |
| Extraction Method | None (assay performed on crude lysate) | Same |
| Detection Technique | Fluorescent DNA probe | Same |
| Internal Control | Yes | Same |
| Assay Result | Qualitative | Same |
4
{5}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Alere i Strep A 2 (K173653) | Alere i Strep A (K141757) |
| Amplification Primers (templates) | Modified to enhance performance | Targeted towards conserved regions of the *S. pyogenes* genome |
| Amplification Reaction Mixtures | Formulation optimized to enhance performance and manufacturability | Separate reaction mixtures for target nucleic acid and Internal Control |
| Internal Control | Sequence modified and concentration increased to improve tolerance to clinical matrix | Present |
| Mixing of Amplification Reactions | Magnetic mixing added to enhance performance | None |
| Sample Receiver Warm Up | Eliminated to reduce time-to-result | 2 min incubation required prior to addition of sample |
| Time to Result | <6 minutes | <8 minutes |
### K. Standard/Guidance Document Referenced (if applicable):
Not applicable.
### L. Test Principle:
Alere i Strep A 2 uses isothermal nucleic acid amplification for qualitative detection of *Streptococcus pyogenes* (Group A *Streptococcus*) DNA from throat swab specimens from patients with signs and symptoms of pharyngitis. Alere i Strep A 2 for use on the Alere i Instrument consists of the following components:
- a) Sample Receiver: containing Elution Buffer;
- b) Transfer Cartridge: for transfer of an aliquot of the eluted sample to the Test Base;
- c) Test Base: comprised of two sealed reaction tubes, each containing a lyophilized reagent pellet; one tube is for detection of the *S. pyogenes* target (if present), the other tube is for detection of the Internal Control.
The reaction tubes in the Test Base contain the reagents required for enzymatic lysis of *S. pyogenes* bacteria and amplification of the target nucleic acid and Internal Control. Isothermal amplification takes place using specific templates (similar to primers) that target a conserved region of the *S. pyogenes* genome. The same template primers are also used to amplify the oligonucleotide Internal Control. Detection of amplified products occurs through hybridization of fluorescently labeled Molecular Beacons that are specific to the intervening
5
{6}
regions between the template primers on the Internal Control and S. pyogenes amplicons. Result interpretation is based upon the gradient and amplitude of the fluorescence curves for the S. pyogenes and Internal Control targets in their respective optical channels and comparison to the appropriate threshold values.
To perform the assay, the Sample Receiver and Test Base are inserted into the Alere i Instrument. A throat swab is then expressed in the Elution Buffer. The operator uses the Transfer Cartridge to transfer a portion of the expressed sample to the Test Base to dissolve the lyophilized reagent pellets and initiate cell lysis and DNA amplification. Heating, mixing and fluorescent detection are conducted automatically by the instrument. Results are displayed on the instrument screen and archived on-board together with the sample identification number and date and time of testing. Results may also be printed.
### M. Performance Characteristics (if/when applicable):
#### 1. Analytical performance:
##### a. Precision/Reproducibility:
A study to evaluate the reproducibility of Alere i Strep A 2 was conducted at three sites, over a period of five days, with three operators per site. Swab samples for testing were prepared with throat swab matrix that was seeded with different levels of S. pyogenes strain ATCC 19615. The target levels tested included a true negative (no target), Low Positive (2X Limit of Detection (LoD) and Moderate Positive (3X LoD). On each day of testing, each operator tested a blinded panel consisting of two replicates of each target level, for a total of 90 data points per target level (5 days x 3 sites x 3 operators x 2 replicates = 90 replicates per target level). In total, 11 Alere i Instruments were included in the study, together with two lots each of the assay components (Sample Receiver, Test Base and Test Cartridge). All results were as expected (Table 1). The reproducibility of Alere i Strep A 2 was therefore considered acceptable.
