K172920 · Philips Medical Systems Nederland B.V. · LNH · Jan 22, 2018 · Radiology
Device Facts
Record ID
K172920
Device Name
3D APT
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LNH · Radiology
Decision Date
Jan 22, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Amide Proton Transfer weighted imaging (3D APT) is a software option intended for use on Ingenia 3.0T and Ingenia 3.0T CX MR Systems. 3D APT is indicated for use in magnetic resonance imaging of the brain. 3D APT consists of an acquisition and reconstruction technique employing frequency-selective magnetization transfer effects to derive images reflecting the spatial distribution of amide protons, and thereby protein density. 3D APT images may assist a trained physician in diagnosis and therapy planning. APTW can be combined with multi-coil acceleration approaches (SENSE).
Device Story
3D APT is a software option for Philips Ingenia 3.0T/3.0T CX MR systems; utilizes frequency-selective magnetization transfer effects to image amide proton distribution/protein density. Input: MR signals acquired via frequency-selective RF pulses (1-2s duration) at ±3.5 ppm offsets; MultiTransmit system alternates excitation across 2 channels. Processing: 3D DIXON TSE readout; B0 inhomogeneity correction; signal subtraction between ±3.5 ppm offsets to isolate amide proton signal. Output: APTw images displayed with color scale. Used in clinical settings by radiologists/physicians; assists in diagnosis and therapy planning by visualizing protein density variations. Benefits: provides non-invasive contrast reflecting molecular protein content.
Clinical Evidence
Bench testing only. Verification and validation testing performed to demonstrate compliance with international and FDA-recognized consensus standards. No clinical data provided.
Technological Characteristics
Software-based MR imaging feature. Employs frequency-selective saturation RF pulses (1-2s) and 3D DIXON TSE readout. Features B0 inhomogeneity correction and MultiTransmit RF system (2-channel). Connectivity: Integrated into Philips Ingenia 3.0T/3.0T CX MR systems. Output: Color-scaled APTw images.
Indications for Use
Indicated for use in magnetic resonance imaging of the brain to assist trained physicians in diagnosis and therapy planning.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 (K163116)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in blue, followed by "U.S. FOOD & DRUG" in a larger font size and "ADMINISTRATION" in a smaller font size below.
Philips Medical Systems Nederland B.V. % Susan Quick Regulatory Affairs Specialist Philips Medical Systems (Cleveland) Inc. 595 Miner Rd CLEVELAND OH 44143
January 22, 2018
Re: K172920
Trade/Device Name: 3d Apt Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LNI Dated: December 22, 2017 Received: December 26, 2017
Dear Susan Ouick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert A. Ochs, Ph.D.
For
For
Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K172920
Device Name 3D APT
#### Indications for Use (Describe)
Amide Proton Transfer weighted imaging (3D APT) is a software option intended for use on Ingenia 3.0T and Ingenia 3.0T CX MR Systems. 3D APT is indicated for use in magnetic resonance imaging of the brain. 3D APT consists of an acquisition and reconstruction technique employing frequency-selective magnetization transfer effects to derive images reflecting the spatial distribution of amide protons, and thereby protein density. 3D APT images may assist a trained physician in diagnosis and therapy planning. APTW can be combined with multi-coil acceleration approaches (SENSE).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
Philips Medical Systems Nederland B.V. Magnetic Resonance Imaging 3D APT K172920
# 3D APT
## Section 005
# 510(k) Summary
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.
