K172404 · GE Healthcare · MUE · Oct 30, 2017 · Radiology
Device Facts
Record ID
K172404
Device Name
SenoBright HD
Applicant
GE Healthcare
Product Code
MUE · Radiology
Decision Date
Oct 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1715
Device Class
Class 2
Indications for Use
SenoBright HD is an extension of the existing indication for diagnostic mammography with Senographe Pristina. The SenoBright HD application shall enable contrast enhanced breast imaging a dual energy technique. This imaging technique can be used as an adjunct following mammography and ultrasound exams to help localize a known or suspected lesion.
Device Story
SenoBright HD is a software-based imaging application for the Senographe Pristina FFDM system. It performs contrast-enhanced spectral mammography (CESM) using a dual-energy technique. Following an iodine-based contrast injection, the system acquires two X-ray exposures: one optimized for general mammography and one optimized for iodine contrast. The software processes and recombines these images to produce both a standard FFDM-like image and a recombined contrast-enhanced image. Used in clinical settings by radiologists, the output allows simultaneous visualization of breast tissue and contrast enhancement. This aids in localizing lesions, potentially improving diagnostic confidence and clinical decision-making for breast cancer assessment.
Clinical Evidence
Clinical image evaluation performed to assess image quality in a clinical setting. Expert radiologists reviewed images and confirmed that the image quality of SenoBright HD is appropriate for clinical use. No specific sensitivity/specificity metrics provided.
Technological Characteristics
Software-based imaging feature for Senographe Pristina FFDM system. Utilizes dual-energy X-ray acquisition (low and high energy spectra). Features AOP (Automatic Optimization of Parameters) based on breast thickness and density. Includes artifact management algorithms for skin and nipple regions. Operates as an extension of existing digital mammography hardware.
Indications for Use
Indicated for patients requiring adjunct contrast-enhanced breast imaging following mammography and ultrasound to localize known or suspected lesions.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
{1}------------------------------------------------
October 30, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare GE Medical Systems SCS % Mounir Zaouali Regulatory Affairs Leader 283 rue de la Miniere Buc, 78530 FRANCE
Re: K172404
Trade/Device Name: SenoBright HD Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: August 4, 2017 Received: August 9, 2017
Dear Mounir Zaouali:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K172404
Device Name SenoBright HD
Indications for Use (Describe)
SenoBright HD is an extension of the existing indication for diagnostic mammography with Senographe Pristina. The SenoBright HD application shall enable contrast enhanced breast imaging a dual energy technique. This imaging technique can be used as an adjunct following mammography and ultrasound exams to help localize a known or suspected lesion.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are four white swirls around the letters. The logo is simple and recognizable.
Section 5: 510(k) Summary
SenoBright HD
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. There are also three white curved lines emanating from the circle, giving the impression of movement or energy.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | August 4, 2017 |
|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>GE Medical Systems SCS<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE |
| Primary Contact Person: | Mounir Zaouali,<br>Regulatory Affairs Leader,<br>GE Medical Systems SCS,<br>283 RUE DE LA MINIERE<br>78530 BUC - FRANCE Phone : + 33 1 30 70 45 39<br>Phone : + 33 1 30 70 45 39<br>Fax : + 33 1 30 70 41 40<br>Email : Mounir.Zaouali@ge.com |
| Secondary Contact Person: | Diane Uriell<br>Sr. Director of Regulatory Affairs, XR and Women's Health,<br>GE Healthcare,<br>Atlanta, Georgia - USA<br>Phone +1 262 290 8212<br>Email: Diane.Uriell@ge.com |
| Device Trade Name: | SenoBright HD |
| Common/Usual Name: | Full Field Digital Mammography System |
| Classification Names: | 21 CFR 892.1715, Class II |
| Product Code: | MUE |
| Predicate Device(s): | K103485<br>Contrast Enhanced Spectral Mammography (CESM) |
| Device Description: | The subject of this submission is a modification of Senographe<br>Pristina FFDM system (cleared in K162268) that will introduce a new |
| | imaging option called SenoBright HD. This imaging option has been<br>previously cleared for Senographe Essential FFDM system in<br>K103485,marketed as SenoBright Contrast Enhanced Spectral<br>Mammography (CESM).<br>The dual energy exposures will be done following an iodine based<br>contrast injection of the patient and with a single breast compression. |
| | The new mode of operation for Senographe Pristina system is<br>referred to as SenoBright HD Contrast Enhanced Spectral<br>Mammography (CESM) due to the nature of taking an exposure with<br>the x-ray spectrum optimized for general mammographic imaging and<br>a second exposure with the x-ray spectrum optimized for the iodine<br>based contrast image. |
| | The main modification of the Senographe Pristina system is the<br>addition of software feature to control the low and high energy<br>images and post processing and recombination of those images to<br>create FFDM like image and recombined image. These two images<br>allow the visualization of the breast tissue in a way that is typical and<br>familiar for mammographic imaging and the x-ray contrast<br>enhancement in the breast at the same time. |
| Intended Use: | SenoBright HD is an extension of the existing indication for diagnostic<br>mammography with Senographe Pristina. The SenoBright application<br>shall enable contrast enhanced breast imaging using a dual energy<br>technique. This imaging technique can be used as an adjunct<br>following mammography and ultrasound exams to help localize a<br>known or suspected lesion. |
| Technology: | SenoBright HD and predicate device, SenoBright Contrast Enhanced<br>Spectral Mammography (CESM), have the same intended use, same<br>fundamental scientific technology, principle of operation and similar<br>operating parameters. |
| Comparison of Technological<br>Characteristics with The<br>Predicate Device: | SenoBright HD |
| | The main changes between SenoBright HD and the predicate device<br>SenoBright Contrast Enhanced Spectral Mammography (CESM)<br>include: |
| | 1) Different hosting mammography system (Senographe<br>Essential vs Senographe Pristina. |
| | 2) Improvements to the acquisition technique, the high energy<br>AOP takes into account both the breast thickness and the breast<br>density to determine the optimal X-rays techniques. This is based on<br>the predicate and Senographe Pristina low energy AOP that takes into<br>account both breast thickness and density. |
| | 3) Additional artefacts management is incorporated to minimize the artefacts that may appear at the breast skin line and nipple. |
| Determination of Substantial<br>Equivalence: | SenoBright HD has the same indications for use as its predicate device. It uses the same fundamental technology as Contrast Spectral Mammography (CESM). Differences between SenoBright HD do not raise new questions of safety and effectiveness. The impact of differences between the predicate and new device has been assessed using applicable FDA Guidance and Standards. In particular, the following testing was performed: • Complete Image Quality Bench Performance testing was completed showing appropriate performance. • A Clinical image evaluation was performed showing that image quality of SenoBright HD in a clinical setting is good as assessed by expert radiologists. • Successful System and Software verification and validation. |
| Conclusion: | GE Healthcare considers the SenoBright HD to be as safe, as effective, and performance is substantially equivalent to the predicate device(s). |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The logo is colored in a light blue hue. The circular frame has small, wave-like embellishments around its perimeter, adding a touch of visual interest to the design. The logo is simple, recognizable, and conveys a sense of tradition and innovation.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. There are also some water droplet-like shapes around the circle.
# GE Healthcare 510(k) Premarket Notification Submission
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.