Table 1. Results from the Alere i Strep A 2 Reproducibility Study by site and overall
| Target Level | Number (%) | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 | | Site 2 | | Site 3 | | Overall | |
| | Positive | Negative | Positive | Negative | Positive | Negative | Positive | Negative |
| Negative | 0 | 30(100) | 0 | 30(100) | 0 | 30(100) | 0 | 90(100) |
| Low Positive(2X LoD) | 30(100) | 0 | 30(100) | 0 | 30(100) | 0 | 90(100) | 0 |
| Moderate Positive(3X LoD) | 30(100) | 0 | 30(100) | 0 | 30(100) | 0 | 90(100) | 0 |
6
{7}
# b. *Linearity/assay reportable range:*
Not applicable.
# c. *Traceability, Stability, Expected values (controls, calibrators, or methods):*
# *Internal Control*
Alere i Strep A 2 contains an oligonucleotide Internal Control that is designed to monitor the amplification/detection process and reagent function. The Internal Control is lyophilized in one of the two reaction tubes in the Test Base and is amplified and detected separately from the *S. pyogenes* target (if present) using a different optical channel of the Alere i Instrument. The Internal Control must be amplified and detected for a negative assay result to be reported. For *S. pyogenes* positive samples in which target amplification occurs, the amplification of the Internal Control is ignored. If *S. pyogenes* DNA is not detected and the Internal Control fails to amplify the test result is reported as “Invalid”.
# *Positive and Negative Control Swabs*
The Alere i Strep A 2 kit contains swabs for use as Positive and Negative Controls that should be tested when the assay is performed on an instrument for the first time, with each new shipment of reagents and by each new operator. Additional controls are available as a separate kit and may be tested as necessary to comply with Quality Control requirements. The Positive Control Swab is coated with inactivated Group A *Streptococcus*. The Negative Control Swab is coated with inactivated Group C *Streptococcus*.
During the prospective Clinical Study described in Section M(3)(a), Positive and Negative Control Swabs were run on each Alere i instrument on each day that clinical specimens were tested. The expected results were obtained from 730/749 (97.5%) Positive Control Swabs and 729/742 (98.2%) Negative Control Swabs.
# *Reagent Stability*
Alere i Strep A 2 reagents and controls should be stored at 2-30°C until the expiration date shown on the outer packaging.
# *Specimen Stability*
During the Clinical Study that was conducted to establish the performance of Alere i Strep A 2 (Section M(3)(a)), the majority of throat swab specimens (898/981, 91.5%) were tested within 1 hour of collection. Those specimens not tested within 2 hours of collection were refrigerated at 2-8°C and tested within 24 hours.
Specimen stability was demonstrated analytically using alternative types of collection swab that were seeded with throat swab matrix either with or without *S. pyogenes*, and stored for different durations prior to testing. Ten *S. pyogenes* positive (~3X LoD) and 10 *S. pyogenes* negative samples were tested at each time point after
7
{8}
storage at either 2-8°C or 22°C. The sample storage conditions that were validated for each swab type are summarized in Table 2.
Table 2. Acceptable transport and storage conditions demonstrated analytically with alternative specimen collection devices
| Collection & Transport Device | Maximum Storage Time | |
| --- | --- | --- |
| | 2-8°C | 22 °C |
| Puritan Foam Swab | 72 hours | 72 hours |
| Puritan Polyester Swab | 72 hours | 72 hours |
| Puritan Hydraflock Swab | 72 hours | 72 hours |
| Copan Nylon Flocked Swab | 72 hours | 72 hours |
| BBL CultureSwab Liquid Amies | 6 hours | 6 hours |
The analytical Specimen Stability Study also showed that rayon swabs and the BBL CultureSwab Liquid Stuart specimen collection and transport system are not compatible with Alere i Strep A 2 and should not be used. This is noted as a Limitation in the device labeling.
### d. Detection limit:
### Limit of Detection
The Limit of Detection (LoD) of Alere i Strep A 2 was estimated for two strains of S. pyogenes by testing various dilutions of enumerated cell stocks in throat swab matrix. The LoD for each strain was then confirmed by testing a further 20 replicates at the lowest target level that produced 100% positive results. The LoD was defined as the lowest concentration tested at which \( \geq95\% \) of assay replicates produced positive results.
