The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for: • viewing the posterior segment of the eye, including two- and three-dimensional imaging • cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT) • fundus imaging • fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA) • autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak) • performing measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various ocular diseases, including: • age-related macular degeneration • macular edema • diabetic retinopathy • retinal and choroidal vascular diseases • glaucoma The device is indicated for viewing geographic atrophy. The SPECTRALIS HRA+OCT and SPECTRALIS OCT include the following reference databases: • a retinal nerve fiber layer thickness reference database, which is used to quantitatively compare the retinal nerve fiber layer in the human retina to values of Caucasian normal subjects - the classification result being valid only for Caucasian subjects • a reference database for retinal nerve fiber thickness and optic nerve head neuroretinal rim parameter measurements, which is used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values of normal subjects of different races and ethnicities representing the population mix of the USA (Glaucoma Module Premium Edition)
Device Story
Non-contact ophthalmic diagnostic imaging device; used in office settings by ophthalmologists. Inputs: optical signals from eye posterior segment via standard, Wide Field (WFO), or Ultra-Widefield objective lenses. Transforms inputs into 2D/3D cross-sectional, fundus, fluorescence, and autofluorescence images. Includes reference databases for quantitative comparison of retinal nerve fiber layer and optic nerve head parameters. Output: visual images and anatomical measurements displayed to clinician. Clinical decision-making: aids detection and management of ocular diseases (e.g., AMD, glaucoma, diabetic retinopathy). Benefits: non-invasive visualization and quantitative assessment of ocular structures.
Clinical Evidence
Bench testing only. Evaluated optical resolution using 1951 USAF resolution test chart (MIL-STD-150A). Compared standard objective (SO) vs. Wide Field Objective (WFO); SO average resolution 13um, WFO 18um. Results consistent with theoretical evaluation. Compliance with ISO 14971, IEC 60601-1, ISO 10993, and IEC 62304 verified.
Technological Characteristics
Non-contact ophthalmic imaging; laser scanning ophthalmoscopy and OCT. Materials/components: WFO/WFO2 objective lenses with anti-reflective coatings. Connectivity: Networked/PC-based (Heyex 2 system, Windows 10). Standards: ISO 14971, AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 10993, IEC 60825-1, IEC 62304.
Indications for Use
Indicated for patients requiring posterior segment eye imaging and measurement of ocular anatomy/lesions to aid in detection/management of age-related macular degeneration, macular edema, diabetic retinopathy, retinal/choroidal vascular diseases, glaucoma, and geographic atrophy. Includes reference databases for retinal nerve fiber layer thickness (Caucasian subjects) and optic nerve head parameters (multi-ethnic USA population).
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 30, 2018
Heidelberg Engineering GmbH % Lena Sattler Consultant Orasi Consulting, LLC. 1655 Forest Dr. Medina, Ohio 44256
Re: K173648
Trade/Device Name: Spectralis HRA + OCT and variants Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO, MYC Dated: May 30, 2018 Received: June 1, 2018
Dear Lena Sattler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Alexander Beylin -S 2018.08.30 13:12:03 -04'00'
for Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173648
### Device Name Spectralis HRA+OCT and variants
### Indications for Use (Describe)
The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for:
- · viewing the posterior segment of the eye, including two- and three-dimensional imaging
- · cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT)
- · fundus imaging
· fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA)
- · autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak)
- · performing measurements of ocular anatomy and ocular lesions.
The device is indicated as an aid in the detection and management of various ocular diseases, including:
- · age-related macular degeneration
- macular edema
- · diabetic retinopathy
- · retinal and choroidal vascular diseases
- glaucoma
The device is indicated for viewing geographic atrophy.
The SPECTRALIS HRA+OCT and SPECTRALIS OCT include the following reference databases:
· a retinal nerve fiber layer thickness reference database, which is used to quantitatively compare the retinal nerve fiber layer in the human retina to values of Caucasian normal subjects – the classification result being valid only for Caucasian subjects
• a reference database for retinal nerve fiber thickness and optic nerve head neuroretinal rim paramements, which is used to quantitatively compare the retinal nerve fiber laver and neuroretinal rim in the human retina to values of normal subjects of different races and ethnicities representing the population mix of the USA (Glaucoma Module Premium Edition)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a stylized, blocky font. The top line reads "HEIDELBERG" and the bottom line reads "ENGINEERING". A maroon square is on the left of the top line and on the right of the bottom line.
