SYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-4 Software Version 1

K172618 · Olympus Medical Systems Corp. · GCJ · Dec 5, 2017 · Gastroenterology, Urology

Device Facts

Record IDK172618
Device NameSYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-4 Software Version 1
ApplicantOlympus Medical Systems Corp.
Product CodeGCJ · Gastroenterology, Urology
Decision DateDec 5, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

This medical control unit for endosurgery has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment.

Device Story

UCES-4 is a medical control unit for endosurgery; functions as a central hub for Olympus endoscopes and ancillary equipment. Inputs include user commands from a touch panel and status signals from connected ancillary devices. Device transforms inputs into control commands and status displays; enables centralized operation, automatic initial settings, and interlocking of equipment. Used in clinical settings (e.g., OR) by physicians or clinical staff. Healthcare providers interact via touch panel to monitor equipment status and adjust settings centrally. Benefits include streamlined workflow, reduced manual adjustments, and improved coordination of endosurgical equipment.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including software validation, electromagnetic compatibility, electrical safety, and human factors/usability engineering.

Technological Characteristics

Central control unit with power supply, main control circuit, communication circuit, and data storage. Connectivity via interface units (MAJ-2191, MAJ-2249, MAJ-2250, MAJ-2251, MAJ-2268). Standards applied: IEC 60601-1:2005+A1, IEC 60601-1-2:2014, IEC 60601-2-18:2009, ISO 14971:2007. No patient-contacting materials.

Indications for Use

Indicated for use with Olympus endoscopes and ancillary equipment to provide central operation, display, automatic initial setting, and interlocking control during endosurgical procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. December 5, 2017 Olympus Medical Systems Corp. % Sheri L. Musgnung Regulatory Affairs Manager Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610 Re: K172618 > Trade/Device Name: SYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-4 Software Version 1 Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ, ODA Dated: November 1, 2017 Received: November 2, 2017 Dear Sheri L. Musgnung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clear and easy to read. Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K172618 Device Name SYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-4 Software Version 1 Indications for Use (Describe) This medical control unit for endosurgery has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |------------------------------------------------------------------------------------------| | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Below the line is the text "Your Vision, Our Future". # 510(k) SUMMARY # SYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-4 Software Version 1 August 30, 2017 ## I. General Information - 1 Applicant: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047 - 1 Official Correspondent: Sheri L. Musgnung Regulatory Affairs Manager Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147Fax: 484-896-7128 Email: sheri.musgnung@olympus.com - 1 Manufacturer: SHIRAKAWA OLYMPUS CO., LTD. 3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura, Nishishirakawa-gun, Fukushima, 961-8061, Japan Establishment Registration Number: 3002808148 #### II. Device Identification - Device Trade Name: SYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-4 Software Version 1 - . Common Name: Endosurgery system - . Regulation Number: 876.1500 - I Regulation Name: Endoscope and accessories - . Regulatory Class: II - . Classification Panel: Gastroenterology and urology {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. A thin, yellow line is underneath the word. Below the line, the words "Your Vision. Our Future." are written in a smaller font. - . Product Code: GCJ and ODA #### III. Predicate Device Information - . Device Name: SYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-3 Software Version 4 - . Common Name: Endosurgery system - . Regulation Number: 876.1500 - . Regulation Name: Endoscope and accessories - . Regulatory Class: II - . Applicant: OLYMPUS MEDICAL SYSTEMS CORP. - . 510(k) Number: K121701 ## IV. Device Description The SYSTEMS INTEGRATION ENDOALPHA MEDICAL CONTROL UNIT FOR ENDOSURGERY UCES-4 Software Version 1 (hereinafter referred to as "UCES-4") has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment. - System Components The UCES-4 consists of the following components. | Model | Device description | |----------|--------------------------------------| | UCES-4 | MEDICAL CONTROL UNIT FOR ENDOSURGERY | | MAJ-2191 | EXTENSION UNIT FOR UCES-4 | | MAJ-2249 | INTERFACE FOR ESU (VIO SERIES) | | MAJ-2250 | INTERFACE FOR ESU (FORCE SERIES) | | MAJ-2251 | UNIVERSAL INTERFACE | | MAJ-2268 | INTERFACE FOR OLYMPUS LINK DEVICE | #### Structure The UCES-4 has built-in power supply circuit, main control circuit, communication control circuit, and data storage device. Through communications with each of the K172618 Page 2 of 4 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, horizontal, yellow line. Under the line are the words "Your Vision. Our Future" in a smaller font. ancillary equipment, it provides central display of the ancillary equipment status and performs centralized operation of the ancillary equipment via the touch panel. - Principle Communicating with the connected ancillary equipment, the UCES-4 displays the information of the communications on the touch panel. It sends input commands from input devices such as the touch panel to the ancillary equipment to perform centralized control. # V. Indications for Use This medical control unit for endosurgery has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment. # VI. Comparison of Technological Characteristics The UCES-4 has the same technological characteristics as the predicate device as follows: - · Operating principle - · Electrical characteristic - · Mechanical characteristic - Communication characteristic . - · Energy source - · Material (no patient contacting material) The UCES-4 has modified the software and hardware, and the following technological characteristics are different from the predicate device. - · Performance - Graphical user interface - · Compatibility with other devices The validation testing demonstrated that these differences do not affect the safety and effectiveness of the subject device. # VII. Summary of Non-Clinical Testing Non-clinical testing was not conducted since the basic technological characteristics of the subject device are identical to the predicate device. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for OLYMPUS. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Underneath the line, the words "Your Vision, Our Future" are written in a smaller, gray font. The differences of technological characteristics between the predicate device and the subject device are confirmed that they are substantially equivalent through the following tests and standards. - · The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". - · The electromagnetic compatibility and electric safety had been confirmed. - · The risk management was performed in accordance with established in-house acceptance criteria based on ISO 14971:2007. - · The human factors activities were performed in accordance with the FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices". The following voluntary standards have been applied to the UCES-4; - · IEC 60601-1: 2005+A1 - · IEC 60601-1-2: 2014 - · IEC 60601-2-18: 2009 - · ISO 14971: 2007 # VIII. Conclusion Compared to the legally marketed predicate device. the UCES-4 does not incorporate any significant changes in the indications for use, method of operation, material, or design that could affect the safety or effectiveness of the subject device. K172618 Page 4 of 4
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%