K172514 · Sandstone Diagnostics, Inc. · POV · Nov 17, 2017 · Hematology
Device Facts
Record ID
K172514
Device Name
Trak Plus Male Fertility Testing System
Applicant
Sandstone Diagnostics, Inc.
Product Code
POV · Hematology
Decision Date
Nov 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Intended Use
The Trak® Male Fertility Testing System is intended for semi-quantitative assessment of sperm concentration at 15 million sperm per milliliter (M/mL) or below, between 15 and 55 M/mL, and above 55 M/mL. It also provides a qualitative assessment of semen volume. Sperm concentration and semen volume are only two factors that could impact a man's fertility status and time to pregnancy. For complete assessment of male reproductive health, the user should consult a physician. For in vitro, over the counter home use.
Device Story
System includes Trak Engine (instrument), disposable Props, Volume Cups, and droppers. User collects semen in Volume Cup; cup markings allow qualitative volume assessment (threshold 1.5 mL). Semen sample introduced into Prop; Engine spins Prop to isolate sperm via density gradient separation. Sperm cells compact into visible column (pellet) in Prop; column height corresponds to sperm concentration (below 15 M/mL, 15-55 M/mL, above 55 M/mL). Contaminants remain suspended above density medium. User visually interprets column height and volume level. Device provides home-based fertility screening; results intended to supplement physician consultation for reproductive health assessment.
Clinical Evidence
Two clinical studies (n=232 and n=127) evaluated lay-user interpretation of semen volume against reference methods. Stage 1: Sensitivity 97.0% (91.6-99.0%), Specificity 95.4% (90.4-97.9%). Stage 2: Sensitivity 95.2% (86.9-98.4%), Specificity 96.9% (89.3-99.1%). Overall accuracy >96%. Bench testing confirmed volume cup precision, liquefaction effectiveness, and concentration category equivalence to previous cup design.
Technological Characteristics
System uses density gradient centrifugation for sperm isolation. Volume cups are coated with digestive enzymes to reduce viscosity. Measurement is visual (fluid height for volume, pellet height for concentration). Battery-operated Engine provides centrifugal force. No software, connectivity, or electronic data transmission.
Indications for Use
Indicated for adult males seeking semi-quantitative assessment of sperm concentration and qualitative assessment of semen volume for fertility monitoring. For OTC home use.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
K153683 — Trak Male Fertility Testing System · Sandstone Diagnostics, Inc. · May 16, 2016
K183602 — SwimCount Sperm Quality Test · Motilitycount Aps · Jun 27, 2019
K241628 — YO Home Sperm Test · Medical Electronic Systems , Ltd. · Nov 29, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 17, 2017
Sandstone Diagnostics, Inc. Ulrich Schaff Chief Technology Officer 6111 Southfront Road, Suite J Livermore, California 94551
Re: K172514
Trade/Device Name: Trak Plus Male Fertility Testing System Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: POV Dated: August 18, 2017 Received: August 21, 2017
Dear Ulrich Schaff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Leonthena R. Carrington -S
Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172514
Device Name Trak® Male Fertility Testing System
#### Indications for Use (Describe)
The Trak® Male Fertility Testing System is intended for semi-quantitative assessment of sperm concentration at 15 million sperm per milliliter (M/mL) or below, between 15 and 55 M/mL, and above 55 M/mL. It also provides a qualitative assessment of semen volume.
Sperm concentration and semen volume are only two factors that could impact a man's fertility status and time to pregnancy. For complete assessment of male reproductive health, the user should consult a physician. For in vitro, over the counter home use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K172514.
| 807.92 (a)(1): Name: | Sandstone Diagnostics, Inc. |
|----------------------|------------------------------------------------------|
| Address: | 6111 Southfront Road, Suite J<br>Livermore, CA 94551 |
| Phone: | 925-315-7246 |
| FAX: | 925-215-2269 |
| Contact: | Ulrich Schaff, PhD |
#### 807.92 (a)(2): Device name- trade name and common name, and classification
Trade name: Trak® Male Fertility Testing System
Common Name: Semen Analysis Test
Classification: 21 CFR § 864.5220
#### 807.92 (a)(3): Identification of the legally marketed predicate devices
Trak® Male Fertility Testing System, Sandstone Diagnostics, Inc., Livermore, CA, K153683
#### 807.92 (a)(4): Device Description
The revised Trak® Male Fertility Testing System (Trak) includes a small instrument (the Engine), disposable units in which liquefied semen sample is introduced and the result is interpreted (the Props), and consumables, including Volume Cups and sample droppers.
