Agfa's DX-D Imaging Package is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The DX-D Imaging Package may be used wherever conventional screen-film systems may be used. Agfa's DX-D Imaging Package is not indicated for use in mammography.
Device Story
Solid state direct radiography (DR) system; captures general radiographic images of human anatomy. Comprises NX workstation and flat-panel detectors (DR 14e, DR 17e). Detectors use Cesium Iodide (CsI) or Gadolinium Oxysulfide (GOS) scintillators to convert X-rays to raw digital data; data transmitted via wired Ethernet or wireless to NX workstation. Workstation performs image processing (MUSICA), QC checks, and display. Used in clinical settings; operated by radiographers/technicians. System synchronizes detector with X-ray generator for exposure. Benefits include digital workflow, image manipulation, and integration with existing X-ray systems; replaces conventional film/screen or CR systems.
Clinical Evidence
Bench testing only. No clinical or animal studies performed. Image quality validation conducted using anthropomorphic phantoms evaluated by independent radiographers; results indicate performance equal to or better than predicate detectors. Usability, functionality, and grid evaluations confirmed performance within acceptance criteria.
Technological Characteristics
Solid state flat-panel detectors (CsI or GOS scintillator); 150 μm pixel size; 16-bit A/D conversion. Wireless/wired communication. NX workstation (Windows 7/10). Standards: IEC 60601-1, IEC 60601-1-2, ISO 14971, ISO 13485, ISO 62366, ISO 62304.
Indications for Use
Indicated for general projection radiographic applications to capture diagnostic quality images of human anatomy. May be used wherever conventional screen-film systems are used. Not indicated for mammography, fluoroscopy, angiography, or tomography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
October 13, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Agfa HealthCare N.V. % Ms. ShaeAnn Cavanagh Regulatory Affairs Manager NA AGFA Healthcare 10 South Academy Street GREENVILLE SC 29601
Re: K172784
Trade/Device Name: DX-D Imaging Package Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: September 14, 2017 Received: September 15, 2017
Dear Ms. Cavanagh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 4. Indications for Use Statement
{3}------------------------------------------------
Page 4 – 2
{4}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--|
| Indications for Use | | | |
| 510(k) Number (if known) | K172784 | | |
| Device Name | DX-D Imaging Package | | |
| Indications for Use (Describe) | Agfa's DX-D Imaging Package is indicated for use in general projection radiographic applications to capture for display | | |
Agfa's DX-D Imaging Package is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The DX-D Imaging Package may be used wherever conventional screen-film systems may be be used.
Agfa's DX-D Imaging Package is not indicated for use in mammography.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
{6}------------------------------------------------
# 5.510(K) Summary
{7}------------------------------------------------
{8}------------------------------------------------
# 510(K) SUMMARY Agfa's DX-D Imaging Package - DR 14e & DR 17e Detectors
#### SUBMITTER I.
Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium
Contact: Wim Govaerts, Prepared: September 14, 2017 Telephone: +32-34444-6246
#### II. DEVICE
Name of Device: DX-D Imaging Package Classification Name: Stationary X-Ray System Regulatory Classification: Class II, 21 CFR 892.1680 Product Code: MQB
#### III. PREDICATE DEVICES
This is a 510(k) for Agfa's DX-D Imaging Package, a solid state, flat panel x-ray imaging device. It is substantially equivalent to systems with Agfa's DX-D Imaging Package (K161368) and Innolux Corporation's Innolux RIC device (K162344).
#### DEVICE DESCRIPTION IV.
Agfa's DX-D Imaging Package is a solid state flat panel x-ray system, a direct radiography (DR) system (product code MOB) intended to capture general radiography images of the human body. It is a combination of Agfa's NX workstation and one or more flat-panel detectors.
This submission is to add the DR14e and DR17e Flat Panel Detectors to Agfa's DX-D Imaging Package portfolio. Agfa's DR 14e and DR 17e wireless panels are currently marketed by Innolux as RIC 35C/G and RIC 43C/G, which is one of the predicates for this submission.
Principles of operation and technological characteristics of the new and predicate devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically identical to both predicates, K161368 and K162344. It uses the same workstation as predicate K161368 and the same flat panel detectors to capture and digitize the images as predicate K162344.
Laboratory data and image quality evaluations conducted with independent radiologists confirm that performance is equivalent to the predicate.