# 510(K) SUMMARY
Date Prepared
June 12, 2018
## SPONSOR/510(K) OWNER/ MANUFACTURER
Heidelberg Engineering GmbH Max-Jarecki-Strasse 8 69115 Heidelberg Germany Telephone: Facsimile: Email: Establishment Registration No.:
+49 6221 64630 +49 6221 / 64 63 62 Christoph.Schoess@HeidelbergEngineering.com 8043762
## OFFICIAL CONTACT PERSON
Lena Sattler Orasi Consulting, LLC. 1655 Forest Dr. Medina, OH 44256 Telephone: Facsimile: E-mail:
(440) 554-3706 (866) 904-4315 lena@orasiconsulting.com
## COMMON/USUAL NAME
Optical Coherence Tomography
## PROPRIETARY OR TRADE NAMES
## SPECTRALIS HRA+OCT and variants
## CLASSIFICATION INFORMATION
| Classification Name: | Ophthalmoscope, AC Powered |
|-----------------------|---------------------------------|
| | Ophthalmoscope, Laser, Scanning |
| Medical Specialty: | Ophthalmic |
| Device Class: | II |
| Classification Panel: | Ophthalmic Device Panel |
| Product Codes: | OBO, MYC |
Heidelberg Engineering GmbH
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The top line reads "HEIDELBERG" and the bottom line reads "ENGINEERING". There is a red square to the left of the top line and to the right of the bottom line.
## PRODUCT CODE: CLASSIFICATION / CFR TITLE
OBO. MYC:
Class II § 21 CFR 886.1570
## LEGALLY MARKETED UNMODIFIED PREDICATE DEVICE
| Trade/Device Name: |
|---------------------------------------|
| Applicant: |
| 510(k) Premarket Notification number: |
| Classification: |
| CFR Title: |
| FDA Product Code(s): |
| Classification Name: |
Common Name: Medical Specialty: Classification Panel:
SPECTRALIS HRA+OCT and variants Heidelberg Engineering GmbH K172649 Class II 21 CFR 886.1570 OBO, MYC Ophthalmoscope, AC Powered Ophthalmoscope, Laser, Scanning Optical Coherence Tomography Ophthalmic Ophthalmic Device Panel
The device name and classification information for the modified device is identical to the cleared device.
## GENERAL DEVICE DESCRIPTION
The Heidelberg Engineering SPECTRALIS HRA+OCT is a device used to image the anterior and posterior segments of the human eye and to aid in the assessment and management of various diseases of the posterior segment, such as age-related macular degeneration, diabetic retinopathy, and glaucoma. The device is capable of acquiring infrared and blue reflectance images (K101223), as well as a composite color image referred to as "MultiColor image" due to addition of a green laser light source (K121993). New reference data for peripapillary retinal nerve fiber layer thickness and optic nerve head parameters have been added in K152205, as well as two (2) new accessory objective lenses (Wide Field, Ultra-Widefield), and OCT Enhanced Depth Imaging mode. In K172649, an updated 55° objective lens (WFO2) with enhanced anti-reflective coating was added to allow for Widefield MultiColor imaging, as well as an updated and faster spectrometer, called OCT2 module, a replacement of the patient and data management system, named Heyex 2, restructured user documentation, and Windows 10 support.
The modification to the SPECTRALIS is:
Widefield OCT - In addition to OCT imaging with the standard objective (for posterior segment imaging; 30° field of view) and the Anterior Segment Objective
Heidelberg Engineering GmbH
{6}------------------------------------------------
(anterior segment imaging; 30° field of view), OCT imaging is enabled with the Wide Field Objective (WFO) and Wide Field Objective 2 (WFO2). Both objectives provide a 55° diameter circular field of view. The field of view can be reduced by software to 35° and 25°. The only difference between WFO and WFO2 is an enhanced antireflective coating to allow for MultiColor imaging on the WFO2 (cleared in K172649). Both coatings cover the OCT wavelength, so OCT imaging can be enabled for both objective lenses.
