K173425 · Sensus Healthcare, Inc. · JAD · Mar 23, 2018 · Radiology
Device Facts
Record ID
K173425
Device Name
SRT-100+
Applicant
Sensus Healthcare, Inc.
Product Code
JAD · Radiology
Decision Date
Mar 23, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5900
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K173425 · Mar 23, 2018
SRT-100+
Sensus Healthcare, Inc.
Published clinical literature
A literature review of multiple clinical studies was conducted to support the safety and efficacy claims for the SRT-100+ in treating primary malignant epithelial neoplasms of the skin and keloids.
Literature Review; Keloids; Malignant Neoplasms; Superficial Radiotherapy
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Summary of multiple clinical studies for the treatment of keloids and supporting literature; Literature review
Patients with primary malignant epithelial neoplasms of the skin and keloids
Not applicable for this study
Safety and efficacy of superficial radiotherapy
Indications for Use
The SRT-100+ System is a low energy x-ray system, intended for superficial radiotherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's Sarcoma; and (t) the treatment of keloids. Keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue. A red-diode laser is employed for the assisting with cone applicator placement.
Device Story
SRT-100+ is a stand-alone superficial x-ray radiation therapy system. It comprises a control console, base unit with high-voltage generator, red-diode laser for applicator positioning, and interchangeable applicators. The system emits filtered low-energy x-rays (20-100 kV) using a Comet MXR-100 tube. Applicators define the source-to-skin distance and beam diameter; each contains a unique magnet binary code for automatic system detection and dose-rate correlation. Operated by physicians or qualified personnel in clinical settings, the device delivers fractionated radiation doses to skin lesions. The control console manages settings; no automated treatment planning is performed. The system provides precise, user-error-free dose rate calculations based on the detected applicator. Benefits include targeted treatment of skin cancers and keloids while sparing healthy tissue through fractionated delivery.
Clinical Evidence
Bench testing only. No clinical trials were conducted for this submission. Safety and efficacy are supported by a review of clinical literature regarding superficial radiotherapy for the indicated skin conditions and keloids, alongside verification and validation testing performed by independent laboratories to ensure compliance with IEC/ISO standards.
Technological Characteristics
System uses a Comet MXR-100 metal ceramic x-ray tube. Energy range: 20, 30, 50, 70, 100 kV. Filtration: 0.10 mm Al (20-30 kV), 0.45 mm Al (50 kV), 0.75 mm Al (70 kV), 1.15 mm Al (100 kV). Includes red-diode laser (FDA Class 3A). Applicators use magnetic binary identification for automatic detection. Stand-alone unit with articulating arm. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-8, IEC 62366, IEC 60825-1.
Indications for Use
Indicated for superficial radiotherapy of primary malignant epithelial neoplasms of the skin (basal cell, squamous cell, metatypic, cutaneous appendage, Kaposi's sarcoma) and keloids in patients requiring radiation therapy.
Regulatory Classification
Identification
An x-ray radiation therapy system is a device intended to produce and control x-rays used for radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Sensus Healthcare, Inc. % Mr. Nicolas Soro QA/RA Manager 851 Broken Sound Parkway NW, Suite 215 BOCA RATON FL 33487
March 23, 2018
Re: K173425
Trade/Device Name: SRT-100+ Regulation Number: 21 CFR 892.5900 Regulation Name: X-ray radiation therapy system Regulatory Class: II Product Code: JAD Dated: February 23, 2018 Received: February 26, 2018
Dear Mr. Soro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173425
Device Name SRT-100+
Indications for Use (Describe)
The SRT-100+ System is a low energy x-ray system, intended for superficial radiotherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's Sarcoma; and (t) the treatment of keloids. Keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue.
A red-diode laser is employed for the assisting with cone applicator placement.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
## 510(k) SUMMARY
[As Required by 21 CFR 807.92(c)]
| Submitter's Name & Address: | Sensus Healthcare<br>851 Broken Sound Parkway NW<br>Suite 215<br>Boca Raton, FL 33487 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kal Fishman, CTO<br>Telephone (561) 922-5808<br>Fax (561) 948-2071<br>kal@sensushealthcare.com |
| Date Summary Prepared: | March 19, 2018 |
| Device Name: | Trade/Proprietary Name – SRT-100+<br>Common/Usual Name – Superficial X-ray Radiation<br>Therapy System<br>Classification Name – X-ray Radiation Therapy Syster<br>(892.5900) |
| Classification: | Class II |
| Product Code: | JAD |
| Regulation Number: | 892.5900 |
| Predicate Device: | Sensus Healthcare SRT-100 Vision (K150037) |
| Reference Devices: | Sensus Healthcare SRT-100 (K123985) |
# Previous FDA Submissions & Clearances
510(k) Summary (Sensus Healthcare SRT-100+)
{5}------------------------------------------------
Table 1.0 contains a compilation of previously cleared versions of the SRT-100 and SRT-100 Vision Systems.
