The T2 STRATOSPHERE™ Expandable Corpectomy System is a vertebral body replacement system intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body due to tumor, or trauma (i.e., fracture). The T2 STRATOSPHERE™ Expandable Centerpiece may be used with or without optional modular end caps which accommodate individual anatomic requirements. The device is to be used with supplemental fixation cleared for use in the thoracolumbar spine. T2 STRATOSPHERE™ Expandable Corpectomy System is intended for use with autograft or allograft, as an adjunct to fusion. The T2 STRATOSPHERE™ Expandable Corpectomy System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage turnors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
Device Story
Vertebral body replacement system; consists of expandable centerpieces, modular end caps, and surgical instruments. Used in thoracolumbar spine (T1-L5) to replace unstable/collapsed vertebrae due to trauma or tumor. Implanted by surgeons during spinal procedures; requires supplemental fixation. Restores spinal column integrity; supports bone graft (autograft/allograft) for fusion or palliative stabilization in advanced tumor cases. Provides mechanical support to spinal column; benefits patients by stabilizing spine and restoring structural integrity.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included static compression, compression fatigue, static torsion, torsion fatigue, and expulsion testing per ASTM F2077 and ASTM Draft Standard F-04.25.02.02. All tests met pre-determined acceptance criteria.
Technological Characteristics
Expandable vertebral body replacement system; materials similar to predicates. Mechanical testing performed per ASTM F2077 and ASTM Draft Standard F-04.25.02.02. Classified as MR-Conditional. System includes centerpieces, modular end caps, and associated surgical instruments.
Indications for Use
Indicated for patients with collapsed, damaged, or unstable vertebral bodies in the thoracolumbar spine (T1-L5) due to tumor or trauma (fracture). Intended for use with autograft/allograft as an adjunct to fusion, or for spinal column integrity restoration in patients with advanced stage tumors where fusion is not feasible due to limited life expectancy.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
T2 ALTITUDE™ Expandable Corpectomy System (K100976)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2017
Medtronic Sofamor Danek Ms. Kanesha Hines Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K173125
Trade/Device Name: T2 STRATOSPHERE™ Expandable Corpectomy System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MQP Dated: September 28, 2017 Received: September 29, 2017
Dear Ms. Hines:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
### Page 2 - Ms. Kanesha Hines
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173125
Device Name
T2 STRATOSPHERE™ Expandable Corpectomy System
#### Indications for Use (Describe)
The T2 STRATOSPHERE™ Expandable Corpectomy System is a vertebral body replacement system intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, or unstable vertebral body due to tumor, or trauma (i.e., fracture). The T2 STRATOSPHERE™ Expandable Centerpiece may be used with or without optional modular end caps which accommodate individual anatomic requirements. The device is to be used with supplemental fixation cleared for use in the thoracolumbar spine.
T2 STRATOSPHERE™ Expandable Corpectomy System is intended for use with autograft or allograft, as an adjunct to fusion. The T2 STRATOSPHERE™ Expandable Corpectomy System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage turnors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
# MEDTRONIC T2 STRATOSPHERE™ Expandable Corpectomy System
## December 2017
| 1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Submitter:<br>Telephone: (901)396-3133<br>Fax: (901) 346-9738<br>Kanesha Hines<br>Regulatory Affairs Specialist<br>Contact Person<br>Direct Telephone: (901)399-2670<br>Date Prepared<br>December 7, 2017<br>T2 STRATOSPHERE™ Expandable Corpectomy System<br>Name of Device<br>Common Name<br>Spinal intervertebral body fixation orthosis<br>Classification Name<br>Spinal intervertebral body fixation orthosis: 21 CFR 888.3060<br>Class II<br>Classification<br>Product Codes<br>MQP 21 CFR 888.3060<br>There are 4 Predicates.<br>Primary Predicate 1- T2 ALTITUDE™ Expandable<br>Corpectomy System (K100976, S.E. 10/21/2010)<br>Predicate 2- Aesculap HydroLift VBR (K083186, S.E.<br>03/04/2010)<br>Predicate Devices<br>Predicate 3- T2 XVBR™ Spinal System (K071033, S.E.<br>08/14/2007)<br>Predicate 4- VERTE-STACK Spinal System® (K052931, S.E.<br>11/16/2005) | | Medtronic Sofamor Danek |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|------------------------------------------------------------------|
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | The predicates have not been subject to a design related recall. |
| The T2 STRATOSPHERE™ Expandable Corpectomy System | Description of Devices | |
| consists of centerpieces, modular end caps, and associated | | |
| instruments. The T2 STRATOSPHERE™ Expandable<br>Corpectomy System is intended for vertebral body replacement to | | |
| | aid in the surgical correction and stabilization of the spine for | |
| | tumor and trauma pathologies. | |
| | The T2 STRATOSPHERE™ Expandable Corpectomy System is a | |
| | vertebral body replacement system intended for use in the | |
| | thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or | |
| | unstable vertebral body due to tumor, or trauma (i.e., fracture). | |
| | The T2 STRATOSPHERE™ Expandable Centerpiece may be used | |
| | with or without optional modular end caps which accommodate | |
| | individual anatomic requirements. The device is to be used with | |
| | supplemental fixation cleared for use in the thoracolumbar spine. | |
| Indications for Use | T2 STRATOSPHERE™ Expandable Corpectomy System is | |
| | intended for use with autograft or allograft, as an adjunct to | |
| | fusion. The T2 STRATOSPHERE™ Expandable Corpectomy | |
| | System is also intended to restore the integrity of the spinal | |
| | column even in the absence of fusion for a limited time in patients | |
| | with advanced stage tumors involving the thoracolumbar spine in | |
| | whom life expectancy is of insufficient duration to permit | |
| | achievement of fusion, with bone graft used at the surgeon's | |
| | discretion. | |
| | T2 STRATOSPHERE™ Expandable Corpectomy System | |
| | Implants | |
| | The primary predicate is T2 ALTITUDE™ Expandable | |
| | Corpectomy System (K100976, S.E. 10/21/2010). The subject | |
| | and predicate implants have identical intended use and similar | |
| | indications and materials as Primary Predicate 1 (K100976, S.E. | |
| Comparison of<br>Technological<br>Characteristics with the<br>Predicate Devices | 10/21/2010) and Predicate 2 K083186, S.E. 03/04/2010). The | |
| | predicate and subject devices have identical function and similar | |
| | scientific fundamental technology. | |
| | T2 STRATOSPHERE™ Expandable Corpectomy System | |
| | Instruments | |
| | The primary predicate is T2 ALTITUDE™ Expandable | |
| | Corpectomy System (K100976, S.E. 10/21/2010). The predicate | |
| | and subject devices have identical function and similar scientific | |
| | fundamental technology. | |
| | Mechanical Testing | |
| Performance Data | | |
| | In accordance with the Guidance for Industry and FDA Staff - | |
| | Spinal System 510(k)'s,Medtronic has evaluated the subject | |
| | devices to demonstrate substantial equivalence to the | |
| | predicate devices. | |
| | | |
| | Design verification testing for the subject implants was completed | |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.