← Product Code LJE · K172527

# Amplatz Renal Dilator Set, Catheter for Use with Amplatz Renal Dilator Set, Amplatz Dilator, Amplatz Renal Introducer, Amplatz Radiopaque TFE Sheath, Clear Amplatz Sheath with Radiopaque Stripe (K172527)

_Cook Incorporated · LJE · Oct 19, 2017 · GU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172527

## Device Facts

- **Applicant:** Cook Incorporated
- **Product Code:** LJE
- **Decision Date:** Oct 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** GU
- **Attributes:** Therapeutic

## Intended Use

Amplatz Renal Dilator Set is intended for nephrostomy tract dilation and the placement of a sheath. Catheter for Use with Amplatz Renal Dilator Set is intended to be used over a pre-positioned wire guide for percutaneous dilation and access of nephrostomy tract. Amplatz Dilator is intended for nephrostomy tract dilation to allow for the placement of a sheath. Amplatz Renal Introducer is intended to be used for nephrostomy tract dilation and the placement of a sheath. Amplatz Radiopaque TFE Sheath is intended to be used as a working channel to maintain a previously established nephrostomy tract. Clear Amplatz Sheath with Radiopaque Stripe is intended to be used as a working channel to maintain a previously established nephrostomy tract.

## Device Story

Device set includes fascial dilators, catheters, and sheaths used in percutaneous nephrostomy procedures. Operates by mechanical dilation of nephrostomy tract over pre-positioned wire guide; sheath provides working channel for subsequent instrumentation. Used in clinical settings by physicians. Provides access to renal system; facilitates minimally invasive urological interventions. Supplied sterile for single-use.

## Clinical Evidence

Bench testing only. Evidence includes biocompatibility, compatibility verification, functional verification, radiopacity verification, and three-year accelerated aging shelf-life testing. All predetermined acceptance criteria were met.

## Technological Characteristics

Components include fascial dilators, catheters, and sheaths (TFE and clear materials with radiopaque stripes). Mechanical dilation principle. Supplied sterile for single-use. No software or electronic components.

## Predicate Devices

- 8/10 Dilator/Sheath Set; Amplatz Type Renal Dilator Sheath; Amplatz Type Renal Sheath; Clear Renal Sheath ([K851144](/device/K851144.md))
- Amplatz Type Graduated Renal Dilator; Fascial Dilators ([K820865](/device/K820865.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 19, 2017

Cook Incorporated Rebecca Odulio (Li-Chun Liu) Regulatory Affairs Specialist 750 Daniels Way. P.O. Box 489 Bloomington, Indiana 47402

Re: K172527

> Trade/Device Name: Amplatz Renal Dilator Set, Catheter for Use with Amplatz Renal Dilator Set, Amplatz Dilator, Amplatz Renal Introducer, Amplatz Radiopaque TFE Sheath, Clear Amplatz Sheath with Radiopaque Stripe

Regulation Number: None Regulation Name: Unclassified Regulatory Class: Class II Product Code: LJE Dated: August 18, 2017 Received: August 21, 2017

Dear Rebecca Odulio (Li-Chun Liu):

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Joyce M. Whang -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

2

510(k) Number (if known)

K172527

Device Name Amplatz Renal Dilator Set

Indications for Use (Describe)

Amplatz Renal Dilator Set is intended for nephrostomy tract dilation and the placement of a sheath.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{3}------------------------------------------------

3

510(k) Number (if known)

K172527

Device Name

Catheter for Use with Amplatz Renal Dilator Set

Indications for Use (Describe)

Catheter for Use with Amplatz Renal Dilator Set is intended to be used over a pre-positioned wire guide for percutaneous dilation and access of nephrostomy tract.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{4}------------------------------------------------

4

510(k) Number (if known)

K172527

Device Name Amplatz Dilator

Indications for Use (Describe)

Amplatz Dilator is intended for nephrostomy tract dilation to allow for the placement of a sheath.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{5}------------------------------------------------

Expiration Date: January 31, 2017

See PRA Statement below.

510(k) Number (if known) K172527

Device Name Amplatz Renal Introducer

Indications for Use (Describe)

Amplatz Renal Introducer is intended to be used for nephrostomy tract dilation and the placement of a sheath.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

6

510(k) Number (if known) K172527

Device Name Amplatz Radiopaque TFE Sheath

Indications for Use (Describe)

Amplatz Radiopaque TFE Sheath is intended to be used as a working channel to maintain a previously established nephrostomy tract.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{7}------------------------------------------------

7

510(k) Number (if known)

K172527

Device Name

Clear Amplatz Sheath with Radiopaque Stripe

Indications for Use (Describe)

Clear Amplatz Sheath with Radiopaque Stripe is intended to be used as a working channel to maintain a previously established nephrostomy tract.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, bold, white letters on a red background. Below the word "COOK" is the word "MEDICAL" in white letters on a smaller red background. The logo is simple and clean, with a focus on the company name.

COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA 235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM

#### Section 2.0 510(k) Summary

# Amplatz Renal Dilator Set Catheter for Use with Amplatz Renal Dilator Set Amplatz Dilator Amplatz Renal Introducer Amplatz Radiopaque TFE Sheath Clear Amplatz Sheath with Radiopaque Stripe Date Prepared: October 19, 2017

### Submitted By:

| Submission:           | Traditional 510(k) Premarket Notification                     |
|-----------------------|---------------------------------------------------------------|
| Applicant:            | Cook Incorporated                                             |
| Contact:              | Rebecca Odulio (Li-chun Liu)                                  |
| Applicant Address:    | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 x104673                                        |
| Contact Fax Number:   | (812) 332-0281                                                |

### Device Information:

| Device Trade Name                                  | Common Name              | Product Code | Regulation Number /Regulation<br>Name/Regulation Class/<br>Classification Panel |
|----------------------------------------------------|--------------------------|--------------|---------------------------------------------------------------------------------|
| Amplatz Renal Dilator Set                          |                          |              |                                                                                 |
| Catheter for Use with Amplatz<br>Renal Dilator Set |                          |              |                                                                                 |
| Amplatz Dilator                                    |                          |              |                                                                                 |
| Amplatz Renal Introducer                           | Catheter,<br>Nephrostomy | LJE          | None,<br>Unclassified,<br>Class II,<br>Gastroenterology/Urology                 |
| Amplatz Radiopaque TFE Sheath                      |                          |              |                                                                                 |
| Clear Amplatz Sheath with<br>Radiopaque Stripe     |                          |              |                                                                                 |

### Predicate Devices:

- K851144: 8/10 Dilator/Sheath Set; Amplatz Type Renal Dilator Sheath; Amplatz Type ● Renal Sheath; Clear Renal Sheath (Van-Tec, Inc.)
- K820865: Amplatz Type Graduated Renal Dilator; Fascial Dilators (Van-Tec, Inc.) ●

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The red background is a rectangle with a diagonal cut on the bottom left corner.

# Device Description:

The Amplatz Renal Dilator Set contains the following components: an 8.0 French Catheter for Use with Amplatz Renal Dilator Set, various fascial dilators, a 10.0 French introducer sheath, various Amplatz Dilators, and various Amplatz Renal Introducers (an assembly of Amplatz Dilator and Amplatz Sheath). The set is intended for nephrostomy tract dilation and the placement of a sheath, which acts as a working channel during percutaneous nephrostomy procedures. It is supplied sterile and intended for one-time use.

The set includes components that may be sold separately with dimensional and material variances compared to the specifications of the original set components. The set components that are available in single sellable unit are as follows:

- Amplatz Renal Dilator Set ●
- Catheter for Use with Amplatz Renal Dilator Set .
- . Amplatz Dilator
- Amplatz Renal Introducer ●
- Amplatz Radiopaque TFE Sheath; ●
- Clear Amplatz Sheath with Radiopaque Stripe

### Indications for use:

The Amplatz Renal Dilator Set is intended for nephrostomy tract dilation and the placement of a sheath.

The Catheter for Use with Amplatz Renal Dilator Set is intended to be used over a pre-positioned wire guide for percutaneous dilation and access of nephrostomy tract.

The Amplatz Dilator is intended for nephrostomy tract dilation to allow for the placement of a sheath.

The Amplatz Renal Introducer is intended to be used for nephrostomy tract dilation and the placement of a sheath.

The Amplatz Radiopaque TFE Sheath is intended to be used as a working channel to maintain a previously established nephrostomy tract.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, bold, sans-serif font in the upper portion of the logo. Below "COOK" is the word "MEDICAL" in white, smaller, sans-serif font.

COOK INCORPORATED COOKMEDICAL.COM

The Clear Amplatz Sheath with Radiopaque Stripe is intended to be used as a working channel to maintain a previously established nephrostomy tract.

### Comparison to Predicates:

The subject devices have similar indications for use, methods of operation, and fundamental technological characteristics as the predicates. The differences between the subject devices and the predicate devices include minor dimensional variations and materials. Characteristics of the subject devices that differ from the predicate devices are supported by testing. The subject devices do not raise new questions of safety or effectiveness as compared to the predicate devices.

### Technological Characteristics:

The following tests have been conducted to demonstrate that the Amplatz Renal Dilator Set, Catheter for Use with Amplatz Renal Dilator Set, Amplatz Dilator, Amplatz Renal Introducer, and Amplatz Sheath (Amplatz Radiopaque TFE Sheath; Clear Amplatz Sheath with Radiopaque Stripe) met applicable design and performance requirements.

- Biocompatibility
- Compatibility Verification ●
- Functional Verification ●
- . Radiopacity Verification
- Shelf Life of Three Years Accelerated Aging ●

# Conclusion:

All predetermined acceptance criteria of the testing were met. Therefore, the results of these tests support a conclusion that the subject devices perform as intended and support a determination of substantial equivalence to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K172527](https://fda.innolitics.com/device/K172527)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
