← Product Code [PEK](/productcode/PEK) · K173334

# SeaSpine Spinous Process System (K173334)

_SeaSpine Orthopedics Corporation · PEK · Feb 22, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173334

## Device Facts

- **Applicant:** SeaSpine Orthopedics Corporation
- **Product Code:** [PEK](/productcode/PEK.md)
- **Decision Date:** Feb 22, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The SeaSpine Spinous Process System is a posterior non-pedicle supplemental fixation system intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: · Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) - · Spondylolisthesis - · Trauma (i.e., fracture or dislocation) - · Spinal tumor The SeaSpine Spinous Process System is not intended for stand-alone use.

## Device Story

Posterior non-pedicle supplemental fixation system; consists of implantable spacer with male/female plates and set screws; clamps bilaterally to spinous processes. Spikes/teeth on plates provide fixation to resist rotation, flexion, and extension. Used by surgeons in clinical settings to achieve supplemental fusion at a single level in non-cervical spine (T1-S1). Implants available in various sizes to match patient anatomy. Device provides mechanical stability during fusion process; not intended for stand-alone use.

## Clinical Evidence

No clinical data. Substantial equivalence determination based on non-clinical bench testing. Mechanical performance (static axial compression, static axial torsion, dynamic axial compression bending) evaluated per ASTM F1717 and ASTM F2624.

## Technological Characteristics

Posterior non-pedicle supplemental fixation system. Materials: Ti-6Al-4V ELI (ASTM F136). Design: Implantable spacer with male/female plates, variable barrel width, fixed wing/plate length, and set screws. Fixation mechanism: Bilateral clamping with integrated spikes/teeth. Sterilization: Not specified. Connectivity: N/A. Software: None.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- SeaSpine Spinous Process System ([K121924](/device/K121924.md))
- Aspen MIS Fixation System ([K071877](/device/K071877.md), [K121316](/device/K121316.md))
- Aurora Spine ZIP Flared System ([K141317](/device/K141317.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 4, 2021

SeaSpine Orthopedics Corporation Gina Flores Sr. Regulatory Affairs Specialist 5770 Armada Drive Carlsbad, California 92008

Re: K173334

Trade/Device Name: SeaSpine Spinous Process System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK

Dear Gina Flores:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 22, 2018. Specifically, FDA is updating this SE Letter to correct the trade name in the Indications for Use Page and Indications for Use section of the 510(k) Summary.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Colin O'Neill, M.B.E., OHT6: Office of Orthopedic Devices, (301) 796-6428, colin.oneill@fda.hhs.gov.

Sincerely,

Colin O'neill -S

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 22, 2018

SeaSpine Orthopedics Corporation Ms. Gina Flores Senior Regulatory Affairs Specialist 5770 Armada Drive Carlsbad, California 92008

Re: K173334

Trade/Device Name: SeaSpine Spinous Process System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: PEK Dated: January 26, 2018 Received: January 29, 2018

Dear Ms. Flores:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{2}------------------------------------------------

# Page 2 - Ms. Gina Flores

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Ronald P. Jean

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K173334

Device Name SeaSpine Spinous Process System

#### Indications for Use (Describe)

The SeaSpine Spinous Process System is a posterior non-pedicle supplemental fixation system intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions:

· Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)

- · Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal tumor

The SeaSpine Spinous Process System is not intended for stand-alone use.

| Type of Use (Select one or both, as applicable) | <div> </div> |
|-------------------------------------------------|--------------|
|-------------------------------------------------|--------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

# Contact Details

| Applicant Name:      | SeaSpine Orthopedics Corporation                       |
|----------------------|--------------------------------------------------------|
| Address:             | 5770 Armada Drive, Carlsbad CA                         |
| Phone number:        | (760) 216-5136                                         |
| Fax number:          | (760) 683-6874                                         |
| Contact person:      | Gina Flores, Sr. Regulatory Specialist                 |
| Email address:       | gina.flores@seaspine.com                               |
| Date Prepared:       | February 21, 2018                                      |
| Device Name          |                                                        |
| Trade Name:          | SeaSpine Spinous Process System                        |
| Common Name:         | Spinous Process Plate                                  |
| Classification Name: | Spinal Interlaminal Fixation Orthosis                  |
| Classification:      | 21 CFR 888.3050, Spinal Interlaminal Fixation Orthosis |
| Class:               | II                                                     |
| Product Code:        | PEK                                                    |

# Legally Marketed Predicate Device

| 510(k)<br>Number      | Product Code | Trade Name                         | Manufacturer                           |
|-----------------------|--------------|------------------------------------|----------------------------------------|
| Primary Predicate     |              |                                    |                                        |
| K121924               | PEK          | SeaSpine Spinous Process<br>System | SeaSpine<br>Orthopedics<br>Corporation |
| Additional Predicates |              |                                    |                                        |
| K071877,<br>K121316   | KWP, MNI     | Aspen MIS Fixation System          | Zimmer/Biomet<br>(formerly Lanx, Inc.) |
| K141317               | PEK          | Aurora Spine ZIP Flared<br>System  | Aurora Spine, Inc.                     |

## Device Description

The SeaSpine Spinous Process System is a posterior, non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). The Spinous Process System consists of an implantable spacer featuring 2 plates (male and female) with a variable barrel width, fixed wing length, and fixed plate width, and set screws that clamp bilaterally to the spinous processes. Each plate contains spikes (teeth) for fixation to the spinous process to aid in resisting 

{5}------------------------------------------------

rotation, flexion, and extension after implantation. Implants are available in a range of sizes to suit the individual pathology and anatomical conditions of the patient, and manufactured from Ti-6Al-4V ELI per ASTM F136.

# Intended Use/Indications for Use

The SeaSpine Spinous Process System is a posterior non-pedicle supplemental fixation system intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions:

· Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)

- Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal tumor

The SeaSpine Spinous Process System is intended for stand-alone use.

# Summary of Technological Characteristics

The Spinous Process System and predicate device have the same operational principle; they are spacers that engage the spinous processes and provide fixation during fusion.

The SeaSpine Spinous Process System is substantially equivalent to the cited predicate devices in areas including intended use/indications for use, technological characteristics (operating principle, design, materials, sterility, manufacturing, etc.) and performance (mechanical safety).

The subject and predicate device are based on the following similar technological elements:

- · Plate lengths
- Plate widths
- · Barrel lengths and widths
- · Same implant materials, Titanium Alloy (6AL-4V ELI per ASTM F136)
- · Utilize the same set screw

## Non-Clinical Testing

Static axial compression, static axial torsion, and dynamic axial compression bending tests were performed using methods based on ASTM F1717 and ASTM F2624. The testing demonstrated substantially equivalent mechanical performance as compared to the predicate.

## Clinical Testing

Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.

## Conclusions

The submitted data demonstrate that the SeaSpine Spinous Process System is as safe, as effective, and performs at least as safely and effectively as the cited legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K173334](https://fda.innolitics.com/device/K173334)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
