K172772 · Dao Health · HGX · Oct 13, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K172772
Device Name
Freemie Independence, Freemie Liberty
Applicant
Dao Health
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Oct 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Freemie® Independence and Freemie® Liberty are powered breast pumps to be used by lactating women to express and collect milk from their breast. The Freemie® Independence and Freemie® Liberty pumps are intended for use by a single user.
Device Story
Powered breast pump for lactating women; used for single or double breast milk expression/collection. Device is lightweight, compact, and clips to clothing for mobile use in home or office. User adjusts suction (10 levels) and cycle speed (11 levels) via tactile buttons; 3 memory settings available. Firmware controls low-voltage DC swash plate vacuum pump and solenoid; provides vacuum -10 to -280 mmHg and 18-150 cycles/minute. Physical barrier prevents milk backflow. Independence model features LED indicators and 40-minute auto shut-off; Liberty model features LCD screen and user-adjustable auto shut-off (5-40 minutes). Powered by rechargeable lithium-ion polymer battery; charged via micro USB. Healthcare provider does not operate; patient self-use. Output (vacuum) facilitates milk expression, potentially benefiting patient by enabling convenient, mobile breastfeeding support.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), home healthcare environment (IEC 60601-1-11), battery safety (IEC 62133), photobiological safety (IEC 62471), and software verification/validation. Performance testing verified vacuum levels (-10 to -280 mmHg), cycle speeds (18-150 cpm), backflow protection, motor life, and drop testing.
Technological Characteristics
Powered breast pump; swash plate vacuum mechanism. Rechargeable 3.7V/2200mAh lithium-ion polymer battery. Tactile switches; LED (Independence) or LCD (Liberty) interface. Micro USB charging. Embedded firmware control. Backflow protection via physical barrier. Dimensions <4 inches diameter, <2 inches tall; weight 9.2 oz. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62133, and IEC 62471.
Indications for Use
Indicated for lactating women to express and collect milk from their breast. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Spectra S1 Plus and Spectra S2 Plus Breast Pump (K150476)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is a stylized graphic of an eagle, while the FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are placed side-by-side, with the HHS logo on the left and the FDA logo on the right. The logos are both in color and are of high resolution.
October 13, 2017
DAO Health % Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K172772
> Trade/Device Name: Freemie® Independence and Freemie® Liberty Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: September 13, 2017 Received: September 14, 2017
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172772
#### Device Name
Freemie® Independence and Freemie® Liberty Breast Pump
Indications for Use (Describe)
The Freemie® Independence and Freemie® Liberty are powered breast pumps to be used by lactating women to express and collect milk from their breast. The Freemie® Independence and Freemie® Liberty pumps are intended for use by a single user.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Freemie® Independence and Freemie® Liberty Breast Pump
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### SUBMITTER:
DAO Health 1345 Easy Lane El Dorado Hills, CA 95762
### DATE PREPARED:
July 12, 2017
# PRIMARY CONTACT PERSON:
Dave Paul Phone: 916-339-7388 FAX: 877-869-1973 e-mail: dave@freemie.com
#### Device:
TRADE NAME: Freemie® Independence and Freemie® Liberty Breast Pump COMMON NAME: Powered Breast Pump CLASSIFICATION NAME: Pump, Breast, Powered REGULATORY CLASS: II PRODUCT CODE: HGX REGULATION NUMBER: 884.5160
### PREDICATE DEVICE:
Spectra S1 Plus and Spectra S2 Plus Breast Pump (K150476) The predicate devices have not been subject to a design-related recall.
### INDICATIONS FOR USE:
The Freemie® Independence and Freemie® Liberty are powered breast pumps to be used by lactating women to express and collect milk from their breast. The Freemie® Independence and Freemie® Liberty pumps are intended for use by a single user.
### DEVICE DESCRIPTION:
The Freemie® Independence and Freemie® Liberty powered breast pumps utilizes the supplied previously cleared device (K130349) Freemie® Breast Pump Collection System to express and collect milk from the breast of a lactating woman. Pumping can be performed on one breast (single pumping) or both breasts (double pumping) at the same time. The system is designed for mobile pumping (home, office, etc.) as the device is light weight, compact in size and can be attached to the user's clothing with a clip.
{4}------------------------------------------------
Ten (10) suction levels, eleven (11) cycle speeds and three (3) memory settings (suction and cycle levels) are user adjustable via tactile switches (referred to as buttons). The pumps are powered by a rechargeable lithium-ion polymer battery which can be charged with a USB to micro USB provided cable (laptop, car, etc.) and alternatively with a supplied wall charge adapter with a USB port for charging the battery with mains power. The sole function of the Micro USB port on the pump is for charging the pump. The power to the pump is always supplied by the battery, even when charging.
Firmware controls a low voltage DC swash plate vacuum pump and solenoid that is capable of providing vacuum levels from -10 to -280 mmHq with cycle speeds from 18 to 150 cycles per minute. A physical barrier on the collection side prevents backflow of milk into the pump.
The Freemie® Independence Pump has a series of LED lights to indicate the pump's current use settings to the user and has a pre-programmed auto shut-off timer that turns the pump off after 40 minutes. The Freemie® Liberty Breast Pump has an LCD screen to indicate the pump's current use settings to the user and has an auto shut-off timer that is user adjustable in five minute increments from 5 to 40 minutes.
# DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICE COMPARED TO THE PREDICATE DEVICE
The indications for use of Freemie® Independence and Freemie® Liberty powered breast pumps are the same as the predicate devices. The subject and predicate devices have the same intended use - to express and collect milk from the breast of a lactating woman.
