The Sera™ with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Sera™ with DPOAE includes all ages. The Sera™ with TEOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Sera™ with TEOAE includes all ages. The Sera™ with ABRIS is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The target population for Sera™ with ABRIS is newborns. The Sera™ System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.
Device Story
Sera™ is a handheld, touchscreen-based audiometric screening device used by trained professionals (audiologists, ENT surgeons, doctors) in clinical settings. It performs three types of hearing screenings: DPOAE (Distortion Product Otoacoustic Emissions) using pure tone pairs; TEOAE (Transient Evoked Otoacoustic Emissions) using click stimuli; and ABRIS (Auditory Brainstem Response) using CE-Chirp™ stimuli. For ABRIS, transcutaneous electrodes measure EEG activity in response to acoustic stimuli. The device automatically processes and analyzes EEG signals; it assigns a 'Pass' result if a response is detected or a 'Refer' result if no response is detected after 3 minutes. Results are displayed on the screen and can be printed via Bluetooth. The device assists in identifying inner ear and auditory brainstem abnormalities, enabling early detection of hearing loss.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including design verification, validation, EMC/safety testing (IEC 60601-1 series), and performance testing per IEC 60645 series for OAE/ABR measurements.
Technological Characteristics
Handheld unit with touchscreen and rechargeable battery. Connectivity: Bluetooth for printing. Sensing: OAE probe, RadioEar IP30 insert earphone, transcutaneous electrodes. Standards: IEC 60601-1 (safety), IEC 60645-6/7 (performance). Software: Moderate level of concern. ABRIS uses CE-Chirp™ stimulus (35 dB nHL).
Indications for Use
Indicated for audiologic evaluation and documentation of ear disorders (DPOAE/TEOAE) in all ages, and ear/nerve disorders (ABRIS) in newborns. Used by trained professionals.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 14, 2018
Interacoustics A/S Amy Yanta Director of Regulatory Affairs Audiometer Alle 1 Middelfart, 5500 Dk
Re: K173567
Trade/Device Name: Sera™ Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO. GWJ Dated: November 20, 2017 Received: November 20, 2017
# Dear Amy Yanta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173567
Device Name Sera™M
#### Indications for Use (Describe)
The Sera™ with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Sera with DPOAE includes all ages.
The Sera™ with TEOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Sera with TEOAE includes all ages.
The Sera™ with ABRIS is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The target population for Sera with ABRIS is newborns.
The Sera™ System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe symbol above the word "Interacoustics" in a bold, sans-serif font. The globe symbol is made up of three curved lines that form a sphere with a horizontal line through the center. The logo is black and white.
FDA 510(k) K173567 Sera Summary as required by 21 CFR 807.92.
#### Administrative Information
Submitter: Interacoustics A/S
Audiometer Allé 1 5500 Middelfart Denmark Tel: +45 6371 3555 Fax: +45 6371 3522
Contact Person: Amy Yanta Director of Regulatory Affairs 10393 West 70th Street Eden Prairie, MN 55344 952-947-6097 amyy@diagnostic-group.us
Date Summary Prepared: August 25, 2017
#### Device Identification
Trade Name: Sera™
Common Name: otoacoustic emission device/evoked response auditory stimulator and measurements Device Classification Name: Audiometer/Stimulator, Auditory, Evoked Response Device classification: Class II Panel: Ear Nose & Throat / Neurology Classification Regulation: 874.1050/882.1900 Product Code: EWO/GWJ
Predicate Device: easyScreen, cleared on 08/29/2017 via K171506
Device Description
The device is audiometric equipment used for assisting of inner ear and auditory brainstem abnormalities.
Sera™ features a touch-screen display and user-friendly software in a compact hardware design. Sera™ can be purchased with various licenses allowing you to perform different hearing screening tests.
Sera™ uses auditory brainstem response (ABRIS) technology to screen patients for hearing loss. A modified click stimulus, the CE-Chirp", of 35 dB nHL is delivered into the patient's ear while electrodes placed on the patient's head measure EEG activity.
Auditory brainstem response (ABRIS) test produces a short acoustic stimulus and measures via transcutaneous electrodes the auditory evoked potentials from the inner ear, the auditory nerve and the brainstem.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe symbol above the word "Interacoustics" in a bold, sans-serif font. Below the company name is a tagline in a smaller font, but the text is not legible. The logo is in black and white.
The EEG is processed and analyzed automatically using the Sera™'s respons a response is detected, the screening is stopped automatically and a Pass result is assig ear. When no response is detected after 3 minutes of EEG activity has been processed, a Refer result is assigned.
Distortion product otoacoustic emissions (DPOAE) technology uses pairs of pure tones presented in sequence to screen patients for cochlear hearing loss. Responses to the stimulus are predictable and therefore can be measured via a sensitive microphone placed in the patient's ear canal.
