K173198 · U&I Corporation · MRW · Jan 3, 2018 · OR
Device Facts
Record ID
K173198
Device Name
Facet Screw Fixation System
Applicant
U&I Corporation
Product Code
MRW · OR
Decision Date
Jan 3, 2018
Decision
SESE
Submission Type
Traditional
Device Class
Class U
Attributes
Therapeutic
Intended Use
The Facet Screw Fixation System is intended to stabilize the spine as an aid to fusion through bilateral immobilization of facet joints. The Facet Screw Fixation System is indicated for posterior surgical treatment with or without bone graft at single or multiple levels from C2 to S1 (inclusive). For transfacet fixation, the screws are inserted posteriorly through the superior side of the facet, across the facet joint, and into the pedicle. For translaminar facet fixation, the screws are inserted posteriorly through the lateral aspect of the spinous the lamina, through the superior side of the facet, across the facet joint, and into the Facet Screw Fixation System is indicated for treatment for any or all of the following: Spondylolisthesis Spondylolysis Pseudoarthrosis or failed previous fusions which are symptomatic Degenerative Disc Disease (DDD) as defined by back pain of discogenic origin with degeneration of the disk confirmed by history and radiographic studies and/or degenerative disease of the facets with instability Trauma including spinal fractures and/or dislocations
Device Story
Facet Screw Fixation System provides spinal stabilization via bilateral facet joint immobilization; adjunct for fusion. Components include cannulated titanium alloy screws and washers; screws available in multiple diameters/lengths. Surgical placement via K-WIRE; transfacet or translaminar approach. Used in posterior surgical procedures by surgeons to treat spinal instability/degeneration. Device provides mechanical fixation to support bone fusion; benefits include stabilization of facet articular process with minimal invasion.
Clinical Evidence
Bench testing only. Performed static torsion, driving torque, static pullout (ASTM F543-13), and static/dynamic bending (ASTM F2193-14) to verify mechanical performance equivalence to predicates.
Technological Characteristics
Materials: Titanium alloy (ASTM F136). Design: Cannulated screws and washers for transfacet/translaminar fixation. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring spinal stabilization as an aid to fusion at levels C2-S1. Conditions include spondylolisthesis, spondylolysis, symptomatic pseudoarthrosis/failed fusions, degenerative disc disease with instability, and spinal trauma (fractures/dislocations).
K101364 — LANX FACET SCREW SYSTEM · Lanx, Inc. · Aug 16, 2010
K250679 — FFX Facet Fixation System · Sc Medica · Dec 4, 2025
K152137 — FacetBRIDGE System · Ldr Spine USA · Jan 25, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized representation of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
U&i Corporation Yeon-Jung Choi Regulatory Affairs Specialist 20, Sandan-ro 76 eon-gil(Rd) Uijeongbu-si, 11781 Kr
Re: K173198
Trade/Device Name: Facet Screw Fixation System Regulatory Class: Unclassified Product Code: MRW Dated: December 20, 2017 Received: December 22, 2017
Dear Yeon-Jung Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Colin O'neill -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173198
Device Name Facet Screw Fixation System
#### Indications for Use (Describe)
The Facet Screw Fixation System is intended to stabilize the spine as an aid to fusion through bilateral immobilization of facet joints. The Facet Screw Fixation System is indicated for posterior surgical treatment with or without bone graft at single or multiple levels from C2 to S1 (inclusive). For transfacet fixation, the screws are inserted posteriorly through the superior side of the facet, across the facet joint, and into the pedicle. For translaminar facet fixation, the screws are inserted posteriorly through the lateral aspect of the spinous the lamina, through the superior side of the facet, across the facet joint, and into the Facet Screw Fixation System is indicated for treatment for any or all of the following:
Spondylolisthesis
Spondylolysis
Pseudoarthrosis or failed previous fusions which are symptomatic
Degenerative Disc Disease (DDD) as defined by back pain of discogenic origin with degeneration of the disk confirmed by history and radiographic studies and/or degenerative disease of the facets with instability
Trauma including spinal fractures and/or dislocations
