Synermed Opiate Enzyme Immunoassay

K172416 · Infrared Laboratory Systems, LLC (Dba Synermed) · DJG · Apr 19, 2018 · Clinical Toxicology

Device Facts

Record IDK172416
Device NameSynermed Opiate Enzyme Immunoassay
ApplicantInfrared Laboratory Systems, LLC (Dba Synermed)
Product CodeDJG · Clinical Toxicology
Decision DateApr 19, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3650
Device ClassClass 2

Intended Use

The Synermed Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of opiates in human urine at a cutoff value of 300 ng/mL when calibrated against morphine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. This assay is for prescription use only. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmatory method such as GCMS or permitting laboratories to establish quality control procedures. The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

Device Story

Synermed Opiate Enzyme Immunoassay; competitive enzyme immunoassay for urine opiate detection. Input: human urine samples. Principle: competition between drug in sample and morphine-labeled G6PDH enzyme for fixed antibody; enzyme activity inversely proportional to drug concentration. Output: spectrophotometric absorbance change at 340 nm; qualitative or semi-quantitative results. Used in clinical laboratories; operated by trained laboratory personnel on Synermed IR-500 Chemistry Analyzer. Output used by clinicians to identify need for confirmatory testing (GC/MS or LC/MS) and clinical decision-making regarding drug use. Benefits: rapid preliminary screening for opiates.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per CLSI EP05-A3 and EP12-A using 11 urine concentrations (80 measurements each). Linearity assessed per NCCLS EP06-A across 10 concentrations. Analytical specificity tested against 10 endogenous compounds at +/- 25% of cutoff. Method comparison performed on 100 samples against Hitachi 717 and LC-MS/MS; demonstrated 91% agreement with LC-MS/MS for positive samples and 100% for negative samples.

Technological Characteristics

Enzyme immunoassay reagent kit; contains mouse monoclonal anti-morphine antibody, G6PDH enzyme-drug conjugate, NAD, and sodium azide preservative. Operates via competitive binding; absorbance measured on automated clinical chemistry analyzers (e.g., IR-500). Storage at 2-8°C. No specific ASTM materials cited.

Indications for Use

Indicated for qualitative and semi-quantitative determination of opiates in human urine at 300 ng/mL cutoff. For professional, prescription use only. Intended for use with automated clinical chemistry analyzers to provide preliminary results requiring confirmation by GC/MS or LC/MS.

