← Product Code [KTT](/productcode/KTT) · K172438

# ARBOR External Fixation System (K172438)

_Globus Medical, Inc. · KTT · Jan 12, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172438

## Device Facts

- **Applicant:** Globus Medical, Inc.
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Jan 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The ARBOR™ External Fixation System is indicated for use in construction of an external fixation frame for the treatment of pediatric and adult fractures and/or reconstruction of long bones (including metacarpal and metatarsal), and the pelvis. The ARBORTM External Fixation System is intended for: - Stabilization of open or closed fractures with soft tissue injuries; - Polytrauma; - Vertically stable pelvic fractures or as a treatment adjunct for vertically unstable pelvic fractures; - Arthrodesis and osteotomies with soft tissue problems; - Revision procedures where other devices have been unsuccessful including failures of total joints; - Neutralization of fractures stabilized with limited internal fixation; - Non-unions/septic non-unions; - Intraoperative reduction/stabilization tool to assist with indirect reduction; - Correction of deformity; and - Unilateral rectilinear bone segment transport or leg lengthening.

## Device Story

ARBOR External Fixation System consists of Schanz pins, external fixation clamps, and bars. Device provides mechanical stabilization for bone fractures and reconstruction. Used by surgeons in clinical settings (OR/hospital) to construct external fixation frames. System components are modular; clamps connect pins to bars to create rigid constructs. Output is a physical fixation frame that stabilizes bone segments, facilitating healing, reduction, or deformity correction. Benefits include stabilization of complex fractures, polytrauma management, and revision of failed internal fixation.

## Clinical Evidence

Bench testing only. Mechanical testing included insertion torque, pullout, static torsion, and static axial compression per ASTM F543 and F1541. Biocompatibility confirmed via material standards. MRI safety evaluated per ASTM F2052, F2182, and F2213. Bacterial endotoxin testing performed per ANSI/AAMI ST-72:2011.

## Technological Characteristics

System comprises metallic Schanz pins, clamps, and bars. Materials meet applicable standards for orthopedic implants. Mechanical testing performed per ASTM F543 (pins) and F1541 (external fixation components). MRI safety tested per ASTM F2052, F2182, and F2213. No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes Large External Fixation System ([K031428](/device/K031428.md))
- Gexfix External Fixation System ([K052605](/device/K052605.md) & [K160972](/device/K160972.md))
- SBi Mini Rail External Fixation System ([K093550](/device/K093550.md))
- Stryker Hoffmann 3 ([K122284](/device/K122284.md))
- Apex Pins ([K061493](/device/K061493.md))
- Synthes Medium External Fixation System ([K040258](/device/K040258.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 12, 2018

Globus Medical Inc. Kelly Baker Senior Vice President/Regulatory and Clinical Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403

Re: K172438

Trade/Device Name: ARBOR External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT, LXT Dated: December 12, 2017 Received: December 13, 2017

Dear Kelly Baker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172438

Device Name ARBORTM External Fixation System

#### Indications for Use (Describe)

The ARBOR™ External Fixation System is indicated for use in construction of an external fixation frame for the treatment of pediatric and adult fractures and/or reconstruction of long bones (including metacarpal and metatarsal), and the pelvis.

The ARBORTM External Fixation System is intended for:

- · Stabilization of open or closed fractures with soft tissue injuries;
- · Polytrauma;
- · Vertically stable pelvic fractures or as a treatment adjunct for vertically unstable pelvic fractures;
- · Arthrodesis and osteotomies with soft tissue problems;
- · Revision procedures where other devices have been unsuccessful including failures of total joints;
- · Neutralization of fractures stabilized with limited internal fixation;
- · Non-unions/septic non-unions;
- · Intraoperative reduction/stabilization tool to assist with indirect reduction;
- · Correction of deformity; and
- · Unilateral rectilinear bone segment transport or leg lengthening.

| Type of Use (Select one or both, as applicable) | <div> <span> <input type="checkbox"/> </span> <span> <input type="checkbox"/> </span> </div> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------|

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary: ARBOR™ External Fixation System

| Company:        | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800                                                                                               |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:        | Kelly J. Baker, Ph.D.<br>Senior Vice President, Regulatory and Clinical Affairs                                                                                                       |
| Date Prepared:  | January 9, 2018                                                                                                                                                                       |
| Device Name:    | ARBOR™ External Fixation System                                                                                                                                                       |
| Common Name:    | External Fixation System                                                                                                                                                              |
| Classification: | Per 21 CFR as follows:<br>§888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories<br>Product Codes: KTT, LXT<br>Regulatory Class: II, Panel Code: 87 |

Primary Predicate: Synthes Large External Fixation System (K031428)

| Additional<br>Predicates: | Gexfix External Fixation System (K052605 & K160972)<br>SBi Mini Rail External Fixation System (K093550)<br>Stryker Hoffmann 3 (K122284) & Apex Pins (K061493)<br>Synthes Medium External Fixation System (K040258) |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

### Purpose:

The purpose of this submission is to request clearance for the ARBOR™ External Fixation System.

## Device Description:

The ARBOR™ External Fixation System is comprised of Schanz pins, external fixation clamps, and bars. The pins and bars are available in various sizes, and the fixation clamps are available in several designs all capable of use with any size pins and bars.

### Indications for Use:

The ARBOR™ External Fixation System is indicated for use in construction of an external fixation frame for the treatment of pediatric and adult fractures and/or reconstruction of long bones, small bones (including metacarpal and metatarsal), and the pelvis.

{4}------------------------------------------------

The ARBOR™ External Fixation System is intended for:

- . Stabilization of open or closed fractures with soft tissue injuries;
- Polytrauma: ●
- Vertically stable pelvic fractures or as a treatment adjunct for vertically . unstable pelvic fractures;
- Arthrodesis and osteotomies with soft tissue problems; ●
- . Revision procedures where other devices have been unsuccessful including failures of total joints;
- . Neutralization of fractures stabilized with limited internal fixation;
- . Non-unions/septic non-unions;
- Intraoperative reduction/stabilization tool to assist with indirect reduction: ●
- Correction of deformity; and ●
- Unilateral rectilinear bone segment transport or leg lengthening. ●

### Performance Data:

Performance of the ARBOR™ External Fixation System was evaluated in accordance with ASTM F543 and F1541. Engineering analysis and mechanical testing of pins, bars, and clamps was performed, including insertion torque, pullout, static torsion, and static axial compression. Performance data demonstrates substantial equivalence to the predicate devices.

Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST-72:2011. Biocompatibility of patient-contacting materials was demonstrated by using materials that meet applicable standards or are used in 510(k) cleared devices. MRI testing was performed on the subject devices per ASTM F2052, F2182, and F2213.

## Technological Characteristics:

The ARBOR™ External Fixation System implants have the same technological characteristics as the predicate devices including design, intended use, material composition, function, and range of sizes.

### Conclusions:

The ARBOR™ External Fixation System has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K172438](https://fda.innolitics.com/device/K172438)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
