K173628 · Acclarent, Inc. · PGW · Mar 10, 2018 · Neurology
Device Facts
Record ID
K173628
Device Name
Acclarent ENT Navigation System
Applicant
Acclarent, Inc.
Product Code
PGW · Neurology
Decision Date
Mar 10, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The ACCLARENT® ENT Navigation System is intended for use during intranasal and paranasal image-guided navigation procedures for patients who are eligible for sinus procedures.
Device Story
System provides real-time tracking of navigated surgical instruments during intranasal and paranasal procedures. Inputs: pre-acquired CT images and electromagnetic signals from sensors in instrument tips. Operation: Field Ring generates magnetic field; induces current in instrument sensor to calculate position; system registers CT images to patient anatomy; displays tool tip icon on registered image. Used in surgical settings by ENT physicians. Output: visual display of instrument location relative to patient anatomy. Benefit: assists surgeon in navigating sphenoid, frontal, and maxillary sinuses; improves procedural accuracy.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety, EMC, mechanical testing, location accuracy verification against a robotic system, software functional testing, and pre-clinical cadaver studies simulating sinuplasty workflows.
Technological Characteristics
Electromagnetic tracking system; includes Navigation Console, Field Ring, Instrument Hub, Patient Tracker, and Registration Probe. Uses pre-acquired CT images for navigation. Connectivity: Workstation-based. Software-driven tracking of magnetic sensors in flexible tools.
Indications for Use
Indicated for patients eligible for sinus procedures requiring intranasal and paranasal image-guided navigation.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
March 10, 2018
Acclarent, Inc. James Patrick Garvey II Associate Director, Regulatory Affairs 1525-B O'Brien Drive Menlo Park, CA 94025
Re: K173628
Trade/Device Name: Acclarent® ENT Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: February 8, 2018 Received: February 9, 2018
Dear James Patrick Garvey II:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
## Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a bold, sans-serif font, with the first four letters in black and the last four in purple. Above the word "ent" is a purple arc. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" in a smaller font.
## ACCLARENT® ENT Navigation System Special 510(k) Premarket Notification
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K173628 |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | ACCLARENT® ENT Navigation System |
| Indications for Use (Describe) | The ACCLARENT® ENT Navigation System is intended for use during intranasal and paranasal image-guided navigation procedures for patients who are eligible for sinus procedures. |
| Type of Use (Select one or both, as applicable) | <div> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a bold, sans-serif font, with the first half in black and the second half in purple. Above the word "Acclarent" is a purple arc. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES", with "Johnson & Johnson" in red.
## ACCLARENT® ENT Navigation System
Special 510(k) Premarket Notification
## 510(k) Summary
| [807.92(a)(1)] Submitter Information | |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Biosense Webster, Inc.<br>33 Technology Drive<br>Irvine, CA 92618, USA<br>Tel.: (800) 729-9010<br>Fax: (949) 450-6886 |
| Contact Person: | James Patrick Garvey II<br>Associate Director, Regulatory Affairs<br>Phone: 949-789-8505<br>Fax: 650-687-5889 |
| Authored by: | Anna Gantman<br>Quality and Regulatory Program Manager<br>Biosense Webster (Israel), Ltd.<br>+972-52-808-9735 |
| Date Summary Prepared: | January 19, 2018 |
| [807.92(a)(2)] Name of Device | |
| Device Trade Name: | ACCLARENT® ENT Navigation System |
| Device Common Name: | Image Guided Surgery System |
| Device Classification: | Class II, 21 CFR 882.4560 |
| Classification Name: | Ear, Nose, and Throat Manual Surgical Instrument (21 CFR 882.4560) |
| Product Code | PGW |
| [807.92(a)(3)] Legally Marketed Devices | |
| Predicate Device: | CARTO® ENT Navigation System<br>510(k)#: K161701 |
| [807.92(a)(4)] Device Description | |