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="12" width="12"> <path d="M1 6 L4 9 L11 2" fill="none" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" width="12"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Manufacturer: | U & I Corporation<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>11781, Korea |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | U & I Corporation<br>20, Sandan-ro 76beon-gil(Rd), Uijeongbu-si, Gyeonggi-do,<br>11781, Korea |
| Sponsor Contact: | Yeon-Jung Choi, Regulatory Affairs Specialist<br>+82 31 860 6838<br>yjchoi@youic.com |
| Date Prepared: | December 26, 2017 |
| Device Name: | Trade Name: Facet Screw Fixation System |
| Classification Name: | Unclassified |
| Common Name: | System, Facet Screw Spinal Device |
| Product Code: | MRW |
| Predicate Devices: | NuVasive™ Triad™ Facet Screw System (K020411) [Primary]<br>Resolute Facet Screw System (K121551)<br>DISCOVERY Facet Screw Fixation System(K012773) |
### Description of Device:
The Facet Screw Fixation System is manufactured by U&I Corporation. The Facet Screw Fixation System is a set of screws used for translaminar or transfacet fixation in the spine as an adjunct for fusion. The Facet Screw Fixation System consists of screws and washer. The screws are provided in two diameters and each diameter screw is provided in multiple lengths to accommodate the various anatomy of the spine. The washer is provided to help compression for bone fusion. The screw is cannulated to aid in placement via K-WIRE. All implant components are made of titanium alloy in accordance with ASTM F 136. The Facet Screw Fixation System is intended to provide immobilization and stabilization of the facet articular process to support fusion with minimal invasion.
{4}------------------------------------------------
#### Indications For Use:
The Facet Screw Fixation System is intended to stabilize the spine as an aid to fusion through bilateral immobilization of facet joints. The Facet Screw System is indicated for posterior surgical treatment with or without bone graft at single or multiple levels from C2 to S1 (inclusive). For transfacet fixation, the screws are inserted posteriorly through the superior side of the facet, across the facet joint, and into the pedicle. For translaminar facet fixation, the screws are inserted posteriorly through the lateral aspect of the spinous process, through the lamina, through the superior side of the facet, across the facet joint, and into the pedicle. The Facet Screw Fixation System is indicated for treatment for any or all of the following:
- Spondylolisthesis
- Spondylolysis
- Pseudoarthrosis or failed previous fusions which are symptomatic ●
- Degenerative Disk Disease (DDD) as defined by back pain of discogenic origin with degeneration of disk confirmed by history and radiographic studies and/or degenerative disease of the facets with instability
- Trauma including spinal fractures and/or dislocations.
### Substantial Equivalence:
Facet Screw Fixation System is substantially equivalent to Resolute Facet Screw System (K121551). NuVasive™ Triad™ Facet Screw System (K020411) and DISCOVERY Facet Screw Fixation System (K012773) in design, material, mechanical performance, function and intended use. The primary predicate device is the NuVasive™ Triad™ Facet Screw System, and the others are additional predicate devices.
The mechanical performance of the Facet Screw Fixation System demonstrated substantial equivalence to predicate devices.
- 1. Comparison Technological Characteristics
The predicate and proposed devices have the similar intended use and basic fundamental scientific technology and share the following similarities;
- The similar indications for use
- Similar design features ●
- Incorporate the same materials
- The equivalent mechanical performance
{5}------------------------------------------------
## 2. Performance Testing
The Facet Screw Fixation System was tested in a non clinical setting (bench testing) to assess that to no new safety and efficiency issues were raised with this device. Performance testing verifies performance of the Facet Screw Fixation System is substantially equivalent to predicate devices.
The following tests were performed:
- (1) Static torsion test according to ASTM F543-13
- (2) Driving torque test according to ASTM F543-13
- (3) Static pullout test according to ASTM F543-13
- (4) Static bending test according to ASTM F2193-14
- (5) Dynamic bending test according to ASTM F2193-14
#### 3. Conclusion
The data and information provided in this submission support the conclusion that the Facet Screw Fixation System is substantially equivalent to predicate devices with respect to indications for use and technological characteristics.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.