| Clearance | Date | Class | Product<br>Code | System | Indications for Use |
|-----------|--------------------|-------|-----------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K150037 | 16 October<br>2015 | II | JAD | SRT-100<br>Vision | The SRT-100 Vision is a low energy x-ray system, with ultrasound imaging and red-diode laser capabilities, intended for superficial radiotherapy treatment of primary malignant epithelial neoplasms of the skin and keloids. |
| K131582 | 28 August<br>2013 | II | JAD | SRT-100<br>Vision | The SRT-100 Vision is a low energy x-ray system, with ultrasound imaging and red-diode laser capabilities, intended for superficial radiotherapy treatment of primary malignant epithelial neoplasms of the skin and keloids. |
| K123985 | 14 May<br>2013 | II | JAD | SRT-100 | The SRT-100 is a low energy x-ray system intended for superficial radiotherapy treatment of primary malignant epithelial neoplasms of the skin and keloids. |
| K063456* | 9 January<br>2007 | II | JAD | SRT-100 | The SRT-100 is a low energy x-ray system intended for superficial radiotherapy treatment of primary malignant epithelial neoplasms of the skin. |
Table 1.0 – Previously Cleared SRT-100 Systems
*Note: This original clearance was provided to Topex, Inc. in 2007. It was also the predicate device referenced for the Esteya System.
{6}------------------------------------------------
## Device Description
The Sensus Healthcare SRT-100+ is a complete, stand-alone, x-ray radiation therapy system. It consists of four separate components: (a) control console; (b) base unit; (c) red-diode laser; and (d) applicators.
- (a) Control Console: Specifically designed module housing the switches and indicators used by the operator to set up and execute x-ray exposures. The controls adjust the machine functions and settings only! There is no treatment planning capability. The Control Console is connected, through a cable, to the Base Unit.
- (b) Base Unit: The base unit consists of a cabinet containing the high voltage generator, power supply components, cooling system, and an arm/positioning mechanism on which the x-ray tube housing assembly is mounted. A series of Applicators are included, which are affixed to the x-ray port on the x-ray tube housing assembly to limit the x-ray beam and provide fixed Source-to-Skin Distance (SSD). The X-ray Tube-Housing Assembly contains a motorized filter mechanism, which moves the appropriate beam filter: (a) 0.10 mm Al at 20 to 30 kV: (b) 0.45 mm A1 at 50kV: (c) 0.75 mm Al at 70 kV; and (d) 1.15 mm Al at 100 kV; into the beam path depending on the kV setting selected by the operator.
- (c) Red-Diode Laser: A red-diode laser is integrated with the SRT-100+ System. The laser is manufactured by U.S. Laser and is classified as FDA Laser Class 3A. The application of the red-diode laser with the Sensus SRT-100 Vision has been tested in accordance with IEC 60825-1.
- (d) Applicators: The system is shipped with a set of interchangeable treatment applicators, which define the source to skin distance (SSD) and the diameter of the
{7}------------------------------------------------
treatment beam's exposure. The applicator size, therefore, determines the amount of total dose delivered per minute to the lesion and the actual area that will be treated by the system's x-ray beam. Each applicator is embedded with a unique magnet binary combination, which allows the system to automatically detect an applicator as it is mounted on the x-ray port. This provides the system with the information about the applicator's SSD and diameters, which allows it to correlate the applicable dose rate for each applicator that is attached to the x-ray port, thus allowing for a precise and user-error-free dose rate per minute calculation. There are a variety of applicator sizes available for use with the Sensus Healthcare SRT-100+ System, driven by the treatment modality.
#### Indications for Use:
The SRT-100+ System is a low energy x-ray system, intended for superficial radiotherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's Sarcoma; and (f) the treatment of keloids. Keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue.
A red-diode laser is employed for the assisting with cone applicator placement.
## Prescriptive Statement
The SRT-100+™ System is intended for use by a physician and other specially-qualified individuals properly trained in the system's use, science, and application.
{8}------------------------------------------------
Caution: Federal law restricts this device to sale by or on the order of a physician.
## Technological Characteristics/Principles of Operation
The SRT-100+ produces and emits filtered, low energy (20, 30, 50, 70, and 100 kV) xray radiation, which is electrically generated using a conventional metal ceramic x-ray tube (Comet MXR-100). Provision is made to limit the x-radiation to a specified treatment field, and to control the radiation dose to the patient through selection and monitoring of energy, emission level and duration of emission. To mitigate effects of ionizing radiation on healthy cells, and to accumulate more damage in the neoplastic cells and keloids associated with scar tissue, the total dose is fractionated, which means distributing the total dose over a period of time. Typically, 8 to 12 fractions at a rate of 1 to 5 per week are used to deliver a total dose of 40-60 Gy, although larger PMENs may require up to 40 fractions over an 8-week period for a total dose of 80 Gy. When treating keloids, typically 1 to 4 fractions are employed, delivering a total dose in the range of 10 to 40 Gy.
