K172414 · Global Mailer Partners, LLC · HIS · Oct 12, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K172414
Device Name
JUD custom fit condom
Applicant
Global Mailer Partners, LLC
Product Code
HIS · Obstetrics/Gynecology
Decision Date
Oct 12, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
The JUD custom fit condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases).
Device Story
JUD custom fit condom is a natural rubber latex sheath designed to cover the penis; features reservoir end and cylindrical shape. Available in 23 distinct length/width combinations to provide a custom fit. User selects appropriate size using provided fitting kit and instructions. Pre-lubricated with silicone-based lubricant; contains no spermicide, flavors, or color additives. Used by individuals for contraception and prophylactic protection against sexually transmitted diseases. Principle of operation involves physical barrier placement on erect penis prior to sexual contact to prevent transmission of bodily fluids and infectious organisms. Benefits include improved fit through size selection, potentially enhancing comfort and effectiveness.
Clinical Evidence
Bench testing only. Testing performed per ISO 4074:2015 and WHO 2010 standards, including dimensional analysis, lubricant quantity, burst volume/pressure, freedom from holes, and package seal integrity. Biocompatibility testing conducted per ISO 10993 series (cytotoxicity, acute systemic toxicity, vaginal irritation, skin sensitization) confirmed no adverse reactivity. Shelf-life testing validated 5-year stability.
Technological Characteristics
Material: Natural rubber latex. Design: Cylindrical sheath with reservoir tip. Dimensions: 23 variants (lengths 130-199mm, widths 45-64mm). Lubricant: Silicone-based. Sterilization: Not specified. Connectivity: None. Software: None. Standards: ISO 4074:2015, ISO 10993.
Indications for Use
Indicated for contraception and prevention of pregnancy and sexually transmitted diseases in individuals using male condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{1}------------------------------------------------
#### 510(k) SUMMARY
| Date of Summary: | September 26, 2017 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Applicant: | Global Mailer Partners, LLC<br>50 South Pointe Drive,<br>Suite 1401,<br>Miami Beach, LF 33139<br>Jud Ireland, CEO<br>516-279-7745<br>jp.ireland@gmail.com |
| Contact Person: | Adam Harris, MM, RAC<br>Regulatory Consultant<br>Associate Director, Regulatory Affairs<br>Target Health Inc.<br>261 Madison Ave 24th Floor<br>New York, NY 10016<br>Phone: 646-218-2009<br>Fax: 212-682-2105<br>aharris@targethealth.com |
| Proprietary/Trade Name: | JUD custom fit condom |
| Common Name: | Male Latex Condom |
| Classification Name: | Condom |
| Device Classification | Class II per 21 CFR §884.5300 |
| Product Code | HIS, Condom |
| Predicate Device | They Fit Condom K150072 |
| Reference Device | They Fit Condom K122219 |
The predicate and reference devices have not been subject to a design-related recall.
#### Device Description:
The JUD custom fit condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. The condom is shaped with a reservoir end and is flat with a cylindrical shape. JUD custom fit condoms are available in 23 sizes of different length/width combinations. There is a fitting kit and instructions included as part of safe and effective usage of the condom.
JUD custom fit condoms are provided pre-lubricated with a silicone-based lubricant. JUD custom fit condoms are not provided with spermicide, and contain no flavors or color additives. The user must use a fitting kit to select the appropriate size of the JUD custom fit condom.
