← Product Code [OCW](/productcode/OCW) · K172775

# MiniLap System with MiniGrip Handle, MiniLap System with ThumbGrip Handle (K172775)

_Teleflexmedical, Inc. · OCW · Feb 12, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172775

## Device Facts

- **Applicant:** Teleflexmedical, Inc.
- **Product Code:** [OCW](/productcode/OCW.md)
- **Decision Date:** Feb 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

MiniLap instruments are a family of minimally invasive devices with the means to penetrate soft tissue to access certain areas of the human anatomy. The devices are used to grasp, hold, and manipulate other soft internal tissues, as well as items such as hernia mesh, during pediatric and adult surgery. The MiniPolar Instruments are used to cut and cauterize soft tissue during pediatric and adult surgery. The MiniLap System with MiniGrip Handles grasping instruments have applications in a variety of general, thoracic, gynecologic, urologic, laparoscopic procedures for manipulation of tissue, as well as items such as hernia mesh, during pediatric and adult surgery. Bipolar Grasper instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic procedures for manipulation and coagulation of tissue and for manipulation of items such as hernia mesh during pediatric and adult surgery. The male bipolar connection is located at the back of the handle and may be utilized for bipolar applications when attached to standard two plus bipolar cables and their generators (a male connection is provided as an integral part of the instrument).

## Device Story

Teleflex MiniLap System comprises percutaneous surgical instruments for minimally invasive access; includes ThumbGrip (needle/cannula shaft with retractable working tip) and MiniGrip (pistol-grip handle with rotating jaws) families. Devices penetrate soft tissue directly without trocars; used for grasping, holding, manipulating tissue/hernia mesh, or electrosurgical cutting/coagulation. Electrosurgical variants connect to standard ESUs (e.g., Valleylab Force FX) via monopolar or bipolar cables. Operated by surgeons in clinical settings; provides reduced trauma by eliminating trocar sites and incision closure. Output is mechanical tissue manipulation or thermal cautery effect. Benefits include reduced surgical trauma and improved access. Single-use, sterile, disposable.

## Clinical Evidence

Bench testing only. Verification and validation testing confirmed device performance against intended use. Electrosurgical components tested to IEC 60601-1. Packaging/sterility validated per ISO 11607-1. Stability testing supported shelf-life extension.

## Technological Characteristics

Disposable manual percutaneous instruments; 2.3mm-2.4mm shaft diameters. Materials compliant with ISO 10993-1. Energy: Monopolar (banana plug) or Bipolar (standard two-plug cable). Features: Retractable working tips, stabilizing pivot disk, pistol-grip handle with ratchet/rotation hub. Sterilization: Terminally sterilized. Connectivity: Standard ESU cable interface.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- MINI LAP Instruments ([K070686](/device/K070686.md))
- MINI LAP Pediatric Instruments with Medusa Clamping System ([K070352](/device/K070352.md))
- MINI LAP Electrocautery Instruments ([K083754](/device/K083754.md))
- MINI-LAP BIPOLAR ELECTROCAUTERY DEVICES ([K113597](/device/K113597.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 12, 2018

Teleflex Medical, Inc. Holly Hallock Regulatory Affairs Manager 3015 Carrington Mill Blvd Morrisville, North Carolina 27560

## Re: K172775

Trade/Device Name: Teleflex MiniLap System with ThumbGrip Handle, Teleflex MiniLap System with ThumbGrip MiniPolar Handle, Teleflex MiniLap System with MiniGrip Handle, Teleflex MiniLap System with MiniGrip BiPolar Handle Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OCW, GEI Dated: September 12, 2017 Received: September 14, 2017

Dear Holly Hallock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Aments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Image /page/1/Picture/5 description: The image shows a signature block with the word "Sincerely" at the top. Below that is the name "Jennifer R. Stevenson -S3" in a larger font. The text appears to be part of a formal document or letter.

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### K172775 - Page 1 of 4

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known)

#### K172775

Device Name

Teleflex MiniLap® System with ThumbGrip Handle

#### Indications for Use (Describe)

Minil ap instruments are a family of minimally invasive devices with the means to penetrate soft tissue to access certain areas of the human anatomy. The devices are used to grasp. hold, and manipulate other soft internal tissues, as well as items such as hernia mesh, during pediatric and adult surgery.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

CONTINUE ON A SEPARATE PAGE IF NEEDED

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

Device Name

Teleflex MiniLap® System with ThumbGrip MiniPolar® Handle

Indications for Use (Describe)

The MiniPolar Instruments are used to cut and cauterize soft tissue during pediatric and adult surgery.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

Device Name

Teleflex MiniLap® System with MiniGrip® Handle

Indications for Use (Describe)

The MiniLap System with MiniGrip Handles grasping instruments have applications in a variety of general, thoracic, gynecologic, urologic, laparoscopic procedures for manipulation of tissue, as well as items such as hernia mesh, during pediatric and adult surgery.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </span>                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{5}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known)

Device Name

Teleflex MiniLap® System with MiniGrip® BiPolar Handle

#### Indications for Use (Describe)

Bipolar Grasper instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic procedures for manipulation and coagulation of tissue and for manipulation of items such as hernia mesh during pediatric and adult surgery.

