← Product Code [MTD](/productcode/MTD) · K172393

# Advisor HD Grid Mapping Catheter, Sensor Enabled (K172393)

_St. Jude Medical, Inc. · MTD · Apr 23, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172393

## Device Facts

- **Applicant:** St. Jude Medical, Inc.
- **Product Code:** [MTD](/productcode/MTD.md)
- **Decision Date:** Apr 23, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Advisor HD Grid Mapping Catheter, Sensor Enabled, is indicated for multiple electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

## Device Story

Sterile, single-use, irrigated, high-density mapping catheter; 7.5F shaft; 8F deflectable distal section; paddle-shaped distal tip with 16 electrodes; two distal shaft ring electrodes; two magnetic sensors. Used in cardiac electrophysiology labs by physicians. Catheter connects to EnSite Velocity or EnSite Precision Cardiac Mapping Systems via electrical connector; includes fluid lumen for irrigation. Magnetic sensors enable 3D visualization and localization within the mapping system. Physician manipulates catheter handle to deflect paddle; electrodes record cardiac electrical signals (electrograms) or provide stimulation. Output displayed on mapping system console; assists physician in identifying cardiac arrhythmias and guiding clinical interventions. Benefits include high-density mapping capabilities for improved diagnostic accuracy in atrial and ventricular regions.

## Clinical Evidence

Bench, in vivo, and biocompatibility testing performed. Bench testing included physical/dimensional characteristics, handling (actuation force, torque, column strength), durability, irrigation performance, electrical/magnetic characteristics, and compliance with ANSI/AAMI/IEC 60601 and ISO 10555-1. Biocompatibility testing per ISO 10993 confirmed safety (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility, thrombogenicity). In vivo studies included a GLP safety study and an accuracy study. Thrombogenicity testing conducted with heparin (ACT > 300).

## Technological Characteristics

7.5F shaft, 8F deflectable distal section; paddle-shaped tip with 16 electrodes; two magnetic sensors for 3D localization; polymer braided shaft. Irrigated tip design. Compatible with EnSite mapping systems. Complies with ANSI/AAMI/IEC 60601 (electrical safety, defibrillation protection) and ISO 10555-1 (radiopacity, tensile strength, corrosion resistance). Biocompatibility per ISO 10993.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- PentaRay NAV High Density Mapping Catheter, Biosense Webster ([K120425](/device/K120425.md))
- Advisor FL Circular Mapping Catheter, Sensor Enabled, St. Jude Medical ([K160335](/device/K160335.md))

## Reference Devices

- EnSite Velocity Cardiac Mapping System v5.2
- EnSite Precision Cardiac Mapping System v2.2

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two main elements: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue.

April 23, 2018

St. Jude Medical, Inc. Jennifer Ruether Regulatory Affairs Manager 5050 Nathan Lane North Plymouth, Minnesota 55442

Re: K172393

Trade/Device Name: Advisor HD Grid Mapping Catheter, Sensor Enabled Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: March 14, 2018 Received: March 16, 2018

Dear Jennifer Ruether:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M.A. Wilhelmsen
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K172393

Device Name

Advisor HD Grid Mapping Catheter, Sensor Enabled

Indications for Use (Describe)

The Advisor HD Grid Mapping Catheter, Sensor Enabled, is indicated for multiple electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

| Type of Use (Select one or both, as applicable)                                  |  |
|----------------------------------------------------------------------------------|--|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> |  |
| <div> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div>   |  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the St. Jude Medical logo. The logo consists of a green square grid with one square rotated, followed by the text "ST. JUDE MEDICAL" in black. Below the logo, the text "St. Jude Medical is now Abbott." is written in blue.

Abbott Laboratories St. Jude Medical 5050 Nathan Lane North Plymouth, MN 55442 USA 651 756 5400 Main 651 756 5355 Fax

510(k) Number K172393

# Submitter Information

| Date Prepared               | April 13, 2018                                                                                               |
|-----------------------------|--------------------------------------------------------------------------------------------------------------|
| Submitter Name &<br>Address | St. Jude Medical<br>5050 Nathan Lane North<br>Plymouth, MN 55442<br>Phone: 651 756 5400<br>Fax: 651 756 5355 |
| Contact Person              | Jennifer Ruether<br>Regulatory Affairs Manager                                                               |

# Device Identification

| Trade Name            | Advisor™ HD Mapping Catheter, Sensor Enabled™                                                                                                                                                                                                                                                                 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name           | Catheter, Intracardiac Mapping, High-Density Array                                                                                                                                                                                                                                                            |
| Classification        | 21 CFR 870.1220; Electrode recording catheter or electrode recording probe;<br>Procode MTD; Class II                                                                                                                                                                                                          |
| Predicate Device      | Primary: PentaRay NAV High Density Mapping Catheter, Biosense Webster;<br>K120425                                                                                                                                                                                                                             |
|                       | Secondary: Advisor FL Circular Mapping Catheter, Sensor Enabled, St. Jude<br>Medical; K160335                                                                                                                                                                                                                 |
| Reference Application | EnSite™ Velocity™ Cardiac Mapping System v5.2 & EnSite Precision™<br>Cardiac Mapping System v2.2 (currently under review with FDA in a<br>concomitant 510(k) submission)                                                                                                                                      |
| Indications for Use   | The Advisor HD Grid Mapping Catheter, Sensor Enabled, is indicated for<br>multiple electrode electrophysiological mapping of cardiac structures in the<br>heart, i.e., recording or stimulation only. This catheter is intended to obtain<br>electrograms in the atrial and ventricular regions of the heart. |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a green square made up of smaller squares, with one square tilted. To the right of the square is the text "ST. JUDE MEDICAL" in a bold, sans-serif font. Below the company name is the text "St. Jude Medical is now Abbott" in a smaller, blue font.