{9}------------------------------------------------
| Performance<br>Characteristics | DX-D 10 Flat-<br>Panel Detector | DX-D 20 Flat-<br>Panel Detector<br>(Handle) | DR 14e Wireless<br>Detector | DR 17e Wireless<br>Detector |
|--------------------------------|------------------------------------------------|------------------------------------------------|----------------------------------------|----------------------------------------|
| Scintillator | CsI, GOS | CsI, GOS | CsI, GOS | CsI, GOS |
| Cassette size | 35x43cm/14x17in | 35x43cm/14x17in | 35x43cm/14x17in | 43x43cm/17x17in |
| Pixel Size | 139 μm | 139 μm | 150 μm | 150 μm |
| A/D Conversion | 14 bits | 14 bits | 16 bits | 16 bits |
| Interface to<br>Generator | Ethernet | Ethernet | AED &<br>Synchronized | AED &<br>Synchronized |
| Communication | Tethered | Tethered | Wireless | Wireless |
| Power | I/O Interface<br>Box: 100-240<br>VAC, 47-63 Hz | I/O Interface Box:<br>100-240 VAC,<br>47-63 Hz | Battery: replaceable<br>& rechargeable | Battery: replaceable<br>& rechargeable |
| Weight | 3.9 kg (8.6 lbs) | 4.9 kg (10.8 lbs) | 2.95 kg (6.50 lbs) | 3.65 kg (8.04 lbs) |
Configuration information can be found in the DR 17e User Manuals. The DR 14e and DR 17e detectors can be integrated in an X-ray system that communicates to a workstation. The DR 14e and DR 17e Service Manual details the possible configurations and integrations with the NX workstation and X-ray generator. All of Agfa HealthCare's DR systems (i.e. DX-D 100-K103597, DX-D 300-K103050, DX-D 600-K112670, DR 400-K141192, and DR 600-K152639) will integrate with the new device. The NX9000 Service Manual, Chapter 4 and associated appendices addresses the installation and configuration with other system components.
#### V. INDICATIONS FOR USE
Agfa's DX-D Imaging Package is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The DX-D Imaging Package may be used wherever conventional screen-film systems may be used.
Agfa's DX-D Imaging Package is not indicated for use in mammography.
Intended use has not changed as a result of any labeling modification(s).
#### VI. COMPARISON OF TEDCHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICES
The new device and the both predicate devices (K162344) are solid state imaging devices. Product Code MOB. Agfa's DX-D Imaging Package is substantially equivalent to both predicate devices (K161368 and K162344) in that it uses precisely the same technology to capture and transmit images.
Principles of operation and technological characteristics of the new and predicate devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically identical to both predicates, K161368 and K162344. It uses the same workstation as predicate K161368 and the same flat panel detectors to capture and digitize the images as predicate K162344. Differences in devices do not alter the intended diagnostic effect.
Performance data, image quality clinical evaluations, grid test, and usability/functionality data are adequate to ensure equivalence.
{10}------------------------------------------------
The DX-D Imaging Package indications for use statement is substantially equivalent to both predicate devices (K161368 and K162344). The new device and both predicate devices (K161368 and K162344) include the statement that the device is not indicated for mammography. The DX-D Imaging Package and predicate device (K161368) describe the imaging applications may be used utilized where screen-film systems exist. The new device and predicate device (K162344) have similar indications for use statements in that both devices are intended to display general radiographic images of the human body and can be used in place of conventional film/screen or CR systems. Intended uses are the same. The devices have the same technological characteristics.
The only difference between the new device and predicate device (K162344) is that the predicate device's Indication For Use statement includes specific contraindications for fluoroscopy. angiography, and tomography. The specific contraindications are applicable to the new device since it is only indicated for general projection applications. Fluoroscopy, angiography and tomography are not considered to be general projection applications. There are no differences between the new device and predicates K161368 and K162344 that impact safety and effectiveness.
The table on the next page compares these technological characteristics.