The Volume Cups are used collect, liquefy, and assess the volume of the semen sample. Markings on the Volume Cup allow the interpretation of volume at/below or above the 1.5 mL threshold. The Trak Engine and Props allow for the semi-quantitative assessments of total sperm concentration: below 15 million/mL, between 15 and 55 million/mL, and above 55 million/mL.
Trak uses the principle of density gradient separation to isolate sperm cells from human semen to provide an estimation of sperm concentration. The Trak Engine spins a test Prop to compact sperm cells within an introduced semen sample into a visible column (or "pellet"). The Prop gives a defined shape to the column, the height of which corresponds to the concentration of sperm cells in the sample. Since semen may also contain cell debris, immature sperm cells, and other contaminant particulates that could contribute to the apparent size of a pellet, it is necessary to filter out the contaminants. Trak achieves this filtering by removing contaminants from view based on density across a predefined liquid density medium.
{4}------------------------------------------------
During operation, approximately 0.17 mL of semen is metered by centrifugal action from the sample inlet into the metering chamber of the Prop. During rotation, the semen floats on "top" of the pre-loaded density medium. Sperm cells pass through the medium due to their high density while contaminants remain floating on the medium. When the spin sequence is complete, the sperm cells form a visible column that is displayed to the user for interpretation. Contaminants that are less dense than the liquid density medium are suspended "above" the medium, substantially separated from the sperm cells and are generally too diffuse to visualize.
#### 807.92 (a)(5): Intended Use
The Trak® Male Fertility Testing System is intended for semi-quantitative assessment of sperm concentration at 15 million sperm per milliliter (M/mL) or below, between 15 and 55 M/mL, and above 55 M/mL. It also provides a qualitative assessment of semen volume.
Sperm concentration and semen volume are only two factors that could impact a man's fertility status and time to pregnancy. For complete assessment of male reproductive health, the user should consult a physician. For in vitro, over the counter home use.
| Comparison | Subject Device<br>Trak® Male Fertility Testing<br>System<br>(revised) | Predicate Device<br>Trak® Male Fertility Testing<br>System<br>(K153683) |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Trak® Male Fertility Testing<br>System is intended for semi-<br>quantitative assessment of sperm<br>concentration at 15 million sperm per<br>milliliter (M/mL) or below, between 15<br>and 55 M/mL, and above 55 M/mL. It<br>also provides a qualitative assessment<br>of semen volume.<br><br>Sperm concentration and semen volume<br>are only two factors that could impact a<br>man's fertility status and time to<br>pregnancy. For complete assessment of<br>male reproductive health, the user<br>should consult a physician. For in vitro,<br>over the counter home use. | The Trak® Male Fertility Testing<br>System is intended for semi-<br>quantitative assessment of sperm<br>concentration at 15 million sperm per<br>milliliter (M/mL) or below, between<br>15 and 55 M/mL, and above 55<br>M/mL.<br><br>Sperm concentration is only one factor<br>that could impact a man's fertility status<br>and time to pregnancy. For complete<br>assessment of male reproductive health,<br>the user should consult a physician. For<br>in vitro, over the counter home use. |
| Class | Class II | Same |
| Regulation Number | 21 CFR 864.5220 | Same |
| Product Code | POV | Same |
| Branch | Hematology (81) | Same |
#### 807.92 (a)(6): Technological Similarities and Differences to the Predicate
The following chart describes similarities and differences between the revised Trak system and
{5}------------------------------------------------
| Comparison | Subject Device<br>Trak® Male Fertility Testing<br>System<br>(revised) | Predicate Device<br>Trak® Male Fertility Testing<br>System<br>(K153683) |
|------------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Class | Class II | Same |
| Test Type | Semi-Quantitative and Qualitative | Semi-Quantitative |
| Test Locale | Home use | Same |
| Sample Type | Human semen | Same |
| Test Reporting | Visual readout of cell column height and<br>Visual readout of fluid height | Visual readout of cell column height |
| Test Principle | Centrifuged packed cell volume and<br>graduated chamber fluid volume | Centrifuged packed cell volume |
| Concentration Primary<br>Cut-off | 15 M/mL (lower reference limit, WHO<br>semen analysis guidelines 5th edition,<br>2010) | Same |
| Concentration<br>Secondary Cut-off | 55 M/mL (indication of reduced time to<br>pregnancy based on Slama et al 2002<br>study) | Same |
| Volume Cut-off | 1.5 mL (lower reference limit, WHO<br>semen analysis guidelines 5th edition,<br>2010) | None |
| Test Control Method | External Quality Control test solution | Same |
### 807.92 (b)(1): Brief Description of Nonclinical Data
#### Precision of Volume Cup
The objective of this study was to assess the precision of the Volume Cup. Three lots of Volume Cups were qualitatively evaluated by three operators over five days on 5 artificial semen samples with different volumes and were compared with the reference volume. Each combination of day and operator/lot were tested in 5 replicates to obtain 75 total replicates per semen volume.
