K173333 · W. L. Gore & Associates, Inc. · OXF · Apr 5, 2018 · General, Plastic Surgery
Device Facts
Record ID
K173333
Device Name
GORE ENFORM Biomaterial
Applicant
W. L. Gore & Associates, Inc.
Product Code
OXF · General, Plastic Surgery
Decision Date
Apr 5, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GORE® ENFORM Biomaterial is indicated for use in the reinforcement of soft tissue. This includes use in patients requiring soft tissue reinforcement in plastic and reconstructive surgery. Examples of applications where GORE® ENFORM Biomaterial may be used include hernia repair as suture line reinforcement, muscle flap reinforcement, and general tissue reconstructions.
Device Story
GORE ENFORM Biomaterial is a porous, 3D bioabsorbable scaffold composed of PGA:TMC copolymer; functions to reinforce soft tissue by filling deficits during wound healing; elicits physiological response to replace scaffold with native tissue; gradually absorbs. Two configurations: one standard, one with added PGA:TMC film layer for visceral protection in intraperitoneal applications. Supplied sterile for single use; trimmed to size by surgeon during procedure. Used in surgical settings for plastic and reconstructive surgery. Benefits include soft tissue reinforcement and gradual replacement by native tissue.
Clinical Evidence
Bench testing and animal studies only. Bench testing evaluated surgical manipulation, suture needle penetration, stiffness, and tensile strength; all met performance criteria. Animal studies compared visceral protection, tissue ingrowth, and tissue response against predicate controls.
Technological Characteristics
Porous, 3D scaffold of bioabsorbable PGA:TMC copolymer. Available in two configurations: standard and with an added PGA:TMC film layer for visceral protection. Supplied sterile for single use. Biocompatibility evaluated per ISO 10993-1. Mechanical properties include tensile strength and stiffness.
Indications for Use
Indicated for patients requiring soft tissue reinforcement in plastic and reconstructive surgery, including hernia repair, suture line reinforcement, muscle flap reinforcement, and general tissue reconstructions.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 5, 2018
W.L. Gore & Associates, Inc. Barbara Smith Regulatory Associate 1505 N. Fourth Street Flagstaff, Arizona 86004
Re: K173333
Trade/Device Name: GORE ENFORM Biomaterial Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OXF, OXC, OWT, OWZ Dated: October 20, 2017 Received: October 23, 2017
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
# Sincerely. Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173333
Device Name GORE® ENFORM Biomaterial
#### Indications for Use (Describe)
GORE® ENFORM Biomaterial is indicated for use in the reinforcement of soft tissue. This includes use in patients requiring soft tissue reinforcement in plastic and reconstructive surgery. Examples of applications where GORE® ENFORM Biomaterial may be used include hernia repair as suture line reinforcement, muscle flap reinforcement, and general tissue reconstructions.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Submitter
W. L. Gore & Associates, Inc. 301 Airport Road Elkton, Maryland 21921 Regulatory contact: Barbara L. Smith, RAC Phone: 410-506-8189 E-mail: blsmith@wlgore.com
## Date Prepared
October 20, 2017
## Device Names/Classification
| Device Name: | GORE® ENFORM Biomateria |
|----------------------|---------------------------|
| Classification Name: | Mesh, surgical, polymeric |
| Regulation: | 21CFR 878.3300 |
| Classification: | Class II |
| Product Code: | OXF, OWT, OWZ, OXC |
## Predicate Devices
- . K163217 GORE® BIO-A® Tissue Reinforcement
- K152609 GORE® SYNECOR Intraperitoneal Biomaterial ●
- K143380 BARD® Phasix™ ST Mesh
## Device Description
As packaged, GORE® ENFORM Biomaterial is a porous, three-dimensional sheet comprised of a bioabsorbable PGA:TMC copolymer in a matrix (scaffold) structure that functions to reinforce soft tissue during the phases of wound healing by filling softtissue deficits. The bioabsorbable, porous scaffold structure of the ENFORM device elicits a physiological response which fills the deficit with native tissue and gradually absorbs the device. There are two configurations of the GORE® ENFORM Biomaterial. One configuration will possess an added PGA:TMC film layer on one side of the device to provide visceral protection in soft tissue reinforcement applications requiring intraperitoneal contact with the viscera. Both ENFORM configurations are available in various sizes and can be trimmed to the desired shape by the surgeon at time of use. The GORE® ENFORM Biomaterial is supplied sterile for single use only.
{4}------------------------------------------------
### Indications for Use
GORE® ENFORM Biomaterial is indicated for use in the reinforcement of soft tissue. This includes use in patients requiring soft tissue reinforcement in plastic and reconstructive surgery. Examples of applications where GORE® ENFORM Biomaterial may be used include hernia repair as suture line reinforcement, muscle flap reinforcement, and general tissue reconstructions.
#### Differences in Technological Characteristics
The GORE® ENFORM Biomaterial with film configuration, a fully absorbable device, contains a visceral protection film layer not present in the predicate GORE® BIO-A® Tissue Reinforcement. However, the GORE® SYNECOR Intraperitoneal predicate possesses a visceral protection film comprised of the same material as the GORE® ENFORM Biomaterial with film. (The permanent PTFE knit middle layer of the predicate GORE® SYNECOR is not present in the GORE® ENFORM Biomaterial devices and therefore, was not compared.) The predicate PHASIX™ ST, also a fully absorbable device, possesses a visceral protection layer comprised of a different material (hvdrogel barrier).
#### Summary of Performance Testing
Biocompatibility: Biocompatibility evaluation appropriate for a tissue contacting implant with permanent duration of exposure was conducted according to "Use of International Standard /SO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Guidance for Industry and Food and Drug Administration Staff. June 16, 2016".
Bench study: The bench testing conducted but was not limited to evaluation of surgical manipulation, suture needle penetration, stiffness, tensile strength. Results demonstrated that both configurations of the GORE® ENFORM Biomaterial met the performance criteria established for the indicated uses of the device.
Animal study: Animal studies were conducted to compare the performance of the ENFORM Biomaterial to the predicate control devices with respect to visceral protection, tissue ingrowth and tissue response.
#### Conclusion
The information and test results contained within this 510(k) premarket notification demonstrate that the subject GORE® ENFORM Biomaterial device is substantially equivalent to the predicate devices in terms of indications for use, design, materials, biocompatibility, sterilization, packaging, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.