{11}------------------------------------------------
| PRODUCT COMPARISON TABLE | | | |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DX-D Imaging Package<br>(New Device) | AGFA DX-D Imaging<br>Package<br>(PREDICATE-K161368) | Innolux RIC<br>(PREDICATE - K162344) |
| Communications | Same as both Predicates | Wireless | Wireless/Wired |
| Flat Panel | Same as both Predicates | Flat Panel Detector | Flat Panel Detector |
| Detector Material | Same as both Predicates | Gadolinium Oxysulfide<br>(GOS) or Cesium Iodide<br>(CsI) scintillator | Gadolinium Oxysulfide<br>(GOS) or Cesium Iodide<br>(CsI) scintillator |
| Detector Sizes | Same as Predicate K162344 | 10x12 in.<br>14x17 in. | 14x17 in<br>17x17in |
| Active Matrix | Same as Predicate K162344 | 2400 x 2880 | 2336x2836<br>2832x2836 |
| Pixel size | Same a Predicate K162344 | 148 µm | 150 µm |
| Dynamic Range | Same as both Predicates | 16 bit | 16 bit |
| Maximum Image<br>Acquisitions/hr. | Same as Predicate K161368 | 240 | -- |
| Power Supply | Same as Predicate K162344 | +12 V 1A DC Battery | 6-12V 2.73A DC Battery |
| Operator Workstation | Same as Predicate K161368 | Agfa NX | N/A, workstation is not<br>included in the system |
| Image processing | Same as Predicate K161368 | MUSICA | 3rd Party Software |
| Operating system | Windows 7 & 10 | Windows 7 | Standard computer operating<br>system |
| Display System | Same as Predicate K161368 | Separately cleared medical<br>display (K051901) | Separately cleared medical<br>display |
| Indications for Use | Agfa's DX-D Imaging<br>Package is indicated for use<br>in general projection<br>radiographic applications to<br>capture for display<br>diagnostic quality<br>radiographic images of<br>human anatomy. The DX-D<br>Imaging Package may be<br>used wherever conventional<br>screen-film systems may be<br>used.<br>Agfa's DX-D Imaging<br>Package is not indicated for<br>use in mammography. | Agfa's DX-D Imaging<br>Package is indicated for use<br>in general projection<br>radiographic applications to<br>capture for display diagnostic<br>quality radiographic images<br>of human anatomy. The DX-<br>D Imaging Package may be<br>used wherever conventional<br>screen-film systems may be<br>used.<br>Agfa's DX-D Imaging<br>Package is not indicated for<br>use in mammography. | The INNOLUX RIC is<br>intended to capture for<br>display radiographic images<br>of human anatomy.<br>It is intended for use in<br>general projection<br>radiographic applications<br>including pediatric and<br>neonatal exams wherever<br>conventional film/screen or<br>CR systems may be used.<br>RIC is not intended for<br>mammography, fluoroscopy,<br>tomography, and<br>angiography applications. |
{12}------------------------------------------------
#### VII. PERFORMANCE DATA
Laboratory testing and software testing (for a moderate level of concern device) using equivalent test protocols as used for the cleared detectors were evaluated by qualified individuals employed by the sponsor to demonstrate that adequate design controls (according to 21 CFR 820.30) were in place.
Performance data, image quality clinical evaluations, grid test, and usability/functionality data have been provided.
In-hospital image quality comparisons have been conducted with qualified independent radiologists as well.
Where patient images were utilized, they were first anonymized to remove all identifying patient information. No animal or clinical studies were performed in the development of the new device. No patient treatment was provided or withheld.
#### Bench Testing
Image quality clinical evaluations, performance/functionality data, usability data, and grid evaluation have been provided. No patient treatment was provided or withheld. No clinical or animal testing was performed in the development of the DX-D Imaging Package.
- . Image Quality Validation testing was conducted using anthropomorphic phantoms and evaluated by qualified independent radiographers. The test results indicated that the DR 14e, and DR17e flat-panel detectors have at least the same if not better image quality than other flat-panel detectors currently on the market (DX-D 10 and DX-D 20).
- . Usability and functionality evaluations were conducted with a qualified internal radiographer. The results of these tests fell within the acceptance criteria for all of the flatpanel detectors; therefore, the DX-D Imaging Package supports a radiographic workflow including calibration, compatibility, linear dose and dynamic ranges.
- . Grid Evaluation was conducted with a qualified internal radiographer. The results of the grid evaluation of the chest, skull, and pelvis remained consistent with other Agfa HealthCare flat-panel detectors currently on the market including the predicate (K161368). The intended use is fulfilled using different flat-panel detectors.
#### Software Validation Testing
Verification and validation plans comprise of test protocols. The software components of the device have been certified and validated. During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with conventional x-ray film previously released to the field.
The software components for the DR 17e comprise of the XRDi18 and NX9000 workstation. There are no risks identified in the Broadly Acceptable Region and no identified residual risks in the ALARP region after mitigation for XRDi18. Zero risks were identified in the Not Acceptable Region. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk.
There are no identified residual risks for the NX9000 in the ALARP region after mitigation. Only
{13}------------------------------------------------
three risks were identified in the Broadly Acceptable Region after mitigation. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk.
The term "Level of Concern" means the level of risk that the software device is determined to be if the software were to fail. The Level of Concern for the NX 9000 device has been determined to be moderate and the Level of Concern for the XRDi18 software has been determined to be minor.
The product, manufacturing and development processes conform to product safety and medical imaging standards including:
### PRODUCT STANDARDS
- IEC 60601-1: 2012 Medical Electrical Equipment: General Requirements for Safety and ● Essential Performance.