#### Summary results
The following includes grand averages for each condition, sum for each category and percent correct calls with respect to the reference method.
| ID # | Reference<br>Volume (mL) | # Results ≤ 1.5<br>mL | # Results > 1.5<br>mL | % Correct | Quant Volume result<br>grand average ± SD (mL) |
|------|--------------------------|-----------------------|-----------------------|-----------|------------------------------------------------|
| 1 | 1.2 mL | 75 | 0 | 100 | 1.32 ± 0.12 |
| 2 | 1.3 mL | 74 | 1 | 98.7 | 1.43 ± 0.09 |
| 3 | 1.5 mL | 17 | 58 | n/a* | 1.61 ± 0.10 |
| 4 | 1.7 mL | 1 | 74 | 98.7 | 1.86 ± 0.16 |
| 5 | 1.8 mL | 0 | 75 | 100 | 1.93 ± 0.10 |
* no performance goal at the threshold; data tabulated for information, only
{6}------------------------------------------------
| I<br>D<br># | N | Mean | Residual | | Between Day | | Between<br>Operator‡ | | Operator*<br>Day† | | Total SD<br>and %CV | |
|-------------|----|------|----------|------|-------------|------|----------------------|------|-------------------|------|---------------------|------|
| | | | SD | CV | SD | CV | SD | CV | SD | CV | SD | CV |
| 1 | 75 | 1.32 | 0.08 | 6.1% | 0.07 | 5.0% | 0 | 0.0% | 0.06 | 4.8% | 0.12 | 9.2% |
| 2 | 75 | 1.43 | 0.06 | 4.0% | 0.02 | 1.6% | 0.02 | 1.3% | 0.07 | 4.7% | 0.09 | 6.5% |
| 3 | 75 | 1.61 | 0.09 | 5.5% | 0.00 | 0.0% | 0.00 | 0.0% | 0.05 | 2.8% | 0.10 | 6.2% |
| 4 | 75 | 1.86 | 0.09 | 5.0% | 0.03 | 1.4% | 0.00 | 0.3% | 0.00 | 0.0% | 0.10 | 5.2% |
| 5 | 75 | 1.93 | 0.10 | 5.1% | 0.02 | 1.2% | 0.01 | 0.6% | 0.00 | 0.0% | 0.10 | 5.3% |
The following includes SD (mL) and %CV between Day, between Operator/Lot, and total for each concentration.
#Operator factor is convolved with Lot: each operator used a unique lot.
*This is an interaction effect, it was found to be significant for some volumes tested.
The Volume Cup meets expected concordance for precision within operators/lots, days, and runs.
#### Functional Equivalence to Previous Cup
A series of studies were performed to prove that the Volume Cup functions equivalently to the previous Collection Cup in (1) promoting semen liquefaction, (2) semen homogenization; and (3) generating the equivalent Trak sperm concentration result.
#### Semen Liquefaction
The objective of this study was to determine the effectiveness of new Volume Cup in promoting semen liquefaction. A total of 20 semen sample aliquots from three semen samples were tested for liquefaction by measuring the volume of semen dispensed by the transfer devices included with the Trak System (250 µL exact volume pipettes) after incubation according to directions.
