← Product Code [LLZ](/productcode/LLZ) · K172998

# uWS-MI (K172998)

_Shanghai United Imaging Healthcare Co., Ltd. · LLZ · Apr 5, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172998

## Device Facts

- **Applicant:** Shanghai United Imaging Healthcare Co., Ltd.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Apr 5, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

uWS-MI is a software solution intended to be used for viewing, manipulation, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications: The PET/CT Oncology application is intended to provide tools to display and analyze the follow-up PET/CT data, with which users can do image registration, lesion segmentation, and statistical analysis. The PET/CT Dynamic Analysis application is intended to display the dynamic PET image data and its associated timeactivity curve. The PET/CT Brain Analysis (NeuroQ™) application is intended to analyze the brain PET scan, give quantitative results of the relative activity of 240 different brain regions, and make comparison of activity of normal brain regions in AC database. The PET/CT Cardiac Analysis (ECTbTM) application is intended to provide cardiac short axis reconstruction, browsing function. And it also performs perfusion analysis and cardiac function analysis of the cardiac short axis.

## Device Story

uWS-MI is a software-only medical imaging workstation for viewing, processing, and analyzing PET, CT, and PET/CT studies. Input data is received via DICOM 3.0 over network or external media. The device performs image registration, 3D visualization (MIP, VR, CPR, MPR), lesion segmentation, and statistical analysis (ROI/VOI). It includes specialized modules for oncology, dynamic PET, brain analysis, and cardiac perfusion/function. Used in clinical settings by trained, licensed physicians to interpret images and statistics for diagnostic support. Output is displayed on-screen for clinical review, aiding in lesion assessment, cardiac function evaluation, and neurological activity quantification. The device facilitates clinical decision-making by providing quantitative metrics and comparative analysis tools.

## Clinical Evidence

No clinical data. Bench testing only. Software verification and validation were performed, including hazard analysis, software requirements specification, architecture description, and cybersecurity documentation. Performance evaluation reports were provided for PET/CT oncology and image fusion modules. Previously cleared applications (NeuroQ, ECTb) were referenced for their respective functionalities.

## Technological Characteristics

Standalone software application running on Windows 7. Supports DICOM 3.0 connectivity. Features include 2D/3D visualization (VR, CPR, MPR, MIP), image fusion (auto/manual/spot), and quantitative analysis tools. Complies with NEMA PS 3.1-3.20 (DICOM), ISO 14971 (risk management), and IEC 62304 (software life cycle).

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Syngo.via ([K123920](/device/K123920.md))

## Reference Devices

- Syngo™TrueD Software ([K101749](/device/K101749.md))
- PET VCAR ([K063324](/device/K063324.md))
- NeuroQ™3.6 ([K130451](/device/K130451.md))
- Emory Cardiac Toolbox™3.2 ([K130902](/device/K130902.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA acronym and the agency's name are in blue.

Shanghai United Imaging Healthcare Co., Ltd. % Shumei Wang QM&RA, VP, Jiading District No. 2258 Chengbei Road Shanghai. 201807 CHINA

April 5, 2018

Re: K172998

Trade/Device Name: uWS-MI Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 30, 2018 Received: April 2, 2018

Dear Shumei Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172998

Device Name uWS-MI

#### Indications for Use (Describe)

uWS-MI is a software solution intended to be used for viewing, manipulation, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:

The PET/CT Oncology application is intended to provide tools to display and analyze the follow-up PET/CT data, with which users can do image registration, lesion segmentation, and statistical analysis.

The PET/CT Dynamic Analysis application is intended to display the dynamic PET image data and its associated timeactivity curve.

The PET/CT Brain Analysis (NeuroQ™) application is intended to analyze the brain PET scan, give quantitative results of the relative activity of 240 different brain regions, and make comparison of activity of normal brain regions in AC database.

The PET/CT Cardiac Analysis (ECTbTM) application is intended to provide cardiac short axis reconstruction, browsing function. And it also performs perfusion analysis and cardiac function analysis of the cardiac short axis.

