K172719 · Healthbeacon, Ltd. · MMK · May 11, 2018 · General Hospital
Device Facts
Record ID
K172719
Device Name
HealthBeacon HB2 Sharps Bin
Applicant
Healthbeacon, Ltd.
Product Code
MMK · General Hospital
Decision Date
May 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The HealthBeacon HB2 Sharps Bin is intended for use by a single patient in the home setting. The HealthBeacon HB2 Sharps Bin comprises an outer enclosure that remains in the home. The outer enclosure contains a replaceable, single-use sharps bin that is intended to provide a receptacle for used, contaminated medical sharps. The single-use sharps bin also acts as an enclosure during transport of the used sharps to ultimate disposal.
Device Story
Device consists of outer enclosure and replaceable, single-use inner sharps bin; used in home setting by patient. Outer enclosure features trap door for syringe deposition; electronic counter tracks number of deposited syringes. Device provides user prompts via display screen regarding injection timing and site rotation; optional text message reminders available. Usage data transmitted to cloud-based tracking database. Power supplied by IEC60601-1 certified supply. Benefits include safe sharps containment, transport, and improved patient adherence to injection schedules.
Clinical Evidence
Bench testing only. Device tested to ISO 23907:2012 for stability, handle strength, penetration resistance, closure permanence, and drop-test leakage/damage resistance. Usage counter accuracy and 5-year functional life verified through bench testing.
Technological Characteristics
Sharps bin: polypropylene (yellow body, blue lid). Enclosure: ABS plastic. Dimensions: 9.5" x 9.4" x 6.9" (enclosure). Power: IEC60601-1 certified supply. Connectivity: Cloud-based tracking database. Standards: ISO 23907:2012.
Indications for Use
Indicated for single patients in home settings for the disposal of used, contaminated medical sharps.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
May 11, 2018
HealthBeacon Ltd Joyce Power Quality Manager Unit 20, Naas Road Business Park, Muirfield Drive Dublin 12 Ireland
Re: K172719
Trade/Device Name: HealthBeacon HB2 Sharps Bin Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: MMK Dated: April 3, 2018 Received: April 4, 2018
Dear Joyce Power:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K172719
# Device Name
HealthBeacon HB2 Sharps Bin
#### Indications for Use (Describe)
The HealthBeacon HB2 Sharps Bin is intended for use by a single patient in the home setting. The HealthBeacon HB2 Sharps Bin comprises an outer enclosure that remains in the home. The outer enclosure contains a replaceable, single-use sharps bin that is intended to provide a receptacle for used, contaminated medical sharps. The single-use sharps bin also acts as an enclosure during transport of the used sharps to ultimate disposal.
The HealthBeacon HB2 Sharps Bin outer enclosure is a neutral gray color body with duck egg sides. The approximate dimensions are: width 9.5 inches by depth 9.4 inches and height 6.9 inches, with a weight of 51b, including an empty sharps bin.
The inner, single-use sharps bin has a yellow semi-translucent body with a blue lid. The approximate dimensions of the bin are: width 9 inches by depth 7 inches and height of 4.5 inches. The gross volume is 2.4 US quarts, with an empty weight of 11oz.
The outer enclosure has a trap door to allow the used sharps to be deposited into the inner sharps bin. The aperture of the trap door, and thus the maximum size of a used sharp, is 6.5 by 1 inches.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY - K172719
# HealthBeacon HB2 Sharps Bin
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
HealthBeacon Ltd Unit 20 Naas Road Business Park Muirfield Drive Dublin 12 Ireland
Phone: +353 1 450 8480
Contact Person: Joyce Power, Quality Manager Email: joyce@healthbeacon.com
Date Prepared: 04 May 2018
### Name of Device
HealthBeacon HB2 Sharps Bin.
### Common or Usual Name
Sharps bin
# Classification Name/Product Code/CFR Reference
Class II, Accessory to hypodermic single lumen needle, 21 CFR 880.5570
Product Code: MMK
### Predicate Device
Medline Sharps Container [K132767]
{4}------------------------------------------------
# Intended Use / Indications for Use
The HealthBeacon HB2 Sharps Bin is intended for use by a single patient in the home setting. The HealthBeacon HB2 Sharps Bin comprises an outer enclosure that remains in the home. The outer enclosure contains a replaceable, single-use sharps bin that is intended to provide a receptacle for used, contaminated medical sharps. The single-use sharps bin also acts as an enclosure during transport of the used sharps to ultimate disposal.