- IEC 60601-1-2: 2014-4 Medical Electrical Equipment - Part 1-2: General Requirements for Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests.
### QUALITY MANAGEMENT STANDARDS
- ISO 14971:2007 Application of Risk Management to Medical Devices ●
- ISO 13485:2003 Medical Devices Quality Management Systems Requirements For ● Regulatory purposes
- ISO 62366:2007 Application of Usability Engineering to Medical Devices ●
- ISO 62304:2006 Medical Device Software Software Lifecycle Processes ●
#### Summarv
Based on the performance data as documented in the above testing. DX-D Imaging Package is found to have a safety and effectiveness profile that is similar to the predicate devices.
#### VIII. CONCLUSIONS
Agfa's DX-D Imaging Package has an Indications For Use statement identical to the predicate devices (K161368 and K162344). Intended uses are virtually the same. The devices have the same technological characteristics.
The new device and the both predicate devices (K162344) are solid state imaging devices. Product Code MOB. Agfa's DX-D Imaging Package is substantially equivalent to both predicate devices (K161368 and K162344) in that it uses precisely the same technology to capture and transmit images.
The DX-D Imaging Package indications for use statement is substantially equivalent to both predicate devices (K161368 and K162344). The new device and both predicate devices (K161368 and K162344) include the statement that the device is not indicated for mammography. The DX-D Imaging Package and predicate device (K161368) describe the imaging applications may be used
{14}------------------------------------------------
utilized where screen-film systems exist. The new device and predicate device (K162344) have similar indications for use statements in that both devices are intended to display general radiographic images of the human body and can be used in place of conventional film/screen or CR systems. Intended uses are the same. The devices have the same technological characteristics.
There are no differences between the device and predicates K161368 and K162344 that impact safety and effectiveness.
Performance data, image quality clinical evaluations, grid test, and usability/functionality data are adequate to ensure equivalence.
This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
{15}------------------------------------------------
{16}------------------------------------------------
# 10. Executive Summary
# Reason for Submission
Agfa HealthCare is a manufacturer of a full line of analog and digital medical imaging devices. In 1990 it introduced its first computed radiography2 system, the ADC 70. That device and each of its successor models used phosphor coated image plates to capture x-ray images, releasing the images upon exposure of the plates to a laser scanner (a digitizer). With these devices, images are previewed and manipulated on its NX computer workstation, and then sent to a printer or PACS system for use by the physician.
In 2009, Agfa introduced its first direct radiography3 system, DX-D Imaging Package (K092669), which did not require a digitizer but captured images with scintillator flat panel detectors of Cesium Iodide or Gadolinium Oxysulfide manufactured for Agfa by Varian. Since then, Agfa has integrated its imaging package with traditional x-ray systems; for example, it's DX-D 100 (K103597), DX-D 300 (K103050) and DX-D 600 (K112670).
Agfa recently received premarket clearance for the DX-D40C/G Flat Panel Detector in the DX-D Imaging Package portfolio (K142184) in October 2014.
Agfa submitted a premarket notification to include the DR 10s and DR 14s Flat Panel Detectors in the DX-D Imaging Package portfolio (K161368) and received clearance in September 2016, which is one of the predicates for this submission.
Agfa is submitting this premarket notification because it plans to include the DR 14e and DR 17e Flat Panel Detectors in the DX-D Imaging Package portfolio. Agfa's DR 14e and DR 17e wireless panels are currently marketed by Innolux as RIC 35C/G and RIC 43C/G, which is one of the predicates for this submission.
Agfa's DX-D Imaging Package is a solid state flat panel x-ray system, a direct radiography (DR) system (product code MQB) intended to capture general radiographic images of the human body. It is a combination of Agfa's NX workstation and one or more flat-panel detectors.
Principles of operation and technological characteristics of the new and predicate devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically identical to both predicates, K161368 and K162344. It uses the same workstation as predicate K161368 and the same flat panel detectors to capture and digitize the images as predicate K162344. Differences in devices do not alter the intended diagnostic effect.
Agfa has tested the device to confirm it meets specifications and operates as planned. Performance data, image quality clinical evaluations, grid test, and usability/functionality data are adequate to ensure equivalence.
NOTE: The DX-D Imaging Package and its components are referred to by a number of names in
<sup>2</sup> Systems utilizing phosphor image plates are referred to as "computed radiography" (or CR) systems.
<sup>3</sup> Systems using photostimulable flat panel detectors are often referred to as "direct" on "digital radiography" (or DR) systems.
{17}------------------------------------------------
this submission and its exhibits: DX-D, DR, DR 14e, DR 17e. As used here, all these terms refer to the same device.
