Spectra Optia Apheresis System
K172590 · Terumobct, Inc. · LKN · Mar 2, 2018 · GU
Device Facts
| Record ID | K172590 |
| Device Name | Spectra Optia Apheresis System |
| Applicant | Terumobct, Inc. |
| Product Code | LKN · GU |
| Decision Date | Mar 2, 2018 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic, Real-World Evidence, Pediatric |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|
| K172590 · Mar 2, 2018 | Spectra Optia Apheresis System | Terumobct, Inc. | Retrospective clinical routine-use data from 3 hospital sites | Retrospective analysis of 58 WBC depletion procedures in 43 subjects to evaluate the performance and safety of the Spectra Optia Apheresis System in a clinical setting. | Retrospective study; WBC depletion; Leukocytosis; Routine clinical use |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|
| WBC Depletion clinical study; Multicenter, single-arm, retrospective data collection study; Follow-up/Duration: During procedure and up to 24 hours post-procedure | Patients with leukocytosis at risk for leukostasis (leukemia patient population); Sample Size: 43 subjects (58 procedures); Number of Sites: 3 | Not applicable for this study | Percent decrease in WBC count; percent of processed WBCs collection efficiency (CE); adverse events (AEs) |
Indications for Use
The Spectra Optia® Apheresis System, a blood component separator, may be used to perform therapeutic plasma exchange. The Spectra Optia® Apheresis System, a blood component separator. may be used to perform Red Blood Cell Exchange (RBCX) procedures for the transfusion management of Sickle Cell Disease in adults and children. The Spectra Optia® Apheresis System, a blood component separator, may be used to reduce White Blood Cells for patients with leukocytosis at risk for leukostasis.
Device Story
Automated centrifugal blood component separator; separates whole blood into cellular and plasma components. System comprises apheresis machine (centrifuge, pumps, valves, control systems), sterile single-use disposable tubing set, and embedded software. Used in clinical settings by trained operators. During white blood cell depletion (WBCD), system pumps patient blood into tubing; centrifuge spins at speed targeting packing factor of 4.5. Automated Interface Management (AIM) system monitors interface, adjusts plasma pump flow rate, and controls collect valve based on cell detection. Collects target cells into bag; returns plasma and RBCs to patient. Provides automated, monitored procedure; reduces operator burden compared to manual systems; effectively decreases circulating WBCs in leukemia patients.
Clinical Evidence
Multicenter, single-arm, retrospective study of 58 WBC depletion procedures in 43 subjects. Primary effectiveness endpoints: percent decrease in WBC count (mean 54.7%, SD 21.3%) and collection efficiency (mean 58.7%, SD 16.1%). Safety endpoint: adverse events (AEs) during and 24 hours post-procedure. No serious adverse events or unanticipated adverse device effects reported. Observed AEs (thrombocytopenia, anemia, hypocalcemia, hypokalemia) consistent with patient population. Results comparable to literature.
Technological Characteristics
Automated centrifugal separator. Components: apheresis machine, sterile single-use disposable tubing set, embedded software. Sensing: Automated Interface Management (AIM) system monitors interface and cell concentration. Connectivity: Standalone. Sterilization: Not specified. Algorithm: Rule-based control of pump flow rates and valve positioning based on real-time interface monitoring.
Indications for Use
Indicated for therapeutic plasma exchange, Red Blood Cell Exchange (RBCX) for sickle cell disease management in adults and children, and white blood cell reduction in patients with leukocytosis at risk for leukostasis.
Predicate Devices
- COBE® Spectra Apheresis System (K831004)
- Spectra Optia® Apheresis System (K153601)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 2, 2018
Terumo BCT, Inc. Nicholas Wong, RAC Sr. Regulatory Affairs Specialist 10811 W. Collins Avenue Lakewood, Colorado 80215-4415
Re: K172590
Trade/Device Name: Spectra Optia® Apheresis System Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LKN Dated: January 31, 2018 Received: February 1, 2018
Dear Nicholas Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172590
Device Name Spectra Optia® Apheresis System
### Indications for Use (Describe)
The Spectra Optia® Apheresis System, a blood component separator, may be used to perform therapeutic plasma exchange.
The Spectra Optia® Apheresis System, a blood component separator. may be used to perform Red Blood Cell Exchange (RBCX) procedures for the transfusion management of Sickle Cell Disease in adults and children.
