K173391 · Osteomed, LLC · KCT · May 10, 2018 · General Hospital
Device Facts
Record ID
K173391
Device Name
OsteoMed QuickFix Hybrid MMF Sterilization Tray
Applicant
Osteomed, LLC
Product Code
KCT · General Hospital
Decision Date
May 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The OsteoMed MMF Sterilization Tray is intended to contain MMF implants and surgical instruments for sterilization, storage and handling. The OsteoMed MMF Sterilization Tray is suitable for dynamic air removal (pre-vacuum) steam sterilization methods. The tray is not intended to maintain sterility; it is intended to be used in conjunction with a validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices. The sterilization tray may also be used in conjunction with a legally marketed rigid container.
Device Story
Device consists of anodized/non-anodized aluminum base, lid, and locking latch; perforated design allows steam penetration. Used in hospital environments to organize, protect, and transport dental MMF implants and surgical instruments during steam sterilization cycles. Operated by hospital staff; requires placement within FDA-cleared sterilization wrap or rigid container to maintain sterility post-process. Benefits include organized, protected, and effective sterilization of surgical instrumentation. Device withstands repeated steam sterilization cycles.
Clinical Evidence
No clinical data. Bench testing only: design validation, ship testing (ASTM D-4169), cleaning validation (AAMI TIR 30), and sterilization validation (3 consecutive tests in wrap and rigid container). Durability confirmed via 100-cycle steam sterilization testing.
Technological Characteristics
Materials: Anodized and non-anodized aluminum. Form factor: Rectangular base with lid and locking latch (9.5 x 5.0 x 1.5 inches). Sterilization: Pre-vacuum steam (270°F/132°C for 4 minutes). Perforated design for steam penetration. Reusable. Maximum load capacity: 2.3 lbs.
Indications for Use
Indicated for containing MMF implants and surgical instruments during sterilization, storage, and handling in hospital environments. Suitable for use with dynamic air removal (pre-vacuum) steam sterilization. Not intended to maintain sterility independently; requires use with FDA-cleared sterilization wrap or rigid container. Maximum load: 2.3 lbs.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 10, 2018
OsteoMed, LLC Kathryn Jayne Manager, Regulatory Affairs 3885 Arapaho Rd. Addison, TX 75001
Re: K173391
Trade/Device Name: OsteoMed MMF Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Tray Regulatory Class: Class II Product Code: KCT Dated: April 09, 2018 Received: April 10, 2018
Dear Kathryn Jayne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173391
Device Name OsteoMed MMF Sterilization Tray
### Indications for Use (Describe)
The OsteoMed MMF Sterilization Tray is intended to contain MMF implants and surgical instruments for sterilization, storage and handling. The OsteoMed MMF Sterilization Tray is suitable for dynamic air removal (pre-vacuum) steam sterilization methods. The tray is not intended to maintain sterility; it is intended to be used in conjunction with a validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices. The sterilization tray may also be used in conjunction with a legally marketed rigid container.
Sterilization validation was performed on MMF implants and as surgical instrumentation. Do not exceed a maximum load of 2.3 lbs. in the sterilization tray.
Validated sterilization parameters for OsteoMed MMF Sterilization Tray: Method: Steam Cycle Pre-Vacuum Temperature: 270°F (132°C) Exposure Time: 4 minutes Minimum Dry Time: 30 minutes
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | |
| <div style="display:flex; align-items:center;"><input checked="checked" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OsteoMed" in a stylized font. The logo is surrounded by an oval shape, with the top half of the oval in blue and the bottom half in gray. The logo is clean and professional, and it is likely used to represent a medical company.
### 510(k) SUMMARY-K173391
#### SUBMITTER I.
OsteoMed 3885 Arapaho Rd. Addison, TX 75001
| Phone: | 972-677-4766 |
|--------|---------------------|
| Fax: | 800-390-2620 |
| Email: | kjayne@osteomed.com |
Contact: Phone: Fax: Email:
Kathryn Jayne 972-677-4766 800-390-2620 kjayne@osteomed.com
Alternate Contact: Alicia Hemphill Phone: 972-677-4783 Email: ahemphill@osteomed.com
Date Prepared:
May 08, 2018
#### II. DEVICE
| Name of the Device: | OsteoMed MMF Sterilization Tray |
|-----------------------|------------------------------------------------------------------------|
| Common or Usual Name: | Sterilization Tray |
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes &<br>Other Accessories |
| Regulation: | 21 CFR 880.6850 |
| Regulatory Class: | II |
| Product Code: | KCT |
#### III. PREDICATE DEVICE
Primary Predicate: NuVasive® Sterilization Trays (K143579)
#### IV. DEVICE DESCRIPTION
The OsteoMed MMF Sterilization Tray is composed from a material commonly used (anodized and non-anodized Aluminum) to enclose, protect, and organize
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OsteoMed" in large, bold letters. The logo has a swoosh design that goes around the word "OsteoMed". The swoosh is blue on top and gray on the bottom.
