K172508 · Ageless Health Industrial Co., Ltd. · FLL · Aug 8, 2018 · General Hospital
Device Facts
Record ID
K172508
Device Name
Ageless Health Medical Digital Thermometer
Applicant
Ageless Health Industrial Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Aug 8, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Ageless Health Medical Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home. lt can be used for axillary measurement, oral measurement and rectal measurement.
Device Story
Digital thermometer uses thermistor sensor to detect body temperature via resistance changes; ASIC processes signal; buzzer provides audible feedback; LCD displays temperature. Used by clinicians or patients in hospital or home. Device provides intermittent temperature readings to assist in monitoring patient health status.
Clinical Evidence
Bench testing only. Evaluated against IEC 60601-1, IEC 80601-2-56, IEC 60601-1-2, IEC 60601-1-11, ASTM E1112-00, ISO 10993-5, and ISO 10993-10.
Technological Characteristics
Thermistor-based sensing; ASIC signal processing; 1.5V button battery power; LCD display; 60s response time; 32.0°C ~ 42.9°C range; accuracy +/- 0.1°C (35.0-39.0°C) and +/- 0.2°C (remainder). Complies with IEC 60601-1, IEC 80601-2-56, IEC 60601-1-2, IEC 60601-1-11, ASTM E1112, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for measurement and monitoring of human body temperature in hospital or home settings via axillary, oral, or rectal routes for all patient populations.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Ageless Health Medical Digital Thermometer (K152508)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 8, 2018
Ageless Health Industrial Co., Ltd % Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou. 510000 China
Re: K172508
Trade/Device Name: Ageless Health Medical Digital Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: July 4, 2018 Received: July 9, 2018
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## K172508
## 510(k) Summary
This summary of 510(K) information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
## 1. Submitter's Information
Sponsor Name: Ageless Health Industrial Co., Ltd. Address: 3/F, A1 Bldg, Dongshen Sima Industrial Area, No.33 Shenbei Road, Sima Village, Changping Town, Dongguan City, Guangdong Province, China Contact Person: Victor Wan (Vice-president) Phone: +86-769-81158038 Fax: +86-769-82289331 E-mail: victor@agelh.com
## Application Correspondent:
Application Correspondent:
Contact Person: Ms. Cassie Lee
Guangzhou GLOMED Biological Technology Co., Ltd.
Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China
Tel: +86-20-61099984
Email: regulatory@glomed-info.com
#### 2. Subject Device Information
| Type of 510(k): | Traditional |
|----------------------|---------------------------------------------------------------------------------------------------------------------|
| Common Name: | Thermometer, electronic |
| Trade Name: | Ageless Health Medical Digital Thermometer (models: BT-301, BT-302, BT-303, BT-305, BT-306, BT-308, BT-311, BT-318) |
| Classification Name: | Thermometer, electronic, clinical |
| Review Panel: | General Hospital |
| Product Code: | FLL |
| Regulation Number: | 880.2910 |
| Regulation Class: | II |
#### 3. Predicate Device Information
{4}------------------------------------------------
## Predicate Device 1
510(K) Number: K152508 Sponsor: Dongguan Ageless Health Industrial Co., Ltd. Common Name: Thermometer, electronic Trade Name: Ageless Health Medical Digital Thermometer Classification Name: Thermometer, electronic, clinical Review Panel: General Hospital Product Code: FLL Regulation Number: 880.2910 Regulation Class: II
## Predicate Device 2
510(K) Number: K101387 Sponsor: Fudakang Industrial Co., Ltd Classification Name: Thermometer, electronic, clinical Trade Name: Digital Thermometer Review Panel: General Hospital Product Code: FLL Regulation Number: 880.2910 Regulation Class: II
#### 4. Device Description
Ageless Health Medical Digital Thermometers comprises a thermistor for getting temperature signal, a reference resistor for comparing the resistance of the thermistor, a buzzer for sounding effect, an ASIC for processing the target temperature digitally, and a LCD for displaying the temperature result. The design principle of thermometer is based on thermos sensor and ASIC technology. A thermistor is used as thermos sensor. The ASIC gets the sensor's signal from human body, then processes the signal and calculates the result, after that displays the temperature result by a LCD.
#### 5. Intended Use
Ageless Health Medical Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home. lt can be used for axillary measurement, oral measurement and rectal measurement.
## 6. Test Summary
Ageless Health Medical Digital Thermometer has been evaluated the safety and performance by lab bench testing according to the following standards:
{5}------------------------------------------------
- . IEC 60601-1 (2005 + CORR.1:2006 + CORR.2:2007 + AM1:2012) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- . IEC 80601-2-56 (2009) Medical Electrical Equipment - Part 2-56: Particular Requirements for Basic Safety and Essential Performance of Clinical Thermometers for Body Temperature Measurement
- . IEC 60601-1-2 (2007) Medical Electrical Equipment - Part 1-2: Collateral Standard: Electromagnetic Compatibility
- . IEC 60601-1-11 (2010) Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used
- . ASTM E1112-00 (2006) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
- . ISO 10993-5 (2009) Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro cytotoxicity
- . ISO 10993-10 (2010) Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization
#### Comparison to Predicate Device 7.
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards.
