K172535 · Olympus Winter & Ibe GmbH · GEI · Nov 16, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172535
Device Name
Bipolar applicator CELON ProCurve V
Applicant
Olympus Winter & Ibe GmbH
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The bipolar applicator, in combination with the compatible electrosurgical generator, is indicated for endovascular coagulation of blood vessels in patients with superficial vein reflux.
Device Story
Bipolar applicator CELON ProCurve V uses Radiofrequency Induced Thermotherapy (RFITT) for endovascular coagulation of varicose veins. Device connects to compatible electrosurgical generator; operator pulls applicator slowly and steadily through vein while continuously activated to achieve controlled tissue coagulation. Used in clinical settings by healthcare providers. Benefits include effective treatment of superficial vein reflux via thermal coagulation. Device is single-use, sterile, and disposable.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-2), biocompatibility (ISO 10993), and preclinical simulated use studies. Simulated use evaluated tissue effects (porcine muscle, liver, kidney) and intraluminal performance across three generator modes and power settings, demonstrating comparable performance to the predicate.
Technological Characteristics
Bipolar electrosurgical applicator; RFITT technology. Materials tested per ISO 10993. Sterilized via gamma radiation (ISO 11137-1). Complies with IEC 60601-1, IEC 60601-2-2, and ISO 10555-1. Single-use, sterile, disposable. No software.
Indications for Use
Indicated for endovascular coagulation of blood vessels in patients with superficial vein reflux.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 16, 2017 Olympus Winter & Ibe GmbH % Dolan Mills Sr. Specialist, Regulatory Affairs Gyrus ACMI, Inc. 136 Turnpike Road Southborough, Massachusetts 01772
Re: K172535
Trade/Device Name: Bipolar applicator "CELON ProCurve V" Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 10, 2017 Received: November 13, 2017
Dear Dolan Mills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director
Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K172535
Device Name
Bipolar applicator "CELON ProCurve V"
Indications for Use (Describe)
The bipolar applicator, in combination with the compatible electrosurgical generator, is indicated for endovascular coagulation of blood vessels in patients with superficial vein reflux.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Notification
# K172535
## 510(k) Summary of Safety and Effectiveness
## November 10, 2017
# 2.1 General Information
| Manufacturer: | Olympus Winter & Ibe GmbH<br>Kuehnstr. 61<br>22045 Hamburg<br>Germany<br>Establishment Registration Number: 9610773 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Dolan Mills<br>Sr. Specialist, Regulatory Affairs<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 901.373.0236<br>Fax: 901.373.0220<br>Email: dolan.mills@olympus-osta.com<br>Establishment Registration Number: 300379030 |
## 2.2 Device Identification
| Proprietary name: | Bipolar applicator "CELON ProCurve V" |
|-----------------------------|---------------------------------------------------------|
| Common name: | CELON ProCurve V |
| Device Classification name: | Electrosurgical, Cutting & Coagulation &<br>Accessories |
| Review Panel: | General & Plastic Surgery |
| Regulation Number: | 21 CFR 878.4400 |
| Device class: | Class II |
| Product code: | GEI |
Model Number: WB990207
## 2.3 Predicate Device
The CELON ProCurve V is considered substantially equivalent to K111887, GEI, 21CFR 878.4400:
Covidien (Medtronic) CLOSUREFAST™ ENDOVENOUS RADIOFREQUENCY ABLATION CATHETER
{4}------------------------------------------------
According to public information this device is currently available in the US market and is designed to be used for endovascular coagulation of blood vessels in patients with superficial vein reflux.
No reference devices were used in this submission.
# 2.4 Device Description
The CELON ProCurve V is a bipolar applicator that uses Radiofrequency Induced Thermotherapy (RFITT) technology for treating varicose veins. The medical purpose of the Radiofrequency Induced Thermotherapy is to achieve controlled tissue coagulation. The CELON ProCurve V is used in conjunction with a compatible generator.
# 2.5 Indications for Use
The subject device has the same intended use as the predicate device.
"The bipolar applicator, in combination with the compatible electrosurgical generator, is indicated for endovascular coagulation of blood vessels in patients with superficial vein reflux."
# 2.6 Comparison of Technological Characteristics
The technological characteristics of the CELON ProCurve V are considered equivalent to the predicate device. Both the CELON ProCurve V and the predicate device share the following characteristics:
- fundamental scientific technology (bipolar electrosurgery)
- principles of operation (endovascular coagulation of blood vessels with a ● catheter electrode)
- used in conjunction with an electrosurgical generator.
