Drive DeVilbiss iGo 2 Portable Oxygen Concentrator
Applicant
Devilbiss Healthcare, LLC
Product Code
CAW · Anesthesiology
Decision Date
Mar 23, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Drive DeVilbiss iGo2 Portable Oxygen Concentrator is indicated for the administration of supplemental oxygen. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
Device Story
Portable oxygen concentrator (POC) for home/institutional use; provides supplemental oxygen via pulse dose. Inputs: ambient air drawn via piston compressor. Principle: pressure-vacuum swing adsorption using two molecular sieve beds; nitrogen adsorbed under pressure, oxygen collected in product tank. Output: pulse-dose oxygen delivered to patient upon inhalation detection. Features 'Smart Dose' algorithm to automatically increase oxygen setting if patient breathing rate increases. Operated by patient or caregiver. Healthcare provider uses output to manage oxygen therapy; benefits patient by providing ambulatory oxygen support. Device includes LCD for status/alerts; powered by AC or DC/battery.
Clinical Evidence
No clinical testing performed. Substantial equivalence demonstrated via bench testing using patient simulation models to verify oxygen output, breathing detection, and compliance with ISO/IEC standards.
Technological Characteristics
Pressure-vacuum swing adsorption; lithium-based zeolite molecular sieve. Piston compressor. Pulse-dose delivery. Power: 100-240 VAC or 12-24 VDC. Dimensions: 9.6"H x 8.75"W x 3.6"D. Weight: <6 lbs. Connectivity: None. Standards: IEC 60601-1 (3rd Ed), IEC 60601-1-2, ISO 80601-2-69, ISO 80601-2-67, ISO 14971, ISO 13485. Software: Microprocessor-based control for valve timing, oxygen monitoring, and Smart Dose algorithm.
Indications for Use
Indicated for adult and pediatric patients (>10kg) requiring supplemental oxygen. Not for life support; no patient monitoring capabilities.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Predicate Devices
DeVilbiss iGo 306DS Portable Oxygen Concentrator (K081468)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health and Human Services logo. The logo on the right is the Food and Drug Administration logo. The FDA logo is in blue and contains the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
March 23, 2018
DeVilbiss Healthcare, LLC Sandy Figueroa Manager, Regulatory Affairs 100 DeVilbiss Drive Somerset, Pennsylvania 15501
Re: K172648
Trade/Device Name: Drive DeVilbiss iGo 2 Portable Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: February 19, 2018 Received: February 20, 2018
Dear Sandy Figueroa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172648
Device Name
Drive DeVilbiss iGo2 Portable Oxygen Concentrator
Indications for Use (Describe)
The Drive DeVilbiss iGo2 Portable Oxygen Concentrator is indication of supplemental oxygen. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Name and address of the manufacturer and<br>sponsor of the 510(k) submission: | DeVilbiss Healthcare LLC<br>100 DeVilbiss Drive<br>Somerset, PA 15501 |
|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Official contact person for all<br>correspondence: | Betty Miller<br>Regulatory/Compliance Manager<br>Phone: 814-443-7606<br>Fax: 814-443-7575 |
| Date Prepared: | March 22, 2018 |
| Device Name: | Drive DeVilbiss iGo2 Portable Oxygen<br>Concentrator |
| Proprietary name of new device: | Drive DeVilbiss iGo2 Portable Oxygen<br>Concentrator |
| Common or usual name of the device: | Generator, Oxygen, Portable |
| DeVilbiss Model Number | 125D, 125K, 125AA |
| Classification | Class II |
| Panel Code: | CAW |
| CFR Regulation Number: | 21 CFR 868.5440 |
| Primary Predicate Device Name and 510(k)<br>number: | DeVilbiss iGo 306DS Portable Oxygen<br>Concentrator (K081468) |
| Reference Device Name and 510(k) number: | Model 350G Gas Conserver (K090421) |
# Description of Device:
The proposed Drive DeVilbiss iGo2 Portable Oxygen Concentrator (POC) is similar to the predicate devices, pressure-vacuum swing adsorption based on molecular sieve technology using two containers filled with molecular sieve (sieve beds). Ambient air is drawn into the concentrator via a piston style compressor and compressed into a single sieve bed. As pressure increases nitrogen is adsorbed and concentrated oxygen exits one end of the bed and is collected in a product tank. Simultaneously in the other sieve bed, nitrogen is desorbed as the pressure decreases to a vacuum and is exhausted into the atmosphere. Once the pressure or vacuum reaches a defined maximum a valve switches the compressor connections to the sieve beds. The sieve bed previously connected to the pressure head will be connected to vacuum. The oxygen collected in the product tank is then dispensed to the patient upon detecting a patient inhalation, much like an oxygen conserving device.
