K172714 · Genesis Biosystems, Inc. · MUU · Dec 8, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172714
Device Name
LipiVage
Applicant
Genesis Biosystems, Inc.
Product Code
MUU · General, Plastic Surgery
Decision Date
Dec 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted. the harvested fat is only to be used without any additional manipulation. LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
Device Story
LipiVage® is a 60cc hollow barrel harvesting syringe with an integrated in-line polyester filter; designed for aspiration, harvesting, filtering, and transferring autologous adipose tissue. Device connects to a user-supplied vacuum/aspirator source via universal tubing and to a cannula via a male nozzle. During lipoplasty, suction draws adipose tissue into the syringe; the in-line filter concentrates and filters the tissue, eliminating the need for centrifugation or decanting. The device is single-use, sterile, and intended for professional use by surgeons in clinical settings. By avoiding additional manipulation of harvested fat, the device maintains tissue viability for immediate re-implantation into the same patient. The system facilitates efficient fat grafting by streamlining the harvesting and processing workflow.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing, including biocompatibility (ISO 10993-1), sterilization (ISO 11137), packaging/shelf-life (ASTM F1980, F1140, F1929), and mechanical performance (ISO 80369). Adipose viability testing confirmed tissue viability was maintained.
Indicated for aspiration, harvesting, filtering, and transferring autologous adipose tissue for aesthetic body contouring in patients undergoing plastic, reconstructive, neuro, GI, general, orthopedic, gynecological, thoracic, or laparoscopic surgery. Contraindicated for use with manipulated fat.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other and are connected by a series of lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 8, 2017
Genesis Biosystems, Incorporated % Mr. Stuart R. Goldman Senior Consultant, Regulatory Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin. Texas 78746
Re: K172714
Trade/Device Name: LipiVage® Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: August 24, 2017 Received: September 8, 2017
Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172714
Device Name LipiVage®
### Indications for Use (Describe)
LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted. the harvested fat is only to be used without any additional manipulation.
LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the word "EMERGO" in black font. To the left of the word is a red circular logo with three white curved lines inside. The logo is simple and modern. The text is clean and easy to read.
## 510(k) Summary
LipiVage®
к 172714
#### Submission Sponsor 5.1
Genesis Biosystems, Inc.
1500 Eagle Court
Lewisville, TX 75057
Phone Number: 972.315.7888
Contact: Jim Lafferty
Email: JLafferty@genesisbiosystems.com>
#### 5.2 Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road, Bldg. 1, Suite 300
Austin, TX 78746
Office Phone: 512.327.9997
Contact: Stuart R. Goldman, Senior Consultant, RA/QA
Email: project.management@emergogroup.com
#### 5.3 Date Prepared
August 24, 2017
#### Device Identification 5.4
Trade/Proprietary Name: LipiVage® Common/Usual Name: Harvesting syringe Classification Name: Suction lipoplasty system Regulation Number: MUU Product Code: 878.5040 Device Class: Class II Classification Panel: General & Plastic Surgery
{5}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for EMERGO. The logo consists of a red circular symbol on the left and the word "EMERGO" in black letters on the right. The circular symbol has three curved white lines inside it, creating a stylized "E" shape.
#### Legally Marketed Predicate Device 5.5
GID 700™ (K120902)
#### 5.6 Device Description
LipiVage® is a 60cc hollow barrel harvesting syringe with plunger and an in-line polyester filter that is intended to receive, concentrate and filter autologous adipose tissue from lipoplasty procedures, for reimplantation of the extracted tissue back into the same patient at other anatomical locations in their body using a piston syringe and injection cannula. At one end of the LipiVage® harvesting syringe there is a male connector (nozzle) for fitting the female connector (hub) of a cannula for the harvesting of fat cells, while at the other end of the device there is a universal tubing connection to attach to a tubing set (not supplied by Genesis) that is connected to a controlled vacuum source, such as an aspirator or wall suction in the user facility. The LipiVage® harvesting syringe incorporates an in-line filter so there is no need to centrifuge, decant or expose fat cells to unnecessary handling. The device is for professional use only, made from medical grade polymers, single use, and provided sterile to the user.
#### 5.7 Indication for Use
LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.
LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, and laparoscopic surgery.
#### 5.8 Substantial Equivalence Discussion
Genesis Biosystems has chosen the GID 700™ harvesting canister system as the predicate device for its LipiVage® harvesting syringe. The GID 700™ was cleared by the FDA under K120902. The following table compares the subject device to the predicate device with respect to their indications for use, technology and performance testing, thus demonstrating the basis for determination of substantial equivalence between these devices.