Type of Use (*Select one or both, as applicable*)

| <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|-----------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark gray.

# SECTION 3

510(k) Summary

SECTION 3 - 1 of 11

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is dark gray. The logo is simple and modern.

## 510 (k) SUMMARY

- Date of Preparation: 1. March 29, 2018
#### 2. Sponsor Identification

Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China

Establishment Registration Number: Not yet registered or the Number

Contact Person: Shumei Wang Position: QM&RA VP Tel: +86-021-67076888-6776 Fax: +86-021-67076889 Email: shumei.wang@united-imaging.com

#### 3. Identification of Proposed Device

Trade Name: uWS-MI Common Name: PET/CT Image Post-Processing Software Model(s): uWS-MI

Regulatory Information Classification Name: Picture archiving and communications system Classification: II Product Code: LLZ Regulation Number: 21 CFR 892.2050 Review Panel: Radiology

#### 4. Identification of Predicate Device(s)

Predicate Device 510(k) Number: K123920 Device Name: Syngo.via

Reference Device#1 510(k) Number: K101749 Device Name: Syngo™TrueD Software

Reference Device#2 510(k) Number: K063324 Device Name: PET VCAR

## SECTION 3 - 2 of 11

{5}------------------------------------------------

Reference Device#3 510(k) Number: K130451 Device Name: NeuroQ"M3.6

Reference Device#4 510(k) Number: K130902 Device Name: Emory Cardiac Toolbox™3.2

#### 5. Device Description

uWS-MI is a comprehensive software solution designed to process, review and analyze PET, CT and PET/CT patient studies. It can transfer images in DICOM 3.0 format over a medical imaging network or import images from external storage devices such as CD/DVDs or flash drives. These images can be functional data, such as PET as well as anatomical datasets, such as CT. It can be at one or more time-points or include one or more time-frames. Multiple display formats including MIP and volume rendering and multiple statistical analysis including mean. maximum and minimum over a user-defined region is supported. A trained, licensed physician can interpret these displayed images as well as the statistics as per standard practice.

#### 6. Indications for Use

uWS-MI is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It supports interpretation and evaluation of examinations within healthcare institutions. It has the following additional indications:

- . The PET/CT Oncology application is intended to provide tools to display and analyze the follow-up PET/CT data, with which users can do image registration, lesion segmentation, and statistical analysis.
- . The PET/CT Dynamic Analysis application is intended to display the dynamic PET image data and its associated time-activity curve.
- . The PET/CT Brain Analysis (NeuroQ 100) application is intended to analyze the brain PET scan, give quantitative results of the relative activity of 240 different brain regions, and make comparison of activity of normal brain regions in AC database.
- . The PET/CT Cardiac Analysis (ECTbTM) application is intended to provide cardiac short axis reconstruction, browsing function. And it also performs perfusion analysis, activity analysis and cardiac function analysis of the cardiac short axis.

# SECTION 3 - 3 of 11

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted.

#### Summary of Technological Characteristics 7.

uWS-MI is a medical diagnostic application for viewing, manipulation, 3D visualization, post-processing and comparison of medical images. After importing the DICOM image data into the system, the operator is able to perform image browsing and processing and can further obtain advanced information for diagnosis. This is identical to the predicate device.

The following tables compare the main features, principles of operation, fundamental scientific technology and intended use of uWS-MI when compared to the predicate devices.

#### Substantially Equivalent (SE) Comparison 8.

The proposed device has similar indications for use, provides similar overall

functions and performs in a similar manner with the predicate device and reference devices.

## SECTION 3 - 4 of 11

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is in bold, sans-serif font and is stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in bold. The logo is simple and modern.

