EZ Plate

K173306 · Zavation Medical Products, LLC · KWQ · Jan 22, 2018 · Orthopedic

Device Facts

Record IDK173306
Device NameEZ Plate
ApplicantZavation Medical Products, LLC
Product CodeKWQ · Orthopedic
Decision DateJan 22, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The EZ Plate™ cervical plate system is intended for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the treatment of the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion.

Device Story

EZ Plate™ is an anterior cervical plate system consisting of plates and self-tapping/self-drilling screws. Used by surgeons in clinical settings to provide spinal stabilization as an adjunct to fusion. The system functions as a mechanical fixation construct to immobilize spinal segments during the healing process. It is implanted via surgical procedure to treat degenerative disc disease, trauma, deformity, and other specified spinal pathologies. The device provides structural support to the cervical spine, facilitating patient recovery and fusion.

Clinical Evidence

Bench testing only. Static compression bending, torsion, and dynamic compression bending were performed per ASTM F1717 on worst-case constructs. Results demonstrated performance equivalent to or better than predicate devices.

Technological Characteristics

Components manufactured from titanium alloy (Ti-6Al-4V) per ASTM F136. System includes plates and self-tapping/self-drilling screws in various lengths and diameters. Mechanical fixation principle. No software or electronic components.

Indications for Use

Indicated for patients requiring anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis, or failed previous fusion.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. The words "U.S. Food & Drug Administration" are written in blue to the right of the square. January 22, 2018 Zavation Medical Products LLC Matt Jones Design Engineer 220 Lakeland Parkway Flowood, Mississippi 39232 Re: K173306 Trade/Device Name: EZ Plate™ Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: December 18, 2017 Received: December 19, 2017 Dear Matt Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Vincent J. Devlin -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173306 Device Name EZ Plate™M ### Indications for Use (Describe) The EZ Plate™ cervical plate system is intenior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the treatment of the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510K Summary | Date: | December 18, 2017 | |----------------------------|---------------------------------------------------------------------------------------------------------------------| | Submitter: | Zavation Medical Products LLC<br>220 Lakeland Pkwy<br>Flowood, MS 39232<br>Phone: 601-919-1119<br>Fax: 800-447-1302 | | Contact Person: | Matt Jones | | Type of 510(k) submission: | Traditional | | Trade name: | EZ Plate™ | | Common name: | Anterior Cervical Plate | | Classification regulation: | 888.3060 | | Device classification: | Class II | | Classification Panel: | Orthopedic | | Product code: | KWQ | ## Device Description: The EZ Plate™ cervical plate system consists of self-tapping/self-drilling screws and plates. Screws are available in a variety of diameter and length combinations. Plates are available in a variety of lengths. ## Intended Use: The EZ Plate™ cervical plate system is intended for anterior screw fixation of the cervical spine (C2-C7) as an adjunct to fusion. These implants have been designed to provide stabilization for the treatment of the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumor, pseudarthrosis or failed previous fusion. ### Materials: The EZ Plate™ cervical plate system components are manufactured from titanium alloy (Ti-6Al-4V) as described by ASTM F136. {4}------------------------------------------------ K173306 page 2 of 2 ### Primary Predicate Device: K112533 Zavation Cervical Plate System ### Additional Predicate Devices: K133518 Biomet MaxAn K021761 Medtronic Premier Plate K001794 EBI Cervical Plate System ### Technological Characteristics: The EZ Plate™ cervical plate system possesses the same technological characteristics as the predicate. These include: basic design (plate designed fixation system having various screw diameters and lengths), material (titanium alloy), sizes (variety of plate and screw sizes), and intended use (as described above). ### Performance Data: Static compression bending and torsion, and dynamic compression bending were performed according to ASTM F1717 on a worst-case, cervical plate construct. The mechanical test results demonstrated that the EZ Plate™ cervical plate system performs as well as or better than the predicate devices. ### Conclusion: The EZ Plate™ cervical plate system is substantially equivalent to the devices referenced above and is therefore safe and effective for its intended use.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%