{2}------------------------------------------------
# Indications for Use
The JUD custom fit condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases).
| | SIZE<br>CODE | LENGTH<br>(mm) | WIDTH (mm) |
|-----|--------------|----------------|------------|
| 1. | 54W | 130 | 45 |
| 2. | 54E | 130 | 47 |
| 3. | 54J | 130 | 51 |
| 4. | 36J | 150 | 51 |
| 5. | 75J | 173 | 51 |
| 6. | 54M | 130 | 49 |
| 7. | 36M | 150 | 49 |
| 8. | 54N | 130 | 57 |
| 9. | 36N | 150 | 57 |
| 10. | 75N | 173 | 57 |
| 11. | 29N | 199 | 57 |
| | SIZE CODE | LENGTH (mm) | WIDTH (mm) |
| 12. | 36P | 150 | 60 |
| 13. | 75P | 173 | 60 |
| 14. | 29P | 199 | 60 |
| 15. | 75R | 173 | 64 |
| 16. | 29R | 199 | 64 |
| 17. | 54S | 130 | 53 |
| 18. | 36S | 150 | 53 |
| 19. | 75S | 173 | 53 |
| 20. | 54U | 130 | 55 |
| 21. | 36U | 150 | 55 |
| 22. | 75U | 173 | 55 |
| 23. | 29U | 199 | 55 |
# Comparison of Technological Characteristics
| Characteristic | Subject Device | Predicate Device | Reference Device | Comparison |
|---------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------|------------|
| Device Name | Jud custom fit condom | They Fit Condom | They Fit Condom | N/A |
| 510(k) Number | K172414 | K150072 | K122219 | N/A |
| Common Name | Male Latex Condom | Male Latex Condom | Male Latex Condom | N/A |
| Manufacturer | Richter Rubber Technology, Sdn. Bhd.<br>Plot 33 Kuala Ketil Industrial Estate, 09300 Kuala Ketil, Malaysia | They Fit Condom Michael Cecil, MD<br>4140 Tate Street Covington, GA 30014 | They Fit Condom Michael Cecil, MD<br>4140 Tate Street Covington, GA 30014 | N/A |
| Indications for Use | The JUD custom fit condom is used for contraception and for prophylactic purposes (to help | The condom is used for contraception and for prophylactic purposes to help | The condom is used for contraception and for prophylactic purposes (to help | Same |
{3}------------------------------------------------
| | prevent pregnancy<br>and the transmission<br>of sexually<br>transmitted<br>diseases). | prevent pregnancy<br>and the<br>transmission of<br>sexually<br>transmitted<br>diseases. | prevent pregnancy<br>and the<br>transmission of<br>sexually<br>transmitted<br>diseases). | |
|----------------------|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Size(s) | Varied sizes with<br>different<br>length/width<br>combinations. | Varied sizes with<br>different<br>length/width<br>combinations. | Varied sizes with<br>different<br>length/width<br>combinations. | Same |
| Length | 130mm, 150mm,<br>173mm, 199mm | 123mm, 133mm,<br>143mm, 153mm | 163mm, 178mm,<br>193mm, 208mm | Different;<br>subject<br>device<br>lengths fall<br>within the<br>range of the<br>predicate<br>and<br>reference<br>devices. |
| Width | 45 mm, 47mm,<br>49mm, 51mm,<br>53mm, 55mm,<br>57mm, 60mm,<br>64mm | 45mm, 47mm,<br>49mm, 51mm,<br>53mm, 57mm,<br>60mm, 64mm | 49mm, 51mm,<br>53mm, 55mm,<br>57mm, 60mm,<br>64mm | Same |
| Reservoir tip | yes | yes | yes | Same |
| Single use<br>device | yes | yes | yes | Same |
| Natural Latex | yes | yes | yes | Same |
The principle of operation of the JUD custom fit condoms, placing condom on erect penis before sexual contact to avoid transmission of bodily fluid and infectious organisms within, is identical to predicate. As noted in the table above, JUD custom fit condoms have the same indications for use and basic technological characteristics as the predicate and reference condoms. The JUD custom fit condoms are made of natural rubber latex, are prelubricated, and have similar sizes to the predicate. The JUD custom fit condom dimensions fall within the range of sizes of the predicate and reference devices, though they are offered in different size configurations. The difference in available sizes of the JUD custom fit condoms does not raise different questions of safety and effectiveness.
