← Product Code [MQK](/productcode/MQK) · K173075

# ZyMot ICSI Sperm Separation Device, ZyMot Multi Sperm Separation Device (850µl, 3ml) (K173075)

_Dxnow, Inc. · MQK · Mar 8, 2018 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173075

## Device Facts

- **Applicant:** Dxnow, Inc.
- **Product Code:** [MQK](/productcode/MQK.md)
- **Decision Date:** Mar 8, 2018
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 884.6160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The ZyMot ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures. The ZyMot Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.

## Device Story

Devices (ZyMōt ICSI and ZyMōt Multi) prepare motile sperm from semen for ART procedures; operate via sperm motility (swim-up) principle; used in clinical laboratory settings by embryologists/technicians. ZyMōt ICSI: 5 micro-channels; semen loaded into inlet; motile sperm collected from outlet after 30-minute incubation at 37°C. ZyMōt Multi: 850µl or 3ml volumes; semen added to lower chamber; separation medium added to upper chamber; microporous filter separates chambers; motile sperm swim through filter into upper chamber during 30-minute incubation at 37°C. Output is a concentrated sample of motile sperm for clinical use in fertilization procedures; enhances sperm quality for ART; single-use, radiation-sterilized.

## Clinical Evidence

Bench testing only. Sterilization validated per ISO 11137-2:2012. Shelf-life supported by accelerated aging (ASTM F1980-07), dye penetration (ASTM F1929-98), and seal strength (ASTM F88/F88M-15). Endotoxin levels ≤2.15 EU/device (USP<85). Human sperm survival assay (HSSA) confirmed ≥80% motility compared to control after 24 hours. Performance testing on donor semen showed significant increases in motile and progressively motile sperm percentages post-separation across all device versions.

## Technological Characteristics

Materials: Polymethylmethacrylate, borosilicate glass, polycarbonate, flash-spun high-density polyethylene fibers. Principle: Passive sperm motility (swim-up) through micro-channels or microporous filters. Energy: None (passive). Connectivity: None. Sterilization: Radiation (SAL 10^-6). Form factor: Disposable culture dish. Software: None.

## Regulatory Identification

Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- Qualis ([K133295](/device/K133295.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 8, 2018

DxNow, Inc. Kevin Sly Senior Advisor to DxNow, Inc. 401 Professional Drive, Suite 130 Gaithersburg, Maryland 20879-3429

Re: K173075

Trade/Device Name: ZyMot ICSI Sperm Separation Device, ZyMot Multi Sperm Separation Device (850µl, 3ml) Regulation Number: 21 CFR 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: Class II Product Code: MQK Dated: February 6, 2018 Received: February 6, 2018

Dear Kevin Sly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173075

Device Name ZyMōt ICSI Sperm Separation Device ZyMot Multi Sperm Separation Device (850ul, 3ml)

### Indications for Use (Describe)

The ZyMot ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures.

The ZyMot Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (IVF) and intrauterine insemination (IUI) procedures.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary K173075

#### 1. Submitter

DxNow, Inc. 401 Professional Drive, Suite 130 Gaithersburg, MD 20879-3429 Contact: Bill Sharp, CEO Email: bsharp@dxnowinc.com Phone: 240-454-9893

#### 2. Correspondent

Kevin Sly Senior Advisor, DxNow, Inc. Email: ksly@dxnowinc.com Phone: 240-751-6521

#### 3. Date Prepared: March 6, 2018

#### 4. Device Identification:

| Trade name:          | ZyMōt ICSI Sperm Separation Device<br>ZyMōt Multi Sperm Separation Device (850µl, 3ml) |
|----------------------|----------------------------------------------------------------------------------------|
| Common name:         | Sperm Separation Device                                                                |
| Classification name: | Assisted Reproduction Labware (21 CFR 884.6160)                                        |
| Product code:        | MQK (Labware, Assisted Reproduction)                                                   |
| Regulatory class:    | II                                                                                     |

#### ട. Predicate Device

Qualis (K133295). This predicate device has not been subject to any design related recalls.

#### 6. Device Description

The ZyMōt ICSI Sperm Separation Device and ZyMōt Multi Sperm Separation Device (850μl, 3ml) are used to prepare motile sperm for assisted reproductive technology (ART) procedures. They separate motile sperm from the semen based on the mobility (i.e., swimup nature) of motile sperm.

{4}------------------------------------------------

The ZyMōt ICSI Sperm Separation Device has five micro-channels, each accommodating 2ul of semen and processes up to five separations from one semen sample. Each channel has an inlet port for applying the semen sample and an outlet port for collecting the motile sperm. The ports are connected by a fluid-filled micro-channel in which separation occurs.

The ZyMōt Multi Sperm Separation Device has two processing volumes, 850µl and 3ml. Each version has an inlet port that communicates with the lower sample chamber. The sample chamber is separated from the upper collection chamber by a microporous filter. Untreated semen is added through the inlet port. After incubation, the separated motile sperm are collected from the upper chamber through the outlet port.

The subject devices are radiation-sterilized devices with a sterility assurance level (SAL) of 10-6. They are individually packaged and for single-use only.

#### 7. Indications for Use

The ZyMōt ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures.

