ThermiX Temperature Control Radiofrequency System, Accessories - RF Electrodes
K173582 · Thermigen, L.L.C · GEI · Jan 17, 2018 · General, Plastic Surgery
Device Facts
Record ID
K173582
Device Name
ThermiX Temperature Control Radiofrequency System, Accessories - RF Electrodes
Applicant
Thermigen, L.L.C
Product Code
GEI · General, Plastic Surgery
Decision Date
Jan 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Thermi Temperature Controlled Radiofrequency (RF) System (generator, electrodes/hanpieces and accessories) are indicated: - for use in dermatological and general surgical procedures for electrocoagulation and hemostasis - to create lesions in nervous tissue
Device Story
Thermi Temperature Controlled RF System comprises RF generator, percutaneous/transcutaneous electrodes, foot pedal, and power cord. Device delivers RF energy to soft tissue via electrode antenna, inducing heating. Integrated thermocouple provides temperature feedback; system automatically adjusts energy output to maintain user-defined set point (35-90°C). Used by physicians in clinical settings for electrocoagulation, hemostasis, and nervous tissue lesioning. Real-time temperature/impedance monitoring allows precise control over thermal effects, aiming to achieve desired clinical outcomes while ensuring safety through automated energy regulation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional, functional, performance, shelf life, biocompatibility, tensile strength, and thermal effect on tissue (muscle, liver, kidney, porcine skin). Electrical safety and essential performance verified per IEC 60601-1, 60601-1-2, 60601-1-6, and 60601-2-2.
Technological Characteristics
RF generator with temperature/impedance feedback loop. Operating temperature range: 35-90°C. Includes percutaneous and transcutaneous electrodes. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60601-2-2. Biocompatibility per ISO 14971 risk assessment.
Indications for Use
Indicated for patients requiring electrocoagulation, hemostasis in dermatological/general surgical procedures, or creation of lesions in nervous tissue.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2017
ThermiGen. L.L.C. Ms. Suzanne Cheang Regulatory Affairs Director 3131 West Royal Lane, Suite 100 Irving, Texas 75063
Re: K173582
Trade/Device Name: Thermi Temperature Controlled Radiofrequency (RF) System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 17, 2017 Received: November 20, 2017
Dear Ms. Cheang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173582
#### Device Name
Thermi Temperature Controlled Radiofrequency (RF) System
Indications for Use (Describe)
The Thermi Temperature Controlled Radiofrequency (RF) System (generator, electrodes/hanpieces and accessories) are indicated:
- for use in dermatological and general surgical procedures for electrocoagulation and hemostasis
- to create lesions in nervous tissue
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 6: 510(K) SUMMARY
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date is November 15, 2017 [21 CFR 807.92(a)(1)].
#### General Information
| Trade Name | ThermiX Temperature Controlled Radiofrequency (RF) System |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Classification Name | 21 CFR §878.4400 Electrosurgical Cutting and Coagulation and<br>Accessories |
| Regulatory Class | Class II |
| Product Code | GEI |
| Submitter | ThermiGen, L.L.C.<br>3131 West Royal Lane Suite 100<br>Irving, TX 75063 |
| Contact | Ms. Suzanne Cheang<br>Regulatory Affairs, Director<br>Phone: (214) 888-0683<br>Fax: (214) 279-0101<br>Email: scheang@thermi.com |
#### Predicate Device
| K130689 | ThermiGen Symphony RF System |
|---------|-------------------------------------------|
| K171094 | Thermi Reusable Non-invasive RF Electrode |
| K170116 | Thermi Injectable RF Electrode |
#### Indications for Use
The Thermi Temperature Controlled Radiofrequency (RF) System (generator, electrodes/handpieces and accessories) are indicated:
- . for use in dermatological and general surgical procedures for electrocoagulation and hemostasis
- to create lesions in nervous tissue.