# Required Information: (510(K) Summary, Integrity Statement, and Indications for Use)
The statement of Indications for Use is included as Section 4.
In accordance with 21 CFR 807.87(h), Agfa HealthCare NV has prepared a 510(k) Summary which is included as Section 5.
Also required by 21 CFR 807.87(k) is a statement certifying that the information contained in a 510(k) notification is truthful and accurate. This statement is included as Section 6.
# Relevant Guidance
Agfa is aware of FDA guidance documents relevant to this submission. The influence of the following FDA documents is discussed briefly below:
- . Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (August 1999). The device is of the scintillator-photodetector type described in this guidance. The guidance requests certain imaging performance data for the imaging plate/digitizer system and flat-panel detectors. The detector used in the new device is identical to the detector in predicate (K161368). Test data is provided in Section 18 and in the referenced Exhibits 5 & 7.
- Guidance for the Content of Premarket Submissions for Software Contained in Medical ● Devices (May 2005). This guidance requests manufacturers to assess the software "Level of Concern" and to provide certain information about the software. The software information presented in this submission is for the latest version of the NX workstation (NX9000). The software information is included in this submission 16, and in the referenced exhibits.
- Off-The-Shelf software Use in Medical Devices (September 1999). FDA guidance for Off-● the-Shelf Software asks device manufacturers consider carefully the risks imposed by use of off-the-shelf software and to provide certain basic information for the OTS software used in their devices. OTS software is included in the NX workstation. This information is provided in Section 16.
- Cybersecurity of Networked Medical Devices Using Off-the-Shelf Software (January ● 2005). FDA guidance raises concerns about cybersecurity threats of networked medical devices. Agfa addresses those concerns during development, manufacturing, installation and maintenance activities. OTS software is included in the NX workstation. Information is provided in Section 16.
{18}------------------------------------------------
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices . (October 2014). This guidance requires manufacturers to identify assets, threats, and vulnerabilities, to assess their likelihood and impact on device functionality and end users/patients, to determine risk levels and suitable mitigation strategies and to assess residual risk and risk acceptance criteria. Section 16 provides information on how Agfa addressed item. functional and system level security including cyber-security for NX Workstation and DR 14e and DR 17e detectors.
# Predicate Devices
The predicate devices are:
- 1. K161368 Agfa's latest submission for the DX-D Imaging Package, and
- 2. K162344 Innolux Corporation's Innolux RIC
The new device and the both predicate devices (K162344) are solid state imaging devices, Product Code MQB. Agfa's DX-D Imaging Package is substantially equivalent to both predicate devices (K161368 and K162344) in that it uses precisely the same technology to capture and transmit images.
The DX-D Imaging Package indications for use statement is substantially equivalent to both predicate devices (K161368 and K162344). The new device and both predicate devices (K161368 and K162344) include the statement that the device is not indicated for mammography. The DX-D Imaging Package and predicate device (K161368) describe the imaging applications may be used utilized where screen-film systems exist. The new device and predicate device (K162344) have similar indications for use statements in that both devices are intended to display general radiographic images of the human body and can be used in place of conventional film/screen or CR systems. Intended uses are the same. The devices have the same technological characteristics.
The only difference between the new device and predicate device (K162344) is that the predicate device's Indication For Use statement includes specific contraindications for fluoroscopy, angiography, and tomography. The specific contraindications are applicable to the new device since it is only indicated for general projection applications. Fluoroscopy, angiography and tomography are not considered to be general projection applications. There are no differences between the new device and predicates K161368 and K162344 that impact safety and effectiveness.
Principles of operation and technological characteristics of the new and predicate devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically identical to both predicates, K161368 and K162344. It uses the same workstation as predicate K161368 and the same flat panel detectors to capture and digitize the images as predicate K162344. Differences in devices do not alter the intended diagnostic effect.
Laboratory data and image quality evaluations conducted with independent radiologists confirm that performance is equivalent to the predicates.
{19}------------------------------------------------
# Testing Overview
Laboratory testing and software testing (for a moderate level of concern device) using equivalent test protocols as used for the cleared detectors were evaluated by qualified individuals employed by the sponsor to demonstrate that adequate design controls (according to 21 CFR 820.30) were in place.
Performance data, image quality clinical evaluations, grid test, and usability/functionality data are adequate to ensure equivalence.
Where patient images were utilized, they were first anonymized to remove all identifying patient information. No animal or clinical studies were performed in the development of the new device. No patient treatment was provided or withheld.