OsteoMed's dental non-sterile devices, which meet national or international standards. The OsteoMed MMF Sterilization Tray is intended to provide storage for appropriate dental (MMF) devices and accessories during sterilization, storage, and transportation within the hospital environment.
The tray is composed of a base, a lid, and locking latch that secure the device and accessories. The tray is perforated to allow for steam sterilization. An FDAcleared wrap or FDA-cleared rigid sterilization container must be used for sterilization and to maintain the sterility of the contents. The OsteoMed MMF Sterilization Tray is designed to be used with standard autoclaves used in hospitals and healthcare facilities. As such, the OsteoMed MMF Sterilization Tray is effective for containing devices during sterilization and have been designed to withstand repeated steam sterilization cycles.
#### V. INDICATIONS FOR USE
The OsteoMed MMF Sterilization Tray is intended to contain MMF implants and surgical instruments for sterilization, storage and handling. The OsteoMed MMF Sterilization Tray is suitable for dynamic air removal (pre-vacuum) steam sterilization methods. The tray is not intended to maintain sterility; it is intended to be used in conjunction with a validated, FDA-cleared sterilization wrap in order to maintain sterility of the enclosed devices. The sterilization tray may also be used in conjunction with a legally marketed rigid container.
Do not exceed a maximum load of 2.3 lbs. in the sterilization tray.
Validated sterilization parameters for OsteoMed MMF Sterilization Tray:
Method: Steam Cycle Pre-Vacuum Temperature: 270°F (132°C) Exposure Time: 4 minutes Minimum Dry Time: 30 minutes
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
At a high level, the subject and predicate devices are substantially equivalent based on the following same technological elements as demonstrated below.
| Characteristics | OsteoMed MMF Sterilization<br>Tray<br>Subject Device | NuVasive Sterilization Trays<br>Predicate Device (K143579) | Comparison |
|-----------------|------------------------------------------------------|------------------------------------------------------------|------------|
| Product Code | KCT | KCT | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OSTEOMED" in large, bold, blue letters. The logo is surrounded by a gray and blue swoosh design. The logo is clean and professional, and it is likely used on the company's website and marketing materials.
| 21 CFR | 880.6850 | 880.6850 | Same |
|-----------------------------------------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| System<br>Components | Base, lid, locking latch | Base, lid, insert trays, brackets,<br>caddies | Similar |
| Material<br>Composition | Aluminum | Aluminum, Nylon, Silicone,<br>Stainless Steel, Polypropylene,<br>Polyphenylsulfone | Similar |
| Physical<br>Properties | Evenly distributed perforated<br>hole | Evenly distributed perforated<br>hole | Same |
| Configurations/<br>Dimensions<br>(L<br>x W x H) in. | Rectangle base with lid<br>(9.5 x 5.0 x 1.5) | Rectangle base with lid, inserts,<br>brackets and caddies; Multiple<br>dimensions to accommodate<br>different product configurations<br>(9.8 x 6.3 x 3.6) | Similar |
| Sterilant<br>Penetration | Sterilant (steam) penetration<br>through perforations in tray | Sterilant (steam) penetration<br>through perforations in tray | Same |
| Sterilization<br>method | Steam | Steam | Same |
| Sterilization cycle | Pre-vacuum | Pre-vacuum | Same |
| Exposure<br>temperature | 270°F (132°C) | 270°F(132°C) | Same |
| Exposure time | 4 minutes | 4 minutes | Same |
| Dry time | 30 minutes | 30 minutes | Same |
| Vent to Volume<br>Ratio | 0.050 in²/in³ | unknown | unknown |
| Reusable | Yes | Yes | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OsteoMed" in large, bold letters. The word "Osteo" is in blue, and the word "Med" is in gray. The logo is surrounded by a blue and gray oval shape. The blue part of the oval is on top, and the gray part is on the bottom.