{4}------------------------------------------------
# Nonclinical Testing Summary
#### Bench testing
Dimensional tests, burst volume and pressure tests, and freedom from holes and visible defects tests have been conducted per WHO 2010 and ISO 4074:2015. Package seal integrity test has been conducted per ISO 4074:2015. The dimensions and burst volume and pressure of the condoms tested conform to the aforementioned standards. There were no visible defects or damaged package seal identified in the tested samples.
| Description | Summary and Conclusion |
|----------------------------------------|-----------------------------------|
| Dimensional | ISO 4074: Acceptance criteria met |
| Lubricant quantity | ISO 4074: Acceptance criteria met |
| Burst Volume and Pressure | ISO 4074: Acceptance criteria met |
| Freedom from Holes and Visible Defects | ISO 4074: Acceptance criteria met |
| Package Seal Integrity | ISO 4074: Acceptance criteria met |
## Shelf Life
The JUD custom fit condom has a shelf-life of 5 years. The results of shelf life testing show that each condom variant maintained its specifications for the proposed shelf-life.
## Biocompatibility
The JUD custom fit condom is considered a surface contacting device of category A – limited contact of <24 hours on intact skin/mucosal membrane. The skin contacting component is the latex condom itself made from natural latex rubber. These condoms contain no flavors or color additives.
| The following biocompatibility testing was conducted on the JUD custom fit condom: | |
|------------------------------------------------------------------------------------|--|
|------------------------------------------------------------------------------------|--|
| Biological<br>Endpoint | Description | ISO Standard | Results | Test article |
|-------------------------------|----------------------------------------------------------|-----------------------|-------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Cytotoxicity | In-vitro cytotoxicity_using<br>the direct contact method | ISO 10993-<br>5:2009 | No cytotoxic<br>reactivity was found | Identical to<br>finished<br>devices* |
| Acute<br>systemic<br>toxicity | Systemic toxicity in Swiss<br>albino mice | ISO 10993-<br>11:2006 | There was no<br>mortality or evidence<br>of systemic toxicity<br>resulted from the test<br>article extract. | Identical to<br>finished<br>devices |
| Vaginal<br>irritation | Vaginal irritation in rabbits | ISO 10993-<br>10:2010 | The test article was<br>found to be non-<br>irritating, i.e., did not<br>show any signs of<br>vaginal irritation. | Identical to<br>finished<br>devices |
| Skin | Skin sensitization | ISO 10993- | The test article | Identical to |
{5}------------------------------------------------
| sensitization | (maximization) test in<br>Guinea Pigs | 10:2010 | extract did not<br>induce skin<br>sensitization. | finished<br>devices |
|---------------|---------------------------------------|---------|--------------------------------------------------|---------------------|
|---------------|---------------------------------------|---------|--------------------------------------------------|---------------------|
*The test article utilized for the biocompatibility testing was a single condom variant. All JUD custom fit condoms undergo the same manufacturing process, have the same chemical, physical, and surface properties, and have the same ratio of component materials as the test article.
## Substantial Equivalence Conclusion
The subject device is identical to the predicate devices in intended use. The subject device, predicate and reference devices are constructed out of the same materials (natural rubber latex), and have the same basic technological characteristics. The subject device dimensions fall within those of the predicate and reference devices, though the subject device has slightly different size configurations. Both the subject device and predicate devices conform to the same standards for performance specifications. Differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The non-clinical testing conducted on the subject device, including dimensional testing, lubricant quantity, burst volume and pressure testing, freedom from holes and visible defects testing, package seal integrity, shelf life, cytotoxicity, acute systemic toxicity, vaginal irritation and skin sensitization demonstrate that the subject device is as safe and effective as the predicate. Therefore, the JUD custom fit condom is substantially equivalent to the predicate, TheyFit.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, forming a symbolic representation of people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 12, 2017
Global Mailer Partners, LLC % Adam Harris. MM. RAC Associate Director, Regulatory Affairs Target Health, Inc. 261 Madison Avenue, 24th Floor New York, NY 10016
Re: K172414 Trade/Device Name: JUD custom fit condom Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: August 9, 2017 Received: August 10, 2017
Dear Adam Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.