- . Image Quality Validation testing was conducted using anthropomorphic phantoms and evaluated by qualified independent radiographers. The test results indicated that the DR 14e, and DR17e flat-panel detectors have at least the same if not better image quality than other flat-panel detectors currently on the market (DX-D 10 and DX-D 20). Refer to Exhibit 7 for the Image Quality Validation plan, report and completed scoresheets. Refer to Exhibit 6 for Sample Images used in the image quality evaluations. Section 18 contains the complete summary.
MISC Files: Raw images are provided electronically on the eCopy CD-ROM.
- Usability and functionality evaluations were conducted with a qualified internal radiographer. The results of these tests fell within the acceptance criteria for all of the flatpanel detectors; therefore, the DX-D Imaging Package supports a radiographic workflow including calibration, compatibility, linear dose and dynamic ranges. Refer to Exhibit 5-2, Design Verification Report, Exhibit 5-4, Design Validation Report and Exhibit 5-7, User Observation Sheet Report and Section 18 for the results and referenced exhibits.
- . Grid Evaluation was conducted with a qualified internal radiographer. The results of the grid evaluation of the chest, skull, and pelvis remained consistent with other Agfa HealthCare flat-panel detectors currently on the market including the predicate (K161368). The intended use is fulfilled using different flat-panel detectors. Refer to Exhibit 5-8, Grid Evaluation on IQ and Section 18 for the complete results.
# Device Description Overview
Section 11 of this submission, Device Description, contains a detailed description of Agfa's
DX-D Imaging Package. This section is intended as a brief introduction.
Agfa's DX-D Imaging Package is a solid state flat panel x-ray system, a direct radiography (DR) system (product code MQB) intended to capture genral radiographic images of the human body. It is a combination of Agfa's NX workstation and one or more flat-panel detectors. It is integrated with compatible x-ray systems such as the DX-D 100 (K103597), DX-D 300 (K103050), DX-D 600 (K112670). DR 400 (K141192). and DR 600 (K152639).
{20}------------------------------------------------
Principles of operation and technological characteristics of the new and predicate devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically identical to both predicates. K161368 and K162344. It uses the same workstation as predicate K161368 and the same flat panel detectors to capture and digitize the images as predicate K162344. There are no differences between the device and predicates K161368 and K162344 that impact safety and effectiveness.
The flat-panel Detector (DR 14e and DR 17e) is used to capture the images. The have been cleared by the FDA and are used by a number of device manufacturers.
The detectors are identical to those used in the predicate, Innolux RIC flat panel detector (K162344). Detectors transmit the raw image to the NX workstation over a wired gigabit Ethernet connection or a wireless signal via an access point. They are provided in two sizes (a 17x17 in. and 14x17 in.), each with either a Cesium Iodide (CsI) or Gadolinium Oxysulfide (GOS) scintillator layer. The 17x17 in. detectors are used with fixed x-ray systems like the DX-D 600 (K112670) and DX-D 300 (K103050). The 14x17 in. detectors are provided with a handle for easy portability and are used with mobile x-ray systems like the DX-D 100 (K103597).
All of the detectors operate with the same underlying technology. X-rays are absorbed by the detector and converted to a raw digital data. That data is normalized, pixel-by-pixel before it is provided to the NX workstation for further processing and display. A real-time interpolation is also performed for any pixels identified as defective. Table 2 on the following page summarizes this data:
| Performance<br>Characteristics | DX-D 10 Flat-Panel<br>Detector | DX-D 20 Flat-Panel<br>Detector (Handle) | DR 14e Wireless<br>Detector | DR 17e Wireless<br>Detector |
|--------------------------------|----------------------------------------------|----------------------------------------------|----------------------------------------|----------------------------------------|
| Scintillator | CsI, GOS | CsI, GOS | CsI, GOS | CsI, GOS |
| Cassette size | 35x43cm/14x17in | 35x43cm/14x17in | 35x43cm/14x17in | 43x43cm/17x17in |
| Pixel Size | 139 μm | 139 μm | 150 μm | 150 μm |
| A/D Conversion | 14 bits | 14 bits | 16 bits | 16 bits |
| Interface to<br>Generator | Ethernet | Ethernet | AED &<br>Synchronized | AED &<br>Synchronized |
| Communication | Tethered | Tethered | Wireless | Wireless |
| Power | I/O Interface Box: 100-<br>240 VAC, 47-63 Hz | I/O Interface Box: 100-<br>240 VAC, 47-63 Hz | Battery: replaceable<br>& rechargeable | Battery: replaceable<br>& rechargeable |
| Weight | 3.9 kg (8.6 lbs) | 4.9 kg (10.8 lbs) | 2.95 kg (6.50 lbs) | 3.65 kg (8.04 lbs) |
#### Table 2: Summary of Detector Performance Characteristics
The NX Workstation is a commercially available computer with Agfa software. All Agfa computed radiography and direct radiography systems us the NX workstation. The NX workstation is cleared under the following 510(k) premarket submissions; K053634 (NX1.0), K071162 (NX 2.X for MUSICA²), K081963 (NX2008 add MUSICA2 Platinum), and K090672 (NX add neonatal and pediatric).
