← Product Code [NEZ](/productcode/NEZ) · K173459

# Olympus Small Intestinal Capsule Endoscope System (K173459)

_Olympus Medical Systems Corp. · NEZ · Mar 13, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173459

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [NEZ](/productcode/NEZ.md)
- **Decision Date:** Mar 13, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa. - It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia(IDA) not detected by upper and lower endoscopy. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas. The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM may be used as a tool in the detection of abnormalities of the small intestine and is intended for use in adults only.

## Device Story

System consists of ingestible capsule endoscope, wearable antenna unit/recorder, and workstation software. Capsule captures and transmits small intestine images to recorder worn by patient. Data downloaded to workstation for physician review. Features OMNI mode algorithm to assist review by identifying and storing redundant images, and Red Color Detection function to highlight suspected blood/red areas. Used in clinical settings by physicians to detect small intestine abnormalities. Benefits include improved reading efficiency and reduced review time for clinicians, facilitating diagnosis of obscure GI conditions.

## Clinical Evidence

Multicenter, randomized trial evaluated OMNI mode reading efficiency. Seven expert physicians reviewed cases in randomized order comparing OMNI mode to Normal mode. OMNI mode mean review time was 27.3 minutes vs. 75.1 minutes for Normal mode (p<0.001), representing a 64% reduction in review time. Study supports claims for improved reading efficiency and reduction of redundant images.

## Technological Characteristics

System includes ingestible capsule, antenna unit, recorder, and workstation software. Connectivity via radio signal communication. Complies with ISO 14971:2007 for risk management. Includes software-based OMNI mode for image processing and Red Color Detection function. Powered by battery pack. Lead-type antenna added to system configuration.

## Regulatory Identification

An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”

## Predicate Devices

- OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM ([K163069](/device/K163069.md))
- OLYMPUS CAPSULE ENDOSCOPE SYSTEM ([K090210](/device/K090210.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 13, 2018

Olympus Medical Systems Corp. % Daphney Germain-Kolawole Senior Project Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610

Re: K173459

Trade/Device Name: OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM Regulation Number: 21 CFR§ 876.1300 Regulation Name: Ingestible Telemetric Gastrointestinal Capsule Imaging System Regulatory Class: II Product Code: NEZ Dated: January 31, 2018 Received: February 1, 2018

Dear Daphney Germain-Kolawole:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Charles Viviano -S

Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K173459

#### Device Name

OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM

Indications for Use (Describe)

The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa.

- It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.

- It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.

- It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia(IDA) not detected by upper and lower endoscopy.

The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas.

The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM may be used as a tool in the detection of abnormalities of the small intestine and is intended for use in adults only.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Underneath the line is the text "Your Vision, Our Future" in a smaller, blue font.

K173459 Page 1 of 4

March 6, 2018

# 5. 510(k) SUMMARY

#### 5.1 General Information

- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Contact Position: ● Daphney Germain-Kolawole Senior Project Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-5691 Fax: 484-896-7128 Email: daphney.germain-kolawole@olympus.com

### 5.2 Device Identification

| ● Device Trade Name:    | OLYMPUS SMALL INTESTINAL CAPSULE<br>ENDOSCOPE SYSTEM             |
|-------------------------|------------------------------------------------------------------|
| ● Common Name:          | Capsule Imaging System                                           |
| ● Regulation Number:    | 876.1300                                                         |
| ● Regulation Name:      | Ingestible telemetric gastrointestinal capsule imaging<br>system |
| ● Regulatory Class:     | II (Special Controls)                                            |
| ● Classification Panel: | Gastroenterology and urology                                     |
| ● Product Code:         | NEZ                                                              |

### 5.3 Predicate Device Information

| Primary Predicate Device (PD1)                       |                                  |            |
|------------------------------------------------------|----------------------------------|------------|
| Model name                                           | Applicant                        | 510(k) No. |
| OLYMPUS SMALL INTESTINAL<br>CAPSULE ENDOSCOPE SYSTEM | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K163069    |
| Additional Predicate Device (PD2)                    |                                  |            |
| Model name                                           | Applicant                        | 510(k) No. |
| OLYMPUS<br>CAPSULE ENDOSCOPE SYSTEM                  | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K090210    |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Underneath the line is the text "Your Vision, Our Future" in a smaller font.