#### Table 1 SE discussion for basic functions

| Item                             | Proposed Device                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                | Remark |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| General                          | uWS-MI                                                                                                                                                                                                             | syngo.via(K123920)                                                                                                                                                                                              |        |
| Device<br>Classification<br>Name | Picture Archiving and<br>Communications System                                                                                                                                                                     | Picture Archiving and<br>Communications System                                                                                                                                                                  | Same   |
| Product Code                     | LLZ                                                                                                                                                                                                                | LLZ                                                                                                                                                                                                             | Same   |
| Regulation<br>Number             | 21 CFR 892.2050                                                                                                                                                                                                    | 21 CFR 892.2050                                                                                                                                                                                                 | Same   |
| Device Class                     | II                                                                                                                                                                                                                 | II                                                                                                                                                                                                              | Same   |
| Classification<br>Panel          | Radiology                                                                                                                                                                                                          | Radiology                                                                                                                                                                                                       | Same   |
| Specification                    |                                                                                                                                                                                                                    |                                                                                                                                                                                                                 |        |
| Image<br>communication           | Standard network protocols like<br>TCP/IP and standard<br>communication protocol DICOM.<br>Additional fast image.                                                                                                  | Standard network protocols like TCP/IP<br>and standard communication protocol<br>DICOM. Additional fast image.                                                                                                  | Same   |
| Hardware /OS                     | Windows 7                                                                                                                                                                                                          | Windows 7                                                                                                                                                                                                       | Same   |
| Patient<br>Administration        | Display and manage the image data<br>information of all patients stored in<br>the database.                                                                                                                        | Display and manage the image data<br>information of all patients stored in the<br>database.                                                                                                                     | Same   |
| Review 2D                        | 2D Review consists of basic<br>processing of the 2D images, e.g.<br>rotation, scaling, translation,<br>windowing, and measurements.                                                                                | 2D Review consists of basic processing<br>of the 2D images, e.g. rotation, scaling,<br>translation, windowing, and<br>measurements.                                                                             | Same   |
| Review 3D                        | 3D Review consist of                                                                                                                                                                                               | 3D Review consist of functionalities for                                                                                                                                                                        | Same   |
| Item                             | Proposed Device                                                                                                                                                                                                    | Predicate Device<br>syngo.via(K123920)                                                                                                                                                                          | Remark |
|                                  | uWS-MI<br>functionalities for displaying and<br>processing image in the 3D form,<br>e.g. VR, CPR, MPR, MIP, etc. The<br>module also includes tools for VOI<br>analysis.                                            | displaying and processing image in the<br>3D form, e.g. VR, CPR, MPR, MIP, etc.<br>The module also includes tools for VOI<br>analysis.                                                                          |        |
| Filming                          | Filming is a module dedicated for<br>image printing. The print tools<br>provide layout editing for both<br>single images and series.                                                                               | Filming is a module dedicated for image<br>printing. The print tools provide layout<br>editing for both single images and series.                                                                               | Same   |
| Fusion                           | Including Auto registration,<br>Manual registration, Spot<br>registration                                                                                                                                          | Including Auto registration, Manual<br>registration, Spot registration                                                                                                                                          | Same   |
| Inner View                       | Inner view the vessel , colon ,<br>trachea                                                                                                                                                                         | Inner view the vessel , colon , trachea                                                                                                                                                                         | Same   |
| Visibility                       | User-defined the display property of<br>fused image: Adjustment of preset<br>of T/B value; Adjustment of the<br>fused rate; Adjustment of pse                                                                      | User-defined the display property of<br>fused image: Adjustment of preset of T/B<br>value; Adjustment of the fused rate;<br>Adjustment of pse                                                                   | Same   |
| ROI/VOI                          | Plotting ROI or VOI, and obtaining<br>the maximum activity value, the<br>minimum activity value, mean<br>activity value, the volume/area of<br>region, and the maximum diameter<br>of volume, peak activity value; | Plotting ROI or VOI, and obtaining the<br>maximum activity value, the minimum<br>activity value, mean activity value, the<br>volume/area of region, and the maximum<br>diameter of volume, peak activity value; | Same   |
| MIP Display                      | The image can be displayed as MIP<br>and rotating play                                                                                                                                                             | The image can be displayed as MIP and<br>rotating play                                                                                                                                                          | Same   |
| Compare                          | Load two studies to compare                                                                                                                                                                                        | Load two studies to compare                                                                                                                                                                                     | Same   |