Target levels were determined by measuring the optical density of stock suspensions of S. pyogenes at 630nm (OD \( _{630} \) ) and using a quantitative laboratory-developed PCR assay to estimate genomic equivalents. \( ^{1} \) Chamber counts performed by microscopy showed that an OD \( _{630} \) of 0.1 was equivalent to 2 x 10 \( ^{8} \) cells/mL. The results of the study are shown in Table 3. The higher of the two LoD values expressed in terms of genomic copies was used as the basis for testing the inclusivity of Alere i Strep A 2 (Table 4).
\( ^{1} \) Morozumi, M., et al. J Clin Microbiol 2006 44: 1440-1446
8
{9}
Table 3. Limit of detection (LoD) for Alere i Strep A 2
| ATCC Strain | Limit of Detection(per mL of Elution Buffer) | |
| --- | --- | --- |
| | Cells1 | Genomic Copies2 |
| 12344 | 147 | 223 |
| 19615 | 25 | 512 |
ATCC: American Type Culture Collection
\( ^{1} \) As determined by \( OD_{630} \) and correlation with chamber counts
\( ^{2} \) As determined by quantitative PCR
### Inclusivity
The inclusivity of Alere i Strep A 2 was evaluated by testing 14 strains of S. pyogenes in addition to those included in the LoD Study. Testing was performed in throat swab matrix and the lowest dilution of each strain at which each of 10 assay replicates produced a positive result is shown in Table 4.
Table 4. Reactivity titers of different strains of S. pyogenes
| Source | Strain | Genomic copies Per mL Elution Buffer1 | Multiple of LoD for ATCC 19615 |
| --- | --- | --- | --- |
| ATCC | 8135 | 1537 | 3 |
| | 12384 | 1537 | 3 |
| | 12202 | 1537 | 3 |
| | 12203 | 1537 | 3 |
| | 12204 | 1537 | 3 |
| | 12365 | 1537 | 3 |
| | 14289 | 1537 | 3 |
| | 49399 | 3074\( ^{2} \) | 6 |
| | 51339 | 1537 | 3 |
| | 700294 | 1537 | 3 |
| | 12357 | 2049\( ^{2} \) | 4 |
| | 12385 Type 4 | 1537 | 3 |
| | 12385 Loomis | 1537 | 3 |
| Zeptometrix | Z018 | 3074\( ^{2} \) | 6 |
ATCC: American Type Culture Collection; LoD: Limit of Detection
\( ^{1} \) Level at which 10/10 assay replicates produced positive results; the lowest level tested was 1537 genomic copies/mL, approximately 3X the LoD determined for strain ATCC 19615
\( ^{2} \) <10/10 positive results were obtained at lower target levels
### Bioinformatic analysis:
The inclusivity of Alere i Strep A 2 assay was also assessed in silico by searching the National Center for Biotechnology Information (NCBI) Nucleotide database (https://www.ncbi.nlm.nih.gov/nuccore) using the Basic Local Alignment Search Tool (BLAST), with the S. pyogenes amplicon sequence as the query sequence. The region of the S. pyogenes genome targeted by the assay was shown to be well conserved, with no evidence of sequence heterogeneity that could lead to false
9
{10}
negative results. This analysis supports the ability of Alere i Strep A 2 to detect different strains of S. pyogenes and is acceptable.
### e. Analytical specificity:
#### Cross-Reactivity
The analytical specificity of Alere i Strep A 2 was evaluated by testing a variety of commensal and pathogenic species of bacteria and yeast (Table 5). Suspensions of each organism were diluted in phosphate buffered saline and tested in triplicate at a final concentration \( \geq2\times10^{6} \) cells/mL of Elution Buffer. All results were as expected and no evidence of cross-reaction or assay interference was observed.