The HealthBeacon HB2 Sharps Bin outer enclosure is a neutral gray color body with duck egg sides. The approximate dimensions are: width 9.5 inches by depth 9.4 inches and height 6.9 inches, with a weight of 5lb, including an empty sharps bin.
The inner, single-use sharps bin has a yellow semi-translucent body with a blue lid. The approximate dimensions of the bin are: width 9 inches by depth 7 inches and height of 4.5 inches. The gross volume is 2.4 US quarts, with an empty weight of 11oz.
The outer enclosure has a trap door to allow the used sharps to be deposited into the inner sharps bin. The aperture of the trap door, and thus the maximum size of a used sharp, is 6.5 by 1 inches.
# Device Description
The HealthBeacon HB2 Sharps Bin device is the combination of the:
- . Outer Enclosure, which remains in the home, and
- The single use Inner Sharps Bin, which is disposed of, once full. .
The HealthBeacon HB2 Sharps Bin comprises:
- A custom-shaped sharps bin molded in puncture-resistant polypropylene . plastic.
- . An integral sliding and permanently locking lid to close over after use and prior to dispatch to the destruction facility.
- A custom enclosure (the "HealthBeacon HB2 enclosure") for the sharps bin. ● The HealthBeacon HB2 Enclosure provides:
- o A neutral and attractive cover for the sharps bin whilst it is being used in a domestic environment.
- o A double trap door access path to the sharps bin to prevent fingeraccess to the sharps bin and its contents of used syringes while in the Enclosure with the side door closed.
{5}------------------------------------------------
- o A counter of syringes being deposited that will give the user a warning when the sharps bin is getting close to full.
The HealthBeacon HB2 Sharps Bin is molded from 2mm (0.08 inch) polypropylene in 3 snap-together sections. The sharps bin door is a sliding door that once fully closed, locks and cannot be re-opened. The HealthBeacon HB2 Enclosure is a neutrally colored enclosure for the sharps bin. The enclosure is molded from ABS plastic and has interlocks so that the sharps bin can only be fitted in the correct orientation within the enclosure. The enclosure door flap lines up with the sharps bin opening, and there is a spring-loaded trap-door within the enclosure that prevents direct access into the sharps bin itself.
The HealthBeacon HB2 Enclosure has a small electronics module to count and record the used hypodermic syringes being deposited. The usage information is passed to a cloud-based tracking database to provide compliance information. The tracking database also provides prompts to the user on a display screen in the HealthBeacon HB2 Enclosure. The prompts inform the user when the next selfadministered injection is due and the injection site to be used, so that they are rotated in time and thus lower the chance of adverse reactions. The patient can also elect to have text message reminders sent their cell phone.
The HealthBeacon HB2 Enclosure is powered by a Medical Device Certified (to IEC60601-1) power supply. This supplies electrical power for the small electronic counter unit.
{6}------------------------------------------------
# Comparison of Technological Characteristics
The HealthBeacon HB2 Sharps Bin has the same intended use, technological characteristics, and principles of operation as its predicate device. The minor implementation differences between the HealthBeacon HB2 Sharps Bin and its predicate device raise no new issues of safety or effectiveness.
| Feature | HealthBeacon HB2<br>Sharps Bin<br>(K172719) | Medline Sharps<br>Container<br>(K132767) | SE Comparison |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The HealthBeacon<br>HB2 Sharps Bin is<br>intended for use by a<br>single patient in the<br>home setting. The<br>HealthBeacon HB2<br>Sharps Bin comprises<br>an outer enclosure<br>that remains in the<br>home. The outer<br>enclosure contains a<br>replaceable, single-<br>use sharps bin that is<br>intended to provide a<br>receptacle for used,<br>contaminated medical<br>sharps. The single-<br>use sharps bin also<br>acts as an enclosure<br>during transport of the<br>used sharps to<br>ultimate disposal | Medline Sharps<br>Containers are<br>intended to provide a<br>receptacle for used,<br>contaminated medical<br>sharps and act as an<br>enclosure during<br>transport to ultimate<br>disposal | Same intended use of<br>receptacle and safe<br>enclosure for sharps<br>transport.<br>HB2 is specifically<br>intended for home use. |