## INDICATIONS FOR USE – SPECTRALIS PREDICATE DEVICE
The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and threedimensional imaging, cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT), fundus photography, fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA), autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak) and to perform measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various ocular diseases, including age-related macular degeneration, macular edema, diabetic retinopathy, retinal and choroidal vascular diseases, glaucoma, and for viewing geographic atrophy as well as changes in the eye that result from neurodegenerative diseases. The SPECTRALIS HRA+OCT and SPECTRALIS OCT include reference databases for retinal nerve fiber layer thickness and optic nerve head neuroretinal rim parameter measurements, which are used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values found in normal subjects.
{7}------------------------------------------------
# INDICATIONS FOR USE – MODIFIED SPECTRALIS
The Indications for Use for the modified SPECTRALIS is only slightly modified to the Indications for Use of the cleared SPECTRALIS predicate device for clarification. The modified Indication for Use statement is:
The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for:
· viewing the posterior segment of the eye, including two- and three-dimensional imaging
- · cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT)
- · fundus imaging
· fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA)
· autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT with BluePeak)
· performing measurements of ocular anatomy and ocular lesions.
The device is indicated as an aid in the detection and management of various ocular diseases, including:
- age-related macular degeneration
- macular edema
- · diabetic retinopathy
- · retinal and choroidal vascular diseases
- glaucoma
The device is indicated for viewing geographic atrophy.
The SPECTRALIS HRA+OCT and SPECTRALIS OCT include the following reference databases:
· a retinal nerve fiber layer thickness reference database, which is used to quantitatively compare the retinal nerve fiber layer in the human retina to values of Caucasian normal subjects - the classification result being valid only for Caucasian subjects
· a reference database for retinal nerve fiber thickness and optic nerve head neuroretinal rim parameter measurements, which is used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values of normal subjects of different races and ethnicities representing the population mix of the USA (Glaucoma Module Premium Edition)
{8}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE
The modified SPECTRALIS HRA+OCT and variants is a device modification to the cleared SPECTRALIS HRA+OCT and variants (K172649) predicate device. The modified SPECTRALIS has a slightly modified Intended Use/Indications for Use and the same fundamental scientific technology as the predicate device.
The Substantial Equivalence Summary tables below illustrate the comparisons of the modified SPECTRALIS to the predicate device.
The modified SPECTRALIS HRA+OCT and variants measures the same ophthalmic features and parameters as the cleared SPECTRALIS HRA+OCT and variants in K172649.
The changes applied to the SPECTRALIS since the clearance in K172649 do not change the subject population for whom the device is used, the type of images to be acquired, the disease or condition to be diagnosed, or the usefulness of the SPECTRALIS as a quantitative aid to an ophthalmologic examination of the eye.
Technological detail characteristics of the device are unchanged except for the modifications as stated in the General Device Description.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". A red square is on the left side of the first line and the right side of the second line.