Table 5. Species tested for cross-reaction in Alere i Strep A
| Bacteria | |
| --- | --- |
| Arcanobacterium haemolyticum | Staphylococcus aureus |
| Bacillus cereus | Staphylococcus epidermidis |
| Bordetella pertussis | Streptococcus agalactiae |
| Burkholderia cepacia | Streptococcus anginosus |
| Campylobacter rectus | Streptococcus canis |
| Corynebacterium diphtheriae | Streptococcus constellatus subsp. pharyngis |
| Enterococcus faecalis | Streptococcus dysgalactiae subsp. equisimilis |
| Escherichia coli | Streptococcus gallolyticus |
| Fusobacterium necrophorum | Streptococcus intermedius |
| Haemophilus influenzae | Streptococcus mitis |
| Klebsiella pneumoniae | Streptococcus mutans |
| Lactobacillus acidophilus | Streptococcus pneumoniae |
| Moraxella catarrhalis | Streptococcus salivarius |
| Neisseria gonorrhoeae | Streptococcus sanguinus |
| Peptostreptococcaceae\( ^{1} \) | Treponema denticola |
| Prevotella oralis\( ^{2} \) | Veillonella parvula |
| Pseudomonas aeruginosa | |
| Yeast | |
| Candida albicans | |
\( ^{1} \) Formerly Peptostreptococcus micros
\( ^{2} \) Formerly Bacteroides oralis
#### Bioinformatic Analysis
To evaluate the potential for cross-reaction of Alere i Strep A 2 with non-target organisms, BLAST analyses were performed against the NCBI Nucleotide database using the assay target region as the query sequence. The search showed no evidence of homology between the Alere i Strep A 2 target sequence and clinically relevant species that is likely to result in cross-reaction and generation of false-positive results.
#### Contamination Study
The potential for contamination between samples tested on the same Alere i Instrument was assessed by testing an alternating series of Alere i Strep A (K141757) Positive and Negative Control Swabs. The Positive Control Swabs resulted in a final
10
{11}
concentration of S. pyogenes of 1.5-2.5 CFU/mL of Elution Buffer while the Negative Controls delivered Group C Streptococcus at a concentration of \( 5 \times 10^{6} \) CFU/mL. Thirty 30 Positive and 30 Negative Control Swabs were tested on each of two Alere i Instruments (i.e., 60 Positive and 60 Negative Swabs in total). All the controls produced the expected results and no false positive results were observed, indicating that the risk of contamination between runs is acceptably low.
### f. Assay cut-off:
Alere i Strep A 2 is comprised of two separate amplification reactions that occur in separate tubes; one for the S. pyogenes target and one for the Internal Control. The products of each amplification reaction are detected in separate optical channels by monitoring the accumulation of fluorescent signal in real-time. The gradient and amplitude of the fluorescence curve generated in each optic are analyzed and used in a decision algorithm to determine the final assay result. Once a definitive result for a sample is obtained, the assay is terminated. The gradient thresholds (cut-offs) for the assay were established through analysis of known positive and negative samples and were validated in the Clinical Study described in Section M(3)(a).
### g. Assay Interference
The potential for interference with Alere i Strep A 2 was evaluated with 21 endogenous and exogenous substances that may be present in throat swab specimens. Each substance was tested in triplicate in the presence and absence of S. pyogenes ATCC 19615 at a concentration of 3X LoD (1537 genomic copies/mL of Elution Buffer). False negative results were obtained with both human saliva and mucin at concentrations of 10% v/v and 2% w/v, respectively. In addition, a false positive result was observed in the presence of Listerine Antiseptic Mouthwash at 20% v/v. In each case, the expected results were obtained when the substances were tested at lower concentration. These results are reflected in the device labeling. No interference was observed with any of the substances when tested at the concentrations listed in Table 6.
Table 6. Substances tested for potential interference with Alere i Strep A 2
| Substance | Concentration Tested1 |
| --- | --- |
| Acetaminophen | 60.4 mg/mL |
| Acetylsalicylic Acid | 0.65 mg/mL |
| ACT Total Care | 20% (v/v) |
| Albuterol | 0.40 mg/mL |
| Cepacol Mouthwash | 20% (v/v) |
| Cepacol Sore Throat Lozenges | 20% (w/v) |
| Children's Dimetapp Cough & Cold | 20% (v/v) |
| Chloraseptic Max Sore Throat Relief + Coating Action | 20% (v/v) |
| Contact Cold & Flu Tablets | 20% (w/v) |
11
{12}
| Substance | Concentration Tested1 |
| --- | --- |
| Crest Complete Multi-Benefit Whitening + Deep Clean Toothpaste | 20% (w/v) |
| Diphenhydramine HCl | 1 mg/mL |
| Halls Plus | 20% (w/v) |
| Human Saliva | 5% (v/v)2 |
| Ibuprofen | 20 mg/mL |
| Listerine Antiseptic Mouthwash | 10% (v/v)3 |
| Mucin | 1% (w/v)4 |
| Robitussin Maximum Strength Nighttime Cough DM | 20% (v/v) |
| Sucrets Sore Throat & Cough | 20% (w/v) |
| Tylenol Cold Multi-Symptom Liquid | 20% (v/v) |
| Whole Blood | 5% (v/v) |
| Zicam Oral Mist | 20% (v/v) |
\( ^{1} \) Highest concentration in throat swab matrix at which no interference was observed (i.e., no false positive, false negative or invalid results obtained)
\( ^{2} \) 1/3 replicates at 10% v/v saliva produced a false negative result
\( ^{3} \) 1/3 replicates at 20% v/v Listerine Antiseptic Mouthwash produced a false positive result
\( ^{4} \) 1/3 replicates at 2% w/v mucin produced a false-negative result
### 2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
The compatibility of alternative specimen collection and transport devices with Alere i Strep A 2 was demonstrated in the context of the Specimen Stability Studies described in Section M(1)(c).
