← Product Code [JOJ](/productcode/JOJ) · K172472

# Micro-MEC 1.8% Glutaraldehyde Monitor Strip (K172472)

_STERIS Corporation · JOJ · Nov 17, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172472

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [JOJ](/productcode/JOJ.md)
- **Decision Date:** Nov 17, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Micro-MEC 1.8% Glutaraldehyde Monitor Strip is a semi-quantitative chemical indicator for use in determining whether the concentration of glutaraldehyde, the active ingredient, is greater than the minimum effective concentration (MEC) of 1.8% glutaraldehyde established for Micro-Cide™ 28 High Level Disinfectant (Micro-Cide™ 28 HLD).

## Device Story

Device is a single-use chemical indicator strip; consists of an indicator pad on a polymer substrate handle. Used to monitor glutaraldehyde concentration in Micro-Cide™ 28 HLD working solution. User dips strip into solution for 2 seconds; reads result at 120 seconds. Indicator pad changes color from yellow to purple if concentration >1.8%; remains yellow or patchy if ≤1.8%. Provides visual feedback to healthcare personnel regarding whether disinfectant solution requires replacement. Ensures solution efficacy for high-level disinfection; prevents use of sub-therapeutic germicide concentrations.

## Clinical Evidence

Bench testing only. Performance studies confirmed ≥80% pass rate at 2.1% glutaraldehyde and 100% fail rate at 1.8% glutaraldehyde. Additional testing included specificity (tap water exposure), contaminant interference, chemical resistance, shelf-life stability, and blind user testing with inexperienced operators. All tests met pre-defined acceptance criteria.

## Technological Characteristics

Chemical indicator strip; absorbent paper pad on polypropylene substrate. Principle: colorimetric sodium sulfite pH-based reaction. Color change: yellow to purple. Dimensions/form factor: strip. Shelf-life: 24 months unopened, 90 days opened. No energy source or software.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- MetriTest 1.8 Glutaraldehyde Concentration Monitor ([K012335](/device/K012335.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION".

November 17, 2017

STERIS Corporation Jennifer Nalepka Senior Regulatory Affairs Specialist 5960 Heisley Road Mentor. Ohio 44060

Re: K172472

Trade/Device Name: Micro-MEC 1.8% Glutaraldehyde Monitor Strip Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: August 30, 2017 Received: September 1, 2017

Dear Jennifer Nalepka:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Tara A. Ryan -S

for

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

### K172472

Device Name

Micro-MEC 1.8% Glutaraldehyde Monitor Strip

Indications for Use (Describe)

The Micro-MEC 1.8% Glutaraldehyde Monitor Strip is a semi-quantitative chemical indicator for use in determining whether the concentration of glutaraldehyde, the active ingredient in the solutions listed below, is greater than the minimum effective concentration (MEC) of 1.8% glutaraldehyde established for Micro-Cide™ 28 High Level Disinfectant (Micro-Cide™ 28 HLD)

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in large, bold, sans-serif font, with the registered trademark symbol to the right of the word. Below the word "STERIS" is a graphic of several horizontal, wavy lines in blue, resembling water waves.

# 510(k) Summary For Micro-MEC 1.8% Glutaraldehyde Monitor Strip

### Sponsor:

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

### Manufacturer:

Albert Browne, Ltd. Chancery House 190 Waterside Rd. Hamilton Industrial Park Leicester United Kingdom LE5 1QZ

Contact:

Jennifer Nalepka Senior Regulatory Affairs Specialist Telephone: (440) 392-7458 Fax No: (440) 357-9198 E-mail: Jennifer_Nalepka@steris.com

Summary Date: November 15, 2017

Premarket Notification Number: K172472

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

#### 1. Device Name

| Trade Name:            | Micro-MEC 1.8% Glutaraldehyde Monitor Strip                                                  |
|------------------------|----------------------------------------------------------------------------------------------|
| Common/usual Name:     | Chemical Indicator                                                                           |
| Device Classification: | Class II                                                                                     |
| Classification Name:   | Physical/chemical sterilization process indicator<br>(21 CFR 880.2800 (b), Product Code JOJ) |