The ZyMōt Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.

| Device/Predicate<br>Device(s): | K173075 (subject devices)                                                                                                                                                                                                                                                                                                                                                                                                                    | K133295 (predicate device)                                                                                                                                                  |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use         | The ZyMōt ICSI is intended for preparing<br>motile sperm from semen for use in the<br>treatment of infertile couples by<br>intracytoplasmic sperm injection (ICSI)<br>procedures.<br>The ZyMōt Multi is intended for preparing<br>motile sperm from semen for use in the<br>treatment of infertile couples by<br>intracytoplasmic sperm injection (ICSI), in<br>vitro fertilization (IVF) and intrauterine<br>insemination (IUI) procedures. | The Qualis is intended for preparing<br>motile human sperm for use in the<br>treatment of infertile couples by<br>intracytoplasmic sperm insertion<br>(ICSI) fertilization. |
| Design                         | ZyMōt ICSI – A disposable culture dish<br>containing five micro-channels, each<br>connected to an inlet port and an outlet<br>port<br>ZyMōt Multi – A disposable culture dish<br>containing a separation chamber and an<br>inlet port. The separation chamber has a<br>lower sample chamber and an upper                                                                                                                                     | A disposable culture dish with four<br>chambers connected by micro-<br>channels                                                                                             |

#### 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

{5}------------------------------------------------

|                        | collection chamber. The two chambers<br>are separated by a microporous filter.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanism of<br>action | ZyMōt ICSI — The inlet port is preloaded<br>with sperm separation medium and then<br>loaded with semen. After incubation at<br>37°C for 30 min, the motile sperm are<br>collected from the outlet port.<br>ZyMōt Multi – The semen is added to the<br>inlet port to fill the lower sample<br>chamber; then, the separation medium is<br>added to the upper collection chamber.<br>The loaded device is incubated at 37°C for<br>30 min to allow motile sperm to swim up<br>and cross the filter to migrate into the<br>over-laying separation medium in the<br>upper collection chamber. | The semen sample is placed in<br>Chamber A and separation medium<br>is placed in Chamber B.  Fluids from<br>both chambers flow via the micro-<br>channels into the central micro-<br>channel where the two fluids pass<br>side-by-side in laminar flow. Motile<br>sperm can swim across the interface<br>of the laminar flow streams and<br>pass into the separation medium<br>stream but non-motile sperm and<br>debris cannot. Motile sperm that<br>cross into the separation medium<br>flow are carried into Chamber C<br>where they are collected.  Non-<br>motile sperm and debris remain in<br>the semen sample flow from<br>Chamber A into Chamber D. |
| Material               | ZyMōt ICSI – Polymethylmethacrylate,<br>borosilicate glass, flash-spun high-density<br>polyethylene fibers<br>ZyMōt Multi - Polymethylmethacrylate,<br>polycarbonate, flash-spun high-density<br>polyethylene fibers                                                                                                                                                                                                                                                                                                                                                                     | Cyclo-olefin polymer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

The subject and predicate devices have different Indications for Use statements; however, they have the same intended use – preparing motile sperm from semen for use in ART procedures.

The subject and predicate devices have the same fundamental design incorporating a culture dish with an inlet port/chamber for loading of semen and an outlet port/chamber for collection of motile sperm. Both devices rely on sperm motility to separate motile sperm from non-motile sperm or non-sperm cells. The predicate device incorporates microfluidic technology, whereas the subject devices do not. However, this difference does not raise different questions of safety and effectiveness as compared to the predicate device, as both devices (and other cleared devices) have the same fundamental mode of action (i.e., using motility of sperm to prepare pools of motile sperm for use in ART procedures). In addition, the subject and predicate devices have different materials. However, differences in materials do not raise different questions of safety and effectiveness as compared to the predicate device.

#### g. Summary of Non-Clinical Performance Testing:

The following studies have been performed to support substantial equivalence to the predicate devices:

{6}------------------------------------------------

- . Sterilization validation study per ISO 11137-2:2012
- . Transportation simulation study per ASTM D5276-98
- . Package integrity testing following accelerated aging per ASTM F1980-07 to support a 12-month shelf-life:
	- * Dye penetration testing per ASTM F1929-98
	- * Seal strength testing per ASTM F88/F88M-15
- Endotoxin testing per USP<85>: ≤2.15 EU/device .
- . Human sperm survival assay (HSSA) before and after accelerated aging to support a 12-month shelf-life:

Donor sperm were exposed to the subject devices for 30 minutes and then, incubated at room temperature for 24 hours. The rate of motile sperm after incubation was compared to that of the control (no exposure to the subject device). The acceptance criterion is ≥80% of the control motility at 24 hours after exposure for 30 minutes.

- Performance testing: .
Each version of the subject device was used to separate motile sperm from donor semen samples. The separation procedures followed the Instructions for Use, and the percentage of motile sperm and progressively motile sperm in the output samples were compared to those of the input samples. A summary of the results is provided in the table below:

| Device            | Percentage of motile sperm<br>(before/after separation) | Percentage of progressively motile<br>sperm (before/after separation) |
|-------------------|---------------------------------------------------------|-----------------------------------------------------------------------|
| ZyMōt ICSI        | 29.7% / 87%                                             | 26.3% / 74.2%                                                         |
| ZyMōt Multi 850µl | 67% / 83.6%                                             | 44.7% / 68.6%                                                         |
| ZyMōt Multi 3ml   | 67% / 86.5%                                             | 44.7% / 75.4%                                                         |

#### 10. Conclusion

The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K173075](https://fda.innolitics.com/device/K173075)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