# SECTION 3 - 5 of 11

{8}------------------------------------------------

# SECTION 3 - 6 of 11

{9}------------------------------------------------

Table 2 SE discussion for advanced applications

| Application                                                   | Function name                                        | Propose<br>d device<br>uWS-MI | Reference<br>device#1:<br>SyngoTM TrueD<br>Software<br>(K101749) | Reference<br>device#2:<br>GE PET<br>VCAR<br>(K063324) | Reference<br>device#3:<br>NeuroqTM3.6<br>(K130451) | Reference<br>device#4:<br>Syntermed<br>Emory<br>Cardiac<br>ToolboxTM3.<br>2 (K130902) | Remark                                                                                                                                                                                                         |
|---------------------------------------------------------------|------------------------------------------------------|-------------------------------|------------------------------------------------------------------|-------------------------------------------------------|----------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PET/CT<br>Dynamic<br>Analysis                                 | Reframe/Rebin                                        | Yes                           | Do not have the<br>function of data<br>reframe                   | /                                                     | /                                                  | /                                                                                     | The Reframe/Rebin<br>module is applied<br>to generate<br>combined data of<br>any time phase<br>defined by the user.<br>This difference will<br>not impact the<br>safety and<br>effectiveness of the<br>device. |
|                                                               | ROI Analysis                                         | Yes                           | Yes                                                              | /                                                     | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Pseudo color                                         | Yes                           | Yes                                                              | /                                                     | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Automatic cine                                       | Yes                           | Yes                                                              | /                                                     | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Curve Analysis                                       | Yes                           | Yes                                                              | /                                                     | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Table Statistics                                     | Yes                           | Yes                                                              | /                                                     | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Save                                                 | Yes                           | Yes                                                              | /                                                     | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Filming                                              | Yes                           | Yes                                                              | /                                                     | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
| PET/CT<br>Oncology                                            | Compare<br>display                                   | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Auto<br>registration                                 | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
| Application                                                   | Function name                                        | Propose<br>d device<br>uWS-MI | Reference<br>device#1:<br>SyngoTMTrueD<br>Software<br>(K101749)  | Reference<br>device#2:<br>GE PET<br>VCAR<br>(K063324) | Reference<br>device#3:<br>NeuroqTM3.6<br>(K130451) | Reference<br>device#4:<br>Syntermed<br>Emory<br>Cardiac<br>ToolboxTM3.<br>2 (K130902) | Remark                                                                                                                                                                                                         |
|                                                               | Manual<br>registration                               | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Fix<br>Segmentation                                  | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Adaptive<br>Segmentation                             | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Spread                                               | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Statistical<br>Analysis                              | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Save                                                 | Yes                           | /                                                                | Yes                                                   | /                                                  | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Reformat                                             | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Quality Control                                      | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Slice Display                                        | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
| PET/CT<br>Brain<br>Analysis                                   | Compare                                              | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | PET/CT Fusion                                        | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | EQuAL analysis                                       | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
| (NeuroqTM3.<br>6)                                             | AmyQ                                                 | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
|                                                               | Save results,<br>Capture<br>region/display ,<br>Exit | Yes                           | /                                                                | /                                                     | Yes                                                | /                                                                                     | Same                                                                                                                                                                                                           |
| PET/CT                                                        | Reconstruction                                       | Yes                           | /                                                                | /                                                     | /                                                  | Yes                                                                                   | Same                                                                                                                                                                                                           |
| Application                                                   | Function name                                        | Propose<br>d device<br>uWS-MI | Reference<br>device#1:<br>SyngoTMTrueD<br>Software<br>(K101749)  | Reference<br>device#2:<br>GE PET<br>VCAR<br>(K063324) | Reference<br>device#3:<br>NeuroqTM3.6<br>(K130451) | Reference<br>device#4:<br>Syntermed<br>Emory<br>Cardiac<br>ToolboxTM3.<br>2 (K130902) | Remark                                                                                                                                                                                                         |
| Cardiac<br>Analysis<br>(Emory<br>Cardiac<br>ToolboxTM3.<br>2) | SSS                                                  | Yes                           | /                                                                | /                                                     | /                                                  | Yes                                                                                   | Same                                                                                                                                                                                                           |
|                                                               | Polor Maps                                           | Yes                           | /                                                                | /                                                     | /                                                  | Yes                                                                                   | Same                                                                                                                                                                                                           |
|                                                               | Perfusion<br>Analysis                                | Yes                           | /                                                                | /                                                     | /                                                  | Yes                                                                                   | Same                                                                                                                                                                                                           |
|                                                               | Viability<br>Analysis                                | Yes                           | /                                                                | /                                                     | /                                                  | Yes                                                                                   | Same                                                                                                                                                                                                           |
|                                                               | Functional<br>Analysis                               | Yes                           | /                                                                | /                                                     | /                                                  | Yes                                                                                   | Same                                                                                                                                                                                                           |
|                                                               | Save results,<br>Capture<br>region/display ,<br>Exit | Yes                           | /                                                                | /                                                     | /                                                  | Yes                                                                                   | Same                                                                                                                                                                                                           |

