← Product Code [JDR](/productcode/JDR) · K172405

# SNIPER STAPLE SYSTEM, NON-STERILE (K172405)

_Trilliant Surgical · JDR · Dec 11, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172405

## Device Facts

- **Applicant:** Trilliant Surgical
- **Product Code:** [JDR](/productcode/JDR.md)
- **Decision Date:** Dec 11, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Trilliant Sniper Staple System is indicated for fixation of fractures and osteotomies of the hand, foot, and bones appropriate for the size of the device.

## Device Story

The Sniper Staple System consists of U-shaped Nitinol (Nickel-Titanium alloy) bone staples. These devices are used by orthopedic surgeons for the fixation of bone fractures and osteotomies in the hand and foot. The staples function via shape-memory properties to provide compression across bone segments. The system is provided non-sterile and requires sterilization at the user's facility prior to surgical implantation. The device is intended to stabilize bone segments to facilitate healing.

## Clinical Evidence

No clinical data. Bench testing only, including elastic static bending strength (ASTM F564-10), pull-out fixation strength (ASTM F564-10), transformation temperature testing (ASTM F2082), and corrosion resistance testing (ASTM F2129).

## Technological Characteristics

Material: Nitinol per ASTM F2063. Design: U-shaped metallic fixation device with two legs and a single bridge. Energy source: Shape-memory alloy (passive). Connectivity: None. Sterilization: User-facility sterilization required. Software: None.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Sniper Staple System, Sterile ([K162354](/device/K162354.md))

## Reference Devices

- Tiger Cannulated Screw System ([K081510](/device/K081510.md))
- Gridlock Plating System ([K121452](/device/K121452.md))
- Tiger Cannulated Screw System ([K153338](/device/K153338.md))
- Metasurg Nitinol Staple ([K123926](/device/K123926.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 11, 2017

Trilliant Surgical Jon Olson President, CEO 6721 Portwest Drive. Suite 160 Houston, Texas 77024

# Re: K172405

Trade/Device Name: Sniper Staple System. Non-sterile Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDR Dated: November 8, 2017 Received: November 8, 2017

Dear Jon Olson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

K172405

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K172405

Device Name Trilliant Sniper Staple System

Indications for Use (Describe)

The Trilliant Sniper Staple System is indicated for fixation of fractures and osteotomies of the hand, foot, and bones appropriate for the size of the device.

| Type of Use (Select one or both, as applicable)                                                 |
|-------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">     Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="font-family: sans-serif;">     Over-The-Counter Use (21 CFR 801 Subpart C)</span>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Sniper Staple System

In accordance with 21 CFR 807.92(c) of the Federal Code of Regulations, the following 510(k) summary is submitted for the Sniper Staple System.

| I. GENERAL INFORMATION |                                                                                                                                                                       |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared          | December 11, 2017                                                                                                                                                     |
| Submitted By           | Trilliant Surgical LTD<br>6721 Portwest Dr.<br>Suite 160<br>Houston, TX 77024<br>Telephone: 713-388-6055<br>Contact: Jon Olson<br>Email: jolson@trilliantsurgical.com |
| Trade Name             | Sniper Staple System                                                                                                                                                  |
| Common Name            | Bone staple                                                                                                                                                           |
| Device                 | Staple, Fixation, Bone                                                                                                                                                |
| Regulatory Description | Single/multiple component metallic bone fixation appliances and<br>accessories.                                                                                       |
| Class                  | II                                                                                                                                                                    |
| Product Code           | JDR                                                                                                                                                                   |
| CFR Section            | 21 CFR Section 888.3030                                                                                                                                               |
| Device Panel           | Orthopedic                                                                                                                                                            |
| Predicate Device       | Sniper Staple System, Sterile (Trilliant Surgical LTD, K162354)                                                                                                       |

#### -GENERAL INFORMATION

#### II. DEVICE DESCRIPTION

The Sniper Staple System consists of different-sized staples composed of Nitinol (Nickel-Titanium alloy) intended for fixation of fractures and osteotomies of the hand, foot, and bones appropriate for the size of the device. The system will be offered non-sterile in a tray for sterilization at the user's facility prior to use.

# Materials

Nitinol per ASTM F2063

## Substantial Equivalence Claimed to Predicate Devices

The Sniper Staple System is substantially equivalent to the primary predicate device in terms of intended use, design, materials used, mechanical safety and performance. The Sniper Staple System is sterilized to the same sterility assurance level as the predicate device. The technological characteristics of the proposed device and the predicate are identical: both are U-shaped metallic fixation devices made from Nitinol that use two legs and a single bridge to fixate bone segments.

#### lll. Indications for Use

The Trilliant Sniper Staple System is indicated for fixation of fractures and osteotomies of the hand, foot, and bones appropriate for the size of the device.

{4}------------------------------------------------

#### IV. Non-Clinical Test Summary

The following tests were performed:

- Elastic Static Bending Strength per ASTM F564-10 (2015), Standard o Specification and Test Methods for Metallic Bone Staples, specifically Annex IV - Test Method for Elastic Static Bending of Metallic Bone Staples.
- . Pull-Out Fixation Strength per ASTM F564-10 (2015), Standard Specification and Test Methods for Metallic Bone Staples, specifically Annex II - Test Method for Pull-Out Fixation Strength of Metallic Bone Staples.
- . Transformation temperature testing per ASTM F2082 (2016), Standard Test Method for Determination of Transformation Temperature of Nickel-Titanium Shape Memory Alloys by Bend and Free Recovery
- o Corrosion resistance testing per ASTM F2129 (2017), Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices

The results of these evaluations indicate that the Sniper Staples are safe and equivalent to the predicate device.

#### V. Clinical Test Summary

No clinical studies were performed.

#### VI. Conclusions: Non-Clinical and Clinical

Trilliant Surgical LTD considers the Sniper Staple System to be equivalent to the predicate device listed above. This conclusion is based upon the devices' similarities in principles of operation, technology, materials, and indications for use.

#### VII. Other device submission references

Multiple device submissions from Trilliant Surgical were used to demonstrate acceptable biocompatibility of the subject device. The reference submissions are as follows:

| 510(K) Number | Device Name                   | Decision Date | Manufacturer |
|---------------|-------------------------------|---------------|--------------|
| K081510       | Tiger Cannulated Screw System | 08/04/2008    | Trilliant    |
| K121452       | Gridlock Plating System       | 08/20/2012    | Trilliant    |
| K153338       | Tiger Cannulated Screw System | 01/14/2016    | Trilliant    |
| K123926       | Metasurg Nitinol Staple       | 05/03/2013    | Metasurg     |

---

**Source:** [https://fda.innolitics.com/device/K172405](https://fda.innolitics.com/device/K172405)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