| Regulatory Information | | | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | LipiVage® | GID 700™ | Similarities/Differences |
| Manufacturer | Genesis Biosystems | Gid Group | NA |
| 510(k) | Pending | K120902 | NA |
| Product Code | MUU | MUU | Same |
| Regulation | §878.5040 | §872.5040 | Same |
| Class | II | II | Same |
| Indications for Use | LipiVage® is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.<br><br>The LipiVage® is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery. | The GID 700™ is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.<br><br>The GID 700™ is intended for use in the following surgical specialties when aspiration of soft tissue is desired: plastic and reconstructive surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery. | Same |
| Prescription use | Yes | Yes | Same |
| Fundamental Scientific Technology | | | |
| System Concept | Closed loop | Closed loop | Same |
| Source of Energy | User supplied vacuum. | User supplied vacuum. | Same |
| Device Size<br>(fill volume) | Syringe: 60 cc | Canister: 700 ml | The differences in the size of the subject and predicate devices are a function of their overall differences in design geometries (syringe vs. canister), and not considered to have any significant difference in the overall quality of the |
| | | | harvested fat. |
| Device Materials | Syringe: Molded, polycarbonate<br>- Plunger Tip: Polyolefin<br>- O-Rings: Silicone | Canister: Polymer (unknown) | The subject and predicate<br>device are made of similar<br>medical grade polymer<br>materials. |
| Device Filter | Image: Estene Polyester thread | Image: Similar woven fabric material | The subject device uses<br>Estene® Polyester thread<br>(natural color) for its woven<br>filter fabric with a calculated<br>mesh opening of 180 μm,<br>while the predicate device<br>uses a similar woven fabric<br>material with a reported<br>mesh opening of 200 μm. |
| | Estene® Polyester thread (natural<br>color) woven filter fabric with a<br>calculated mesh opening of<br>180 μm | Similar woven fabric material<br>with a reported mesh opening<br>of 200 µm | |
| Provided Sterile | Yes (gamma radiation) | Yes (gamma radiation) | Same |
| Single Patient<br>Use | Yes | Yes | Same |
| Implantable | No | No | Same |
| | | Relevant Testing Performed | |
| Biocompatibility<br>Testing | Per ISO 10993-1:<br>- Part 5 Cytotoxicity<br>- Part 10 Intracutaneous<br>Reactivity<br>- Part 10 Irritation-Sensitization<br>- Part 11 Systemic Toxicity<br>- Part 11 Pyrogenicity | ISO 10993-1 | Same |
| Device<br>Sterilization | ISO 11137-1,-2 | ISO 11137 | Same |
| Device Packaging<br>and Shelf-Life | ASTM F1980<br>ASTM F1140-07<br>ASTM F1929-15 | Not known. | Device packaging and shelf<br>have been validated for the<br>subject device in accordance<br>with the referenced<br>ASTM standards. |
| Device<br>Performance<br>Testing | ISO 80369-1<br>ISO 80369-7<br>ISO 80369-20 | Implosion<br>Tubing Connection<br>Tensile Strength<br>Mechanical Plugging | |
| Animal Testing | NA | NA | Same |
| Clinical Testing | NA | NA | Same |
| Risk Analysis | ISO 14971 | Not Known. | An appropriate Risk Analysis |
| | | | has been performed on the |
| | | | subject device. |
## Table 5-1 – Device Comparison: LipiVage® vs. GID 700™
{6}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for EMERGO. The logo consists of a red circular icon with three curved lines inside, followed by the word "EMERGO" in black, sans-serif font. The logo is simple and modern, with a focus on the company name.
{7}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the Emergo logo. The logo consists of a red circular icon with three curved white lines inside, followed by the word "EMERGO" in black, sans-serif font. The logo is simple and modern, with a focus on the company name.
{8}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Emergo. The logo consists of a red circular symbol with white curved lines inside, followed by the word "EMERGO" in black, sans-serif font. The logo is simple and modern, with a focus on the company name.
#### Non-Clinical Performance Data 5.9
Genesis Biosystems has submitted LipiVage® and its packaging system for the appropriate biocompatibility testing per ISO 10993-1 and its applicable parts, sterilization per ISO 11137, accelerated aging per ASTM F1980, and performance testing per ISO 80369-1/-7. Adipose viability testing demonstrated that fat tissue viability was maintained. The subject device meets all internal Genesis Biosystems specification requirements to support substantial equivalence to the predicate device.
## 5.10 Clinical Performance Data
The performance and use of Suction Lipoplasty Systems classified under product code MUU and regulated within CFR 878.5040 has been well established in the clinical environment since the first such devices were cleared by the FDA. Therefore, there was no clinical testing required to support LipiVage®, as the indications for use is equivalent to the predicate device, which was not subjected to any clinical testing. The substantial equivalence of LipiVage® to the GID 700™ is supported by the non-clinical testing that was performed, along with other supporting documentation, and presented in this submission.
## 5.11 Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared device, or has the same intended use and different technological characteristics, but it can be demonstrated that the new device is substantially equivalent to the predicate device, and that the new device does not raise any questions regarding its safety and effectiveness when compared to the predicate device.
LipiVage® has the same intended use and indications for use as the predicate device. The conclusions drawn from the data included in this submission, demonstrate that LipiVage® is as safe, as effective, and are substantially equivalent to the predicate device in intended use, indications for use, technological characteristics, mode of operation, mechanical performance, materials, biocompatibility and is therefore substantially equivalent to the GID 700™.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.