# INTENDED USE/INDICATIONS FOR USE STATEMENT CHART
| K172649 PREDICATE DEVICE | SUBJECT DEVICE | Same or<br>Different | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| The SPECTRALIS is a non-contact<br>ophthalmic diagnostic imaging device. It is<br>intended for viewing the posterior segment<br>of the eye, including two- and three-<br>dimensional imaging, cross-sectional<br>imaging (SPECTRALIS HRA+OCT and<br>SPECTRALIS OCT), fundus photography,<br>fluorescence imaging (fluorescein<br>angiography, indocyanine green<br>angiography; SPECTRALIS HRA+OCT,<br>SPECTRALIS HRA), autofluorescence<br>imaging (SPECTRALIS HRA+OCT,<br>SPECTRALIS HRA and SPECTRALIS<br>OCT with BluePeak) and to perform<br>measurements of ocular anatomy and ocular<br>lesions. The device is indicated as an aid in<br>the detection and management of various<br>ocular diseases, including age-related<br>macular degeneration, macular edema,<br>diabetic retinopathy, retinal and choroidal<br>vascular diseases, glaucoma, and for<br>viewing geographic atrophy as well as<br>changes in the eye that result from<br>neurodegenerative diseases. The<br>SPECTRALIS HRA+OCT and<br>SPECTRALIS OCT include reference<br>databases for retinal nerve fiber layer<br>thickness and optic nerve head neuroretinal<br>rim parameter measurements, which are<br>used to quantitatively compare the retinal<br>nerve fiber layer and neuroretinal rim in the<br>human retina to values found in normal | The SPECTRALIS is a non-contact ophthalmic<br>diagnostic imaging device. It is intended for:<br>• viewing the posterior segment of the eye,<br>including two- and three-dimensional The<br>SPECTRALIS is a non-contact ophthalmic<br>diagnostic imaging device. It is intended for:<br>• viewing the posterior segment of the eye,<br>including two- and three-dimensional imaging<br>• cross-sectional imaging (SPECTRALIS<br>HRA+OCT and SPECTRALIS OCT)<br>• fundus imaging<br>• fluorescence imaging (fluorescein<br>angiography, indocyanine green angiography;<br>SPECTRALIS HRA+OCT, SPECTRALIS HRA)<br>• autofluorescence imaging (SPECTRALIS<br>HRA+OCT, SPECTRALIS HRA and<br>SPECTRALIS OCT with BluePeak)<br>• performing measurements of ocular anatomy<br>and ocular lesions.<br>The device is indicated as an aid in the detection<br>and management of various ocular diseases,<br>including:<br>• age-related macular degeneration<br>• macular edema<br>• diabetic retinopathy<br>• retinal and choroidal vascular diseases<br>• glaucoma<br>The device is indicated for viewing geographic<br>atrophy.<br>The SPECTRALIS HRA+OCT and<br>SPECTRALIS OCT include the following | The Indications<br>for Use statement<br>has been updated<br>for clarification:<br>1.) Formatting for<br>improved<br>readability<br>2.) "fundus<br>photography"<br>changed to<br>"fundus imaging"<br>for clarity<br>3.) Removed<br>statement related<br>to neuro-<br>degenerative<br>diseases upon<br>FDA request, to<br>not imply the use<br>of the device to<br>diagnose<br>neurological-based<br>diseases/<br>conditions.<br>4.) Clarification<br>on RDBs included<br>in the device | |
| subjects. | reference databases:<br>• a retinal nerve fiber layer thickness reference<br>database, which is used to quantitatively compare | | |
| | the retinal nerve fiber layer in the human retina to<br>values of Caucasian normal subjects - the<br>classification result being valid only for<br>Caucasian subjects<br>• a reference database for retinal nerve fiber<br>thickness and optic nerve head neuroretinal rim<br>parameter measurements, which is used to<br>quantitatively compare the retinal nerve fiber | | |
| | layer and neuroretinal rim in the human retina to<br>values of normal subjects of different races and<br>ethnicities representing the population mix of the<br>USA (Glaucoma Module Premium Edition) | | |
| | PREDICATE DEVICE<br>K172649 | SUBJECT DEVICE | Same or Different |
| | SPECTRALIS HRA+OCT | | |
| User | Ophthalmologist, office<br>setting | Ophthalmologist, office<br>setting | Same |
| Eye contact<br>required? | No | No | Same |
| Dilation of<br>pupil required? | No | No | Same |
| Working<br>position | The patient is sitting straight<br>in front of the device. The<br>examiner is sitting opposite<br>the patient. | The patient is sitting straight<br>in front of the device. The<br>examiner is sitting opposite<br>the patient. | Same |
| Images of<br>posterior<br>segment of eye | Yes, using standard,<br>WFO/WFO2, or UWF<br>objective lens | Yes, using standard,<br>WFO/WFO2, or UWF<br>objective lens | Same |
| Images of<br>anterior<br>segment of eye | With Anterior Segment<br>Module (K113129) | With Anterior Segment<br>Module (K113129) | Same |
| OCT imaging<br>of the anterior<br>and posterior<br>segment | Anterior segment: With<br>Anterior Segment Module<br>(K113129)<br>Posterior Segment:<br>30° OCT field of view with<br>standard objective | Anterior segment: With<br>Anterior Segment Module<br>(K113129)<br>Posterior Segment:<br>30° OCT field of view with<br>standard objective, 55° OCT<br>field of view with<br>WFO/WF02 | Different; optional<br>OCT imaging with<br>WFO/WFO2 was<br>introduced |
| RNFL<br>Reference<br>Database | RNFL thickness in reference<br>database for 12° circle scans<br>(K101223)<br>RNFL thickness in reference<br>database for fixed diameter<br>circle scans (K152205) | RNFL thickness in reference<br>database for 12° circle scans<br>(K101223)<br>RNFL thickness in reference<br>database for fixed diameter<br>circle scans (K152205) | Same |
| ONH Reference<br>Database | ONH parameters (BMO-<br>MRW, BMO-MRA)<br>reference database<br>(K152205) | ONH parameters (BMO-<br>MRW, BMO-MRA)<br>reference database<br>(K152205) | Same |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image contains the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a dark red square on either side of the text.