SECTION 3 - 7 of 11

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is dark teal in color. The logo is simple and modern in design.

SECTION 3 - 8 of 11

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is dark teal in color. The logo is simple and modern in design.

# SECTION 3 - 9 of 11

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized, dark gray icon that resembles a shield or a letter "U" with a horizontal line through the middle. The overall design is clean and modern.

#### 9. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

### Biocompatibility

Not Applicable to the proposed device, because the device is stand-alone software.

### Electrical Safety and Electromagnetic Compatibility (EMC)

Not Applicable to the proposed device, because the device is stand-alone software.

### Software Verification and Validation

Software verification and validation testing was provided to demonstrate safety and efficacy of the proposed device. This includes a hazard analysis, and the potential hazards have been classified as a moderate level of concern (LOC). These documentations include:

- Software description
- Hazard Analysis
- · Software requirements specification (SRS)
- · Software Architecture Description
- · Software Development Environment Description
- · Software Verification and Validation
- · Cyber security Documents

## Animal Study

No animal study was required.

### Clinical Studies

No clinical study was required.

### Performance Verification

- Performance evaluation Report for PET/CT Oncology and Image Fusion
- . The PET/CT Brain Analysis (NeuroQ100) application was cleared by FDA under K130451 on May 17th 2013
- The PET/CT Cardiac Analysis (ECTb™) application was cleared by FDA under . K130902 on June 14th 2013.

## Product Standards and Guidance

- NEMA PS 3.1 - 3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2016).
# SECTION 3 - 10 of 11

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in a bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" shape, which is also in a bold, sans-serif font. The color of the text and the "U" shape is a dark teal.

- ISO 14971 Medical devices - Application of risk management to medical devices (Edition 2.0, corrected version, 2007).
- . IEC 62304 Medical device software - Software life cycle processes (Edition 1.1, 2015).

#### Summary

The features described in this premarket submission are supported with the results of the testing mentioned above. The uWS-MI was found to have a safety and effectiveness profile that is similar to the predicate device.

### 10. Substantially Equivalent (SE) Conclusion

The proposed device is equivalent to the predicate device with regard to safety and efficacy. This conclusion is based upon a comparison of intended use, technological characteristics, performance specification, device hazards as well as verification and validation results.

In summary, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

## SECTION 3 - 11 of 11

---

**Source:** [https://fda.innolitics.com/device/K172998](https://fda.innolitics.com/device/K172998)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
