PMS (Mode 1) It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance. TENS (Mode 2) To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve.
Device Story
Physiotherapy instrument (models AST-2011A, AST-2011B, AST-802B) is an over-the-counter electronic stimulator. Device delivers electrical pulses to skin via two electrode pads connected to a main unit. User operates device via ON/OFF switch, mode selection, intensity keys, and time selection keys. LCD screen displays mode, intensity, and remaining time. Device operates in two modes: PMS for muscle conditioning and TENS for pain relief. Healthcare provider consultation recommended before use. Output is voltage-controlled, symmetric, biphasic pulsed current. Device provides temporary pain relief or muscle performance improvement, benefiting patients by allowing self-administered therapy for muscle strain or soreness.
Clinical Evidence
Bench testing only. Evaluated for electrical safety (IEC 60601-1, 60601-1-11, 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-5, 10993-10), usability (IEC 62366), and software validation. Waveform output verified against FDA guidance for TENS and muscle stimulators.
Indicated for healthy individuals to improve/facilitate muscle performance (PMS) and for temporary relief of pain associated with sore/aching muscles in shoulder, waist, back, neck, arm, leg, and foot due to exercise or household work (TENS).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out as "U.S. Food & Drug Administration".
November 26, 2018
Shenzhen OSTO Technology Co., Ltd. % Cecilia Ceng Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou. 510006 Cn
Re: K172837
Trade/Device Name: Low-frequency Multi-function physiotherapy instrument Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: October 17, 2018 Received: October 26, 2018
Dear Cecilia Ceng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Timothy A. Marjenin -S
Carlos L. Peña, PhD, MS For Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172837
Device Name
Physiotherapy instrument (Models: AST- 2011A, AST- 2011B, and AST-802B)
Indications for Use (Describe)
PMS (Mode 1)
It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance.
TENS (Mode 2)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Chapter 5. 510(k) Summary
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
510(k) Owner's Name: Shenzhen OSTO Technology Co., Ltd. Establishment Registration Number: Applying Address: No.43 Longfeng Road, Xinsheng Community, Longgang Street Longgang District, Shenzhen City Guangdong Province, CHINA Tel: +86-755-29769546 Fax: +86-755-29769540 Contact Person: Li Yang (General Manager) Email: annaosto@163.com
### Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 Email: regulatory@glomed-info.com
# 2. Subject Device Information
Type of 510(k): Traditional Trade/Device Name: Physiotherapy instrument, Models: AST-2011A, AST-2011B, AST-802B Common Name: Electronic Simulator Classification Name: 1. Stimulator. Nerve. Transcutaneous. Over-The-Counter (NUH) 2. Stimulator, Muscle, Powered, For Muscle Conditioning (NGX) Review Panel: Neurology Product Code: NUH, NGX Regulation Number: 21 CFR 882.5890, 890.5850 Regulatory Class: II
### 3. Predicate Device Information
Sponser: Shenzhen OSTO Technology Company Limited Trade Name: Health Expert Electronic Stimulator Common Name: Electronic Stimulator Classification name: Stimulator, Nerve, Transcutaneous, Muscle, Powered, For Muscle Conditioning, Over-The-Counter 510(k)number: K133929 Review Panel: Neurology, Physical Medicine Product Code: NUH, NGX Regulation Number: 882.5890, 890.5850
{4}------------------------------------------------
Requlation Class: II
# 4. Device Description
Physiotherapy instrument has 2 operation modes and one channel, after insert the electrode wire of the electrode pads, just put the electrode pads on the same side of your body, for shoulder, waist, back, back of the neck, arm, leg or foot (depends on your need, please consult your doctor before using if need), which can give certain electrical pulse through 2 pieces of electrode pads on the skin.
The electronic stimulatory module has the operating elements of ON/OFF Switch, Display screen, Mode Selection key, Intensity enhancement/weakening keys, language switch button and time selection keys. The LCD display screen can show selected mode, output intensity of body and/or sole, time remaining of an application mode, language display systemy.
The device is equipped with accessories of electrode pads, electrode wire, adapter.
The electrode wire is used to connect the pads to the main unit; the adapter wire is used to connect the adapter to the device.
# 5. Intended Use / Indications for Use
PMS (Mode 1)
It is intended to stimulate healthy muscles in order to improve and facilitate muscle performance. TENS (Mode 2)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, back of the neck, arm, leg, and foot due to strain from exercise or normal household work activities by applying current to stimulate nerve.
