4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB, 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA

K172817 · Olympus Medical Systems Corp. · FET · Oct 17, 2017 · Gastroenterology, Urology

Device Facts

Record IDK172817
Device Name4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB, 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA
ApplicantOlympus Medical Systems Corp.
Product CodeFET · Gastroenterology, Urology
Decision DateOct 17, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The camera head has been designed to be used with Olympus endoscopes, camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

Device Story

Camera heads capture optical images from endoscopes via an integrated coupler; CMOS sensor converts optical images into electronic signals; signals transformed into laser via laser diodes; transmitted via optical fiber to VISERA 4K UHD Camera Control Unit for processing. Used in clinical settings by physicians for endoscopic diagnosis, treatment, and observation. Output displayed to healthcare providers to facilitate clinical decision-making during endoscopic procedures. Benefits include high-resolution 4K visualization for improved diagnostic and therapeutic accuracy.

Clinical Evidence

Bench testing only. Design verification tests conducted in accordance with ISO 14971:2007 and 21 CFR 820.30.

Technological Characteristics

Components: endoscopic coupler, camera cable, video connector. Sensing: CMOS image sensor. Energy: laser diodes for signal transmission via optical fiber. Connectivity: connects to VISERA 4K UHD Camera Control Unit. Sterilization: supports user-performed sterilization methods.

Indications for Use

Indicated for use with Olympus endoscopes, camera control unit, and ancillary equipment for endoscopic diagnosis, treatment, and observation in patients undergoing endoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized representation of a human figure, while the FDA acronym and full name are written in blue. October 17, 2017 Olympus Medical Systems Corp. % Sheri L. Musgnung Regulatory Affairs Manager Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610 Re: K172817 Trade/Device Name: 1) 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB 2) 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FET, NWB Dated: September 15, 2017 Received: September 18, 2017 Dear Sheri L. Musgnung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Charles Viviano -S For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172817 Device Name 1) 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB 2) 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA Indications for Use (Describe) The camera head has been designed to be used with Olympus endoscopes, camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. | Type of Use ( <i>Select one or both, as applicable</i> ) | |----------------------------------------------------------| |----------------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Olympus. The logo is in blue and features the company name in a bold, sans-serif font. A yellow line is underneath the company name. Below the line is the company slogan, "Your Vision, Our Future." ## 510(k) SUMMARY ## 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA ## September 15, 2017 ## I. General Information Olympus Medical Systems Corp. Applicant: . 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan Establishment Registration No: 8010047 Sheri L. Musgnung, . Official Correspondent: Regulatory Affairs Manager Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 FAX: 484-896-7128 Shirakawa Olympus Co., Ltd. . Manufacturer: 3-1 Okamiyama, Odakura, Nishigo-mura, Nishirakawa-gun, Fukushima 961-8061, Japan Registration Number: 3002808148 #### Device Identification II. | ■ | Device Trade Name: | 1) 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB<br>2) 4K AUTOCLAVABLE CAMERA HEAD<br>OLYMPUS CH-S400-XZ-EA | |---|-----------------------|----------------------------------------------------------------------------------------------------| | ■ | Common Name: | Camera Heads | | ■ | Regulation Number: | 876.1500 | | ■ | Regulation Name: | Endoscope and Accessories | | ■ | Regulatory Class: | II | | ■ | Classification Panel: | Gastroenterology/Urology | Section 6 510(k) Summary Page 2 of 4 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, horizontal, yellow line. Below the line is the text "Your Vision, Our Future" in a smaller, light-blue font. K172817 FET: Endoscopic video imaging system/component, 1 Product Code: gastroenterology-urology NWB: Endoscope, Accessories, Narrow Band Spectrum #### Predicate Device Information III. | Predicate Device | Predicate Device<br>510(k) No. | |----------------------------------------------------------------------------------|-----------------------------------| | 4K CAMERA HEAD<br>OLYMPUS CH-S400-XZ-EB<br>(as cleared via VISERA 4K UHD System) | K151011<br>(VISERA 4K UHD SYSTEM) | #### Device Description IV. Olympus intends to introduce two new models of 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB and 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA to mainly add sterilization methods to the predicate device 4K CAMERA HEAD which has been cleared as part of VISERA 4K UHD System (K151011). Both 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB and 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA are structurally consisted of endoscopic coupler end, camera cable and video connector end. The 4K Camera Heads function as receiving optical image from a variety of endoscopes that are attached to it via coupler and using incorporated complementary metal oxide semiconductor (CMOS) image sensor to convert optical image into electronic signal. The signal is subsequently transformed into laser by laser diodes and input to VISERA 4K UHD CAMERA CONTROL UNIT by optical fiber for further processing with other ancillary equipment on endoscopic diagnosis, treatment, and observation. #### Indications for Use V. ## 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB The camera head has been designed to be used with Olympus endoscopes, camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Below the line are the words "Your Vision, Our Future" in a smaller, blue font. K172817 observation. ## 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA The camera head has been designed to be used with Olympus endoscopes, camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. ## VI. Comparison of Technological Characteristics The subject devices, 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB and 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA, have the same fundamental scientific technology; indications for use; principal of operation; and energy source as the legally marketed 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB cleared via the VISERA 4K UHD SYSTEM in K151011. The modifications between subject devices and predicate device are indicated as follows: 1) Addition of sterilization methods for reprocessing the device by user - 2) Minor changes on material, dimension, appearance, weight specification ## VII. Summary of non-clinical testing The design verification tests were conducted and validated to be in accordance with ISO 14971:2007, and the design control procedure complies with requirements as specified in 21 CFR 820.30. ## VIII. Conclusion 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB and 4K AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S400-XZ-EA have the same indications for use; fundamental scientific technology; principal of operation; and energy source. The addition of sterilization methods as well as the other minor modifications have been validated and raised no new issues of safety or effectiveness as compared to the predicate device. Therefore, both of the subject devices are considered to be substantially equivalent to their predicate 4K CAMERA HEAD in VISERA 4K UHD system.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...