The Freemie® Independence and Freemie® Liberty powered breast pumps generate vacuum in a similar manner as the predicate device and are capable of providing a similar level of vacuum as the predicate device. The subject and predicate devices have similar technological characteristics such as tactile switches, LCD screen, rechargeable lithium-ion polymer battery, adjustable suction and cycle levels, auto shut-off timer, embedded software (firmware) and backflow protection.
Table 1 below identifies key similarities and differences of the proposed Freemie® Independence and Freemie® Liberty to the legally marketed predicated device, the Spectra S1 Plus and Spectra S2 Plus breast pump (K150476).
| | Subject Device | Predicate Device (K150476) |
|--------------------------------|-----------------------------------------------|-------------------------------------|
| GENERAL DEVICE CHARACTERISTICS | | |
| Product Name | Freemie® Independence and<br>Freemie® Liberty | Spectra S1 Plus and Spectra S2 Plus |
| Manufacturer | DAO Health | Uzinmedicare Co. |
| Product Code | HGX | HGX |
| Regulation Number | 21 CFR 884.5160 | 21 CFR 884.5160 |
| Class | II | II |
| Patient Population | Breastfeeding Women | Breastfeeding Women |
| Indications for Use | The Freemie® Independence and Freemie® | The Spectra S1 Plus and Spectra S2 |
### TABLE 1. COMPARISON OF SUBJECT DEVICE TO PREDICATE DEVICE
{5}------------------------------------------------
| | Liberty are powered breast pumps to be<br>used by lactating women to express and<br>collect milk from their breast. The<br>Freemie® Independence and Freemie®<br>Liberty pumps are intended for use by a<br>single user. | Plus are single-user, powered breast<br>pumps intended to express and collect<br>milk from the breasts of lactating women. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| | User Interface and Controls | |
| Pump Options | Single or Double | Single or Double |
| Power Button | Tactile Switch | Tactile Switch |
| Suction Levels | 10 levels | Massage Mode: 5 levels<br>Expression Mode: 12 levels |
| Cycle Levels | 11 levels | Massage Mode: 1 level<br>Expression Mode: 5 levels |
| Timer | Fixed auto shut-off (40 minutes for<br>Independence only)<br>Adjustable auto shut-off (5-40 minutes for<br>Liberty only) | Fixed auto shut-off (30 minutes) |
| Memory Setting | 3 user programmable settings | No |
| Visual Indicator | LED (Independence only)<br>LCD (Liberty only) | LCD |
| Means of Attaching to<br>Clothing | Clip | No |
| Charging Port | Micro USB | DC Jack (S1 Plus only) |
| | Specifications | |
| Pump Type | Swash Plate | Diaphragm |
| Power Source | 100 - 240V AC | 100 - 240V AC |
| | 50/60 Hz | 50/60 Hz |
| | 0.2A | 600mA |
| | Only when using mains with wall power<br>adapter to charge battery | (Only when using mains with wall power<br>adapter to charge battery for S1 Plus) |
| Battery | 3.70V/2200mAh<br>Rechargeable Lithium-ion Polymer Battery | 11.1V/2000mAh Rechargeable Lithium-<br>ion Polymer Battery (S1 Plus only) |
| Vacuum Range | -10 to -280 mmHg | -50 to -280 mmHg |
| Cycle Speed | 18 to 150 cycles/minute | 38 to 70 cycles/minute |
| Backflow Protection | Yes | Yes |
| Software | Embedded | Embedded |
| Weight | 9.2 ounces | 42.5 ounces |
| Size | <4 inches in diameter, <2 inches tall | <8.5 inches in diameter, <6.5 inches tall |
The technological characteristics of the subject device that are different – number of suction and number of cycle levels, cycle speed range, auto shut-off timer, and user adjustability of the auto shut-off timer, memory settings, means of attaching the pump to the user's clothing, charging port, weight and size do not raise safety and effectiveness questions by these differences.
# SUMMARY OF NON-CLINICAL TESTS:
The Freemie® Independence and Freemie® Liberty powered breast pumps complies with voluntary standards for electrical safety, electromagnetic compatibility, use in the home healthcare environment, usability, portable battery safety requirements and photobiological safety of lamps. The following data were provided in support of the substantial equivalence determination:
- Risk Analysis developed in accordance with ISO 14971:2007
- Electrical safety testing in accordance with IEC 60601-1:2005 (2012 reprint) ●
- Electromagnetic compatibility testing in accordance with IEC 60601-1-2:2014 (4th ● edition)
- Electrical safety usability testing in accordance with IEC 60601-1-6:2010 (30 edition) + ●
{6}------------------------------------------------
### A1:2013
- . Electrical safety testing for use in home in accordance with IEC 60601-1-11:2015 (201 edition)
- Safety testing for portable sealed secondary cells and batteries made from them in ● accordance with IEC 62133:2012 (2nd edition)
- Photobiological safety testing of lamps in accordance with IEC 62471:2006 (1st edition) ●
- Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005)
- . Performance testing to determine minimum and maximum vacuum levels and cycle speeds, backflow protection (fill test), motor life cycle, drop test, voltage verification and Spectra S1 Plus and S2 Plus comparison testing.
- . Biocompatibility: the subject device (the pump itself) does not have direct or indirect user contact. The subject device utilizes the supplied previously cleared device (K130349) Freemie® Breast Pump Collection System.
# CONCLUSION:
The differences between the Freemie® Independence and Freemie® Liberty powered breast pumps and its predicate device do not introduce a new intended use and do not raise new issues of safety and effectiveness. Based on non-clinical testing, DAO Health concludes that the Freemie® Independence and Freemie® Liberty powered breast pumps perform as intended and are substantially equivalent to the legally marketed predicate device, the Spectra S1 Plus and Spectra S2 Plus.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.