The Spectra Optia® Apheresis System, a blood component separator, may be used to reduce White Blood Cells for patients with leukocytosis at risk for leukostasis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary for the Spectra Optia® Apheresis System is provided below.
#### SUBMITTER 1.
Terumo BCT, Inc. 10811 W. Collins Avenue Lakewood, Colorado 80215 Phone: 720-480-6702
Contact Person: Nicholas Wong Title: Sr. Regulatory Affairs Specialist Phone: 303-239-2384 Fax: 303-231-4756
Date Prepared: March 2, 2018
#### 2. DEVICE
| Trade Name of Device: | Spectra Optia® Apheresis System |
|-----------------------|---------------------------------|
| Common or Usual Name: | Apheresis Device or System |
| Regulation Number: | N/A |
| Regulation Name: | N/A |
| Classification Name: | N/A |
| Regulatory Class: | Unclassified |
| Product Code: | LKN |
#### 3. PREDICATE AND REFERENCE DEVICE
Terumo BCT's COBE® Spectra Apheresis System (K831004) Predicate Device: Terumo BCT's Spectra Optia® Apheresis System (K153601) Reference Device:
#### DEVICE DESCRIPTION 4.
As a whole, the Spectra Optia® Apheresis System has been cleared for the following therapeutic apheresis, cell collection and cell processing procedures:
- Therapeutic Apheresis Procedures: ●
- -Therapeutic Plasma Exchange (TPE) - K071079, K153601
- -Red Blood Cell Exchange (RBCX) - K132429, K153601
- Collection/Processing Procedures (cleared by CBER):
- -Mononuclear Cell (MNC) Collection - BK120012 & BK150251
- -Granulocyte Collection - BK130065
{4}------------------------------------------------
- Bone Marrow Processing BK140191 -
The Spectra Optia® Apheresis System is an automated centrifugal system that separates whole blood into its cellular and plasma components. The cleared system is comprised of three major subsystems:
- 1. the apheresis machine itself (centrifuge, centrifuge filler, pumps, valves, computerized safety and control systems, etc.)
- 2. a sterile, single-use, disposable blood tubing set
- 3. embedded software
This 510(k) involves no change in the components (design or material) of the Spectra Optia® Apheresis System. Instead, the purpose of this 510(k) is to expand the indications for use to include reduction of White Blood Cells (WBCs) for patients with leukocytosis at risk of leukostasis.
During a WBCD procedure, the system pumps the patient's blood into the tubing set and spins the centrifuge at a speed required to target the optimal packing factor (4.5) for a WBCD procedure using an anticoagulant (ACD-A). The Automated Interface Management (AIM) system adjusts the flow rate of the plasma pump to control the concentration of cells that flow through the collect port. This step can be controlled via the collection preference. When the AIM system detects cells in the collect port, the system sounds a tone, the collect valve moves into the collect position, and the collect pump pumps the cells into the collection bag. The plasma pumps plasma out of the channel and into the reservoir. The red blood cells (RBC) are pushed out of the channel and into the reservoir, where they combine with the plasma for return to the patient.
#### INTENDED USE/INDICATIONS FOR USE 5.
The Spectra Optia® Apheresis System has been previously cleared for therapeutic plasma exchange (TPE) and Red Blood Cell Exchange (RBCX). This 510(k) involves an expansion of the indications for use, which is reflected below in italics.
# Indication Statement:
The Spectra Optia® Apheresis System, a blood component separator, may be used to perform therapeutic plasma exchange.
The Spectra Optia® Apheresis System, a blood component separator, may be used to perform Red Blood Cell Exchange (RBCX) procedures for the transfusion management of Sickle Cell Disease in adults and children.
The Spectra Optia® Apheresis System, a blood component separator, may be used to reduce White Blood Cells for patients with leukocytosis at risk of leukostasis.
Compared to the predicate. COBE Spectra, the device has the same intended use as a blood component separator. The predicate COBE Spectra device was previously cleared for a specific indication of therapeutic cellular depletion. The proposed modification to the indications for use falls within the general cleared indication of the predicate device.