The male bipolar connection is located at the back of the handle and may be utilized for birolar applications when attached to standard two plus bipolar cables and their generators (a male connection is provided as an integral part of the instrument).

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{6}------------------------------------------------

# 510(k) SUMMARY

# MiniLap® Percutaneous Surgical System: ThumbGrip Instruments and MiniGrip® Graspers

# A. Name, Address, Phone and Fax Number of Applicant

Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560

Phone: 919-433-4918 919-433-4996 Fax:

## B. Contact Person

Holly Hallock Regulatory Affairs Manager

# C. Date Prepared

February 9, 2018

## D. Device Name

| Trade Name             | Teleflex MiniLap® System with ThumbGrip Handle                 |
|------------------------|----------------------------------------------------------------|
| Common Name            | Endoscopic Tissue Approximation Device                         |
| Classification<br>Name | Endoscopic and accessories                                     |
| Trade Name             | Teleflex MiniLap® System with ThumbGrip MiniPolar®<br>Handle   |
| Common Name            | Electrosurgical, Cutting & Coagulation & Accessories           |
| Classification<br>Name | Electrosurgical cutting and coagulation device and accessories |
| Trade Name             | Teleflex MiniLap® System with MiniGrip® Handle                 |
| Common Name            | Endoscopic Tissue Approximation Device                         |
| Classification<br>Name | Endoscopic and accessories                                     |

{7}------------------------------------------------

| Trade Name          | Teleflex MiniLap® System with MiniGrip® BiPolar Handle         |
|---------------------|----------------------------------------------------------------|
| Common Name         | Electrosurgical, Cutting & Coagulation & Accessories           |
| Classification Name | Electrosurgical cutting and coagulation device and accessories |

## E. Device Description

The Teleflex MiniLap® Percutaneous Surgical System is a family of minimally invasive surgical devices that provide a surgeon with the means to penetrate soft tissue to access certain areas of the human anatomy, then grasp, hold and manipulate other soft internal tissue within the body that are related to the surgical procedure.

Please note that this is a bundled submission: the MiniLap Percutaneous Surgical System has two families presented in this 510(k). The first is the MiniLap ThumbGrip Instruments, which have a thumb grip handle to control the instrument tips. These instruments are available in electrosurgical and non-electrosurgical configurations. The second product family discussed in this 510(k) is the MiniLap MiniGrip Graspers, which have a pistol grip handle that allows single handed device operation to control the jaws. While these graspers are currently only cleared for electrosurgical configurations (K113597), the intent of this 510(k) is to support non-electrosurgical configurations of the MiniGrip Graspers as well.

ThumbGrip Instruments and MiniGrip Graspers are disposable manual instruments designed for direct introduction to the surgical site without the need for a traditional insertion conduit. These instruments are used for grasping, holding and manipulating soft internal tissues. These percutaneous instruments provide the benefit of reduced trauma due to no surgical incision closure and the elimination of one or more trocars.

# ThumbGrip Instruments

MiniLap ThumbGrip Instruments consist of an integrated needle/cannula shaft that houses a retractable working instrument. The shaft can be introduced percutaneously to the surgical site, after which the working instrument can be deployed for grasping and manipulating or cauterizing tissue. The ThumbGrip Instruments are available in electrosurgical and nonelectrosurgical configurations, with shaft diameters of 2.4mm and 2.3mm, respectively. A stabilizing pivot disk is available on the non-electrosurgical configurations that can be stuck to the skin of the patient to provide entry-depth control, and a banana plug is available on the electrosurgical configurations for monopolar energy. The working instrument is controlled through the thumb handle and a locking mechanism on the proximal end of the device.

| ThumbGrip Instruments |                             |                     |
|-----------------------|-----------------------------|---------------------|
| Model Number          | Name                        | Configuration       |
| GBC250                | ThumbGrip Alligator Grasper |                     |
| CLC250                | ThumbGrip Clutch Grasper    | Non-electrosurgical |

