K173576 · GE Healthcare · MUE · May 14, 2018 · Radiology
Device Facts
Record ID
K173576
Device Name
Pristina Serena
Applicant
GE Healthcare
Product Code
MUE · Radiology
Decision Date
May 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1715
Device Class
Class 2
Indications for Use
Pristina Serena is an optional accessory of Senographe Pristina intended to provide accurate location of lesions in the breast in three dimensions.
Device Story
Pristina Serena is a hardware/software add-on for the Senographe Pristina mammography platform. It uses stereotaxy to determine 3D coordinates (X, Y, Z) of breast lesions. Input: stereo pair of +/-15° angulated 2D low-dose X-ray images of compressed breast. Operation: user identifies lesion in both images; system computes 3D coordinates and calculates required position for biopsy device holder. Hardware includes Biopsy Positioner (BP) replacing Imaging Bucky, dedicated compression paddles, and mechanical adaptors. Used in clinical settings by healthcare professionals. Output: mechanical guidance for biopsy or hookwire placement. Benefits: enables accurate targeting for minimally invasive breast interventions.
Clinical Evidence
Bench testing only. Performance demonstrated through engineering bench testing, including biopsy accuracy (verification of geometric accuracy between target lesion on X-ray and actual needle tip position) and image quality/dose performance testing. Software verified via unit, integration, and system-level testing. Simulated use testing performed for validation.
Technological Characteristics
Hardware add-on for Senographe Pristina; includes Biopsy Positioner, compression paddles, and mechanical adaptors. Uses stereotactic X-ray imaging principle. Software-managed. Manufactured under 21 CFR 820 and ISO 13485 quality standards.
Indications for Use
Indicated for providing 3D location of breast lesions using stereotactic pairs of 2D X-ray images to guide minimally invasive or interventional procedures, including vacuum-assisted biopsy, core biopsy, pre-surgical localization (e.g., hookwire), and fine needle aspirations (FNA).
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 14, 2018
GE Healthcare Gregory Pessato Regulatory Affairs Program Manager 283 rue de la Miniere Buc, 78530 FRANCE
Re: K173576
Trade/Device Name: Pristina Serena Regulation Number: 21 CFR 892.1715 Regulation Name: Full-Field Digital Mammography System Regulatory Class: Class II Product Code: MUE Dated: April 11, 2018 Received: April 13, 2018
Dear Gregory Pessato:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRHs Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K173576
Device Name Pristina Serena
Indications for Use (Describe)
The Pristina Serena option provides the three-dimensional locations, using information obtained from stereotactic pairs of two-dimensional X-ray images. This information provides guidance for a variety of minimally invasive or interventional procedures in the breast such as: vacuum assisted biopsy, core biopsy, pre-surgical localization (e.g. hookwire), and fine needle aspirations (FNA).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized, cursive font in the center. The circle is surrounded by a thin white border, and there are four small white teardrop shapes evenly spaced around the perimeter of the circle.
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | March 28, 2018 |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE |
| Primary Contact Person: | Gregory Pessato,<br>Regulatory Affairs Program Manager,<br>GE Medical Systems SCS,<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE<br>Phone : + 33 1 30 70 93 16<br>Email : gregorypessato@ge.com |
| Secondary Contact Person: | Diane Uriell<br>Sr. Director of Regulatory Affairs, XR and Women's Health,<br>GE Healthcare,<br>Atlanta, Georgia - USA<br>Phone +1 262 290 8212<br>Email: Diane.Uriell@ge.com |
| Device Trade Name: | Pristina Serena |
| Common/Usual Name: | Stereotaxy biopsy guidance application for a Full Field Digital<br>Mammography System |
| Classification Names: | 21 CFR 892.1715, Full-field digital mammography system, Class II |
| Product Code: | MUE |
| Predicate Device(s): | Senographe Stereo (K040125) |
| Reference Predicate(s) | Senographe Pristina (K162268) |
| Device Description: | Pristina Serena is a hardware and software add-on to the Senographe<br>Pristina mammography platform (cleared in K162268). |
| | Pristina Serena uses the Stereotaxy principle to determine the three-<br>dimensional (3D) location (X, Y and Z coordinates) of a region of<br>interest in the breast (such as a suspicious lesion).<br><br>The stereotaxy biopsy process uses a stereo pair of +/-15° angulated<br>2D low-dose X-ray images of the compressed breast. The user<br>identifies the point of interest in each image of the stereo pair,<br>Pristina Serena computes the three dimensions coordinates (X,Y,Z) of<br>the user-specified location in the stereo pair.<br><br>Pristina Serena uses the target lesion 3D coordinates and information<br>on the geometry of the biopsy device to compute the position of a<br>biopsy device holder that will allow intervention in the breast at the<br>targeted position (biopsy of sample tissue or placement of a hookwire<br>for guidance of surgical interventions).<br><br>The Pristina Serena add-on includes the following items:<br>Hardware: a Biopsy Positioner (BP) the main hardware<br>component of the Pristina Serena option. It can be<br>mounted on the Senographe Pristina in lieu of the<br>Imaging Bucky. The purpose of the BP is to allows<br>breast positioning and mechanical guidance to the<br>target lesion for Biopsy medical application; dedicated breast compression paddles for Biopsy<br>application; mechanical adaptors: pieces of hardware to allow<br>mounting of several kind of biopsy devices on the<br>Biopsy positioner arm. Software: A new software version for the Senographe Pristina<br>platform which includes software to manage the Pristina<br>Serena option. Labeling for the Biopsy Medical application. |