The proposed Drive DeVilbiss iGo2 POC provides pulse dose oxygen in the same manner as primary predicate device (DeVilbiss Model 306 POC, K081468). The "Smart Dose" feature has been added to the proposed device, which is substantially equivalent to the feature in the reference device (Model 350G gas
{4}------------------------------------------------
conserver, K090421). Smart Dose will automatically increase the patient's prescription setting by one if the system determines the patient has increased rate of breathing.
This device is intended for Adult and Pediatric patients with a bodyweight >10kg.
## Indication for Use:
The Drive DeVilbiss iGo2 Portable Oxygen Concentrator is indicated for the administration of supplemental oxygen. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
## Substantial Equivalence:
The proposed (modified) Drive DeVilbiss iGo2 Portable Oxygen Concentrator has the following similarities to those which previously received 510(k) concurrence:
- Indications for Use Statement
- Operating Principle
- Incorporates similar materials, and
- Is manufactured and packaged using similar materials and processes.
Design and verification activities were performed on the proposed Drive DeVilbiss iGo2 Portable Oxygen Concentrator according to design requirements and risk analysis outputs. All bench tests confirmed that the predetermined acceptance criteria were met. This includes oxygen output and breathing detection.
{5}------------------------------------------------
## Comparison of Substantial Equivalence and Differences:
| Characteristics | Predicate Device<br>DeVilbiss iGo 306DS Portable<br>Oxygen Concentrator -<br>K081468 | Reference Device<br>Model 350G Gas<br>Conserver<br>K090421 | Subject Device | Nature of Change | Legally<br>Marketed<br>Product with<br>same feature<br>or<br>specification |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| ltem<br>Description/<br>Method of<br>Operation | 1-3 lpm continuous flow (1-6<br>pulse dose), pressure-Vacuum<br>swing adsorption, oxygen<br>concentrator based on<br>molecular sieve technology.<br>Ambient air is drawn into the<br>concentrator via a piston style<br>compressor. The air then<br>passes through a series of<br>filters that remove dust,<br>bacteria, and other particulate.<br>A poppet-style valve directs air<br>into one of two sieve beds.<br>Nitrogen is adsorbed in the<br>bed as the pressure increases<br>while oxygen flows through,<br>thereby producing a highly<br>enriched oxygen product.<br>Simultaneously in the other<br>bed, nitrogen is absorbed as<br>the pressure decreases and is<br>exhausted into the<br>atmosphere. A momentary<br>intermediate pneumatic<br>sequence ties the beds<br>together with the exhaust<br>blocked for an enhanced<br>nitrogen purge. The cycle<br>continues providing a flow of<br>oxygen to the patient. Oxygen<br>is provided to the patient on a<br>continuous or pulse-dose<br>basis. | The Inspired<br>Technologies 350G<br>Gas Conserver is<br>intended as a<br>delivery device for<br>medial-grade<br>oxygen from high-<br>pressure oxygen<br>cylinders. | (1-5 pulse dose), pressure-<br>Vacuum swing adsorption,<br>oxygen concentrator based on<br>molecular sieve technology.<br>Ambient air is drawn into the<br>concentrator via a piston style<br>compressor. The air then<br>passes through a series of<br>filters that remove dust,<br>bacteria, and other particulate.<br>A valve directs air into one of<br>two sieve beds. Nitrogen is<br>adsorbed in the bed as the<br>pressure increases while<br>oxygen flows through, thereby<br>producing a highly enriched<br>oxygen product.<br>Simultaneously in the other<br>bed, nitrogen is absorbed as<br>the pressure decreases and is<br>exhausted into the<br>atmosphere. A momentary<br>intermediate pneumatic<br>sequence ties the beds<br>together with the exhaust<br>blocked for an enhanced<br>nitrogen purge. The cycle<br>continues providing oxygen to<br>the patient. Oxygen is<br>provided to the patient on a<br>pulse-dose basis. | Minor<br>constructional<br>differences. POC<br>will be smaller and<br>not support<br>continuous flow.