Regulatory Classification

Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION". April 19, 2018 Infrared Laboratory Systems, LLC (dba Synermed) Julie Paschal Regulatory Affairs Specialist 17408 Tiller Court, Suite 1900 Westfield, IN 40674 Re: k172416 Trade/Device Name: Synermed Opiate Enzyme Immunoassay Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG Dated: February 28, 2018 Received: March 5, 2018 Dear Julie Paschal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR {1}------------------------------------------------ 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Kellie B. Kelm -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) k172416 Device Name Synermed Opiate Enzyme Immunoassay #### Indications for Use (Describe) The Synermed Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of opiates in human urine at a cutoff value of 300 ng/mL when calibrated against morphine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. This assay is for prescription use only. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmatory method such as GCMS or permitting laboratories to establish quality control procedures. The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/ MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive. | Type of Use (Select one or both, as applicable) | | | | | |--------------------------------------------------|---------------------------------------------|--|--|--| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # k172416 510(k) Summary #### 1. Company Information Infrared Laboratory Systems, LLC 17408 Tiller Court Suite 1900 Westfield, Indiana 46074 Telephone; (317) 896-1565 FAX: (317) 896-1566 #### 2. Contact Information Julie Paschal Regulatory Affairs Specialist Telephone: 336-269-5069 Email: jpaschal@slplabs.com - 3. Date Prepared: August 3rd 2017 - 4. Device Trade Name: Synermed Opiate Enzyme Immunoassay - 5. Common Name: Opiate Enzyme Immunoassay - 6. Classification Name: Enzyme Immunoassay, Opiate - 7. Classification Regulation: 21CFR862.3650 - 8. Classification Product Code: DJG - 9. Panel: Toxicology (91) #### 10. Reagent Device Classification: | Pro Code | Classification Regulation | Classification Name | Device Class | Panel | |----------|---------------------------|----------------------------|--------------|-------| | DJG | 862.3650 | Enzyme Immunoassay, Opiate | Class II | 91 | {4}------------------------------------------------ #### 11. Identification of Predicates: | K-number | Manufacturer | Product | Method | |----------|--------------------------|----------------------------------|-----------------------| | K110298 | Lin-Zhi<br>International | Enzyme<br>Immunoassay,<br>Opiate | Enzyme<br>Immunoassay | #### 12. Device Description #### Synermed Opiate Enzyme Immunoassay The Synermed Opiate Enzyme Immunoassay is ready to use. The composition of the Synermed Opiate Enzyme Immunoassay Reagent is as follows: Antibody/Substrate Reagent (R1): Contains mouse monoclonal anti-morphine antibody, opiate-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09 %) as a preservative. Enzyme-drug Conjugate Reagent (R2): Contains morphine-labeled opiate-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide (0.09 %) as a preservative. ## Intended Use #### Synermed Opiate Enzyme Immunoassay The Synermed Enzyme Immunoassay is intended for the qualitative and semi-quantitative determination of opiates in human urine at a cutoff value of 300 ng/mL when calibrated against morphine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. This assay is for prescription use only. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmatory method such as GCMS or permitting laboratories to establish quality control procedures. The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or Liguid Chromatography/Mass Spectrometry (GC/MS or LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive. ## 13. Comparison of Technological Characteristics with the Predicate Device The new device is substantially equivalent to the predicate because it has the same intended use and has the same or similar technological characteristics including safety and effectiveness. The same specimen type can be analyzed to detect the amount of absorbance which is proportional to the concentration of the analyte in the specimen. | Items | | |-------|--------------------------------------------------------------| | | Candidate Device: Synermed IR-500 Opiate Enzyme Immunoassay | | | Predicate Device: Lin-Zhi Opiate Immunoassay | {5}------------------------------------------------ | Similarity/Difference | | | |-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Intended Use | The Synermed Opiate Enzyme Immunoassay The<br>Opiate Enzyme Immunoassay from<br>Immunoassay, when used in conjunction with<br>Synermed IR series analyzers is intended fordetermination of morphine in human<br>urine at a cutoff value of 300 ng/mL in urine.<br>The assay is designed for professional use. | Same | | Analyte | Morphine | Same | | Specimen | Urine | Same | | Cutoff | 300 ng/mL | Same | | Matrix | Urine | Same | | Storage | 2-8°C until expiration date | Same | | Calibration/QC | Programmable Cal/OC | Same | ## 15. Summary of Performance Testing The purpose of the performance studies was to validate that the previously cleared reagents have the same performance characteristics on the proposed new analyzer (Synermed IR-500) as compared to the previously cleared Hitachi 717. The sponsor has chosen the representative analyte Opiate using urine as the representative sample matrix. ## Analytical Performance ## a. Precision/Reproducibility Verification of semi-quantitative precision on the Synermed IR-500 was accomplished by implementing the study protocol laid out in NCCLS document EP05-A3, Chapter 3. Qualitative verification was accomplished by implementing the study protocol laid out in CLSI EP12-A, Chapter 7. Eleven concentrations of pooled human urine were run for opiate. Each aliquot was run in duplicate twice a day for twenty days for a total of 80 measurements at each concentration. The mean, standard deviation and coefficient