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Company | Ageless Health<br>Industrial Co., Ltd. | Ageless Health<br>Industrial Co., Ltd. | Fudakang Industrial<br>Co., Ltd. | -- |
| Name and<br>Model | Ageless Health Medical<br>Digital Thermometer,<br>models: BT-301, BT-<br>302, BT-303, BT-305,<br>BT-306, BT-308, BT-<br>311, BT-318 | Ageless Health Medical<br>Digital Thermometer,<br>models: BT-301, BT-<br>302, BT-303, BT-305,<br>BT-306, BT-308, BT-<br>311, BT-318 | Digital Thermometer<br>Model: BT-A11CN,<br>BT-A21CN, BT-A41<br>CN | -- |
| 510(k)<br>Number | K172508 | K152508 | K101387 | -- |
| Thermometer<br>Type | Digital Thermometer | Digital Thermometer | Digital Thermometer | SE |
| Intended Use<br>/ Indication<br>for Use | Ageless Health Medical<br>Digital Thermometers<br>are intended for the<br>measurement and<br>monitoring of human<br>body temperature by<br>doctor or consumers in<br>the hospital or home. | Ageless Health Medical<br>Digital Thermometers<br>are intended for use in<br>measuring human body<br>temperature (Armpit or<br>Oral). | Fudakang Digital<br>Thermometers are<br>intended for the<br>measurement and<br>monitoring of human<br>body temperature by<br>doctor or consumers<br>in the hospital or | SE<br>Note 1 |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark |
| | It can be used for<br>axillary measurement,<br>oral measurement and<br>rectal measurement. | | home.<br>BT-A11CN, BT-<br>A21CN, BT-A41CN<br>can be used for<br>axillary measurement,<br>oral measurement<br>and rectal<br>measurement. | |
| Sensor | Thermistor | Thermistor | Thermistor | SE |
| Signal<br>Processing<br>and Display | Using the resistance<br>change of thermal<br>resistor to detect body<br>temperature and<br>displayed through the<br>LCD. | Using the resistance<br>change of thermal<br>resistor to detect body<br>temperature and<br>displayed through the<br>LCD. | Using the resistance<br>change of thermal<br>resistor to detect body<br>temperature and<br>displayed through the<br>LCD. | SE |
| Power<br>Requirement | 1.5V button battery | 1.5V button battery | 1.5V button battery | SE |
| Measurement<br>Temperature<br>Range | 32.0°C ~ 42.9°C | 32.0°C ~ 42.9°C | 32.0°C ~ 42.9°C | SE |
| Measurement<br>Accuracy | 35.0 ~ 39.0°C : +/-<br>0.1°C<br>The rest: +/- 0.2°C | 35.0 ~ 39.0°C : +/-<br>0.1°C<br>The rest: +/- 0.2°C | 35.0 ~ 39.0°C: +/-<br>0.1°C<br>The rest: +/- 0.2°C | SE |
| Ambient<br>Temperature | 10~35°C | 10~35°C | 0~40°C | SE |
| Response<br>Time | 60s | 60s | 60s | SE |
| Complied<br>Standard | IEC 60601-1, IEC<br>80601-2-56, IEC<br>60601-1-2, IEC 60601-<br>1-11, ASTM E1112,<br>ISO 10993-5, ISO<br>10993-10 | IEC 60601-1, IEC<br>80601-2-56, IEC 60601-<br>1-2, IEC 60601-1-11,<br>ASTM E1112, ISO<br>10993-5, ISO 10993-10 | IEC 60601-1, IEC<br>80601-2-56, IEC<br>60601-1-2, IEC<br>60601-1-11, ASTM<br>E1112, ISO 10993-5,<br>ISO 10993-10 | SE |
{6}------------------------------------------------
## Note 1
The subject device is exactly the same as predicate device K152508, which is also our company cleared device. The only difference for subject device is added rectal measurement function in Indications for Use.
So, we add predicate device K101387 which included rectal measurement function as additional predicate device. As the parameters of subject device are also very similar to additional predicate device K101387, and their Indications for Use are the same, so differences between them do not raise different questions of safety and effectiveness for the subject device.
{7}------------------------------------------------
## Conclusion
The subject device Ageless Health Medical Digital Thermometer has all features of the predicate devices. The few differences amongst the predicates and the subject do not raise different questions of safety and effectiveness.
- 8. Summary Prepared Date
- 15 August 2017
{8}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.