<br>No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Target<br>Population/<br>Indications for<br>Use | The Model 306 Portable<br>Oxygen Concentrator is<br>indicated for the administration<br>of supplemental oxygen. The<br>device is not intended for life<br>support, nor does it provide<br>any patient monitoring<br>capabilities. | This is an<br>ambulatory device,<br>which allows<br>patients to<br>ambulate longer<br>than they would<br>with a continuous<br>flow regulator on<br>the same cylinder. | The Portable Oxygen<br>Concentrator is indicated for<br>the administration of<br>supplemental oxygen. The<br>device is not intended for life<br>support, nor does it provide<br>any patient monitoring<br>capabilities. | Same. Provides<br>pulse oxygen to<br>patient. No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Where Used/<br>Intended Use | Oxygen concentrators are<br>intended to provide<br>supplemental oxygen to<br>persons requiring low flow<br>oxygen therapy. The patient<br>typically receives the oxygen<br>through a nasal cannula. The<br>device delivers continuous<br>oxygen or oxygen pulses<br>during a patients inhalation<br>phase. The pulse flow<br>provides over 87% oxygen<br>concentration. It is used at a<br>patient's home or for their<br>portable needs outside the<br>home and can also be used in<br>institutions such as nursing<br>homes or sub-acute care<br>facilities. Oxygen<br>concentrators are not<br>considered life supporting nor<br>life sustaining. The device has<br>no contraindications. | The 350G Gas<br>Conserver is<br>intended to be<br>used in the<br>hospital, healthcare<br>facilities, or home<br>care environments. | Oxygen concentrators are<br>intended to provide<br>supplemental oxygen to<br>persons requiring low flow<br>oxygen therapy. The patient<br>typically receives the oxygen<br>through a nasal cannula. It is<br>used at a patient's home or for<br>their portable needs outside<br>the home and can also be<br>used in institutions such as<br>nursing homes or sub-acute<br>care facilities. Oxygen<br>concentrators are not<br>considered life supporting nor<br>life sustaining. The device has<br>no contraindications. | Same. No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>- K081468 |
| Characteristics | Predicate Device<br>DeVilbiss iGo 306DS Portable<br>Oxygen Concentrator –<br>K081468 | Reference Device<br>Model 350G Gas<br>Conserver<br>K090421 | Subject Device | Nature of Change | Legally<br>Marketed<br>Product with<br>same feature<br>or<br>specification |
| Operating<br>Principle | Pressure-vacuum-swing<br>adsorption / pressure<br>controlled oxygen separation<br>process | Integral high<br>pressure regulator<br>Dosing device and<br>algorithm Valves | Pressure-vacuum-swing<br>adsorption / pressure<br>controlled oxygen separation<br>process | Same as iGo<br>306DS. No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Standards<br>Certification | IEC 60601-1 2nd Edition<br>IEC 60601-1-1<br>IEC 60601-1-2<br>ISO 8359<br>ISO 14971<br>ISO 13485 | FCC CFR47 Parts<br>15B and 18<br>Industry Canada<br>BS EN 55011:2007<br>IEC 60601-1<br>IPX I | IEC 60601-1 3rd Edition<br>IEC 60601-1-1<br>IEC 60601-1-2<br>ISO 80601-2-69<br>ISO 80601-2-67<br>ISO 14971<br>ISO 13485 | New device<br>conforms to newer<br>revisions of<br>standards. No<br>different questions<br>of safety or<br>effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Output<br>Capacity and<br>Delivery Modes | 3 LPM maximum continuous<br>flow.<br>Continuous flow settings: 1 to<br>3.0 LPM in 0.5 LPM<br>increments<br>Pulse flow settings: 1.0 to 3.0<br>in 0.5 increments; 3.0 to 6.0 in<br>1.0 increments | 16.0 cc/lpm x<br>setting value( i.e.<br>16x 1=16, 16 x 2<br>=32 etc) at 1.0, 1.5,<br>2.0, 2.5, 3.0, 4.0,<br>5.0, and 6.0 Lpm<br>flow settings Sport<br>mode I is setting<br>value + 16 cc Sport<br>Mode 2 is setting<br>value +32 cc/lpm | Pulse flow settings: 1.0 to 5.0<br>in 1.0 increments | Does not support<br>Continuous flow.<br>Pulse mode like<br>iGo 306DS but<br>supports a smaller<br>set of the patient<br>population because<br>of smaller bolus<br>volume.<br>Supports Smart<br>Dose (Sport Mode) | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468<br>Model 350G<br>Gas<br>Conserver<br>K090421 |
| Voltage<br>Required | AC operation: 100-240 VAC,<br>50-60 Hz<br>DC operation: 12-24 VDC | 4 "AA" batteries | AC operation: 100-240 VAC,<br>50-60 Hz<br>DC operation: 10 - 15 VDC | Minimal difference.<br>No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>— K081468 |
