← Product Code [FMI](/productcode/FMI) · K172483

# Kopac Sterile Needle (K172483)

_Poonglim Pharmatech, Inc. · FMI · Jun 21, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172483

## Device Facts

- **Applicant:** Poonglim Pharmatech, Inc.
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Jun 21, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.

## Device Story

Kopac Sterile Needle is a single-use, hypodermic single lumen needle; consists of a beveled metal cannula joined to a female connector (hub) with a protective cap. Used by clinicians to inject or withdraw fluids from below the skin surface. Device is sterilized via E.O. gas and packaged in blister paper. Provides physical access to subcutaneous or deeper tissues for fluid administration or aspiration. Benefits include sterile, single-use delivery of fluids or collection of samples.

## Clinical Evidence

Bench testing only. No clinical data presented. Testing included biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, hemolysis), mechanical testing (stiffness, resistance to breakage, corrosion, elasticity, flexural strength, pullout force), and sterility/endotoxin validation.

## Technological Characteristics

Materials: SUS304 stainless steel cannula, polypropylene (PP) hub and protector, epoxy adhesive. Design: Single lumen, beveled tip. Dimensions: 30G, lengths 1/2" to 1 1/2". Sterilization: Ethylene Oxide (E.O.) gas per ISO 11135:2014. Standards: ISO 7864 (needle performance), ISO 9626 (tubing requirements), ISO 10993 (biocompatibility).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- IMC hypodermic Needle ([K102584](/device/K102584.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 21, 2018

Poonglim Pharmatech Inc. % Peter Chung President Plus Global 300 Atwood Pittsburgh, Pennsylvania 15213

Re: K172483

Trade/Device Name: Kopac Sterile needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: June 20, 2018 Received: June 20, 2018

Dear Peter Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Carolyn Dorgan -

Digitally signed by Carolyn C. Dorgan -S
DN: c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
0.9.2342.19200300.100.1.1=200180081
4, cn=Carolyn C. Dorgan -S
Date: 2018.06.21 16:21:51 -04'00'

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172483

Device Name Kopac Sterile Needle

Indications for Use (Describe)

This device is intended for use to inject fluids into or withdraw fluids from parts of the surface of the skin

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

### [as required by 807.92(c)]

#### 1. Applicant

- Company: POONGLIM Pharmatech Inc. 1)
- 2) Address: 21, Jayumuyeok 1-gil, Gunsan-si, Jeollabukdo-do, Korea
- Tel : 82-63-451-8141 3)
- 4) Fax : 82-63-451-8145
- Prepared date : July 11, 2017 5)
- Contact person : Peter Chung, 412-512-8802 6)
- 7) Contact person address : 300 Atwood Street, Pittsburgh, PA, 15213, USA
- 8) Submission date: August 11, 2017

### 2. Device Information

- 1) Trade name : Kopac Sterile Needle
- 2) Common name : Single-use needle, hypodermic needle
- 3) Classification name : Hypodermic Single Lumen Needle
- 4) Product code : FMI
- 5) Regulation number : 880.5570
- 6) Class of device : Class II
- 7) Panel : General hospital

## 3. The legally marketed device to which we are claiming equivalence

K102584, IMC hypodermic Needle

#### 4. Device description

The device consists of a metal tube that is beveled at one end and at the other end join to a female connector (hub). The needle cap covers intended to provide physical protection to the needle tube. This product is packed by blister paper and sterilized by E.O. gas.

| Model                 | Gauge | Inch (") | mm    | Length of Cannula (A) | Cover dimension (mm) |    |
|-----------------------|-------|----------|-------|-----------------------|----------------------|----|
| Type<br>Y<br>and<br>I | 30-4  | 30G      | 1 1/2 | 40                    | 40(+1.5/-2.5)        | 60 |
|                       | 30-3  |          | 1     | 25                    | 25(+1.5/-2.5)        | 60 |
|                       | 30-2  |          | 5/8   | 16                    | 16(+1/-2)            | 60 |
|                       | 30-1  |          | 1/2   | 13                    | 13(+1/-2)            | 60 |

Lengh of cannula by type

### 5. Intended Use:

This device is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.