{5}------------------------------------------------
#### TECHNOLOGICAL COMPARISON 6.
Spectra Optia System simplifies the WBC Depletion Protocol when compared to the predicate system with the use of on-screen guidance and an integrated cassette tubing set design. The Spectra Optia System key performance characteristics determined by laboratory evaluation for the WBC Depletion Protocol when compared to the predicate are:
- A Mean Collection Efficiency for WBC of 83.6% (±18.8%) compared to 70.6% ● (±12.9%)
- Can collect under a wide range of patient/procedural parameters including hematocrit, . inlet flow rate, and collect flow rate
- Allows operators to perform other tasks during the procedure as the system . continuously monitors and adjusts the interface
The technology for the Spectra Optia System is not identical to the predicate device; however, the differences do not impact on substantial equivalence.
#### PERFORMANCE DATA 7.
Both bench and clinical data were provided in support of substantial equivalence of the modified indications for use for the Spectra Optia® Apheresis System.
#### 7.1. Bench Testing
Terumo BCT performed verification bench testing that demonstrates that the system works as intended. Testing was done in comparison to the predicate device.
The testing was performed over a broad range of operational setting/parameters using blood products that vary with respect to clinical conditions for WBC content. The testing demonstrated that the Spectra Optia® WBC Depletion Protocol is non-inferior to the predicate the COBE Spectra WBC Depletion Protocol. The software, filler, and disposable are able to reduce WBCs effectively over a wide range of patient/procedural parameters including hematocrit (HCT), inlet flow rate (Qin), and collect flow rate.
#### 7.2. Clinical Data
Clinical testing of the Spectra Optia® Apheresis System WBC Depletion Protocols was performed.
The WBC Depletion clinical study was a multicenter, single-arm, retrospective data collection study to evaluate the routine-use performance and safety of WBC Depletion conducted with the Spectra Optia® Apheresis System. Clinical routine-use data were collected retrospectively from 58 WBC Depletion procedures performed on 43 subjects.
#### 7.2.1. Location of Study
The retrospective study was conducted at 3 sites in Europe: UZ Leuven, Belgium, Institute for Transfusion Medicine and Immunohematology, Frankfurt, Germany and Saint Istvan and Saint Laszlo Hospital, Budapest, Hungary.
{6}------------------------------------------------
#### 7.2.2. Primary Effectiveness Endpoint
The primary effectiveness endpoint of the study included evaluation of Device Performance Parameters including percent decrease in WBC count in subject blood following apheresis procedures and percent of processed WBCs collection efficiency (CE) for WBC achieved by the Spectra Optia® Apheresis System.
#### 7.2.3. Primary Safety Endpoint
The primary safety endpoint consisted of an evaluation of adverse events (AEs) during the WBC Depletion apheresis procedure for up to 24 hours post-procedure.
#### 7.2.4. Effectiveness
Data were collected for 43 subjects who underwent a total of 58 WBC Depletion procedures at 3 sites.
The criteria for WBC Depletion are determined by the patient's WBC counts and clinical status. The mean percent decrease in subject WBC counts post-procedure was 54.7% (SD: 21.3%). The observed mean percent decrease in WBC count was consistent with American Society for Apheresis (ASFA) guidelines, where a single leukocytapheresis procedure is reported to reduce WBC count by 30% to 60%.
The CE of the WBC Depletion procedures as measured from the waste bag contents was 58.7% (SD: 16.1%).
#### Safety 7.2.5.
The frequency and type of AEs observed is reflective of the very sick leukemia patient population in which WBC Depletion procedures are typically performed, with no observed safety signals associated with the Spectra Optia® device. The most common Treatment Emergent Adverse Events (TEAEs) (i.e., those reported for > 5% of subjects) were related to hematology or serum chemistry abnormalities, which are expected AEs within this leukemia patient population, including thrombocytopenia and/or decreased platelet count, anemia, hypocalcemia, and hypokalemia. No serious adverse events (SAEs) or unanticipated adverse device effects (UADEs) were reported. One subject (2.3%) terminated the WBC Depletion procedure prematurely due to a moderate TEAE of vascular access complication.
#### Summary 7.2.6.
The data collected within this study indicated that WBC Depletion procedures using the Spectra Optia® Apheresis System were well tolerated and effective at decreasing circulating WBCs in a sick leukemia patient population, as evaluated by percent decrease in WBC count, CE for WBC, and AEs during and within 24 hours post-procedure. Percent decrease in WBC count observed in the study was comparable to results reported in the literature.
#### CONCLUSIONS 8.
Based on the non-clinical and clinical testing, the proposed Spectra Optia Apheresis System with the modified indication for use has been determined to be as safe and effective as the legally marketed predicate device and therefore, may be found substantially equivalent.