## Device Description
The Thermi Temperature Controlled Radiofrequency (RF) System consists of a RF generator with integral temperature and impedance feedback, automatically adjusting energy delivery to maintain set temperature and percutaneous and transcutaneous electrodes accessories that deliver the RF to the targeted location on the patient's anatomy when use with Thermi RF generator. It is operating with set temperature range between 35-90° C that allows the physician to control treatment temperature to achieve the desired clinical outcome.
{4}------------------------------------------------
The Thermi RF System consist of:
- 1. RF Generator
- 2. Percutaneous electrodes
- 3. Disposable and re-usable transcutaneous electrodes/handpieces
- 4. Foot pedal
- 5. Power cord
Thermi RF generator produces an oscillating electric field in the antenna (electrode). The oscillating electrical field is transmitted to the surrounding soft tissue, causing heating of the tissue. A thermocouple in the electrode measures this increase in temperature and maintains a feedback loop to ensure a set point temperature in the tissue.
### Technological Characteristics
The Thermi Temperature Controlled Radiofrequency System is similar with regards to indications for use, design, operation principle and technological characteristics to the predicate devices that were cleared in K130689, ThermiRF System (previously named the Symphony RF System), K170116, Thermi Injectable RF Electrode and K171094, Thermi Reusable Noninvasive RF Electrode. Results of bench testing demonstrate Thermi RF Temperature Controlled Radiofrequency System is as safe and effective as the predicate devices.
### Performance Data
The Thermi Temperature Controlled Radiofrequency System is similar with regards to indications for use and technological characteristics to the predicate devices that were cleared in K130689, ThermiRF System (previously named the Symphony RF System), K170116, Thermi Injectable RF Electrode and K171094, Thermi Reusable Non-invasive RF Electrode. Results of bench testing demonstrate Thermi RF Temperature Controlled Radiofrequency System works within the device parameters ranges of the legally marketed Thermi RF System with the same indications for use.
All testing performed on the Thermi Temperature Controlled Radiofrequency System derived from the risk assessment in accordance with ISO 14971 which evaluated the safety and effectiveness of the design modification in accordance with Thermi Design & Development procedures. The test methodology and acceptance criteria were developed from within Thermi and from related standards.
A series of bench testing was identified and conducted on the subject Thermi Temperature Controlled Radiofrequency System in accordance with protocols to verify design specifications as follows. Testing inclued:
- Dimensional testing ●
- Functional testing ●
- Performance testing ●
- Design features confirmation ●
- Shelf life testing ●
{5}------------------------------------------------
- Biocompatibility testing ●
- Tensile strength
- Materials confirmation
- Thermal effect on tissue testing including muscle, liver, kidney, and porcine skin ●
- Software functional and performance verification
- Software Validation
- Max Output Energy ●
Electrical safety and essential performance testing was completed on the subject Thermi Temperature Controlled Radiofrequency System in accordance with FDA Guidance "Premarket Notification 510(k) Submission for Electrosurgical Devices for General Surgery. The following electrical safety and essential performance testing was completed:
- 1. IEC 60601-1, Medical Electrical Equipment Part 1: General requirements for basic safety and essential performances.
- 2. IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances requirements and test
- 3. IEC 60601-1-6, Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
- 4. IEC 60601-2-2, Medical Electrical Equipment Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
The results of testing show that the subject Thermi Temperature Controlled Radiofrequency (RF) System, met all performance and functional testing and performed as intended; and did not raise new safety or performance questions.
#### Conclusions
The subject Thermi Temperature Controlled Radiofrequency System is similar with regards to indications for use, design, operation principle and technological characteristics to the predicate devices that were cleared in K130689, ThermiRF System (previously named the Symphony RF System), K170116, Thermi Injectable RF Electrode and K171094, Thermi Reusable Noninvasive RF Electrode. The results of testing demonstrate that the Thermi Temperature Controlled Radiofrequency System is safe and effective; and no new questions of safety or efficacy are raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.