of variation were determined for opiate at each concentration. For qualitative analysis, the results were compared to the expected result (negative expected results should yield negative results in the precision study). Each level of pooled human urine was confirmed by Ultra High-Performance Liquid Chromatography Tandem Mass Spectrometry, for verification chromatography refer to Appendix A: UHPLC-MSMS Raw Data. {6}------------------------------------------------ Table 3 – Semi-Quantitative Opiate Precision Comparison to Lin-Zhi Results (per Package Insert) | Opiate Tested<br>Concentrations | IR-500<br>Within-Run<br>Expected | Results<br>Pos/Neg | IR-500<br>Run-To-Run<br>Expected | Results<br>Pos/Neg | |---------------------------------|----------------------------------|--------------------|----------------------------------|--------------------| | 0 ng/mL | - | - | - | - | | 75 ng/mL | - | - | - | - | | 150 ng/mL | - | - | - | - | | 225 ng/mL | - | - | - | - | | 300 ng/mL | + | + | + | + | | 375 ng/mL | + | + | + | + | | 450 ng/mL | + | + | + | + | | 525 ng/mL | + | + | + | + | | 600 ng/mL | + | + | + | + | | 800 ng/mL | + | + | + | + | | 1000 ng/mL | + | + | + | + | # Table 4 – Qualitative Opiate Precision Comparison to Lin-Zhi Results (per Package Insert) | Opiate Tested<br>Concentrations | Within-Run<br>Results<br>Pos/Neg | Run-to-Run<br>Results<br>Pos/Neg | Expected<br>Result<br>(Pos/Neg) | |---------------------------------|----------------------------------|----------------------------------|---------------------------------| | 0 ng/mL | - | - | - | | 75 ng/mL | - | - | - | | 150 ng/mL | - | - | - | | 225 ng/mL | - | - | - | | 300 ng/mL | + | + | + | | 375 ng/mL | + | + | + | | 450 ng/mL | + | + | + | | 525 ng/mL | + | + | + | | 600 ng/mL | + | + | + | | 800 ng/mL | + | + | + | | 1000 ng/mL | + | + | + | {7}------------------------------------------------ ## b. Linearity/Reportable Range Linearity studies were designed using NCCLS EP06-A. Samples were prepared by intermixing a high urine pool with a low urine pool to obtain ten concentrations across the measuring range with four replicates at each concentration. The observed values were compared to the expected values and are summarized below. Each level of pooled human urine was confirmed by Ultra High-Performance Liquid Chromatography Tandem Mass Spectrometry. | "Expected" Value<br>(ng/mL) | Mean Observed Value<br>(ng/mL) | Recovery<br>(%) | |-----------------------------|--------------------------------|-----------------| | 1250 | 1032.5 | 82.6 | | 1100 | 907.3 | 82.5 | | 1000 | 875.0 | 87.5 | | 800 | 662.0 | 82.8 | | 600 | 513.8 | 85.6 | | 450 | 369.5 | 82.1 | | 300 | 240.5 | 80.2 | | 225 | 192.3 | 85.4 | | 150 | 135.0 | 90.0 | | 75 | 67.3 | 89.7 | | 25 | 15.5 | 62.0 | | 0 | 11.0 | #DIV/0! | | Table 3 - Opiate Analytical Recovery | | | | |--------------------------------------|--|--|--| | | | | | {8}------------------------------------------------ ## c. Analytical Specificity The following endogenous compounds were spiked into urine spiked with morphine to +/- 25% of cutoff (225 and 375 ng/mL). The spiked solutions were evaluated on the IR-500. The substances listed in the following table were determined not to interfere at the concentrations tested. | Compound | Concentration<br>ng/mL | -25% Cutoff | | +25% Cutoff | | |------------------------|------------------------|-------------|----------------|-------------|----------------| | | | Qual | Semi-<br>Quant | Qual | Semi-<br>Quant | | Acetone | 1000 | Negative | Negative | Positive | Positive | | Ascorbic Acid | 1500 | Negative | Negative | Positive | Positive | | Creatinine | 500 | Negative | Negative | Positive | Positive | | Ethanol | 1000 | Negative | Negative | Positive | Positive | | Glucose | 3000 | Negative | Negative | Positive | Positive | | Hemoglobin | 300 | Negative | Negative | Positive | Positive | | Human Serum<br>Albumin | 500 | Negative | Negative | Positive | Positive | | Riboflavin | 0.3 | Negative | Negative | Positive | Positive | | Sodium<br>Chloride | 6000 | Negative | Negative | Positive | Positive | | Urea | 6000 | Negative | Negative | Positive | Positive | {9}------------------------------------------------ #### d. Comparison Studies Method comparison was performed according to CLSI EP09-A3, 100 samples for opiate were tested on the IR-500, Hitachi 717 chemistry analyzers, and Ultra High-Performance Liquid Chromatography Tandem Mass Spectrometry (LC-MSMS). The study results are summarized in the table below. Each level of pooled human urine was confirmed by Ultra High-Pressure Liquid Chromatography Tandem Mass Spectrometry, for verification chromatography refer to Appendix A: UHPLC-MSMS Raw Data. ## Table 2 - IR-500 Summary of Semi-Quantitative Comparison Data vs LCMS | Candidate<br>Device<br>Results | Negative | Low<br>Negative<br>(< 50% of the<br>cutoff by<br>LC/MS) | Near Cutoff<br>Negative<br>(50% below<br>the cutoff to<br>the cutoff by<br>LC/MS) | Near Cutoff<br>Positive<br>(50% above<br>the cutoff to<br>the cutoff by<br>LC/MS) | Percent<br>Agreement<br>with LC/MS | |--------------------------------|----------|---------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------| | Positive | 0 | 0 | 0 | 11 | 91% | | Negative | 5 | 19 | 30 | 4 | 100% | | Discrepant<br>Sample # | Drug/Metabolite LC/MS value based on<br>cross reactivity profile | POS/NEG<br>Result | POS/NEG<br>Result | |------------------------|------------------------------------------------------------------|-------------------|-------------------| | 50 | 321.82 ng/mL (Morphine) | Positive | Negative | | 53 | 306.20 ng/mL (Morphine) | Positive | Negative | | 57 | 314.80 ng/mL (Morphine) | Positive | Negative | | 58 | 327.67 ng/mL (Morphine) | Positive | Negative | {10}------------------------------------------------ #### 16. Conclusions The new device is substantially equivalent to the predicate because it has the same intended use and has the same or similar technological characteristics that do not raise new types of questions of safety and effectiveness.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%