{8}------------------------------------------------

| BCK250  | ThumbGrip Babcock Grasper         |                                                     |
|---------|-----------------------------------|-----------------------------------------------------|
| BLC250  | ThumbGrip Bowel Grasper           |                                                     |
| GBC200  | ThumbGrip Alligator Grasper Short |                                                     |
| ECMC300 | ThumbGrip Curved Spatula Probe    | Electrosurgical with monopolar energy (“MiniPolar”) |
| ECMS300 | ThumbGrip Straight Spatula Probe  |                                                     |
| ECMH300 | ThumbGrip Hook Probe              |                                                     |
| ECMP300 | ThumbGrip Conical Probe           |                                                     |

# MiniGrip Graspers

The MiniLap MiniGrip Graspers consist of an integrated 2.4mm needle/cannula shaft that houses a retractable grasper. The shaft, which is insulated for electrosurgical configurations, can be introduced percutaneously to the surgical site, after which the working portion of the instrument can be deployed to approximate and grasp soft tissue.

The grasper jaws are controlled by a pistol-grip handle. The orientation of the jaws relative to the handle can be adjusted by the rotation hub. The ratchet function of the handle locks the position of the jaws in a closed or free mode. The "ON" position locks the jaws together for insertion and removal of the jaws. The "OFF" position of the ratchet allows for free motion of the jaws. The arming button allows the surgeon to arm the device (insertion needle out). A red indication will be visible under the arming button when the needle is exposed.

| MiniGrip Graspers |                                  |                                        |
|-------------------|----------------------------------|----------------------------------------|
| Model Number      | Name                             | Configuration                          |
| BPS300            | MiniGrip BiPolar Grasper         | Electrosurgical with bipolar<br>energy |
| PGAC300           | MiniGrip Alligator Grasper       |                                        |
| PGCC300           | MiniGrip Clutch Grasper          |                                        |
| PGBK300           | MiniGrip Babcock Grasper         | Non-electrosurgical                    |
| PGBC300           | MiniGrip Tong Grasper            |                                        |
| PGAC200           | MiniGrip Alligator Grasper Short |                                        |

The ThumbGrip MiniPolar Instruments connect to most electrosurgical units (ESUs) via a standard monopolar cable, as they provide monopolar energy. The MiniGrip BiPolar Graspers connect to most ESUs via a standard bipolar cable, as they provide bipolar energy. An example of an ESU that may be used with both of these products families is Valleylab Force FX.

# F. Indications for Use and Contraindications

# ThumbGrip Instruments

Indications:

{9}------------------------------------------------

MiniLap instruments are a family of minimally invasive devices with the means to penetrate soft tissue to access certain areas of the human anatomy. The devices are used to grasp, hold, and manipulate other soft internal tissues, as well as items such as hernia mesh, during pediatric and adult surgery.

#### Contraindications:

The device is not for use when laparoscopic techniques generally are contradicted.

## ThumbGrip MiniPolar Instruments

#### Indications:

The MiniPolar Instruments are used to cut and cauterize soft tissue during pediatric and adult surgery.

#### Contraindications:

The device is not intended for use when endoscopic techniques are generally contraindicated.

Do not resterilize. The device is provided sterile and is intended for use in a single procedure. Discard after use.

The MiniPolar instruments with monopolar cautery are not intended for contraceptive coagulation of fallopian tissue, but may be used to achieve hemostasis following transection of the fallopian tube.

## MiniGrip Graspers

#### Indications:

The MiniLap System with MiniGrip Handles grasping instruments have applications in a variety of general, thoracic, gynecologic, urologic, laparoscopic and endoscopic procedures for manipulation of tissue, as well as items such as hernia mesh, during pediatric and adult surgery.

#### Contraindications:

Blind insertion in laparoscopic surgery.

The device is provided STERILE and it is intended for use in a SINGLE procedure. DISCARD AFTER USE. DO NOT RE-STERILIZE.

## MiniGrip BiPolar Graspers

#### Indications:

Bipolar Grasper instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic, and endoscopic procedures for manipulation and coagulation of tissue and for manipulation of items such as hernia mesh during pediatric and adult surgery.

{10}------------------------------------------------

The male bipolar connection is located at the back of the handle and may be utilized for bipolar applications when attached to standard two plug bipolar cables and their generators (a male connection is provided as an integral part of the instrument).

Contraindications: The device is not intended for use when endoscopic techniques are generally contraindicated.

Blind insertion in laparoscopic surgery.

The device is provided STERILE and is intended for use in a SINGLE procedure, DISCARD AFTER USE. DO NOT RESTERILIZE.

This device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures.