Transient otoacoustic emissions (TEOAE) technology uses a click stimulus to screen patients for cochlear hearing loss. Responses to the stimulus are predictable and therefore can be measured via a sensitive microphone placed in the patient's ear canal. The response can be divided into frequency bands for assessment.
### Device Intended Use
The Sera™ with DPOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion Product Otoacoustic Emissions. The target population for Sera™ with DPOAE includes all ages.
The Sera™ with TEOAE is intended for use in the audiologic evaluation and documentation of ear disorders using Transient Evoked Otoacoustic Emissions. The target population for Sera™ with TEOAE includes all ages.
The Sera™ with ABRIS is intended for use in the audiologic evaluation and documentation of ear and nerve disorders using auditory evoked potentials from the inner ear, the auditory nerve and the brainstem. The target population for Sera™ with ABRIS is newborns.
The Sera™ System is to be used by trained personnel only, such as audiologists, ENT surgeons, doctors, hearing healthcare professionals or personnel with a similar level of education. The device should not be used without the necessary knowledge and training to understand its use and how results should be interpreted.
# Technological Characteristics
The Sera™ consists of a handheld unit that utilizes a touchscreen display and a rechargeable battery. A simple cradle is included to support charging of the device's battery. The device supports Bluetooth communication with a label printer for the purpose of printing screening results.
A comparison between the new and predicate devices shows that the technological characteristics and indications for use are equivalent. The device employs similar technology to accomplish the same tasks as the predicates. A detailed table is provided below.
| Description | Predicate: easyScreen<br>(K171506) | Device: Sera™ |
|-------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| Type | Audiometer/ Auditory<br>Brainstem Response –<br>Audiometric equipment | Audiometer/ Auditory Brainstem<br>Response – Audiometric<br>equipment |
Equivalence Predicate Chart 1:
{5}------------------------------------------------
| | | Image: Interacoustics logo |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR 874.1050<br>(otoacoustic emission device)<br>21 CFR 882.1900<br>(Evoked response auditory<br>stimulator) | 21 CFR 874.1050<br>(otoacoustic emission device) 21<br>CFR 882.1900<br>(Evoked response auditory<br>stimulator) |
| Classification Product Code | EWO, GWJ | EWO, GWJ |
| Regulatory Class | Class II | Class II |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Interacoustics. The logo consists of a stylized globe above the word "Interacoustics". The word "Interacoustics" is in a bold, sans-serif font. There is a line underneath the word "Interacoustics".
| Indications for Use | | Interacoustics |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The device DPOAE and TEOAE<br>modules are intended for use in<br>the audiologic evaluation and<br>documentation of ear disorders<br>using Distortion Product<br>Otoacoustic Emissions or<br>Transient Otoacoustic Emissions<br>technology. The target<br>population for the modules<br>includes all ages. | The Sera™ with DPOAE is<br>intended for use in the<br>audiologic evaluation and<br>documentation of ear disorders<br>using Distortion Product<br>Otoacoustic Emissions. The<br>target population for Sera™<br>with DPOAE includes all ages. |
| | The device ABR module is<br>intended for use in the<br>audiologic evaluation and<br>documentation of ear and nerve<br>disorders using auditory evoked<br>potentials from the inner ear,<br>the auditory nerve and the<br>brainstem. The target<br>population for the ABR module<br>is newborns and infants up to 6<br>months of age. | The Sera™ with TEOAE is<br>intended for use in the<br>audiologic evaluation and<br>documentation of ear disorders<br>using Transient Evoked<br>Otoacoustic Emissions. The<br>target population for Sera™<br>with TEOAE includes all ages.<br><br>The Sera™ with ABRIS is<br>intended for use in the<br>audiologic evaluation and<br>documentation of ear and nerve<br>disorders using auditory evoked<br>potentials from the inner ear,<br>the auditory nerve and the<br>brainstem. The target<br>population for Sera™ with ABRIS<br>is newborns.<br><br>The Sera™ System is to be used<br>by trained personnel only, such<br>as audiologists, ENT surgeons,<br>doctors, hearing healthcare<br>professionals or personnel with<br>a similar level of education. The<br>device should not be used<br>without the necessary<br>knowledge and training to<br>understand its use and how<br>results should be interpreted.<br><br>The intended use differs<br>slightly because the Sera<br>separates each function's<br>intended use instead of<br>grouping. |
{7}------------------------------------------------