### 6. Performance data:

- Bench test were performed. Bench testing included biocompatibility, mechanical testing, sterility 1) testing including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.

| Requirement – Test (ISO 7864)                                                                                                                                                                             | Result |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Inner and Outer Surface<br>There must be bumps and flaws on the outer surface. The finished surface must be smooth,<br>and the entire needle tuve must have undergone elecrolytic polishing or equivalent | Pass   |

{4}------------------------------------------------

| polishing treatment.<br>There must be no hazardous oxides, dusts and powders on the inner surface                                                                                                                                                                        | Pass |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| The blade of the needle tuve must be sharply polished without visible bending and have<br>uniform cross section and thickness.                                                                                                                                           | Pass |
| When injecting glycerin into the needle tuve or needle hub, coloring of glycerin should not<br>take place.                                                                                                                                                               | Pass |
| When a lubricant is applied on the needle tube, the lubricant must be non-toxic and satisfy<br>ISO 7864. When observed by unaided eyes, no lubricant drops should be found on the inner<br>and outer surfaces of the needle tube.                                        | Pass |
| <b>Dimensions</b><br>Measured using digital vernier calipers and digital thickness gauge<br>Outer diameter of the needle tube must be satisfied.<br>Length of the needle tube must be satisfied.                                                                         | Pass |
| <b>The needle hub and hub hole</b><br>When the test gauge indicated on the figure below is inserted into the hub hole at low<br>pressure, the taper must fit with the taper gauge of the hub hole. Also, the nozzle of the<br>needle hub must be within the gauge limit. | Pass |
| <b>Elasticity</b><br>Return to original position when bent to 12° using a protractor                                                                                                                                                                                     | Pass |
| <b>Flexural strength</b><br>No snapping when bent to 90° using a protractor                                                                                                                                                                                              | Pass |
| <b>Pullout</b><br>Measurement of maximum value using a tension and compression tester                                                                                                                                                                                    | Pass |

| Requirement - Test (ISO 9626)                                                                                                                                                                                                                                                                                                                                      | Result |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Stiffness<br>When tested in accordance with the method given in ISO 9626 Annex B, apply a downward force given<br>in table at a speed 1 mm/min. Measure the deflection of the tubing at the point of application of the<br>force.                                                                                                                                  | Pass   |
| Resistance to breakage<br>The cannula edg When tested in accordanc with Annex C, the tubing shall not show visible<br>breakage when examined by normal or corrected vision.<br>e shall be sharply ground to a regular cross-section and thickness, with no visible bending                                                                                         | Pass   |
| Resistance to corrosion<br>When tested in accordance with the method given in ISO 9626 Annex D, the tubing is partially immersed<br>in sodium chloride solution( $c(NaCl) = 0.5mol/L$ ) at (23±2)°C, for 7h±5min. The immersed half of the<br>tubing shall show no evidence of corrosion resulting from the test compared visually with the<br>unimmersed portion. | Pass   |

### 2) Biocompatibility

| # | Test item                         | Test method / Test criteria                                                                                | Test result |
|---|-----------------------------------|------------------------------------------------------------------------------------------------------------|-------------|
| 1 | Cytotoxicity                      | ISO 10993-5 Tests for in vitro cytotoxicity                                                                | Pass        |
| 2 | Skin Sensitization<br>Test        | ISO 10993-10 irritation and skin sensitization                                                             | Pass        |
| 3 | Intracutaneous<br>Reactivity Test | ISO 10993-10 Test for irritation and skin sensitization, maximization<br>test for delayed hypersensitivity | Pass        |
| 4 | Acute Systemic<br>Toxicity Test   | ISO 10993-11 Test for systemic toxicity - Acute Systemic Toxicity                                          | Pass        |
| 5 | Pyrogen Test                      | ISO 10993-11 Tests for systemic toxicity, Annex(F) Information on<br>material-mediated pyrogens.           | Pass        |
| 6 | Hemolysis Test                    | ISO 10993-4 Selection of tests for interactions with blood                                                 | Pass        |