| Materials | Sharps bin:<br>polypropylene: Body<br>yellow, lid blue.<br>Biohazard symbol on<br>sharps bin labelling | Translucent<br>polypropylene: Body<br>red, lid white. | Sharps bin body to be<br>red or any color<br>is allowed as long as it<br>is appropriately<br>labelled with bio<br>hazard symbol |
| | Enclosure: Molded<br>ABS | Optional Enclosure:<br>ABS plastic | Same materials |
| Feature | HealthBeacon HB2<br>Sharps Bin<br>(K172719) | Medline Sharps<br>Container<br>(K132767) | SE Comparison |
| Capacities | 1 size: nominal 2.4<br>US quart (2.3 litres) | 2 sizes: nominal 1 US quart and 5 US quart | Similar, both devices<br>easily lifted. HB2 is<br>midway in capacity<br>between the Medline<br>device capacities |
| Sharps Bin<br>Dimensions | Inches: 9 x 7x 4.5<br>mm: 226 x 172 x 110 | 1 US quart:<br>Inches:3.75 x 3.75 x 6<br>mm: 95 x 95 x 153<br>5 US quart:<br>Inches: 10.4 x 4.4 x<br>9.2<br>mm: 265 x 113 x 233 | Reflects the device<br>capacities, differences<br>do not affect the safety<br>or effectiveness |
| Weight | 11 oz, (316 g) | 1 US quart: 3.78 oz,<br>107 g<br>5 US quart: 12.7oz,<br>360 g | Similar |
| Enclosure<br>Dimensions | Width: 9.5 inch (242<br>mm)<br>Depth: 9.4 inch<br>(235 mm)<br>Height: 6.9 inch (171<br>mm)<br>Weight (including<br>empty sharps bin):<br>5lb (2.3 kg) | Not applicable,<br>optional accessory | HealthBeacon HB2<br>Enclosure is part of the<br>sharps bin device |
| Fill indication | 1. Fill line on side of<br>body. Fill level<br>ascertained by<br>shadow through<br>translucent yellow<br>side wall.<br>2. Transparent inner<br>trap door enables<br>direct line of sight into<br>the top of the sharps<br>bin. | Fill line on side of<br>body. Fill level<br>ascertained by<br>shadow through<br>translucent red side<br>wall. | Items#2 &3 allow the<br>patient to have prior<br>warning of the<br>impending need for the<br>replacement of the<br>inner sharps bin,<br>without the need to<br>remove the sharps bin<br>from the enclosure for<br>inspection. Differences<br>do not affect the safety<br>or effectiveness |
| Feature | HealthBeacon HB2<br>Sharps Bin<br>(K172719) | Medline Sharps<br>Container<br>(K132767) | SE Comparison |
| | 3. Electronic counter<br>in outer enclosure<br>provides "bin nearly<br>full" and "bin full"<br>messages on the<br>enclosure display<br>screen. | | |
| Lid | Opening door in outer<br>enclosure, with<br>integral inner trap<br>door | Counter balanced lid<br>with trap door | Similar: Both have a<br>trap door arrangement |
| Access path | The HealthBeacon<br>HB2 Enclosure<br>sharps-disposal<br>aperture door has an<br>integral trap door so<br>that there is no direct<br>path into the inner<br>sharps bin. The outer<br>door has to be closed<br>to open the inner trap<br>door | Counter balanced lid<br>provides a torturous<br>path for safe disposal<br>of syringe into the<br>sharps bin | Same end effect -<br>there is no direct<br>access to the sharps<br>bin contents. |
| Lid locking for<br>transport to<br>disposal site | Sharps bin removed<br>from enclosure and<br>the sharps bin access<br>door is closed; when<br>self- locking latches<br>permanently seal the<br>sharps bin | Counter balanced lid<br>has plastic spring<br>clips to keep it closed<br>during transport | Similar: The<br>HealthBeacon HB2<br>Sharps Bin solution is<br>permanent and can<br>only be opened by<br>deliberate destruction<br>of the door |
| Standards | ISO23907:2012 | Not publicly declared | Differences do not<br>affect the safety or<br>effectiveness |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# Summary of Performance Testing
# Compliance to Consensus Standards
The HealthBeacon HB2 Sharps Bin has been tested to ISO 23907:2012 by an independent testing laboratory and passed all the ISO test requirements for:
- . Stability
- Strength of the handles .
- . Resistance to penetration
- Permanence of the final closure, and ●
- Resistance to damage and leakage after dropping.
# Usage Counter
The used syringe counter system has been tested to prove that it will accurately count the number of syringes deposited, no matter what the size of syringe and that the mechanism remains functional for the device life of 5 years.
# Substantial Equivalence - Conclusion
Based on the intended use, technological characteristics, performance data, and nonclinical tests performed, the HealthBeacon HB2 Sharps Bin is substantially equivalent to and is as safe and as effective as the legally marketed predicate device, Medline Sharps Container (K132767).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.