Abbott Laboratories St. Jude Medical 5050 Nathan Lane North Plymouth, MN 55442 USA 651 756 5400 Main 651 756 5355 Fax

The Advisor HD Grid Mapping Catheter, Sensor Enabled, is a sterile, singleuse, irrigated, high-density mapping catheter with a 7.5F shaft and an 8F distal shaft deflectable section. It is available in a D-F bi-directional curve model that is deflected using the actuator located on the catheter handle. The catheter working length is 110 cm. The device consists of a paddle-shaped distal tip with 16 electrodes, two distal shaft ring electrodes, two magnetic sensors, polymer braided shaft, handle, fluid lumen extension with a luer, and an electrical Device Description connector. The catheter also has an introducer tool intended to compress and guide the distal paddle into, and withdraw from, the hemostasis valve of an introducer sheath. The catheter is compatible with the EnSite Velocity and EnSite Precision Cardiac Mapping Systems and other accessories, including the connecting cable and commercially available irrigation pumps. The Advisor HD Grid Mapping Catheter, Sensor Enabled, is similar to the predicate devices in design and performance. All catheters use multiple electrodes to facilitate electrophysiological mapping and are available in similar dimensions and curve shapes; additionally, all incorporate magnetic sensors to accommodate use with the respective compatible mapping systems. Like the PentaRay catheter, the subject device incorporates a fluid lumen for irrigation of Technological the distal tip and is indicated for use in the atria and ventricles; the catheters Characteristics differ in terms of distal tip shape, and compatible mapping systems. The Advisor HD Grid and the Advisor FL catheters share the same shaft and handle design, and are compatible with the same mapping system; due to the difference in distal tip designs, the subject device differs from the Advisor FL catheter in irrigation and indications for use. Technological differences between the subject and predicate devices in indications for use, irrigation, and system-level accuracy resulted in the same questions of safety and effectiveness. The Advisor HD Grid Mapping Catheter, Sensor Enabled, is substantially equivalent to the predicate devices based on comparisons of the device functionality, technological characteristics, and intended use. The identified differences between the technological characteristics have been evaluated through the following bench, in vivo, and biocompatibility testing: Substantial . Physical and Dimensional Characteristics: Outer dimension Equivalence and O Summary of Studies Electrode spacing O Paddle length O Paddle alignment O Deflection curve shape, angle, and planarity O

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a green square made up of smaller squares, with one square tilted. To the right of the square is the text "ST. JUDE MEDICAL" in a bold, sans-serif font. Below the logo is the text "St. Jude Medical is now Abbott" in a smaller, blue font.

#### Abbott Laboratories

St. Jude Medical 5050 Nathan Lane North Plymouth, MN 55442 USA Main 651 756 5400 651 756 5355 Fax

- . Handling Characteristics:
	- Actuation force/torque O
	- O Freedom from play
	- Manual/adjustable lock O
	- Torque response O
	- O Column strength
	- O Paddle stiffness
- Durability Characteristics:
	- O Torque and tensile strength (after conditioning)
- . Performance Characteristics:
	- Irrigation pressure, leakage, flow rate, luer compliance to ISO O 594-2
	- Electrical resistance, isolation, fluid ingress O
	- Magnetic resistance, isolation, signal characteristics, O compatibility with EnSite systems
- Compliance to ANSI/AAMI/IEC 60601: ●
	- Patient leakage ೧
	- Dielectric strength O
	- Defibrillation protection O
	- Anchorage flex O
	- O Handle push
- Compliance to ISO 10555-1:
	- Radiopacity O
	- Surface/visual inspection O
	- Corrosion resistance O
	- 0 Tensile strength
- Compliance to ISO 1099993: ●
	- Cytotoxicity O
	- Sensitization O
	- Irritation O
	- Acute Systemic Toxicity O
	- Pyrogenicity O
	- Hemocompatibility O
	- o Thrombogenicity
- In vivo studies: ●
	- GLP safety study O
	- Accuracy study O

All thrombogenicity testing was conducted using heparin to maintain an ACT value higher than 300; therefore, Advisor HD Grid requires use of anticoagulation for both right and left side procedures, and is labeled accordingly. The resulting evidence obtained from the design verification and validation testing did not raise new questions of safety and effectiveness, and demonstrates that the subject device is as safe and effective as the predicate devices.

Substantial Equivalence and Summary of Studies

---

**Source:** [https://fda.innolitics.com/device/K172393](https://fda.innolitics.com/device/K172393)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