K173459 Page 2 of 4

### 5.4 Device Description

## The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is a

capsule imaging system used for visualization of the small intestine mucosa. This system consists of capsule endoscopes which capture images and transmit the data, an antenna unit and a recorder which are secured around the patient and receive data from the capsule, and workstation software which downloads the image data from the recorder and processes images for visualization. After the visualization, the diagnostic review (also known as reading) of images would start with user selectable OMNI mode feature. OMNI mode is an algorithm that assists the physician in reviewing the Capsule Endoscopy(CE) case data. The components of this SYSTEM are listed below.(Table 5-1)

| Trade or Proprietary or Model Name<br>for The Subject Device | Model<br>Number               |                  |
|--------------------------------------------------------------|-------------------------------|------------------|
| ENDOCAPSULE SMALL INTESTINAL CAPSULE ENDOSCOPE SET           | MAJ-2027                      |                  |
| (Components)                                                 | CAPSULE ENDOSCOPE (5 pcs.)    | EC-S10           |
| ENDOCAPSULE RECORDER SET                                     | MAJ-2029                      |                  |
| (Components)                                                 | CAPSULE ACTIVATOR (2 pcs.)    | MAJ-1478         |
|                                                              | ENDOCAPSULE RECORDER          | OLYMPUS<br>RE-10 |
|                                                              | BATTERY PACK                  | MAJ-2030         |
|                                                              | ANTENNA UNIT                  | MAJ-2031         |
|                                                              | CRADLE                        | MAJ-2032         |
|                                                              | RECORDER HOLDER               | MAJ-2033         |
|                                                              | ANTENNA HOLDER                | MAJ-2034         |
|                                                              | LEAD ANTENNAUNIT              | MAJ-2294         |
|                                                              | ANTENNA LEAD COVER            | MAJ-1470         |
|                                                              | ENDOCAPSULE SOFTWARE 10       | MAJ-2188         |
|                                                              | ENDOCAPSULE SOFTWARE 10 LIGHT | MAJ-2189         |
| ENDOCAPSULE SOFTWARE 10 UPGRADE PACKAGE                      | MAJ-2190                      |                  |

### Table 5-1: ENDOCAPSULE SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Olympus. The logo is in blue and features the word "OLYMPUS" in a bold, sans-serif font. Below the wordmark is a thin, horizontal yellow line. Underneath the line is the tagline "Your Vision, Our Future" in a smaller, lighter font.

K173459 Page 3 of 4

### 5.5 Indications for Use

The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small

intestine mucosa.

 । It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.

It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.

It may be used in the visualization and monitoring of lesions that may be । potential causes of iron deficiency anemia(IDA) not detected by upper and lower endoscopy.

The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas.

The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM may be used as a tool in the detection of abnormalities of the small intestine and is intended for use in adults only.

### 5.6 Comparison of Technological Characteristics

The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM has the same technological characteristics and design as the predicate device except for Addition of Lead Type Antenna.

All other technological characteristics of both the subject and primary predicate devices are identical.

### 5.7 Performance Data

### 5.7.1 Summary of Non-Clinical Testing

The following performance data were provided in support of the substantial equivalence determination.

### -Performance testing - Bench

Performance testing was conducted to demonstrate that the OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM performs according to specifications and functions as intended. Test results verified the safety and effectiveness of the devices in accordance with design specifications and applicable standards. Conducted performance testings are listed below.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. A thin, yellow line is underneath the word. Below the line, the words "Your Vision. Our Future" are written in a smaller font.

K173459 Page 4 of 4

- Software verification and validation testing
- Electrical safety and electromagnetic compatibility (EMC) test
- Performance testing of radio signal communication
- Radio Interference testing
- FCC Test

### -Risk Management

Risk management was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

### 5.7.2 Summary of Clinical Testing

The clinical testing was performed to support the claims for improvement of capsule reading efficiency and the reduction of redundant images. A multicenter, randomized trial was conducted to examine the hypothesis that OMNI mode can remove and store redundant images of the same region of the small bowel and is superior to conventional review of all CE images (Normal mode) in terms of the time required to review the case data. Seven different expert CE physicians, blinded to the reading mode, reviewed the cases in randomized order . The review time was compared between Normal mode and OMNI mode. OMNI mode required, on average, 27.3 minutes for review while Normal mode required 75.1 minutes, on average (p<0.001), which is a 64% reduction. The result supports the claims for improvement of capsule reading efficiency and the reduction of redundant images. The conducted performance testing is shown below.

- Evaluation of performance of the Omni mode for detecting video capsule endoscopy images:A multicenter randomized controlled trial

### 5.8 Conclusion

Compared to the primary predicate device, the OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM does not incorporate any significant changes in the indications for use, method of operation, material, or design that could affect the safety or effectiveness of the subject device. Therefore, the subject device is substantially equivalent to the cited primary predicate device.

---

**Source:** [https://fda.innolitics.com/device/K173459](https://fda.innolitics.com/device/K173459)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