{5}------------------------------------------------

#### 3) Sterility and LAL test

| # | Test item                    | Test standard                                                                                                                          | Test result |
|---|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------|
| 1 | LAL test                     | USP39 <85>, Bacterial Endotoxins Test (Unit : EU/Device)                                                                               | Pass        |
| 2 | E.O sterilization validation | According to ISO 11135:2014<br>E.O 30%, CO2 70%<br>Temperature : 55°C<br>Exposure time : 5 hours                                       | Pass        |
| 3 | Sterility test               | According to ISO 11737-2                                                                                                               | Pass        |
| 4 | E.O Residual test            | Under the conditions of ISO 10993-7:2008, Ethylene oxide sterilization residuals, the test articles should meet the test requirements. | Pass        |

The performance tests demonstrated that this device is performs in a substantially equivalent manner to the predicate device.

### 7. Comparison Table

| Manufacturer       | POONGLIM Pharmatech Inc.                                                                                                     | International Medsurg Connection<br>(IMC)                                                                                       | Remark                     |       |      |
|--------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------|------|
| 510(K) No.         |                                                                                                                              | K102584                                                                                                                         |                            |       |      |
| Intended use       | This device is intended for use to inject<br>fluids into or withdraw from parts of<br>the body below the surface of the skin | This device is intended for use to inject<br>fluids into or withdraw from parts of<br>the body below the surface of the skin    | Same                       |       |      |
| Material           |                                                                                                                              |                                                                                                                                 |                            |       |      |
| Hub of needle      | Polypropylene (PP)                                                                                                           | Polypropylene (PP)                                                                                                              | Same                       |       |      |
| Protector          | Polypropylene (PP)                                                                                                           | Polypropylene (PP)                                                                                                              |                            |       |      |
| Cannula            | SUS304                                                                                                                       | SUS 304                                                                                                                         |                            |       |      |
| Adhesive           | Epoxy                                                                                                                        | Epoxy Resin                                                                                                                     |                            |       |      |
| Length             | 1/2", 5/8", 1" & 1⁄2"                                                                                                        | 1/2", 5/8", 3/4", 7/8", 1", 1⁄4" & 1⁄2"                                                                                         | Similar                    |       |      |
|                    |                                                                                                                              |                                                                                                                                 | Including predicate device |       |      |
| Gauge              | 30G                                                                                                                          | 16G, 17G, 18G, 19G, 20G, 21G, 22G,<br>23G, 24G, 25G, 26G, 27G, 28G, 29G &<br>30G                                                | Similar                    |       |      |
|                    |                                                                                                                              |                                                                                                                                 | Including predicate device |       |      |
| Cover<br>dimension | 1/2" size: 60mm<br>5/8" size : 60mm<br>1" size: 60mm<br>1⁄2" size: 60mm                                                      | 1/2" size: 40mm<br>5/8" size: 40mm<br>3/4" size: 40mm<br>7/8" size: 40mm<br>1" size: 40mm<br>1⁄4" size: 56mm<br>1⁄2" size: 56mm | Similar                    |       |      |
|                    | Tip<br>configuration                                                                                                         | Bevel                                                                                                                           |                            | Bevel | Same |

### 8. Conclusion

The device has completed testing to show that the device meets its intended use and demonstrates substantial equivalence to the predicate device, K102584. Therefore, it is concluded that the subject device, Kopac Sterile Needle, is substantially equivalent to the legally marketed predicate device, K102584.

---

**Source:** [https://fda.innolitics.com/device/K172483](https://fda.innolitics.com/device/K172483)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