# G. Substantial Equivalence

The proposed MiniLap ThumbGrip Instruments (non-electrosurgical) are substantially equivalent to the predicate devices:

| Predicate Device                                                 | Manufacturer          | 510(k) No. | Date Cleared   |
|------------------------------------------------------------------|-----------------------|------------|----------------|
| MINI LAP Instruments                                             | Teleflex              | K070686    | April 05, 2007 |
| MINI LAP Pediatric<br>Instruments with Medusa<br>Clamping System | Mini-Lap Technologies | K070352    | April 05, 2007 |

The proposed MiniLap ThumbGrip MiniPolar Instruments are substantially equivalent to the predicate devices:

| Predicate Device                                                 | Manufacturer          | 510(k) No. | Date Cleared      |
|------------------------------------------------------------------|-----------------------|------------|-------------------|
| MINI LAP Electrocautery<br>Instruments                           | Teleflex              | K083754    | November 30, 2009 |
| MINI LAP Pediatric<br>Instruments with Medusa<br>Clamping System | Mini-Lap Technologies | K070352    | April 05, 2007    |

The proposed MiniLap MiniGrip Graspers (non-electrosurgical) are substantially equivalent to the predicate devices:

| Predicate Device                                                 | Manufacturer          | 510(k) No. | Date Cleared   |
|------------------------------------------------------------------|-----------------------|------------|----------------|
| MINI-LAP BIPOLAR<br>ELECTROCAUTERY<br>DEVICES                    | Mini-Lap Technologies | K113597    | June 03, 2012  |
| MINI LAP Pediatric<br>Instruments with Medusa<br>Clamping System | Mini-Lap Technologies | K070352    | April 05, 2007 |

{11}------------------------------------------------

The proposed MiniLap MiniGrip BiPolar Graspers are substantially equivalent to the predicate devices:

| Predicate Device                                                 | Manufacturer          | 510(k) No. | Date Cleared   |
|------------------------------------------------------------------|-----------------------|------------|----------------|
| MINI-LAP BIPOLAR<br>ELECTROCAUTERY<br>DEVICES                    | Mini-Lap Technologies | K113597    | June 03, 2012  |
| MINI LAP Pediatric<br>Instruments with Medusa<br>Clamping System | Mini-Lap Technologies | K070352    | April 05, 2007 |

# H. Comparison To Predicate Devices

The proposed MiniLap Percutaneous Surgical System, including the ThumbGrip Instruments and MiniGrip Graspers, has the same technology and functional characteristics as the predicate devices.

Additionally, the devices included in this submission have been on the market for several years with no substantial modifications. The manufacturer of the devices included in this submission is the same manufacturer of the predicate devices referenced. (Note: Teleflex Medical previously acquired the assets of Mini-Lap Technologies.)

## I. Materials

The proposed MiniLap Percutaneous Surgical System, including the ThumbGrip Instruments and MiniGrip Graspers, has the same patient contacting materials as the predicate devices. All materials are in compliance with ISO 10993-1.

## J. Technological Characteristics

A comparison of the technological characteristics of the proposed MiniLap Percutaneous Surgical System, including the ThumbGrip Instruments and MiniGrip Graspers and their predicate devices, has been performed. The results of this comparison demonstrate that the ThumbGrip Instruments and MiniGrip Graspers are equivalent to the marketed predicate devices in technological and performance characteristics.

# K. Performance Data

Ship testing, which has been performed in accordance with ISO 11607-1 Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems, demonstrates that the ThumbGrip Instruments and MiniGrip Graspers are substantially equivalent to their predicate devices.

Additionally, design verification and validation of the ThumbGrip Instruments and MiniGrip Graspers was conducted to document that the device performed to the intended use. All verification and validation results support substantial equivalence.

{12}------------------------------------------------

Electrosurgical devices have been tested to IEC 60601-1 Medical Electrical Equipment -Part 1: General requirements for basic safety and essential performance.

Please note that the predicate devices previously all had shelf lives of 2 years. However, recent stability testing supported extending the shelf life of the MiniGrip Graspers (nonelectrosurgical) and the ThumbGrip Instruments (electrosurgical) from 2 years.

## L. Conclusion

Based upon the discussions above comparing the materials, the technological characteristics, and the performance data, the proposed MiniLap Percutaneous Surgical System, including the ThumbGrip Instruments and MiniGrip Graspers, is substantially equivalent to the predicate devices cleared to market via 510(k)s K070352, K070686, K083754, and K113597.

---

**Source:** [https://fda.innolitics.com/device/K172775](https://fda.innolitics.com/device/K172775)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