#### 2. Predicate Device

| K012335 | MetriTest 1.8 Glutaraldehyde Concentration Monitor - the predicate is identical to the proposed device. |
|---------|---------------------------------------------------------------------------------------------------------|
|---------|---------------------------------------------------------------------------------------------------------|

#### 3. Description of Device

The Micro-MEC 1.8% Glutaraldehyde Monitor Strip is a chemical indicator strip consisting of an indicator pad attached to a polymer substrate which serves as a handle. The test strip has been developed to monitor the concentration of glutaraldehyde in Micro-Cide" 28 HLD working solution. The indicator pad is impregnated with an indicator solution that changes color from vellow to purple if the concentration of the germicide exceeds the minimum effective concentration (MEC) of the solution.

The paper pad on an unused indicator strip is yellow in color. After an immersion time (dip time) of two seconds in a Micro-Cide™ 28 HLD solution, it is read at 120 seconds to verify the glutaraldehyde concentration of >1.8%. At 120 seconds the indicator strip will exhibit a stable uniform purple color if the concentration of glutaraldehyde is >2.1%. The strip will appear yellow or patchy yellow/purple if the solution has ≤1.8% glutaraldehyde, signaling to the customer that the solution must be replaced.

The glutaraldehyde concentration of the neat, unused Micro-Cide™ 28 HLD solution is 3%. To ensure an adequate margin of safety over the MEC of the germicide, the indicator was formulated to undergo a complete color change at glutaraldehyde concentrations of 2.1% in at least 80% of the tests conducted. The concentration of glutaraldehyde that will induce a color change is sufficiently higher than the MEC for the Micro-Cide " 28 HLD solution to ensure that the indicator (test strip) will not indicate a PASS in solutions with glutaraldehyde at or below the MEC of 1.8%. However, due to the kinetics of the chemical reaction that produces the color change, a test conducted on a Micro-Cide™ 28 HLD solution

{5}------------------------------------------------

with a glutaraldehyde concentration of 1.83-2.10% may indicate a PASS or a FAIL result.

The proposed test strip is used in the same way and performs the same as the predicate device to indicate the presence of the same glutaraldehyde concentration, >1.8%.

#### 4. Intended Use

The Micro-MEC 1.8% Glutaraldehyde Monitor Strip is a semi-quantitative chemical indicator for use in determining whether the concentration of glutaraldehyde, the active ingredient, is greater than the minimum effective concentration (MEC) of 1.8% glutaraldehyde established for Micro-Cide™ 28 High Level Disinfectant (Micro-Cide™ 28 HLD).

#### 5. Description of Safety and Substantial Equivalence

The proposed and predicate devices are single use chemical indicator strips that monitor the MEC of glutaraldehyde-containing germicide solutions.

Table 5-1 summarizes the comparison between the proposed device, Micro-MEC 1.8% Glutaraldehyde Monitor Strip, and the predicate.

|              | Proposed K172472<br>Micro-MEC 1.8%<br>Glutaraldehyde Monitor<br>Strip                                                                                                                                                                                                                                                                                                                                  | Predicate K012335<br>MetriTest 1.8<br>Glutaraldehyde<br>Concentration Monitor                                                                                                                                                                                                                                                                                              | Comparison                                                                                           |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Intended Use | The Micro-MEC 1.8%<br>Glutaraldehyde Monitor<br>Strip is a semi-<br>quantitative chemical<br>indicator for use in<br>determining whether the<br>concentration of<br>glutaraldehyde, the active<br>ingredient, is above or<br>below the minimum<br>effective concentration<br>(MEC) of 1.8%<br>glutaraldehyde<br>established for Micro-<br>Cide™ 28 High Level<br>Disinfectant (Micro-<br>Cide™ 28 HLD) | The MetriTest 1.8<br>Glutaraldehyde<br>Concentration Monitor is<br>a glutaraldehyde<br>concentration monitor for<br>use in glutaraldehyde-<br>containing germicide<br>solutions with an MEC of<br>1.8% glutaraldehyde.<br>The MetriTest 1.8<br>Glutaraldehyde<br>Concentration Monitor is<br>dedicated for use with<br>Metricide 28 and<br>Metricide Plus 30<br>Solutions. | Both indicators<br>monitor the<br>concentration<br>greater than 1.8%<br>glutaraldehyde<br>germicide. |
| Substrate    | Absorbent Paper                                                                                                                                                                                                                                                                                                                                                                                        | Absorbent Paper                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                 |