# 6. Test Summary
Physiotherapy instrument, Models: AST-2011B, AST-2011B, AST-802B has been evaluated the safety and performance by lab bench testing as following:
- Electrical safety test according to IEC 60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards �
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
- � Usability test according to IEC 62366 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- � The waveform test has also been conducted to verify the output specifications of the device according to "Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use and Guidance for Powered Muscle Stimulator for Muscle Conditioning".
# 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, mode of operation, and intended use of Electronic Muscle Stimulator is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
# Comparison in Detail(s):
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict | |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------------|
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict | |
| Company | Shenzhen OSTO<br>Technology Co., Ltd | Shenzhen OSTO Technology<br>Co., Ltd | -- | |
| Trade Name | Physiotherapy instrument | Health Expert Electronic<br>Stimulator | -- | |
| Model | AST-2011A, AST-2011B,<br>AST-802B | AST-300C, AST-300D | -- | |
| Classification<br>Name | Stimulator, Nerve,<br>Transcutaneous, Over-<br>The-Counter | Stimulator, Nerve,<br>Transcutaneous, Over-The-<br>Counter | -- | |
| 510(k) Number | Applying | K133929 | -- | |
| Product Code | NUH, NGX | NUH, NGX | SE | |
| Intended Use /<br>Indications for Use | PMS (Mode 1)<br>It is intended to stimulate<br>healthy muscles in order<br>to improve and facilitate<br>muscle performance.<br>TENS (Mode 2)<br>To be used for temporary<br>relief of pain associated<br>with sore and aching<br>muscles in the shoulder,<br>waist,<br>back, back of the neck,<br>arm, leg, and foot due to<br>strain from exercise or<br>normal household work<br>activities<br>by applying current to<br>stimulate nerve. | PMS (Mode 1~8)<br>It is intended to stimulate<br>healthy muscles in order to<br>improve and facilitate muscle<br>performance.<br>TENS (Mode 9~25)<br>To be used for temporary<br>relief of pain associated with<br>sore and aching muscles in<br>the shoulder, waist,<br>back, back of the neck, arm,<br>leg, and foot due to strain<br>from exercise or normal<br>household work activities<br>by applying current to<br>stimulate nerve. | SE | |
| Power Source | Adaptor Input: 100-<br>240Vac, 50-60Hz, 0.1A<br>Output: 5Vdc, 1A<br>Unit Input: 5Vdc, 1A | Adaptor Input: 100-240Vac,<br>50-60Hz, 0.1A<br>Output: 5Vdc, 1A<br>Unit Input: 5Vdc, 1A | SE | |
| -Method of Line<br>Current Isolation | Type BF Applied Part | Type BF Applied Part | SE | |
| Patient<br>Leakage<br>Current | Normal Condition:<br>AC: 54.5μA, DC: 0.5μA<br>Single Fault Condition:<br>AC:120.0μA, DC: 0.6μA | Normal Condition:<br>AC: 54.5μA, DC: 0.5μA<br>Single Fault Condition:<br>AC:120.0μA, DC: 0.6μA | SE | |
| Number of Output<br>Modes | 2 | 25 | SE<br>Note 2 | |
| Number of<br>Output Channels: | 1 | 2 | SE<br>Note 2 | |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict | |
| - Synchronous or<br>Alternating? | N/A | Synchronous | SE<br>Note 2 | |
| - Method of<br>Channel Isolation | N/A | Voltage Transform<br>Isolation<br>"BODY▼" and<br>"BODY▼" buttons for<br>body channel,<br>"SOLE▲" and<br>"SOLE▼" buttons for<br>feet channel | SE<br>Note 2 | |