{21}------------------------------------------------
Workstations provide their users with these basic functions:
- Identify and select patients and exams from work lists,
- Match the digitized images with the appropriate patient and examination. ●
- . Preview images and perform QC checks (e.g.: check if the patient moved during exposure, or confirm the exposure parameters used),
- Process images according to the exam type and other criteria,
- . View images or send them to the appropriate location for viewing, printing and storage; be it a printer, a softcopy capable display such a Picture Archiving and Communications System (PACS) or an archive.
- When included as a component of an x-ray system; set and review patient exposure settings with a software x-ray console.
- When used with direct radiography systems, the workstation synchronizes the detectors ● with the x-ray system so that the exposure is not initiated until the detectors are ready.
- The ability to scan multiple images and stitch them together into a single image: Full Leg
- Full Spine.
- A full screen mode for utilizing the entire screen to display images.
- A configurable user interface button that makes it easy to call up third party applications,
- such as a RIS client, procedure book, etc.
- . Enhanced security with the option to require a configurable ID before use.
- Configurable mandatory patient fields, for better data consistency. .
- Support for departmental quality control efforts with the ability to export technical images ●
- . to Agfa's user quality control tool, AUTO QC2 SP1. Multiple DICOM export destinations.
- The ability to print a study in its entirety by pressing the F7 key. .
NX workstations are compatible with the complete line of Agfa direct radiography systems: DX-D Imaging Package, DX-D100, DX-D300, DX-D400, DR 400, DX-D 600, and DR 600. Users who have other Agfa DR systems can connect them to the NX workstation.
{22}------------------------------------------------
Image /page/22/Figure/1 description: The image shows two AED detectors, one 14x17 and one 17x17, along with an NX workstation. The 14x17 AED detector is on the left side of the image, and the 17x17 AED detector is on the right side of the image. The NX workstation is at the bottom of the image and shows a person using the touch screen.
Figure 1: DX-D Imaging Package Components
# Indications for Use
Agfa's DX-D Imaging Package is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of human anatomy. The DX-D Imaging Package may be used wherever conventional screen-film systems may be used.
Agfa's DX-D Imaging Package is not indicated for use in mammography.
Intended use of the DX-D Imaging Package, as described in the labeling, has not changed as a result of any labeling modification(s).
{23}------------------------------------------------
# FDA Review Issues
Table 3 below summarizes the characteristics of the proposed device with respect to several FDA review issues.
| 510(k) Review Issues | Yes | No | N/A |
|----------------------------------------------------------------------|-----|----|-----|
| Is the device life-supporting or life sustaining? | | X | |
| Is the device implanted (short-term or long-term)? | | X | |
| Does the device use software? | X | | |
| Is the device shipped sterile? | | | X |
| Is the device single use? | | | X |
| Is the device home use? | | X | |
| Is the device for prescription only? | X | | |
| Does the device contain a drug or biological product as a component? | | X | |
| Is this device a kit? | | X | |
| Is the device subject to post-market surveillance | | X | |
| Is the device subject to the Radiation Control Act | X | | |
| Is a Class III Summary and Certification required? | | X | |
| Is a financial certification or disclosure statement required? | | X | |
#### Table 3: 510(k) Review Issues
{24}------------------------------------------------
# Regulatory History
Agfa digital radiography systems have been the subject of numerous 510(k) submissions. These are summarized in the Table 4 below, beginning with the most recent:
| Clearance | Change | Detector Model(s) |
|----------------|---------------------------------------------|-------------------------------|
| K161368 | DX-D Imaging Package4 | DR 10s and DR 14s |
| K152639 | DR600 | |
| K142184 | DX-D Imaging Package | DX-D 40 |
| K141602 | DX-D Imaging Package | Dose Reduction indication |
| K141192 | DR400 | |
| K131408 | CR 12-X | |
| K122736 | DX-D Imaging Package | Pediatric/neonatal indication |
| K121095 | DX-D Imaging Package | DX-D 30 (Wireless) |
| K121948 | CR 10-X | |
| K112670 | DX-D 600 | |
| K111324 | CR Mammography System with DX-M Digitizer | |
| K103597 | DX-D 100 | |
| K103050 | DX-D 300 | |
| K092669 | DX-D Imaging Package5 | DX-D 10 & DX-D 20 |
| K092238 | DX-G Digitizer | |
| K090672 | NX Workstation (NX8000) | Pediatric/neonatal indication |
| K081963 | NX Workstation (NX2008) | MUSICA2 Platinum |
| K071162 | NX 2.0 workstation | MUSICA2 |
| K063421 | DX-Si | |
| K062232 | CR30-X | |
| K062742 | CR85-X | |
| K053634 | NX1.0 workstation | 1st NX with MUSICA |
| Letter-to-file | QS workstation (QS version 3.0) | |
| K050810 | DX-S (previously called CR50) | |
| Letter-to-file | QS workstation (QS version 2.1) | |
| K042779 | Radiotherapy | |
| Letter-to-file | CR75 | |
| K041701 | CR25 | |
| K021542 | ADC Dental | |
| K012750 | ADC Pediatric | |
| K013138 | ADC Compact Plus | |
| K010571 | ADC QS/IPD (workstation software, Uro/Tomo) | |
| K000159 | Full Leg/Full Spine | |
| Letter-to-file | ADC Solo | |
| K974597 | ADC Compact | |
| K904519 | ADC 70 | |
### Table 4: History of Agfa's Radiography 510(k) Submissions
<sup>4</sup> Predicate Device.