### 3. Clinical studies:
a. Clinical Sensitivity:
The clinical performance of Alere i Strep A 2 was established in a multi-center, prospective Clinical Study that was conducted at nine (9) geographically diverse U.S. sites in 2017 among patients who presented with a sore throat or symptoms of pharyngitis. A total of 1028 subjects were initially enrolled in the study of whom 43 were excluded from the analysis of performance for the following reasons: error in sample handling (21), failure to meet eligibility criteria (11), failure to comply with informed consent procedures (11). Of the remaining subjects, 976/985 (99.1%) generated valid Alere i Strep A 2 results on initial testing (initial invalid rate 9/985; 0.9%). Of the nine (9) samples that initially produced invalid results, eight (8) were
12
{13}
retested and five (5/8) produced valid results for a final invalid rate of 4/985 (0.2%).
Of the 981 subjects with evaluable Alere i Strep A 2 test results, 582 (59.3%) were female and 399 (40.7%) were male.
Two throat swabs were collected from each subject. One swab was used for the reference bacterial culture according to standard procedures (inoculation of sheep blood agar followed by identification of \( \beta \) -hemolytic colonies by latex agglutination); while the other swab was tested with Alere i Strep A 2. Swabs for analysis by the reference culture method were shipped to a central laboratory in dry transport tubes. All reference cultures were inoculated within 48 hours of specimen collection and culture plates were incubated for 48 hours before being discarded as negative. The overall performance of Alere i Strep A 2 in relation to the reference culture method is summarized in Table 7 and stratified by site in Table 8. The performance of Alere i Strep A 2 observed in the Clinical Study was considered acceptable.
Table 7. Overall clinical performance of Alere i Strep A 2 versus bacterial culture
| | Reference Culture | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| Alere i Strep A 2 | Positive | 195 | \( 52^1 \) | 247 |
| | Negative | \( 3^2 \) | 731 | 734 |
| | Total | 198 | 783 | \( 981^3 \) |
| Sensitivity | | 98.5% (195/198); 95% CI: 95.6-99.5% | | |
| Specificity | | 93.4% (731/783); 95% CI: 91.4-94.9% | | |
| Positive Predictive Value | | 78.9% (195/247) | | |
| Negative Predictive Value | | 99.6% (731/734) | | |
| Prevalence | | 20.2% (198/981) | | |
95% CI: 95% score confidence interval
\( ^{1} \) 38/52 (73.1%) samples with Alere i Strep A 2 false-positive results relative to culture were positive by an alternative laboratory developed real-time PCR method
\( ^{2} \) 2/3 (66.7%) samples with Alere i Strep A 2 false-negative results relative to culture were positive by an alternative laboratory developed real-time PCR method
\( ^{3} \) 943/981 (96.1%) of specimens were tested with Alere i Strep A 2 within 2 hours of collection
13
{14}
**Table 8.** Alere i Strep A 2 performance stratified by site
| Site | Samples (%) | Culture Positive (% Prevalence) | Percent (%; 95% Score Confidence Interval) | |
| --- | --- | --- | --- | --- |
| | | | Sensitivity | Specificity |
| **001A** | 50 (5.1) | 19 (38.0) | 18/19 (94.7; 75.4-99.1) | 26/31 (83.9; 67.4-92.9) |
| **009A** | 150 (15.3) | 21 (14.0) | 21/21 (100; 84.5-100) | 120/129 (93.0; 87.3-96.3) |
| **024A** | 57 (5.8) | 15 (26.3) | 15/15 (100; 79.6-100) | 36/42 (85.7; 72.2-93.3) |