| Regulated Current<br>or Regulated<br>Voltage? | Voltage Control | Voltage Control | SE | |
| Software/Firmwar<br>e/Microprocessor<br>Control? | Yes | Yes | SE | |
| Automatic<br>Overload Trip? | No | No | SE | |
| Automatic No-Load<br>Trip? | No | No | SE | |
| Automatic Shut<br>Off? | Yes | Yes | SE | |
| Patient Override<br>Control? | No | No | SE | |
| Indicat<br>or<br>Displa<br>y: | - On/Off<br>Status? | Yes | Yes | SE |
| | - Low<br>Battery? | No | No | SE |
| | Voltage/Current<br>Level? | Yes | Yes | SE |
| Timer Range<br>(minutes) | 5-15 min | 25 min | SE<br>Note 3 | |
| Compliance with<br>Voluntary<br>Standards? | No | No | SE | |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict | |
| Compliance* with<br>21 CFR 898? | Yes | Yes | SE | |
| Main Unit Weight | 105 g | 2 kg | SE<br>Note 1 | |
| Main Unit<br>Dimension | AST-802B: 156.6 x 74.5 x<br>17.8 mm<br>AST-2011A: 157.03 X<br>72.87 X 23.89 mm<br>AST-2011B: 156.6 x 74.5<br>x 17.8 mm | 428 x 428.8 x 185mm | SE<br>Note 1 | |
| Housing Materials<br>and Construction | Main unit: ABS plastic | Main unit: ABS plastic | SE | |
| Output Specification | | | | |
| Waveform | Pulsed, symmetric,<br>biphasic | Pulsed, symmetric,<br>biphasic | SE | |
| Shape | Rectangular, with<br>interphase interval | Rectangular, with<br>interphase interval | SE | |
| Maximum Output<br>Voltage | 44V±10% @ 500Ω<br>80V±10% @ 2KΩ<br>112V±10% @ 10KΩ | 44V±10% @ 500Ω<br>80V±10% @ 2KΩ<br>112V±10% @ 10KΩ | SE | |
| Maximum Output<br>Current | 88mA±10% @ 500Ω<br>40mA±10% @ 2KΩ<br>11.2mA±10% @ 10KΩ | 88mA±10% @ 500Ω<br>40mA±10% @ 2KΩ<br>11.2mA±10% @ 10KΩ | SE | |
| Pulse Duration | 120µs | 120µs | SE | |
| Pulse frequency | 77.3Hz | 77.3Hz | SE | |
| Symmetrical<br>phases? | Yes | Yes | SE | |
| Phase Duration | 12.94 ms | 12.94 ms | SE | |
| Net Charge (mC<br>per pulse) | 0µC @ 500Ω<br>Method: Balanced<br>waveform | 0µC @ 500Ω<br>Method: Balanced waveform | SE | |
| Maximum Phase<br>Charge, (mC) | 12.78 µC @ 500Ω | 12.78 µC @ 500Ω | SE | |
| Maximum Current<br>Density, (mA/cm²) | 0.235 mA/cm² @ 500Ω | 0.235 mA/cm² @ 500Ω | SE | |
| Maximum Power<br>Density, (W/cm²) | 1.38 mW/cm² @ 500Ω | 1.38 mW/cm² @ 500Ω | SE | |
| Stimulation<br>parameters | See attached table 1 | See attached table 1 | SE | |
| | Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| Bur<br>st<br>Mo<br>de | Pulses per<br>burst | 400 | 400 | SE |
| | Bursts per<br>second | 0.63 | 0.63 | SE |
| | Burst<br>duration<br>(seconds) | 2.588 | 2.588 | SE |
| | Duty Cycle<br>[Line (b) x<br>Line (c)] | 1.63 | 1.63 | SE |
| ON Time (seconds) | | 0.6s | 0.6s | SE |
| OFF Time<br>(seconds) | | 0.6s | 0.6s | SE |
| Additional Features | | | | |
| | Environment for<br>operation | Temperature: 5 ~ 45<br>Humidity: 15% - 90%<br>Atmospheric pressure<br>range of 700 hPa to 1 060<br>hPa | Temperature: 5 ~ 45° C<br>Humidity: 20 ~ 65% RH | SE<br>Note 4 |
| | Environment for<br>storage | Temperature: -25 - +70°C<br>Humidity: up to 90%<br>Atmospheric pressure<br>range of 700 hPa to 1 060<br>hPa | Temperature: 0 ~ 45° C,<br>Humidity: 10 ~ 90% RH<br>Electrode Pad: 10~20° C | SE<br>Note 4 |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| Table1 | | | | |
|-----------------------------------|------------------------------|------|------------------------------|-------|
| Output Specification | | | | |
| Pulse Duration and Pulse<br>Cycle | | | | |