| Device Description: | The ACCLARENT® ENT Navigation System is intended to be used<br>during intranasal and paranasal surgical procedures to help ENT<br>physicians to track and display the real-time location of the tip of<br>navigated instruments relative to pre-acquired reference images, such<br>as CT.<br><br>The ACCLARENT® ENT Navigation System enables ENT physicians<br>to access sphenoid, frontal, and maxillary sinuses by using the system<br>magnetic tracking technology, identical to the predicate device.<br><br>The system incorporates a Navigation Console, Field Ring, Instrument<br>Hub, Patient Tracker, Registration Probe, Field Ring and Holder,<br>Workstation and accessories. A magnetic field generated by the Field<br>Ring induces a current in the magnetic sensor embedded in the tip of<br>the flexible navigated tool, which helps to accurately calculate the tool<br>tip position. A CT image is imported and registered to the patient<br>coordinates and a tool tip icon is displayed on top of the registered |
| [807.92(a)(5)] Intended Use<br>Indications for Use: | The ACCLARENT® ENT Navigation System is intended for use<br>during intranasal and paranasal image-guided navigation procedures<br>for patients who are eligible for sinus procedures. |
| Difference in Indications<br>from Predicate Device: | The intended use of the device is unchanged from the predicate. |
| [807.92(a)(6)] Technical Characteristics<br>Technological<br>Characteristics: | The substantial equivalence of the ACCLARENT® ENT Navigation<br>System to the predicates is shown by similarity in intended use,<br>indications for use, and performance.<br><br>Like the predicate CARTO® ENT System, the ACCLARENT® ENT<br>Navigation System is an image-guided navigation system intended for<br>use during paranasal surgical procedures. ENT physicians can track<br>and display the real-time location of the tip of navigated instruments<br>relative to pre-acquired CT images.<br><br>Like the predicate, the ACCLARENT® ENT Navigation System<br>utilizes electromagnetic tracking technology for navigation, uses<br>anatomical reference points on the patient's anatomy for intraoperative<br>registration to the image-based model of the anatomy, and uses CT<br>image sets as reference images for the image-based model. The<br>instructions for use and product labels have been updated to reflect the<br>revised device. |
| Non-clinical Performance<br>Data: | The ACCLARENT® ENT Navigation System was tested to ensure that<br>it functions in accordance with the system design specifications related<br>to substantial equivalence in terms of device safety and effectiveness.<br><br>The following nonclinical tests were performed:<br><br>1. Proof of Design electrical tests, to verify all hardware modules<br>perform within specifications.<br><br>2. Location Accuracy tests, where the ACCLARENT® ENT<br>Navigation System electromagnetic locations were compared to the<br>locations provided by a, very accurate robot system over the entire<br>navigation volume, to verify the system precision claim.<br><br>3. Software functional tests, covering the complete system<br>functionality, and including error handling, usability and time |
| 4. | Safety, EMC, and mechanical tests were performed by a nationally recognized testing laboratory to verify compliance with safety and EMC standards for medical devices. |
| 5. | Simulated use accuracy test, in which a complete CT image registration and instrument navigation workflow was performed, to verify the overall accuracy of the system. |
| 6. | Pre-clinical (cadaver) tests were designed to mimic surgical procedures using the ACCLARENT® ENT Navigation System in a simulated clinical environment, to assess the execution of a complete sinuplasty procedure workflow and to qualitatively estimate the system clinical accuracy. |
| | The proposed ACCLARENT® ENT Navigation System passed all tests in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness. |
| Clinical Performance Data: | Clinical data was not necessary to determine that the ACCLARENT® ENT Navigation System was substantially equivalent to the predicate device. The performance data demonstrated that the device performs as intended. |
| Conclusion: | The modified ACCLARENT® ENT Navigation System is substantially equivalent to the currently cleared CARTO® ENT Navigation System based on the completion of non-clinical bench testing as well as similar principles of design, operation and indications for use. |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a bold, sans-serif font, with the first half in black and the second half in purple. A purple arc is above the second half of the word. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" in a smaller font.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Acclarent, a company that is part of the Johnson & Johnson family of companies. The word "Acclarent" is written in a bold, sans-serif font, with the first part in black and the "ent" in purple. Above the purple part of the name is a purple arc. Below the name is the text "PART OF THE Johnson-Johnson FAMILY OF COMPANIES", with "Johnson-Johnson" in red script.
performance (latency).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.