# INTENDED USE CHARACTERISTICS CHART
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Heidelberg Engineering. The logo consists of two lines of text in a bold, sans-serif font. The first line reads "HEIDELBERG" and the second line reads "ENGINEERING". There is a red square to the left of the first line and to the right of the second line.
## NON-CLINICAL PERFORMANCE SUMMARY
The modified SPECTRALIS was evaluated according to the requirements of FDA recognized consensus standards ISO 14971, AAMI / ANSI ES60601-1:2005, IEC 60601-1-2 Edition 3: 2007-03, ISO 10993-1 Fourth Edition 2009-10-15, AAMI/ANSI/ISO 10993-5:2009/(R) 2014, ISO 10993-12 Fourth Edition 2012-07-01, ISO 10993-18:2005, IEC 60825-1 Edition 2.0 2007-03, and AAMI / ANSI / IEC 62304:2006 and was found to meet the requirements of the applicable parts.
## DESIGN CONTROL
Heidelberg Engineering designed and developed the modified SPECTRALIS per the company's Design Control procedure, which complies with the FDA Quality System Regulations CFR Part 820 and ISO 13485:2012. The Design Control procedure also incorporates Risk Management procedures, which comply with ISO 14971:2007.
Risk assessment was conducted on the modified SPECTRALIS, and the impact of the design modifications were assessed on the predicate 510(k) cleared device.
The modified SPECTRALIS is manufactured and tested in the exact manner as the predicate 510(k) cleared device.
Heidelberg Engineering performed bench testing - including light safety, image geometry, and image quality assessment - and software verification and validation, to confirm that the modified SPECTRALIS HRA+OCT functions equivalently to the predicate SPECTRALIS HRA+OCT.
## OPTICAL RESOLUTION TESTING
Bench testing was performed in order to assess all the relevant system properties (resolution, scaling, image quality etc.) for the WFO lens. In particular, a 1951 USAF resolution test chart (MIL-STD-150A) was used to objectively quantify the loss of optical lateral resolution comparing wide-field OCT scans with standard objective (SO) OCT patterns.
The bench testing resulted in an average value of 13um for the SO, and of 18um for the WFO. Compared to the standard lens, the WFO causes a loss in optical lateral resolution of about 3 resolution elements on a USAF 1951 resolution target. These values are in agreement with the theoretical evaluation.
{12}------------------------------------------------
## CONCLUSION
The modified SPECTRALIS is a device modification to the cleared SPECTRALIS predicate device. The modified SPECTRALIS has a slightly modified Intended Use/Indications for Use and the same fundamental scientific technology as the cleared SPECTRALIS predicate device.
In summary, Heidelberg Engineering GmbH is of the opinion that the modified SPECTRALIS HRA+OCT and variants, does not introduce any new potential safety risks, is as effective as the cleared SPECTRALIS HRA+OCT and variants and concludes that the modified SPECTRALIS HRA+OCT and variants, is substantially equivalent to the predicate device.
This 510(k) summary for the Heidelberg Engineering SPECTRALIS HRA+OCT and variants is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR§807.92.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.