| **033A** | 119 (12.1) | 23 (19.3) | 23/23 (100; 85.7-100) | 92/96 (95.8; 89.8-98.4) |
| **060A** | 195 (19.9) | 11 (5.6) | 10/11 (90.9; 62.3-98.4) | 179/184 (97.3; 93.8-98.8) |
| **061A** | 98 (10.0) | 15 (15.3) | 14/15 (93.3; 70.2-98.8) | 74/83 (89.2; 80.7-94.2) |
| **066A** | 163 (16.6) | 39 (23.9) | 39/39 (100; 91.0-100) | 120/124 (96.8; 92.0-98.7) |
| **071A** | 89 (9.1) | 31 (34.8) | 31/31 (100; 89.0-100) | 50/58 (86.2; 75.1-92.8) |
| **074A** | 60 (6.1) | 24 (40.0) | 24/24 (100; 86.2-100) | 34/36 (94.4; 81.9-98.5) |
| **Total** | **981 (100)** | **198 (20.2)** | **195/198 (98.5; 95.6-99.5)** | **731/783 (93.4; 91.4-94.9)** |
# *b. Clinical specificity:*
Refer to Section M(3)(a), above.
# *c. Other clinical supportive data (when a. and b. are not applicable):*
Not applicable.
# 4. Clinical cut-off:
Not applicable.
# 5. Expected values/Reference range:
The performance of the Alere i Strep 2 was evaluated in a prospective Clinical Study conducted at nine (9) sites in the US (Section M(3)(a)). The overall prevalence of *S. pyogenes* (Group A *Streptococcus*) in throat swab specimens was 25.2% (247/981) as determined by Alere i Strep A 2 and 20.2% (198/981) as determined by culture. In Table 9, the prevalence of *S. pyogenes* as determined by Alere i Strep A 2 is stratified by the age and gender of the subjects.
14
{15}
Table 9. Prevalence of S. pyogenes positive subjects by age and gender
| Age/Gender | Number | Alere i Strep A 2 Positive | % Prevalence1 |
| --- | --- | --- | --- |
| <2 years | 25 | 5 | 20.0 |
| 2-11 years | 421 | 155 | 36.8 |
| 12-21 years | 240 | 38 | 15.8 |
| 22-59 years | 254 | 47 | 18.5 |
| ≥60 years | 41 | 2 | 4.9 |
| Male | 399 | 99 | 24.8 |
| Female | 582 | 148 | 25.4 |
| Total | 981 | 247 | 25.2 |
\( ^{1} \) As determined using Alere i Strep A 2
### N. Instrument Name:
Alere i Instrument
### O. System Descriptions:
#### 1. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ____ No ____ X
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ____ No ____ X
#### 2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ____ X ____ No ____
#### 3. Specimen Identification:
Specimens are identified by scanning or typing in the applicable patient identification code.
15
{16}
4. Specimen Sampling and Handling:
Swab specimens are expressed in the Elution Buffer contained within the Sample Receiver. The operator transfers 2 x 100μL aliquots of the expressed specimen to the Test Base using the Transfer Cartridge to rehydrate the reagents for target and Internal Control amplification/detection. Refer to Section I above for additional information.
5. Calibration:
End-user calibration for the Alere i Instrument is not required. Calibration of the optical and thermal systems is performed during the manufacturing process. The instrument is maintenance-free and has no serviceable parts. However, if the instrument is transported or moved, a performance check using Alere i Positive and Negative Controls is recommended by the manufacture prior to testing patient specimens to ensure proper functionality.
6. Quality Control:
Refer to Section M(1)(c).
# P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
The Alere i Instrument was initially cleared under K141520 (Alere i Influenza A&B on the Alere i Instrument).
# Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
# R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.