## Table 5-1. Technological Comparison to Predicate

{6}------------------------------------------------

# STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION Micro-MEC 1.8% Glutaraldehyde Monitor Strip

| Indicator /<br>reaction | Colorimetric sodium<br>sulfite pH-based reaction | Colorimetric sodium<br>sulfite pH-based reaction | Same                                                                         |
|-------------------------|--------------------------------------------------|--------------------------------------------------|------------------------------------------------------------------------------|
| Backing                 | Polypropylene                                    | Polypropylene                                    | Same                                                                         |
| Color change            | Yellow to purple                                 | Yellow to purple                                 | Same                                                                         |
| Detection               | Greater than 1.8%<br>glutaraldehyde              | Greater than 1.8%<br>glutaraldehyde              | Same                                                                         |
| Viewing                 | Paper on front of strip                          | Paper on front of strip                          | Same                                                                         |
| Shelf-life,<br>unopened | Current testing supports 8<br>months             | 24 months                                        | Target of 24 months<br>for proposed strip                                    |
| Shelf-life.<br>opened   | 90 days                                          | 90 days                                          | Same                                                                         |
| Dip and<br>Read Times   | Dip time = 2 second<br>Read time at 120 seconds  | Dip time = 1 seconds<br>Read time at 60 seconds  | Both dip and read<br>times optimized for<br>each glutaraldehyde<br>solution. |

Table 5-2 summarizes the verification activities that were performed, with their respective acceptance criteria and results, to demonstrate that the Micro-MEC 1.8% Glutaraldehyde Monitor Strip is substantially equivalent to the claimed predicate device when used according to its instructions for use. The product effectively determines whether or not the concentration of the use solution of Micro-Cide 28 HLD solution is greater than 1.8% glutaraldehyde. These studies confirm that the device's performance meets the requirements of its pre-defined acceptance criteria and intended use.

| Table 5-2 Performance Test Summary |  |
|------------------------------------|--|
|------------------------------------|--|

| Test                                | Acceptance Criteria                                                                                                                                                           | Result |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Performance                         | • Minimum 80% pass results when tested with 2.1%<br>glutaraldehyde Micro-Cide™ 28 HLD solution<br>• 100% fail results with 1.8% glutaraldehyde Micro-Cide™<br>28 HLD solution | Pass   |
| Specificity                         | Incomplete color change when exposed to tap water only                                                                                                                        | Pass   |
| Contaminants                        | Meets performance specifications in the presence of<br>organic and inorganic contaminants in the test solution                                                                | Pass   |
| Exposure to aggressive<br>chemicals | Meets performance specifications after exposure to<br>aggressive chemicals.                                                                                                   | Pass   |
| Blind study testing                 | Meets performance specifications with inexperienced users                                                                                                                     | Pass   |
| Shelf life                          | Meets performance specifications at each time point after<br>storage in different environments                                                                                | Pass   |
| In-use (open bottle)                | Meets performance specifications at each time point after<br>storage in different environments                                                                                | Pass   |
| Functional stability                | Meets performance specifications at each time point after<br>storage outside bottle                                                                                           | Pass   |

{7}------------------------------------------------

#### 6. Conclusion

The Micro-MEC 1.8% Glutaraldehyde Monitor Strip intended for use with Micro-Cide™ 28 HLD solution is substantially equivalent to the predicate device, MetriTest 1.8 Glutaraldehyde Concentration Monitor. Based on the nonclinical tests performed the subject device is as safe, as effective and performs at least as well as the legally marketed predicate device, K012335, MetriTest 1.8 Glutaraldehyde Concentration Monitor (21 CFR 880.2800 (b), Product code JOJ).

---

**Source:** [https://fda.innolitics.com/device/K172472](https://fda.innolitics.com/device/K172472)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