| Dimensions | 15"H x 12"W x 9"D<br>Cube volume: 1620 cu. in. | 6"H x 3.5"W x 4"D | 9.6"H x 8.75"W x 3.6"D<br>Cube volume: 302.4 cu. in. | Smaller more<br>portable device<br>than iGo 306DS.<br>No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>- K081468 |
| Weight | 17.9 lbs, including battery,<br>19.9 lbs with lightweight cart | 1.39 lbs | Less than 6 lbs, including a<br>standard battery pack. | Lighter more<br>portable device<br>than iGo 306DS.<br>No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| 02<br>Performance<br>Specification | 90% +/-3% at 1 Ipm to 3 Ipm<br>at sea level | N/A - Uses bottled<br>gas | 87-94% at all settings | Same as iGo<br>306DS. No<br>different questions<br>of safety or<br>effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>- K081468 |
| Noise Level @<br>2 Ipm | 40 dBA at pulse flow setting 3;<br>47 dBA overall maximum at 3<br>LPM continuous flow | N/A – Uses bottled<br>gas | 46 dBA at pulse flow setting 2 | Louder than iGo<br>306DS but no<br>different questions<br>of safety or<br>effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>- K081468 |
| Power<br>Consumption | 42 Watts at 1.0 LPM<br>continuous flow<br>104 Watts at 3.0 LPM<br>continuous flow<br>31 Watts at 1.0 pulse flow<br>setting (20 BPM)<br>62 Watts at 6.0 pulse flow<br>setting<br>(20 BPM) | | Approximately 30 Watts at<br>setting of 2 (20 BPM) | Pulse dose<br>efficiency is better.<br>No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Characteristics | Predicate Device<br>DeVilbiss iGo 306DS Portable<br>Oxygen Concentrator –<br>K081468 | Reference Device<br>Model 350G Gas<br>Conserver<br>K090421 | Subject Device | Nature of Change | Legally<br>Marketed<br>Product with<br>same feature<br>or<br>specification |
| Patient Oxygen<br>Outlet<br>Pressure | 5 psig | 19-25 PSIG | 15 psig | Higher pulsed<br>output pressure<br>than iGo but lower<br>than 350G. Will<br>deliver oxygen<br>faster than iGo but<br>there is no different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Filtration | Patient-maintainable gross<br>particle filter | None | Gross particle filter | Same as iGo<br>306DS. No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Battery Run<br>Time | Tested times:<br>4.1 hours at 1.0 LPM<br>1.7 hours at 3.0 LPM<br>5.8 hours at 1.0 LPM (10<br>BPM)<br>2.0 hours at 6.0 LPM (30<br>BPM) | NA | 4.5 hours at a setting of 2 (20<br>BPM) | Same as iGo<br>306DS at<br>equivalent pulse<br>output. No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Compressor<br>Pressure Relief<br>Valve | Pressure Relief valve | NA | Over-pressure protection<br>achieved through software<br>and electronic control. | No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Pneumatic<br>Valve | Two Three way valves | NA | Two Three way valves | Off the shelf valves<br>– No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Molecular<br>Sieve Material | Zeochem lithium-based zeolite | NA | Zeochem lithium-based zeolite | Same as iGo<br>306DS. No<br>different questions<br>of safety or<br>effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Flow Controller | Proportional control valve<br>used for continuous flow and<br>pulse flow | Pulse Valve | Pulse Valve only. Use<br>software to regulate output. | Same as iGo<br>306DS. No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Insulation<br>Class | IEC Class 1, Type BF | Not Published | IEC Class 2, Type BF | Improved required<br>by ISO standard.<br>No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Water Ingress | IPX0 – bare unit<br>IPX1 – Unit with battery | IPX1 | IPX1<br>Or IPX22 | Improved liquid<br>ingress capability.<br>No different<br>questions of safety<br>or effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Hour meter | Mechanical Hour meter | None | Microprocessor or other<br>nonvolatile memory. | Used for service<br>and dealer. No<br>different questions<br>of safety or<br>effectiveness. | DeVilbiss iGo<br>306DS<br>Portable<br>Oxygen<br>Concentrator<br>– K081468 |
| Characteristics | Predicate Device<br>DeVilbiss iGo 306DS Portable<br>Oxygen Concentrator –<br>K081468 | Reference Device<br>Model 350G Gas<br>Conserver<br>K090421 | Subject Device | Nature of Change | Legally<br>Marketed<br>Product with<br>same feature<br>or<br>specification |
| Oxygen<br>Monitor | Ultrasonic oxygen and flow<br>sensing device (FDA<br>approved Re: K913392/B)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.