#### Summary results
A total of three semen samples were used in this study, with no more than 10 repeat measurements from a single sample. All samples were confirmed as liquefied and non-viscous at 30 minutes. The total average for these volume measures was 253.05 µL, with a 95% CI of 249.60 - 256.00 µL, meeting the acceptance criteria.
The data supports that the Volume Cups are effective at promoting semen liquefaction.
#### Trak Sperm Concentration Category Result Equivalence Study
The objective of this study was to determine whether samples incubated in the new Volume Cup give equivalent Trak category results to samples incubated in the previous Collection Cup. Two sperm concentrations in the vicinity of the WHO threshold cutoff (15 M/mL) were tested in 20 replicates for each cup type on Trak Props, and results were categorized as ≤ 15 M/mL (Low), 15 - 55 M/mL (Moderate), and > 55 M/mL (Optimal).
{7}------------------------------------------------
#### Summary results
The following includes summary of results in this study. All conditions were categorized correctly with respect to the reference method at least 90% of the time. Both semen pools were first formulated as a single pool, then divided evenly over the two cup types, and then CASA confirmed for both cups types just before Trak assay initiation.
| Nominal Pool Condition | Measured CASA results | | | | Average<br>Concentration<br>of sample | Trak<br>Results ≤<br>15 M/mL | Trak<br>Results 15<br>– 55<br>M/mL | Trak<br>Results ><br>55 M/mL |
|------------------------|-----------------------|------|------|------|---------------------------------------|------------------------------|------------------------------------|------------------------------|
| 10 M/mL, control | 10.2 | 11.0 | 11.9 | 12.3 | 11.4 | 20 | 0 | 0 |
| 10 M/mL, volume cup | 10.8 | 11.4 | 11.4 | 13.0 | 11.7 | 20 | 0 | 0 |
| 20 M/mL, control | 21.0 | 17.9 | 21.8 | 21.9 | 20.7 | 0 | 20 | 0 |
| 20 M/mL, volume cup | 20.0 | 19.1 | 19.5 | 20.2 | 19.7 | 0 | 20 | 0 |
Trak results for both the new Volume Cup and the previous Collection cup were correctly categorized with respect to the reference method 100% of the time, exceeding the study acceptance criteria.
#### Semen Homogenization study
The objective of this study was to determine whether sperm concentration is evenly distributed within the new Volume Cup vs the previous Collection Cup when agitated. Semen samples were formulated to two sperm concentrations differing by more than 50%. Twenty (20) aliquots were taken from the top and bottom of each cup for each pool, and its sperm concentration was accessed.
The two cups performed equivalently.
#### Summary results
The following includes average results by unique condition, cup type, and grand mean for pools. Sample deviations (SD) are also shown.
| Pool ID | Cup Type | Location | n | Mean ± SD<br>(M/mL) | Cup type<br>Mean ± SD<br>(M/mL) | Grand<br>Mean ± SD<br>(M/mL) |
|---------|------------------------------|----------|----|---------------------|---------------------------------|------------------------------|
| A | Collection Cup<br>(Previous) | Top | 20 | 17.1 ± 0.7 | 17.1 ± 0.7 | 17.3 ± 0.8 |
| | | Bottom | 20 | 17.2 ± 0.7 | | |
| | Volume Cup<br>(New) | Top | 19 | 17.5 ± 0.7 | 17.5 ± 0.8 | |
| | | Bottom | 20 | 17.5 ± 0.9 | | |
| B | Collection Cup<br>(Previous) | Top | 20 | 36.4 ± 2.2 | 35.3 ± 2.5 | 36.1 ± 2.5 |
| | | Bottom | 20 | 34.3 ± 2.5 | | |
| | Volume Cup<br>(New) | Top | 20 | 36.7 ± 2.6 | 36.9 ± 2.2 | |
| | | Bottom | 20 | 37.1 ± 1.8 | | |
### 807.92 (b)(2): Brief Description of Clinical Data
Two studies were performed to assess Lay Subject validation for the 1.5 mL reference marking on the new Volume Cup. Users were asked to determine semen volume with respect to 1.5 mL reference mark, where a volume less than 1.5 mL is considered a positive result for hypospermia.