### 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | September 30, 2017 | |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6, 5684 PC, Best, The Netherlands<br>Establishment Registration Number: 3003768277 | |
| Primary Contact<br>Person: | Susan Quick<br>Regulatory Affairs Specialist<br>Phone: (440) 483-2291<br>E-mail: susan.quick@philips.com | |
| Secondary Contact<br>Person | Jan van de Kerkhof<br>Sr Manager Regulatory Affairs<br>Phone: +31 6 13300542<br>E-mail: jan.van.de.kerkhof@philips.com | |
| Device Name: | Amide Proton Transfer weighted MRI (3D APT) | |
| Classification: | Classification name: | Magnetic Resonance Diagnostic<br>device (MRDD) |
| | Classification Regulation: | 21CFR 892.1000 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Primary Product Code: | LNH, LNI |
| Primary Predicate Device: | Trade name: | Ingenia 1.5T, Ingenia 1.5T S and<br>Ingenia 3.0T R5.3 |
| | Manufacturer: | Philips Medical Systems<br>Nederland B.V. |
| | 510(k) Clearance: | K163116, January 6, 2017 |
| | Classification Regulation: | 21CFR 892.1000 |
| | Classification name: | Magnetic Resonance Diagnostic<br>Device (MRDD) |
| | Classification Panel: | Radiology |
| | Device class | Class II |
| | Product Code: | LNH, LNI |
| | Product Code: | LNH, LNI |
| Device Description: | 3D Amide Proton Transfer (3D APT) is an extension of off-resonance magnetization transfer imaging.<br><br>3D APT images display a spatial distribution of amide protons. The RF-shimmed saturation is generated by alternating excitation from the 2 channels of the MultiTransmit system.<br><br>3D APT images are obtained by subtracting the saturated image intensity at the 2 mirrored frequency offsets relative to the water frequency (±3.5 ppm). Parameters are optimized to null the difference signal for normal brain tissue. | |
| Indications for Use: | 3D Amide Proton Transfer weighted MRI (3D APT) is a software feature intended for use on Ingenia 3.0T (70cm) and Ingenia 3.0T CX (60cm) MR Systems. It's indicated for use in magnetic resonance imaging of the brain.<br><br>Amide Proton Transfer weighted imaging (3D APT) is a software option intended for use on Ingenia 3.0T and Ingenia 3.0T CX MR Systems. 3D APT is indicated for use in magnetic resonance imaging of the brain. 3D APT consists of an acquisition and reconstruction technique employing frequency-selective magnetization transfer effects to derive images reflecting the spatial distribution of amide protons, and thereby protein density. 3D APT images may assist a trained physician in diagnosis and therapy planning. APTW can be combined with multi-coil acceleration approaches (SENSE). | |
| Design Features and<br>Fundamental Scientific<br>Technology: | 3D Amide Proton Transfer (3D APT) is an MR imaging technology where saturation of the water signal using the magnetization transfer effect generates the contrast. Protons bound to macromolecules, specifically amide protons, exchange with the free water pool. Selective narrow-band excitation at the amide proton resonance frequency is used to detect signal differences exclusively associated with the amide protons. This is achieved by a controlled, frequency-selective RF irradiation to saturate protons resonating at +3.5 ppm (the amide proton resonance frequency), and comparing the resulting signal level to that observed when irradiating at -3.5 ppm.<br>The feature consists of:<br>Frequency selective saturation RF pulses (1-2 s duration), alternating between the two channels of the MultiTransmit RF system, at multiple offset frequencies 3D DIXON TSE readout Signal difference calculation between +3.5 ppm and -3.5 ppm to derive the APTw image B0 inhomogeneity correction embedded into the APTw calculation Color scale (rainbow) to display APTw signals between | |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters against a white background. The font is sans-serif and the letters are evenly spaced. The word is horizontally oriented and centered in the image.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and aligned horizontally. The font is sans-serif and appears to be a standard typeface.
| Substantial Equivalence | The proposed 3D Amide Proton Transfer (3D APT) is<br>substantially equivalent to the Magnetization Transfer<br>functionality (MTC and MTR) of the legally marketed primary<br>predicate device, Ingenia 1.5T, Ingenia 1.5T S and Ingenia<br>3.0T R5.3 (K163116, January 6, 2017) and has the same<br>primary indications for use with respect to the following:<br>• Software feature intended for use on the Ingenia 3.0T<br>(70cm) and Ingenia 3.0T CX (60cm) MR Systems<br>• Use in magnetic resonance imaging of the brain<br>• Consists of an image acquisition and reconstruction<br>technique with B0 correction and difference signal<br>determination |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The proposed 3D Amide Proton Transfer (3D APT) is<br>substantially equivalent to the legally marketed primary<br>predicate device, Ingenia 1.5T, Ingenia 1.5T S and Ingenia<br>3.0T R5.3 (K163116, January 6, 2017) in terms of design<br>features, fundamental scientific technology, indications for<br>use, and safety and effectiveness. Additionally, substantial<br>equivalence was demonstrated with non-clinical performance<br>(verification and validation) tests, which complied with the<br>requirements specified in the international and FDA-<br>recognized consensus standards and device-specific<br>guidance. The results of these tests demonstrate that the<br>proposed 3D Amide Proton Transfer (3D APT) meets the<br>acceptance criteria and is adequate for its intended use. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.