| Characteristics | OsteoMed QuickFix Hybrid<br>MMF Sterilization Tray | NuVasive Sterilization Trays | Comparison |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | The OsteoMed MMF Sterilization<br>Tray is intended to contain MMF<br>implants and surgical instruments<br>for sterilization, storage and<br>handling. The OsteoMed MMF<br>Sterilization Tray is suitable for<br>dynamic air removal (pre-vacuum)<br>steam sterilization methods. The<br>tray is not intended to maintain | The NuVasive Sterilization Trays<br>are intended to contain NuVasive<br>implants and surgical instruments<br>for sterilization, storage and<br>handling. The NuVasive<br>Sterilization Trays are suitable for<br>dynamic air removal (pre-vacuum)<br>steam sterilization methods. The<br>trays are not intended to maintain | |
| | sterility; it is intended to be used in<br>conjunction with a validated,<br>FDA-cleared sterilization wrap in<br>order to maintain sterility of the<br>enclosed devices. The sterilization<br>tray may also be used in | sterility; they are intended to be<br>used in conjunction with a<br>validated, FDA-cleared sterilization<br>wrap in order to maintain sterility<br>of the enclosed devices. | |
| | conjunction with a legally<br>marketed rigid container. | Sterilization validations for three<br>level worst case tray (20.5" x 10" x<br>7.375") included three different<br>loading configurations: implant<br>only content (26.4 lbs.), instrument | |
| Indications of Use | Sterilization validation was<br>performed on MMF implants and<br>accessories such as surgical<br>instrumentation. Do not exceed a | only content (39.95 lbs.) and mixed<br>instruments and implants content<br>(30.45 lbs.) | Similar |
| | maximum load of 2.3 lbs. in the<br>sterilization tray. | Validated worst case loading<br>configurations included 23<br>instruments with lumens including<br>the following worst case lumen | |
| | Validated sterilization parameters<br>for OsteoMed MMF Sterilization<br>Tray: | dimensions: - 1.6 mm x 306.8 mm -<br>1.1 mm x 166.4 mm - 7.2 mm x<br>331.9 mm - 5.5 mm x 356.9 mm | |
| | Method: Steam Cycle<br>Pre-Vacuum Temperature: 270°F<br>(132°C)<br>Exposure Time: 4 minutes | Do not exceed a maximum load of<br>25 lbs in the sterilization tray.<br>Validated sterilization parameters<br>for NuVasive Sterilization Trays: | |
| | Minimum Dry Time: 30 minutes | Method: Steam Cycle: Pre-Vacuum<br>Temperature: 270°F (132°C)<br>Exposure Time: 4 minutes<br>Minimum Dry Time: 30 minutes<br>Minimum Cool Down Time: 40<br>minutes | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OSTEOMED" in large, bold letters. The logo has a blue swoosh above the text and a gray swoosh below the text. The logo is simple and modern.
#### PERFORMANCE DATA VII.
### Bench Testing
The following table provides a summary of the bench testing conducted.
| Test<br>Description | Test Method Description | Summary of Test<br>Results |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Design<br>Validation | Surgeons evaluated the<br>functionality of the tray. | Pass; met all<br>requirements |
| Ship Testing | Ship packaged plate per ASTM<br>D-4169 and evaluate package<br>contents for damage upon return | Pass |
| Cleaning<br>Validation<br>(reusable<br>instruments) | Use the ratchet handle as worst-<br>case device to verify that re-<br>useable instruments can be cleaned<br>per instructions to achieve<br>standard requirements for reusable<br>devices per AAMI TIR 30. | Pass |
| Sterilization<br>Validation | Validate sterilization parameters<br>with 3 consecutive tests. Conduct<br>test for both system wrapped in<br>polypropylene sterile wrap and<br>system enclosed in rigid container. | Passed for both sterile<br>wrap and rigid container<br>configurations |
| Sterilization<br>Cycles<br>Validation | The device was steam sterilized<br>for 100 cycles with the<br>sterilization parameters described<br>in the IFU The subject device was<br>then evaluated to determine if it<br>met pre-determined acceptance<br>criteria. | Passed functional and<br>visual criteria after 100<br>cycles |
Bench testing was conducted on production equivalent devices. The devices passed the pre-acceptance criteria.
A tolerance study was conducted for the lid and latch of the sterilization tray.
## Animal Study
No animal studies were performed to demonstrate safety and efficacy.
## Clinical Studies
No clinical studies were performed to demonstrate safety and efficacy.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OsteoMed" in bold, sans-serif font. The word "Osteo" is in blue, and the word "Med" is in gray. The logo is surrounded by a blue and gray swoosh design.
# VIII. CONCLUSIONS
Based on the intended use, technological characteristics, performance data, and nonclinical tests performed, subject Instrument trays, are substantially equivalent to, and are as safe and as effective as, the legally marketed predicate device, NuVasive® Sterilization Trays cleared under K143579 under regulation 21 CFR 880.6850, product code KCT.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.