#### Review of Clinical Literature
A summary of multiple clinical studies for the treatment of keloids and supporting literature has been collected, reviewed, and a clinical evaluation report scripted in support of this 510(k). The literature collected and reviewed supports Sensus Healthcare's claims of safety and efficacy for the application of superficial radiotherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's
{9}------------------------------------------------
Sarcoma; and (f) the treatment of keloids. Keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue.
## Summary of Non-Clinical Performance Testing
All of the Sensus Healthcare SRT-100+ capabilities are also found in the previously cleared SRT-100 Vision System predicate device; and similar capabilities. The Sensus Healthcare SRT-100+ System, as configured, has been engineered and tested to meet Sensus Healthcare product requirements, required electrical and mechanical safety standards, and meet clinical expectations. All testing of the SRT-100+ System was performed in accordance with defined test cases with clearly delineated acceptance criteria. Additionally, FDA consensus standards and recognized ISO and IEC standards (e.g., IEC 60601-1 3rd edition) were employed for the bench testing, functional testing, and overall system performance testing of the SRT-100+. Furthermore, all testing was performed by qualified and accredited independent laboratories.
#### Non-clinical Safety Tests
The Sensus Healthcare SRT-100+ has been designed and constructed to meet the following electrical and mechanical safety standards:
- o IEC 60601-1:2005 Ed.3+A1; C1: 2014 Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance
- o AAMI ES60601-1:2005 +A1:2012 Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance
- CSA C22.2#60601-1:2014 Ed.3 Medical Electrical Equipment Part 1: General o Requirements For Basic Safety And Essential Performance
{10}------------------------------------------------
- IEC 60601-1-2:2007 Part 1: General requirements for safety. Collateral standard: o electromagnetic compatibility - requirements and tests
- IEC 60601-1-6:2010 Ed.3+A1 Medical Electrical Equipment Part 1-6: General o Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
- IEC 60601-2-8:2010 Ed.2+A1 Medical Electrical Equipment Part 2-8: Particular o Requirements For Basic Safety And Essential Performance Of Therapeutic X-Ray Equipment Operating In The Range 10 Kv To 1 Mv
- IEC 62366:2007 Ed.1+A1 Medical Devices Application Of Usability Engineering o To Medical Devices
- IEC 60825-1:2007 Safety of laser products Part 1: Equipment o classification and requirements
### Substantial Equivalence Discussion
## Sensus Healthcare SRT-100 Vision (K150037)
The SRT-100+ was designed using the critical components of the SRT-100 Vision system. including the same X-ray tube for delivery of radiation and the same control boards/PCBs that control the unit. The SRT-100+ has a contracted indication for use when compared to the SRT-100 Vision with the removal of the imaging capabilities (ultrasound module) and a removal of the electronic brachytherapy treatment. Functionally, the unit can perform radiation emissions in the 20, 30, 50, 70, or 100 kV range as opposed to the Vision's larger 20, 30, 50, 60, 70, 80, 90, or 100 kV range. Furthermore, the SRT-100+ has four (4) filtration options vs six (6) filtration options on the SRT-100 Vision. The unit has an articulating arm similar to the predicate device that is used to position the X-ray head for delivery of therapy. During treatment, the same applicators and applicator tips that are attached to the X-ray head of the SRT-100 Vision are used on the SRT-100+. The device is similar in size, dimensions, and
{11}------------------------------------------------
weight where both devices are mounted on a baseplate attached to castors for ease of mobility and positioning during installation and operation of the device.
### Sensus Healthcare SRT-100 (K123985)
The SRT-100 system was originally designed by Topex Inc. and cleared under 510(k) K063456 on January 09, 2007. The Topex SRT-100 and associated intellectual properties (IP) were acquired by Sensus Healthcare in 2010. The SRT-100 user requirements were reviewed along with the SRT-100 Vision user requirements during the development of the SRT-100+ system. Functionally, the SRT-100 can perform radiation emissions in 50, 70, or 100 kV range as opposed to the SRT-100+'s expanded 20, 30, 50, 70, or 100 kV range. The core functionality of the SRT-100+ is very similar to the SRT-100 including comparable size, weight, and dimensions, but ultimately the SRT-100 Vision was chosen as the predicate device due to similarity of subcomponents used in the device.
## Conclusion Statement
All the SRT-100+ System's intended use can be found in the intended use of the predicate device and reference device. Any technological changes to the device are minor, primarily as it pertains to industrial design, and do not raise new questions of safety or effectiveness. Performance testing, along with verification activities demonstrate that SRT-100+ is as safe and effective, and performs as well as the predicate device. Therefore, SRT-100+ can be considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.