- Comparative testing demonstrated that the subject device has equivalent performance to the predicate in respect to applicable modes.
The following differences to the predicate device exist:
- While using radiofrequency technology for coagulation, the predicate . device uses the heat generated through the radiofrequency current.
- While the CELON ProCurve V is pulled slowly and steadily back through the vein and activated the whole time, the predicate device is placed at the first desired position and activated. Afterwards the predicate devices is
{5}------------------------------------------------
pulled back further until the next desired position within the vein and activated again.
- minor dimensional differences exist between the subject and the predicate device
As stated above, the subject and proposed predicate devices have similar design characteristics and show comparable performance. As demonstrated in the nonclinical testing (e.g., biocompatibility, tissue effects), the different technological characteristics do not negatively affect the safety and effectiveness of the subject device.
## 2.7 Performance Data
The following performance data were provided in support of the substantial equivalence determination. All standards applied are FDA recognized international standards.
Performance data was prepared in accordance with the FDA guidance. Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff, issued on August 15, 2016. The guidance was followed for all relevant sections.
## 2.7.1 Biocompatibility testing
The patient contacting materials of the CELON ProCurve V have been successfully tested or assessed for biocompatibility in the following tests in compliance with ISO 10993 standards:
- -Cytotoxicity
- Chemical Analysis -
- Sensitization assessed by biological-toxicological evaluation -
- -Irritation – assessed by biological-toxicological evaluation
- Acute systemic toxicity assessed by biological-toxicological evaluation
- In vitro Hemocompatibility (thrombogenicity) -
- Hemolysis -
Testing was done in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"", CDRH June 16, 2016.
The device is considered external communicating, blood path / indirect, for a duration of less than 24 hours.
{6}------------------------------------------------
Traditional 510(k) Notification
### 2.7.2 Electrical safety and electromagnetic compatibility (EMC)
The design of the CELON ProCurve V complies with recognized 60601 standards as listed in section 2.7.8.
The FDA guidance "Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices", CDRH July 11, 2016 has been followed.
### Thermal Safety 2.7.3
The design of the CELON ProCurve V complies with recognized standard IEC 60601-2-2 as listed in section 2.7.8.
### 2.7.4 Clinical Studies
Clinical and animal studies were not necessary.
### 2.7.5 Software
The CELON ProCurve V does not contain software.
### Performance Testing Bench 2.7.6
Performance tests were executed to ensure that the device functioned as intended and met design specifications. The following non-clinical and preclinical tests were conducted:
1) non-clinical (electrical, dimensional, functional, biocompatibility, stability) 2) preclinical (simulated use) evaluation and testing of tissue effects and thermal safety
Non-clinical: Basic safety and performance testing was performed in accordance with IEC standards. In addition, verification and comparison bench studies were conducted to evaluate the functional performance.
Stability: Sterile samples were subjected to accelerated aging to confirm that the disposable devices maintain functionality and continue to meet specifications over time. The results of the accelerated age testing demonstrate that the device will be stable for the stated shelf-life. In addition, real time age testing will confirm the results of the accelerated age testing. Samples were also subjected to environmental conditioning and ship testing.
Preclinical: Evidence obtained from preclinical simulated use studies demonstrate that the device performs substantially equivalent to the predicate device in relevant aspects associated with usability, tissue effects, and thermal effects. For simulated use testing, three tissue types (porcine muscle, liver, and kidney) were evaluated in all applicable modes. The quality of intraluminal use was also verified for the applicable modes. The tissue effects testing included quantitative and qualitative assessment.
The ex vivo bench tests were based on three generator modes and three power settings which were applied five times to each of three different types of tissue. The
{7}------------------------------------------------
testing demonstrated comparable performance and tissue effects to the predicate. Additionally, in order to verify the quality of the intraluminal use ten ex vivo bench tests based on three generator modes and three power settings were applied to intraluminal animal tissue which also showed comparable performance and tissue effects. All tests passed.
To demonstrate substantial equivalence tissue effects were compared during comparative validation testing. The simulated use testing established substantial equivalence with the predicate device for the intended use based on an assessment of non-clinical performance data, as per 807.92(b).
These comprehensive validation bench tests support equivalence to the predicate device. Testing confirmed that comparable tissue effects could be achieved for applicable modes of operation.
