← Product Code [JAK](/productcode/JAK) · K172983

# HealthCCS (K172983)

_Zebra Medical Vision, Ltd. · JAK · Jun 13, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172983

## Device Facts

- **Applicant:** Zebra Medical Vision, Ltd.
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Jun 13, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

The HealthCCS Device is intended for use as a non-invasive post-processing software that can be used to evaluate calcified plaques in the coronary arteries, which may be a risk factor for coronary artery disease. The software can be used to generate reports of the total risk category of coronary calcium. This information can then be used by a physician for further analysis and treatment. The HealthCCS Device analyses pre-existing heart or chest ECG-Gated Triggered CT scans. The Device is indicated for use only on patients whose age at the CT scan was taken, was above 20 years old. This device generates a 4-category Agatston-equivalent risk score, and the patient management, especially for the patient with the score from 0-10, will depend on the physician's own judgment. It may require further testing to evaluate the appropriate clinical management.

## Device Story

HealthCCS is non-invasive post-processing software for cardiac CT images; identifies/quantifies coronary artery calcification. Input: pre-existing heart/chest ECG-gated CT scans. Operation: automatic identification of calcification via voxel density threshold (130 HU); CNN-based probability thresholding determines likelihood of association with coronary arterial distribution; calculates Agatston-equivalent score. Output: patient report with 4-category risk score. Used by physicians to assist in clinical analysis/treatment planning. Benefits: provides automated, reproducible risk stratification for coronary artery disease.

## Clinical Evidence

Retrospective performance study (n=249 studies). Compared 4-level risk categorization of HealthCCS against ground truth (GT) established by 3 radiologists using Kodak Carestream PACS. Reported overall agreement of 0.89 (95% CI: 0.85-0.92). Reproducibility assessed on 150 studies read three times; all Agatston-equivalent scores were identical across readings.

## Technological Characteristics

Software-based post-processing tool. Inputs: DICOM CT scans (heart/chest, ECG-gated, non-contrast). Processing: CNN-based algorithm for calcification identification; Agatston-equivalent scoring. Threshold: 130 HU. Spatial threshold: 1.5 mm². Standalone software tool.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- CSCS-001A Calcium Scoring Package ([K072737](/device/K072737.md))

## Reference Devices

- Kodak Carestream PACS ([K053347](/device/K053347.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zebra Medical Vision Ltd. % Matan Neeman Director, Quality and Regulatory Shefayim Commercial Center PO Box 25 Shefayim, 6099000 ISRAEL

June 13, 2018

Re: K172983

Trade/Device Name: HealthCCS Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 19, 2018 Received: May 22, 2018

Dear Matan Neeman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172983

Device Name HealthCCS

#### Indications for Use (Describe)

The HealthCCS Device is intended for use as a non-invasive post-processing software that can be used to evaluate calcified plaques in the coronary arteries, which may be a risk factor for coronary artery disease. The software can be used to generate reports of the total risk category of coronary calcium. This information can then be used by a physician for further analysis and treatment. The HealthCCS Device analyses pre-existing heart or chest ECG-Gated Triggered CT scans. The Device is indicated for use only on patients whose age at the CT scan was taken, was above 20 years old. This device generates a 4-category Agatston-equivalent risk score, and the patient management, especially for the patient with the score from 0-10, will depend on the physician's own judgment. It may require further testing to evaluate the appropriate clinical management.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Zebra Medical Vision logo. The logo consists of a yellow horizontal line on the left, followed by a stylized black and white zebra-striped "Z" symbol. To the right of the symbol is the company name "zebra" in black, with the words "MEDICAL.VISION" in smaller font below.

# 5. 510 (k) Summary

## 510(K) Summary - HealthCCS Zebra Medical Vision Ltd.

## 510(k) Number K172983

| Applicant's Name: | Zebra Medical Vision Ltd.  |
|-------------------|----------------------------|
|                   | Shefayim Commercial Center |
|                   | PO Box 25                  |
|                   | Shefayim, 6099000          |
|                   | ISRAEL                     |
|                   | Telephone: +972-9-8827795  |
|                   | Fax: +972-9-8827795        |

- Date Prepared: 06 June 2018
- Trade Name: HealthCCS

## Classification Name:

JAK - Computed tomography x-ray system

# Classification:

Class II, Radiology

#### Predicate Device:

The HealthCCS device is substantially equivalent to the following device:

| Proprietary Name       | CSCS-001A Calcium Scoring Package |
|------------------------|-----------------------------------|
| Premarket Notification | K072737 (5 Oct. 2007)             |
| Classification Name    | Computed tomography x-ray system  |
| Regulation Number      | 21 CFR 892.1750                   |
| Product Code           | JAK                               |
| Regulatory Class       | II                                |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Zebra Medical Vision logo. The logo consists of a stylized "Z" made of black and white stripes, followed by the word "zebra" in black lowercase letters. Below the word "zebra" is the text "MEDICAL.VISION" in smaller black letters.