| | Note: "T" means Pulse Cycle. | | Note: "T" means Pulse Cycle. | |
| Cycle of<br>Mode 1 | | N/A | | |
| | | | Max | Min |
| Intensity | | | | |
| Peak Voltage (V) (0.5KΩ) | | | 44 V | 0.63 |
| Peak Voltage (V) (2KΩ) | | | 80 V | 2.48 |
| Peak Voltage (V) (10K Ω) | | | 112 V | 11.80 |
| Cycle of<br>Mode 2 | 200T | | 200T | |
| Intensity | Max | Min | Max | Min |
| Peak Voltage (V) (0.5KΩ) | 44 V | 0.63 | 44 V | 0.63 |
| Peak Voltage (V) (2KΩ) | 80 V | 2.48 | 80 V | 2.48 |
{10}------------------------------------------------
| Peak Voltage (V) (10K Ω) | 112 V | 11.80 | 112 V | 11.80 |
|--------------------------|------------------|-------------------------------------------------------|---------------------------------------------|-------|
| Cycle of<br>Mode 3 | | | Image: Sub-mode 5 X 1 1s Sub-mode 6 X 2 | |
| Intensity | | N/A | Max | Min |
| Peak Voltage (V) (0.5KΩ) | | | 44 V | 0.63 |
| Peak Voltage (V) (2KΩ) | | | 80 V | 2.48 |
| Peak Voltage (V) (10K Ω) | | | 112 V | 11.80 |
| Cycle of<br>Mode 4 | | | Image: Sub-mode 6 X 1 0.5s Sub-mode 17 X 10 | |
| Intensity | | N/A | Max | Min |
| Peak Voltage (V) (0.5KΩ) | | | 44 V | 0.63 |
| Peak Voltage (V) (2KΩ) | | | 80 V | 2.48 |
| Peak Voltage (V) (10K Ω) | | | 112 V | 11.80 |
| Cycle of<br>Mode 5 | | | Image: Sub-mode 9 X 1 1s Sub-mode 15 X 10 | |
| Intensity | | N/A | Max | Min |
| Peak Voltage (V) (0.5KΩ) | | | 44 V | 0.63 |
| Peak Voltage (V) (2KΩ) | | | 80 V | 2.48 |
| Cycle of<br>Mode 6 | N/A | Image: Sub-mode 15 X 1 and Sub-mode 16 X 1 with 0.5s | | |
| Intensity | N/A | Max | Min | |
| Peak Voltage (V) (0.5KΩ) | N/A | 44 V | 0.63 | |
| Peak Voltage (V) (2KΩ) | N/A | 80 V | 2.48 | |
| Peak Voltage (V) (10K Ω) | N/A | 112 V | 11.80 | |
| Cycle of<br>Mode 7 | N/A | Image: Sub-mode 5 X 1 and Sub-mode 6 X 1 with 1s | | |
| Intensity | N/A | Max | Min | |
| Peak Voltage (V) (0.5KΩ) | N/A | 44 V | 0.63 | |
| Peak Voltage (V) (2KΩ) | N/A | 80 V | 2.48 | |
| Peak Voltage (V) (10K Ω) | N/A | 112 V | 11.80 | |
| Cycle of<br>Mode 8 | N/A | Image: Sub-mode 18 X 5 and Sub-mode 17 X 5 with 0.05s | | |
| Intensity | N/A | Max | Min | |
| Peak Voltage (V) (0.5KΩ) | N/A | 44 V | 0.63 | |
| Peak Voltage (V) (2KΩ) | N/A | 80 V | 2.48 | |
| Peak Voltage (V) (10K Ω) | N/A | 112 V | 11.80 | |
| Cycle of<br>Mode 9 | N/A | Image: Sub-mode 11 X 5 and Sub-mode 15 X 5 with 0.3s | | |
| Intensity | | Max | Min | |
| Peak Voltage (V) (0.5KΩ) | N/A | 44 V | 0.63 | |
| Peak Voltage (V) (2KΩ) | | 80 V | 2.48 | |
| Peak Voltage (V) (10K Ω) | | 112 V | 11.80 | |
| Cycle of<br>Mode 10 | | Image: Sub-mode 8 X 1, Sub-mode 14 X 10, 0.1s | | |
| Intensity | | Max | Min | |
| Peak Voltage (V) (0.5KΩ) | N/A | 44 V | 0.63 | |
| Peak Voltage (V) (2KΩ) | | 80 V | 2.48 | |
| Peak Voltage (V) (10K Ω) | | 112 V | 11.80 | |
| Cycle of<br>Mode 11 | | Image: Sub-mode 5 X 5, Sub-mode 11 X 5, 1s | | |
| Intensity | | Max | Min | |
| Peak Voltage (V) (0.5KΩ) | N/A | 44 V | 0.63 | |
| Peak Voltage (V) (2KΩ) | | 80 V | 2.48 | |
| Peak Voltage (V) (10K Ω) | | 112 V | 11.80 | |
| Cycle of<br>Mode 12 | | Image: Sub-mode 2 X 1, Sub-mode 6 X 20, 0.5s | | |
| Intensity | | Max | Min | |
| Peak Voltage (V) (0.5KΩ) | | 44 V | 0.63 | |
| | | | | |
| Peak Voltage (V) (2KΩ) | N/A | 80 V | 2.48 | |
| Peak Voltage (V) (10K Ω) | | 112 V | 11.80 | |
| Cycle of<br>Mode 13 | | Image: Sub-mode 11 X 1 Sub-mode 12 X 1 0.3s | | |
| Intensity | | Max…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.