| Intended User | populations including new-<br>born infants.<br>The patient group<br>includes all ages and sexes.<br>The easyScreen System is to be<br>used by trained personnel only<br>such as audiologists, ENT<br>surgeons, doctors, hearing<br>healthcare professionals or<br>personnel with a similar level<br>of education. (or trained user<br>with supervision of a<br>professional) | populations including new- born<br>infants.<br>The patient group<br>includes all ages and sexes.<br>The Sera™ System is to be used by<br>trained personnel only, such as<br>audiologists, ENT surgeons,<br>doctors, hearing healthcare<br>professionals or personnel with a<br>similar level of education. The<br>device should not be used without<br>the necessary knowledge and<br>training to understand its use and<br>how results should be interpreted<br>Intended user specifies further<br>intended user. |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Anatomical Sites | Examination of Ear and hearing<br>nerves | Examination of Ear and hearing<br>nerves |
| Safety Standards | IEC 60601-1 | IEC 60601-1 |
| Performance Standards | IEC 60645-7, IEC 60645-6 | IEC 60645-7, IEC 60645-6 |
| Device Type | Screening device (PASS/REFER) | Screening device (PASS/REFER) |
| System Configuration | 1 -channel ABR system<br>operated through a handheld<br>base unit. The base unit is<br>operated as a stand alone<br>device | 1 -channel ABR system operated<br>through a handheld base unit.<br>The base unit is operated as a<br>stand alone device |
| Display Information | Noise status, detected<br>transducer, artifact %,<br>PASS/REFERstatus | Noise status, detected<br>transducer, artifact %,<br>PASS/REFER status |
| Stimulus | Clicks and chirps, pure tones | Clicks and chirps, pure tones |
| Electrode Quality Check | Yes | Yes |
| Impedance Test | acceptable impedance<br><40kOhm. | acceptable impedance<br><40kOhm. |
| Binaural screening | Yes | Yes |
| Pre-amplifier channels | 1 | 1 |
| Stimulus Rate | 90/s | 90/s |
| Pre-amplifier Gain | ABRIS - 64 dB (fixed) + 64dB<br>(Variable) | ABRIS - 72dB<br>The higher dB allows for a higher<br>ceiling of sound to be exported<br>from sound source |
| Stimulus Level | ABR - 30, 40 and 45db nHL/<br>DPOAE -30 dB SPL to 80 dB SPL | ABR - 30, 40 and 45db nHL/<br>DPOAE -30 dB SPL to 80 dB SPL |
| Masking | NONE | NONE |
| Transducers | OAE probe, RadioEar IP30 insert<br>earphone | OAE probe, RadioEar IP30 insert<br>earphone |
| Frequency range (f2) | DPOAE 500Hz - 10kHz | DPOAE 500Hz - 10kHz |
| TEOAE Stimulus | | |
| Frequency Range | 500 to 5500Hz | 500 to 5500Hz |
| | | Interacoustics |
| Stimuli Type | Non-Linear and Linear<br>Short duration signal (Click)<br>According to IEC 60645-3 | Non-Linear<br>Short duration signal (Click)<br>According to IEC 606045-3<br><br>Default is only Non-linear |
| Level | 30 to 90 dB peSPL | 30 to 90 dB peSPL |
| Level Step | 1 dB SPL | 1 dB SPL |
| Transducer | Dedicated OAE Probe | Dedicated OAE Probe |
| Probe Detection | Auto detection | Auto detection |
| Recording | | |
| A/D Resolution | 24 bit | 24 bit |
| Artifact Reject System | 0 -> +60 dB SPL or off | 0 -> +60 dB SPL or off |
| Automatic test with display<br>of PASS-REFER | Yes | Yes |
{8}------------------------------------------------
# Summary of Non-Clinical Testing
Design verification and validation were performed according to current standards for OAE and ABR to assure the device meets its performance specifications. EMC and Safety was performed in compliance with recognized standards IEC 60601-1 series, Medical Equipment – General requirements for basic safety and essential performance. The product meets the requirements from the international standard for OAE measurements IEC 60645 series. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in medical Devices." The software for this device was considered as a "moderate" level of concern since a malfunction of, or a latent design flaw in, the Software Device could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury. The OAE and ABRIS measurements were divided into 3 phases. Phase 1 included when optimization occurred and involved feedback to the operator so that they could adjust such as probe fit, electrode impedance, ambient electrical and acoustic noise etc. Once the pre-test conditions were optimized, phase 2 of data collection proceeded as rapid as possible to allow the maximum quantity of good quality data to be collected in the shortest possible time. Phase 3 proceeded into the data assessment and decision stage and this ran concurrently with Phase 2 once the predetermined minimum amount of data had been collected. Phase 3 then went into the algorithm descriptions for each TEOAE, DPOAE and ABRIS measurements modes and is provided in detail in Annex 16D of this submission. No clinical tests were performed, but based on the fulfillment of the international standards for OAE and ABR we believe the device is safe and effective. The auditory impedance testing characteristics and safety systems were compared and found to be comparable.
### Summary of Clinical Testing
Not applicable. Not required to establish substantial equivalence.
### Conclusion
We have compared the intended use and performance characteristics with the predicate device. The Sera™ was tested according to current standards and the differences found between the devices were related to functionality, not in relation to safety and efficiency. The Sera™ conforms to the current standards. After analyzing bench testing, safety, EMC, and software validation (with risk analysis) testing we conclude that the Sera™ is found to be substantially equivalent to the predicate devices in technological characteristics and indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.