<sup>5</sup> Reference Device.
{25}------------------------------------------------
# Initial Submission
The initial submission for Agfa's direct radiography (DR) systems was the DX-D Imaging Package (K092669). This submission is to add the DR14e and DR17e Flat Panel Detectors to Agfa's DX-D Imaging Package portfolio. Agfa's DR 14e and DR 17e wireless panels are currently marketed by Innolux as RIC 35C/G and RIC 43C/G, which is one of the predicates for this submission (K162344).
# Substantial Equivalence Summary
This 510(k) will provide information that documents the path to substantial equivalence that is highlighted in the flowchart in Section 12. The detailed substantial equivalence argument is presented there.
The new device and the both predicate devices (K162344) are solid state imaging devices. Product Code MOB. Agfa's DX-D Imaging Package is substantially equivalent to both predicate devices (K161368 and K162344) in that it uses precisely the same technology to capture and transmit images.
Principles of operation and technological characteristics of the new and predicate devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically identical to both predicates. K161368 and K162344. It uses the same workstation as predicate K161368 and the same flat panel detectors to capture and digitize the images as predicate K162344. Differences in devices do not alter the intended diagnostic effect.
The DX-D Imaging Package indications for use statement is substantially equivalent to both predicate devices (K161368 and K162344). The new device and both predicate devices (K161368 and K162344) include the statement that the device is not indicated for mammography. The DX-D Imaging Package and predicate device (K161368) describe the imaging applications may be used utilized where screen-film systems exist. The new device and predicate device (K162344) have similar indications for use statements in that both devices are intended to display general radiographic images of the human body and can be used in place of conventional film/screen or CR systems. Intended uses are the same. The devices have the same technological characteristics.
The only difference between the new device and predicate device (K162344) is that the predicate device's Indication For Use statement includes specific contraindications for fluoroscopy, angiography, and tomography. The specific contraindications are applicable to the new device since it is only indicated for general projection applications. Fluoroscopy, angiography and tomography are not considered to be general projection applications. There are no differences between the new device and predicates K161368 and K162344 that impact safety and effectiveness.
Performance data, image quality clinical evaluations, grid test, and usability/functionality data are adequate to ensure equivalence.
Table 5 on the next page summarizes the technical characteristics between the new device and the predicate devices (K161368 and K162344).
{26}------------------------------------------------
| PRODUCT COMPARISON TABLE | | | | |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | DX-D Img Pkg<br>(New Device) | AGFA DX-D Imaging<br>Package<br>(PREDICATE-K161368) | Innolux RIC<br>(PREDICATE - K162344) | |
| Communications | Same as both Predicates | Wireless | Wireless/Wired | |
| Flat Panel | Same as both Predicates | Flat Panel Detector | Flat Panel Detector | |
| Detector Material | Same as both Predicates | Gadolinium Oxysulfide<br>(GOS) or Cesium Iodide<br>(CsI) scintillator | Gadolinium Oxysulfide<br>(GOS) or Cesium Iodide (CsI)<br>scintillator | |
| Detector Sizes | Same as Predicate K162344 | 10x12 in.<br>14x17 in. | 14x17 in<br>17x17in | |
| Active Matrix<br>(14x17 in.) | Same as Predicate K162344 | 2400 x 2880 | 2336x2836<br>2832x2836 | |
| Pixel size | Same a Predicate K162344 | 148 μm | 150 μm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.