#### Stage 1 Study: Interpretation of Photographs
The objective of the study was to demonstrate the accuracy of the new Volume Cup as interpreted by the intended users. A total of 232 semen sample photographs were interpreted as ≤ 1.5 mL or
{8}------------------------------------------------
> 1.5 mL by Lay Subjects as well as trained technicians (blinded to the result), and were confirmed with the result obtained by the reference method.
The descriptive demographic statistics for the Lay Subjects (n = 52) were as follows. Subjects were 28.0 years on average (range 18.0 - 68.0), with a high prevalence of male gender (57.7%). Caucasians (44.2%) with some college education (40.4%). We also performed lay user interpretation questionnaire, and a majority (90.4%) of Lay Subjects gave answers indicating an understanding of the instructions according to the intended interpretation.
#### Summary results Primary Analysis - Sensitivity and specificity for Lay Subject interpretation
The following includes 2 x 2 contingency table, with study results, for performance estimates of Lay Subjects versus reference method.
Image /page/8/Figure/4 description: This image shows a 2x2 table that compares the lay subject result to a reference result. The table shows the number of results for each category. 98 results were less than or equal to 1.5 mL for both the lay subject and reference results. 125 results were greater than 1.5 mL for both the lay subject and reference results.
Reference result, n = 232
The following includes performance parameters for Lay Subject interpretation versus reference method.
| Sensitivity, %<br>(95% CI) | Specificity, %<br>(95% CI) |
|----------------------------|----------------------------|
| 97.0%<br>(91.6 – 99.0%) | 95.4 %<br>(90.4 – 97.9%) |
Additionally, the overall percent agreement (OPA or "accuracy") was found to be 96.1% (92.8 -97.9%). For the analysis set, 9 (3.9%) of interpreted results were discordant versus the reference method, with 6 (2.6%) indicating false positive results and 3 (1.3%) indicating false negative results. Of the Lay Subject vs. reference discrepant results, 9 (100%) were within 0.5 mL of the 1.5 mL threshold with reference results, often reading above 1.5 mL.
{9}------------------------------------------------
#### Trained Technician volume results versus reference results
The following includes 2 x 2 contingency table, with study results, for performance estimates of Trained Technicians versus reference method.
Image /page/9/Figure/2 description: This image shows a 2x2 confusion matrix comparing technician results to reference results for a sample size of 232. The matrix compares results for values less than or equal to 1.5 mL and greater than 1.5 mL. The top left cell shows 100, the top right cell shows 4, the bottom left cell shows 4, and the bottom right cell shows 124.
The following includes performance parameters for Trained Technician interpretation versus reference method
| Sensitivity, % | Specificity, % |
|----------------|----------------|
| (95% CI) | (95% CI) |
| 96.2% | 96.9% |
| (90.5 - 98.5%) | (92.2-98.8%) |
Additionally, the OPA was found to be 96.6% (93.3 - 98.2%). For the analysis set, 8 (3.4%) of interpreted results were discordant versus the reference method, with 4 (1.3%) indicating false negatives and 4 (1.3%) indicating false positive results. Of the Technician vs. reference discrepant results, 8 (100%) were within 0.5 mL of the 1.5 mL threshold.
#### Trained Technician results versus Lay Subject results
The following includes trained Technician vs Lay Subject interpretation contingency table.
| | | ≤ 1.5 mL | > 1.5 mL |
|--------------------|---------|----------|----------|
| Lay Subject result | ≤1.5 mL | 100 | 1 |
| | >1.5 mL | 4 | 127 |
#### Trained Technician result, n = 232
{10}------------------------------------------------
The following includes descriptive statistics showing positive percent agreement (PPA), negative percent agreement (NPA) and overall agreement between Lay and Technician interpretation.
| OPA, %<br>(95% CI) | PPA, %<br>(95% CI) | NPA, %<br>(95% CI) |
|-------------------------|-------------------------|-------------------------|
| 97.8%<br>(95.1 – 99.1%) | 96.2%<br>(90.5 – 98.5%) | 99.2%<br>(95.7 – 99.9%) |
Trained technician tended to agree with the Lay Subject more than 95% of the time, with the lower bound of all comparison statistics above 90%. In this analysis set, Technician and Lay Subject disagreed on 5 (2.2%) of the samples.