#### Reference Device:

| Proprietary Name       | Kodak Carestream PACS                       |
|------------------------|---------------------------------------------|
| Premarket Notification | K053347                                     |
| Classification Name    | Picture archiving and communications system |
| Regulation Number      | 892.2050                                    |
| Product Code           | LLZ                                         |
| Regulatory Class       | II                                          |

## Performance Standards:

No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act.

#### Indication for Use:

The HealthCCS Device is intended for use as a non-invasive post-processing software that can be used to evaluate calcified plaques in the coronary arteries, which may be a risk factor for coronary artery disease. The software can be used to generate reports of the total risk category of coronary calcium. This information can then be used by a physician for further analysis and treatment. The HealthCCS Device analyses pre-existing heart or chest ECG-Gated/Triggered CT scans. The Device is indicated for use only on patients whose age at the time, when the CT scan was taken, was above 20 years old. This device generates a 4-category Agatston-equivalent risk score, and the patient management, especially for the patient with the score from 0-10, will depend on the physician's own judgment. It may require further testing to evaluate the appropriate clinical management.

## Device Description:

The HealthCCS Device is an automatic non-invasive post processing tool that uses cardiac CT images to identify and quantify calcification in the coronary arteries, known to be a risk factor for coronary disease.

HealthCCS Device quantifies calcification on non-contrast cardiac computed tomography (CT) scans. HealthCCS Device calculates the amount of identified calcification and

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Zebra Medical Vision logo. The logo consists of a yellow horizontal line on the left, followed by a stylized black and white zebra-like symbol. To the right of the symbol is the word "zebra" in black, with the words "MEDICAL.VISION" in smaller font below it.

reports the risk category of coronary calcium. This information can then be used by a physician for further analysis and treatment.

The following quantification and data-reporting functionalities are provided by the HealthCCS Device:

## Quantification:

- Calcification is automatically identified based on voxel density above a predefined threshold (130HU)
- । A CNN-based probability threshold is then applied to the identified calcification to determine the likelihood that the calcium is associated within the coronary arterial distribution.
- Coronary Artery Calcium score is calculated for each study series using the । Agatston method
- । If a single study contains more than one series for which Coronary Artery Calcium score is calculated, the results from these series are averaged to give a Agatston equivalent score for the study

#### Data reporting:

- Generate patients' reports with their respective calcium score category
## Performance Data:

Safety and performance of HealthCCS device has been evaluated and verified in accordance with software specifications and applicable performance standards. Software Development and Validation & Verification Processes have been implemented to ensure performance according to specifications, User Requirements and Federal Regulations and Guidance document: "General Principles of Software Validation".

The performance of the HealthCCS device has been validated in a retrospective performance study (n=249 studies), where the agreement between the 4-level risk categorization of the Health CCS device has been compared to the ground truth (GT) categorization by 3 radiologists using the Kodak Carestream PACS device (K053347). Adequate overall agreement of 0.89 (95% CI: [0.85, 0.92]) as well as adequate agreement per category were reported. The reproducibility was assessed on 150 studies that were

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Zebra Medical Vision logo. The logo consists of a yellow horizontal line on the left, followed by a stylized black and white zebra head, and the word "zebra" in black, with the words "MEDICAL.VISION" in smaller black letters below it. The logo is simple and modern, and it is likely used to represent a company that specializes in medical imaging or related technologies.

read three times. All Agatston equivalent scores per study ID were found identical over all three readings.

With respect to the specified requirements of accuracy, precision, and reproducibility, the obtained results have been found to be acceptable. The performance validation study demonstrated that HealthCCS device, provides accurate calcium scoring as compared to a reference device that shares the same intended use and technological characteristics as the predicate device that is currently marketed.

## Technological Characteristics Compared to Predicate Device:

The technological characteristics, e.g., overall design, mechanism of action, mode of operation, performance characteristics, etc., and the intended use of the HealthCCS device are substantially equivalent to the predicate device cited above.