| | Pristina Serena option is compatible with previously installed<br>Senographe Pristina systems. Pristina Serena does not require any<br>hardware modification on the Senographe Pristina platform. |
| Intended Use: | Pristina Serena is an optional accessory of Senographe Pristina<br>intended to provide accurate location of lesions in the breast in three<br>dimensions. |
| Indications for Use | The Pristina Serena option provides the three-dimensional location of<br>target lesions, using information obtained from stereotactic pairs of<br>two-dimensional X-ray images. This information provides guidance for<br>a variety of minimally invasive or interventional procedures in the |
| | breast such as: vacuum assisted biopsy, core biopsy, pre-surgical<br>localization (e.g. hookwire), and fine needle aspirations (FNA). |
| Device Comparison: | Pristina Serena uses the same fundamental technology as Senographe<br>Stereo in terms of lesion targeting technique (2D stereotactic biopsy),<br>stereotactic angle, needle approach (vertical or lateral), field of view<br>and attainable breast volume. Like its predicate, Pristina Serena<br>supports various breast biopsy and intervention devices. The<br>targeting accuracy of Pristina Serena is equivalent to that of<br>Senographe Stereo.<br>The main differences between Pristina Serena and Senographe Stereo<br>include:<br>- Ability to target a lesion above the compression paddle in<br>vertical approach;<br>- Ability to switch needle approach without patient<br>repositioning;<br>- A tilted needle axis to accommodate future 3D biopsy<br>application.<br>Pristina Serena uses the image chain of the Senographe Pristina<br>mammography platform, therefore delivering images of quality<br>equivalent to that of standard screening/diagnostic images. |
| Performance Testing | The safety and performance of the Pristina Serena option was<br>demonstrated through engineering bench performance testing such<br>as:<br>Biopsy accuracy testing: verification of the geometrical<br>accuracy between the target lesion identified on the X-ray<br>Stereo pair images and the actual position of the biopsy<br>needle tip (or needle notch). Image quality and dose performance testing to confirm that<br>performance in Biopsy mode with Pristina Serena is<br>substantially equivalent to that of Senographe Pristina<br>screening/diagnostic imaging system. In addition, the device has successfully completed required design<br>control testing per GE Healthcare's quality management system. No<br>unexpected test results were obtained. The Pristina Serena option<br>was designed and will be manufactured under the Quality System<br>Regulations of 21CFR 820 and ISO 13485. The following quality<br>assurance measures were applied to the development of the system:<br>Risk Analysis Design Reviews |
| | Software Development Lifecycle Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) These tests were performed to provide the requisite data to substantiate performance, claims, and ultimately substantial equivalence. Testing demonstrated that Pristina Serena performs according to specifications and functions as intended. |
| Substantial Equivalence<br>Determination: | Pristina Serena has the same indications for use as its predicate Device. It uses the same fundamental technology as Senographe Stereo. Differences between Pristina Serena and Senographe Stereo do not raise new questions of safety and effectiveness. The impact of differences between the predicate and new device has been assessed using established methods. Testing including conformance to standards; successful verification/validation; and additional bench performance testing, establish that Pristina Serena is substantially equivalent to its predicate Senographe Stereo (K040125) for biopsy functions and reference predicate Senographe Pristina (K162268) for imaging performance. |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. There are also three droplet-like shapes surrounding the circle, giving the impression of motion or energy.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white swirls around the outside of the circle. The logo is simple and recognizable.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized, cursive font in the center. There are white swirls around the letters, and a white border around the blue circle. The logo is simple and recognizable, and it is associated with a well-known company.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.