For use as a tooth shade resin material in dental procedures, such as: - Direct anterior and posterior restorations - Direct bonded composite veneer - Diastema closure - Repair of porcelain/composite
Device Story
OMNICHROMA is a light-cured, radiopaque composite resin used by dentists for direct anterior and posterior restorations. The device is supplied in a single shade, eliminating the need for shade matching. It is applied by the clinician into the cavity or onto the tooth surface and cured via photo-polymerization using a dental curing light. The material serves as a restorative agent to repair carious lesions, close diastemas, or bond veneers. By providing a single-shade solution that conforms to ISO 4049 standards, it simplifies the restorative workflow for the practitioner and provides a functional, aesthetic tooth-colored restoration for the patient.
Clinical Evidence
No clinical tests were performed. Substantial equivalence is supported by non-clinical bench testing of physical properties in accordance with ISO 4049:2009, including depth of cure, flexural strength, water sorption, solubility, color stability, and radiopacity.
Technological Characteristics
Light-cured, radiopaque composite resin. Materials: Silica-zirconia filler, composite filler, UDMA, and TEGDMA monomers. Form factor: Syringe or pre-loaded tip. Energy source: External dental curing light (photo-polymerization). Conforms to ISO 4049:2009 standards for polymer-based restorative materials. Non-sterile.
Indications for Use
Indicated for dental patients requiring direct anterior and posterior restorations, direct bonded composite veneers, diastema closures, or repair of porcelain/composite restorations.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
February 27, 2018
Tokuyama Dental Corporation % Keith Barritt Fish & Richardson P.C. 901 15th Street, Suite 700 Washington, District of Columbia 20005
Re: K173275
Trade/Device Name: OMNICHROMA Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: November 28, 2017 Received: November 29, 2017
### Dear Keith Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173275
Device Name
OMNICHROMA
Indications for Use (Describe)
For use as a tooth shade resin material in dental procedures, such as:
- Direct anterior and posterior restorations
- Direct bonded composite veneer
- Diastema closure
- Repair of porcelain/composite
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K173275
# 510(k) Summary Tokuyama Dental Corporation OMNICHROMA dental resin material
#### Submitter
#### (i) 510(k) Submitter
Tokuyama Dental Corporation 38-9 Taitou 1-chome, Taitou-ku Tokyo 110-0016 Japan Phone: 011-81-3-3835-2261
# (ii) 510(k) Submitter Contact
Keith A. Barritt Fish & Richardson P.C. 901 15th Street, Suite 700 Washington, DC 20005 (202) 783-5070 Phone: Facsimile: (202) 783-2331 barritt@fr.com Email:
#### (iii) Preparation Date
February 26, 2018
#### De vice
| Trade or Proprietary Name: | OMNICHROMA |
|----------------------------|------------------------------|
| Common Name: | tooth shade resin material |
| Classification Name: | material, tooth shade, resin |
| Class: | 2 |
| Product Code: | EBF |
{4}------------------------------------------------
## Predicate Device
Primary Predicate: Tokuyama Dental Corporation's PALFIQUE ESTELITE LV CLEAR (K#010267).
Reference Devices: Kuraray's CLEARFIL MAJESTY ES-2 (K#121583); Heraeus Kulzer's VENUS PEARL (K#112501); Tokuyama Dental Corporation's ESTELITE SIGMA QUICK (K#080940); Tokuyama Dental Corporation's ESTELITE FLOW QUICK (K#051808); and Tokuyama Dental Corporation's ESTELITE BULK FILL FLOW (K#161353)
# De vice Description
The OMNICHROMA device is a light-cured, radiopaque composite resin for use in anterior and posterior restorations and is indicated for all carious classes. The OMNICHROMA device is available in only one shade.
The device does not come sterilized and is not intended to be sterilized prior to use.
# Indications for Use
For use as a tooth shade resin material in dental procedures, such as:
- Direct anterior and posterior restorations
- Direct bonded composite veneer
- Diastema closure
- Repair of porcelain/composite
# Comparison of Technological Characteristics
The OMNICHROMA device does not have its own energy source. There are no accessories provide with the device. The OMNICHROMA device has the same basic technological characteristics in terms of design, material, and chemical composition as the primary predicate device as shown below:
{5}------------------------------------------------
| | | Subject device | Primary predicate | Difference |
|------------------------|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Device name | | OMNICHROMA | Palfique Estelite LV Clear | - |
| Manufacturer | | Tokuyama Dental | Tokuyama Dental | - |
| 510(k) No. | | K173275 | K010267 | - |
| Classification name | | Material, Tooth Shade,<br>Resin | Material, Tooth Shade,<br>Resin | - |
| Indications for Use | | For use as a tooth shade<br>resin material in dental<br>procedures, such as:<br>- Direct anterior and<br>posterior restorations<br>- Direct bonded composite<br>veneer<br>- Diastema closure<br>- Repair of<br>porcelain/composite | For use as a tooth shade<br>resin material in dental<br>procedures. | Indications for<br>new device are<br>within those of<br>the predicate |
| Component | Container | Syringe or Pre-loaded tip | Syringe | Use of pre-loaded<br>tip |
| | Shade | 1 shade | 1 shade | - |
| Principle of operation | | Tooth shade resin material<br>that is cured by photo<br>polymerization. (Light-<br>cure) | Tooth shade resin material<br>that is cured by photo<br>polymerization. (Light-<br>cure) | - |
| Material | Filler | - Silica-zirconia filler<br>- Composite filler | - Silica-titania filler | As all of the<br>ingredients in the<br>subject device<br>have already<br>been authorized<br>by FDA for use<br>in similar devices |
| | Resin matrix monomer | - 1,6-bis(methacryl-<br>ethyloxy<br>carbonylamino)trimethyl<br>hexane (UDMA)<br>- Triethylene glycol<br>dimethacrylate<br>(TEGDMA) | - Methacrylates | |
| Physical property | Sensitivity to ambient light | Conformed to the<br>requirement of ISO 4049 | Conformed to the<br>requirement of ISO 4049 | Conformance to<br>standard; specific<br>measurements<br>may vary slightly |
| | Depth of cure | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | |
| | Flexural strength | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | |
| | Water sorption | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | |
| | Solubility | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | |
| | Color stability | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | |
| | Radio- opacity | Conformed to the<br>requirement of ISO 4049<br>Dentistry -- Polymer-<br>based restorative materials | - | |
| Sterilization | | Non-sterile | Non-sterile | |
# Comparison of OMNICHROMA device with the primary predicate device
{6}------------------------------------------------
{7}------------------------------------------------
#### Performance Data Summary
Non-clinical testing of the physical properties of the OMNICHROMA device was conducted in accordance with ISO 4049:2009. which specifies requirements for dental polymer-based restorative materials supplied in a form suitable for mechanical mixing, hand-mixing, or intra-oral and extra-oral external energy activation, and intended for use primarily for the direct or indirect restoration of cavities in the teeth and for luting. The polymer-based luting materials covered by ISO 4049:2009 are intended for use in the cementation or fixation of restorations and appliances such as inlays, onlays, veneers, crowns and bridges.
There were no clinical tests performed for the OMNICHROMA device.
# Conclusion
Based on the non-clinical testing conducted of the physical properties of the OMNICHROMA device in comparison to the predicate device identified above, and on the biocompatibility of authorized devices for the same use with the same ingredients, it is concluded that the OMNICHROMA device is substantially equivalent to the predicate device.
41318781.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.