A comparison of the technological characteristics with the predicate and reference devices is summarized below.

| Technological<br>Characteristics                                                                      | Proposed<br>Device:<br>HealthCCS<br>software tool               | Predicate Device:<br>CSCS-001A,<br>Calcium Scoring<br>Package (K072737) | Reference Device<br>Kodak Carestream<br>PACS (K053347)  | Summary                     |
|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------------|-----------------------------|
| General                                                                                               |                                                                 |                                                                         |                                                         |                             |
| Modality                                                                                              | CT                                                              | CT                                                                      | CT                                                      | Same                        |
| Image format                                                                                          | DICOM                                                           | DICOM                                                                   | DICOM                                                   | Same                        |
| Supported<br>Computed<br>Tomography<br>(CT) scan - body<br>part                                       | Heart/Chest                                                     | Heart/Chest                                                             | Heart/Chest                                             | Same                        |
| Supported<br>Computed<br>Tomography<br>(CT) scan - dose                                               | Typically Normal<br>Dose                                        | Typically Normal<br>Dose                                                | Typically Normal<br>Dose                                | Same                        |
| Supported<br>Computed<br>Tomography<br>(CT) scan - Use<br>of IV contrast                              | No                                                              | No                                                                      | No                                                      | Same                        |
| Supported<br>Computed<br>Tomography<br>(CT) scan – ECG<br>Gating/Triggering                           | ECG-Gated                                                       | ECG-Gated                                                               | ECG-Gated                                               | Same                        |
| Quantification                                                                                        |                                                                 |                                                                         |                                                         |                             |
| Calcification<br>location marking                                                                     | Automatic                                                       | Semi-automatic,<br>Manual                                               | Semi-automatic,<br>Manual                               | Substantially<br>equivalent |
| Selection of a<br>calcified plaques<br>based on voxel<br>identification<br>above a known<br>threshold | Yes                                                             | Yes                                                                     | Yes                                                     | Same                        |
| Default threshold<br>of Calcium                                                                       | 130 HU<br>(Hounsfield<br>Units)                                 | 130 HU (Hounsfield<br>Units)                                            | 130 HU (Hounsfield<br>Units)                            | Same                        |
| Spatial threshold                                                                                     | 1.5 mm²                                                         | 1 mm²                                                                   | 1 mm²                                                   | Substantially<br>equivalent |
| Coronary<br>Calcification<br>calculation<br>method                                                    | Agatston<br>equivalent score<br>based on the<br>Agatston method | – Agatston score<br>– Mass score<br>– Volume score                      | – Agatston score<br>– Mass score<br>– Volume score      | Substantially<br>equivalent |
| Computed<br>calcium scoring                                                                           | Total calcium<br>score                                          | Total calcium score,<br>and per-artery<br>calcium score                 | Total calcium score,<br>and per-artery<br>calcium score | Substantially<br>equivalent |
| Data reporting                                                                                        |                                                                 |                                                                         |                                                         |                             |
| Generate patient<br>report                                                                            | Yes                                                             | Yes                                                                     | Yes                                                     | Same                        |
| Printable hard<br>copy reports                                                                        | No                                                              | Yes                                                                     | Yes                                                     | Substantially<br>equivalent |
| Maintain a<br>patient database<br>for future<br>reference                                             | No                                                              | Yes                                                                     | Yes                                                     | Substantially<br>equivalent |
| Report of the<br>calcium score<br>category                                                            | Yes                                                             | Yes                                                                     | Yes                                                     | Substantially<br>equivalent |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Zebra Medical Vision logo. The logo consists of a stylized "Z" symbol on the left and the word "zebra" in lowercase letters on the right. Below the word "zebra" is the text "MEDICAL.VISION" in smaller letters. The logo is black and white.

# Substantial Equivalence:

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, HealthCCS

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Zebra Medical Vision logo. The logo consists of a yellow horizontal line, a stylized black and white "Z" symbol, and the word "zebra" in black, with the words "MEDICAL.VISION" in smaller font below it. The logo is simple and modern, and it is likely used to represent a company that specializes in medical imaging or related technologies.

device raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety, efficacy and performance.

The results of the performance comparison study demonstrated that the HealthCCS device performs as intended, in the specified use conditions, similarly to the reference device. The HealthCCS device is therefore substantially equivalent to the predicate device as the reference device and the predicate device share the same intended use and equivalent technological characteristics.

---

**Source:** [https://fda.innolitics.com/device/K172983](https://fda.innolitics.com/device/K172983)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