#### Stage 2 Study: Interpretation of Simulated Sample in the Volume Cup
The objective of the second study was to assess interpretation and ease of use of the physical Volume Cup filled with known volumes of artificial semen as interpreted by the intended users.
A total of 127 semen sample photographs were interpreted as ≤ 1.5 mL, > 1.5 mL, or Invalid by 32 Lav Subjects. Each Lay Subject interpreted 4 Cups filled with 1 mL, 1.2 mL, or 2 mL of artificial semen in random order.
The descriptive demographic statistics for the Lay Subjects (n = 32) were as follows. Subjects were 35.1 years on average (range 25- 59), with a high prevalence of Caucasians (53.1%) with post-graduate education (53.1%).
#### Summary results Primary Analysis – Sensitivity and specificity for Lay Subject interpretation
The following includes 2 x 2 contingency table for performance estimates versus reference method with Stage 2 results.
Image /page/10/Figure/9 description: This image shows a 2x2 confusion matrix comparing the results of a reference test and a lay subject test. The reference test has 127 samples. The matrix shows that when the reference result is less than or equal to 1.5 mL, the lay subject result is less than or equal to 1.5 mL 60 times, and greater than 1.5 mL 3 times. When the reference result is greater than 1.5 mL, the lay subject result is less than or equal to 1.5 mL 2 times, and greater than 1.5 mL 62 times.
{11}------------------------------------------------
The following includes performance parameters for Lay Subject interpretation versus reference method.
| Sensitivity, %<br>(95% CI) | Specificity, %<br>(95% CI) |
|----------------------------|----------------------------|
| 95.2%<br>(86.9-98.4%) | 96.9%<br>(89.3-99.1%) |
These results suggest that in the hands of Lay Subjects, the Volume Cup with appropriate instructions provides adequate resolution of semen volume in the vicinity of the WHO threshold for hypospermia.
#### Lay Subject Questionnaire results
Lav Subjects, after completing a volume cup reader study, also completed a Tester Questionnaire with 11 multiple choice questions. Five of the questions were designed to capture the subject's perceptions of ease of use of the Trak device, and the remaining six questions were to gauge interpretation and understandability of the instructions. While slightly more than a quarter of the respondents assigned "somewhat difficult" or "difficult" to the task of finding the liquid level, more than 95% of the responses for interpreting cups were assigned correctly with respect to the reference method. Additionally, a majority (90.6%) of testers gave answers indicating an understanding of the instructions according to the intended interpretation.
#### CONCLUSIONS
Sandstone Diagnostics. Inc. has performed two cross-sectional studies of a suitable sample sizes to investigate and determine the substantial equivalence of their device to a recognized reference method of semen volume measurement via a calibrated scale. The results of the prospectively stated primary analyses indicate the lower bound of the 95% confidence interval for sensitivity and specificity meeting the pre-specified acceptance criteria. Specifically, Stage 1 resulted in an observed sensitivity of 97.0% (95% CI: 91.6 - 99.0%) and specificity of 95.4% (95% CI: 90.4 -97.9%) with the lower bound of the 95% confidence intervals meeting the acceptance criteria of > 80% for both sensitivity and specificity. An overall percent agreement ("accuracy") between Lay Subject and Reference method was found to be 96.1% (92.8 - 97.9%). Stage 2 resulted in an observed sensitivity of 95.2% (95% CI: 86.9-98.4%) and specificity of 96.9% (95% CI: 89.3-99.1%), also meeting acceptance criteria. These outcomes provide evidence supporting the accuracy of the Trak volume test when used by lay testers representing the population of intended use.
Further, Trak volume results generated by subjects generally agreed with results from the same specimens tested by trained users, indicating the quality of the instructions for use. Subjects from the Stage 2 study presented with filled cups generally felt the volume cup would be easy to use in terms of procedure and interpretation of results with regard to the clinical threshold of 1.5 mL. In summary, the study demonstrated substantial equivalence of the Trak volume test in comparison to the reference method in the hands of the intended use population.
### 807.92 (b)(3): Brief Description of Clinical Data
The conclusions drawn from the analytical and clinical data demonstrate that the device is safe and effective for its intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.