Usability and user interface was also assessed according to the risk management plan. The assessment was based on Olympus predecessor product. Use-related hazardous situations were assessed and risk mitigation measures in terms of usability design for safety were defined. The residual risk was evaluated as acceptable. Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007.
#### 2.7.7 Reprocessing/ Sterilization
The CELON ProCurve V is sold in sterile condition and is defined as single use only and to be disposed of after use. Sterilization is performed according to ISO 11137-1 and packaging conforms to ISO 11607-1. The ionizing radiation (GAMMA rays) sterilization cycle has been validated. A sterility assurance level (SAL) of 10° was reached during validation and will be used for routine sterilization in compliance with regulations in force for sterile medical devices. Shelf Life testing was conducted, including performance testing and package integrity testing, to support a shelf life of 3.25 years for the CELON ProCurve V.
| Standard No. | Standard Title | FDA-Recognition<br>no + date |
|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| AAMI/ANSI ES<br>60601-1:2005 and<br>A1:2012, C1:2009<br>and A2:2010 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance<br>(IEC 60601-1:2005, MOD) | 19-4<br>07/09/2014 |
#### 2.7.8 Applied standards
{8}------------------------------------------------
| Standard No. | Standard Title | FDA-Recognition<br>no + date |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| IEC 60601-1-2<br>Ed. 3.0 b: 2007 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance -<br>Collateral Standard: Electromagnetic disturbances -<br>Requirements and tests | 19-1<br>06/27/2016 |
| IEC 60601-2-2<br>Ed. 5.0: 2009 | Medical electrical equipment - Part 2-2: Particular<br>requirements for the basic safety and essential<br>performance of high frequency surgical equipment and<br>high frequency surgical accessories [Including:<br>Technical Corrigendum 1 (2014)] | 6-336<br>01/27/2015 |
| IEC 60601-1-6<br>Ed. 3.1: 2013 | Medical electrical equipment Part 1-6: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Usability | 5-89<br>06/27/2016 |
| ISO 14971<br>Second edition 2007 | Medical devices - Application of risk management to<br>medical devices | 5-40<br>06/27/2016 |
| ISO 10993-1 Fourth<br>Edition 2009-10-15 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process | 2-220<br>07/26/2016 |
| ISO 11607-1 First<br>Edition 2006 | Packaging for terminally sterilized medical devices - Part<br>1: Requirements for materials, sterile barrier systems and<br>packaging systems [Including: Amendment 1 (2014)] | 14-454<br>01/27/2015 |
| ISO 11137-1 First<br>Edition 2006 | Sterilization of health care products - Radiation - Part 1:<br>Requirements for development, validation and routine<br>control of a sterilization process for medical devices<br>[Including: Amendment 1 (2013)] | 14-428<br>04/04/2016 |
| IEC 62366 Edition 1.1<br>2014-01 | Consolidated Version Medical Devices - Application Of<br>Usability Engineering To Medical Devices. | 5-87<br>12/23/2016 |
| ISO 10993-4 Second<br>Edition 2002-10-15<br>AMENDMENT 1 | Biological Evaluation Of Medical Devices - Part 4:<br>Selection Of Tests For Interaction With Blood<br>[Including: Amendment 1 (2006)]. | 2-235<br>07/26/2016 |
| AAMI ANSI ISO<br>10993-5:2009/ (R)2014 | Biological Evaluation Of Medical Devices - Part 5: Tests<br>For In Vitro Cytotoxicity | 2-153<br>07/26/2016 |
| ISO 10993-12 Fourth<br>Edition 2012-07-01 | Biological Evaluation Of Medical Devices - Part 12:<br>Sample Preparation And Reference Materials | 2-191<br>07/26/2016 |
| ISO 10555-1<br>Second Edition<br>2013.06 15 | Intravascular Catheters -- Sterile And Single-Use<br>Intravascular Catheters -- Part 1: General Requirements.<br>(General Plastic Surgery/General Hospital) | 6-301<br>08/14/2015 |
Table 2.1.8: Applied standards
# 2.8 Conclusion
The performance data supports the safety of the device and demonstrates that the subject device complies with the recognized standards as specified.
In summary, we believe the CELON ProCurve V is substantially equivalent to the predicate device with respect to the general design approach, function, and the intended use. Differences between subject device and predicate do not negatively affect the safety and effectiveness of the subject device and raise no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.