The MicroSTAAR® Injector System Cattridge, Model CQ is a device intended to fold and insert STAAR Surgical Company Collamer® aphakic 3-piece Intraocular Lenses for surgical placement in the eye.
Device Story
Single-use polypropylene cartridge; used with reusable MicroSTAAR MSI-PM or MSI-TM injector. Facilitates loading, folding, and insertion of STAAR Surgical Collamer® 3-piece intraocular lenses (IOL) into the human eye following cataract extraction. Surgeon places lens into cartridge loading area; cartridge is loaded onto injector; injector advancement ejects lens through nozzle into eye. Device ensures lens recovery to specifications after folding/deformation. Terminally sterilized via Ethylene Oxide.
Clinical Evidence
Bench testing only. Validation included mechanical property recovery testing per ISO 11979-3:2012, shelf-life/stability testing per ISO 11979-6:2014, and biocompatibility testing per ISO 10993-1:2009. All tests met acceptance criteria.
Technological Characteristics
Single-use polypropylene cartridge; Ethylene Oxide sterilization. Dimensions and design identical to predicate. Standards: ISO 11979-3:2012 (mechanical properties), ISO 11979-6:2014 (shelf-life), ISO 10993-1:2009 (biocompatibility).
Indications for Use
Indicated for folding and inserting STAAR Surgical Company Collamer® aphakic 3-piece Intraocular Lenses for surgical placement in the eye.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 6, 2017
STAAR Surgical Company Kate Ueda, JD Manager of Regulatory Affairs 1911 Walker Avenue Monrovia, CA 91016
Re: K172375
Trade/Device Name: MicroSTAAR Injector System Cartridge, Model CQ Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS Dated: August 4, 2017 Received: August 7, 2017
Dear Kate Ueda, JD:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172375
Device Name
MicroSTAAR Injector System Cartridge, Model CQ
Indications for Use (Describe)
The MicroSTAAR® Injector System Cattridge, Model CQ is a device intended to fold and insert STAAR Surgical Company Collamer® aphakic 3-piece Intraocular Lenses for surgical placement in the eye.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This summary document has been prepared in accordance with section 21 CFR 807.92(c).
The submitter of the 510(k) is:
| Owner's Name: | STAAR Surgical Company |
|------------------------|--------------------------------------------------|
| Address: | 1911 Walker Avenue<br>Monrovia, California 91016 |
| Contact Person: | Kate Ueda |
| Title: | Manager of Regulatory Affairs |
| Phone: | 626.303.7902 ext. 2316 |
| Email: | kueda@Staar.com |
| Date Summary Prepared: | August 4, 2017 |
#### A. Subject Device:
| Trade name: | MicroSTAAR® Injector System Cartridge, Model CQ |
|----------------------|-------------------------------------------------|
| Common Name: | Intraocular Lens Injector Cartridge |
| Classification Name: | Folders & Injectors, Intraocular Lens (IOL) |
| Class: | Class I – MSS, 21 CFR 886.4300 |
#### B. Predicate Device:
MicroSTAAR108 Injector System Cartridge, K101782 is used as the predicate device for this Special 510(k) Notification. There are three (3) models cleared in K101782, SFC-45, SFC-25, and the CQ Cartridge. This Special 510(k) Notification addresses a design modification of the model CQ Cartridge only.
#### C. Device Description:
The MicroSTAAR Injector System Cartridge, Model CQ (also referred as "MicroSTAAR Injector CQ Cartridge" or "CQ Cartridge") is intended to facilitate the loading, folding and insertion of the STAAR Surgical Collamer® 3-piece Intraocular Lens (IOL) into the human eye following cataract extraction. The model CQ Cartridge is a single use device designed specifically to be used with the reusable MicroSTAAR MSI-PM or MSI-TM injector to deliver the STAAR Surgical Collamer® 3-piece (IOL). The CQ Cartridge is composed of polypropylene and is terminally sterilized using Ethylene Oxide.
#### D. Indications for Use:
The MicroSTAAR® Injector System Cartridge, Model CQ is a device intended to fold and insert STAAR Surgical Company Collamer® aphakic 3-piece Intraocular Lenses for surgical placement in the eye. These are the same indications for use as the predicate model CQ Cartridge.
{4}------------------------------------------------
#### E. Summary of Technological Characteristics
The MicroSTAAR Injector System Cartridge, Model CQ is a single use polypropylene cartridge designed to be used with the reusable MicroSTAAR MSI-PM or MSI-TM injector to fold and deliver the STAAR Surgical Collamer® 3-piece (IOL) in the human eye. It is identical in design, dimensions, processing and packaging as the predicate device. The only difference is the polypropylene formulation composing the device. The following table provides a comparison to the predicate device.
| Characteristic | Predicate Device | Subject Device |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MicroSTAARTM Injector System Cartridges<br>K 101782 | Proposed MicroSTAAR®<br>Injector System Cartridge<br>Model CQ |
| Indications for use | The MicroSTAARTM Injector System Cartridges is a device<br>intended to fold and insert STAAR Surgical Company<br>Collamer® phakic and aphakic one-piece and 3-piece Intraocular<br>Lenses for surgical placement in the eye. | Same<br>(As modified for the model<br>CQ cartridge only-<br>The MicroSTAAR® Injector<br>System Cartridge, Model<br>CQ is a device intended to<br>fold and insert STAAR<br>Surgical Company<br>Collamer® aphakic 3-piece<br>Intraocular Lenses for<br>surgical placement in the<br>eye.) |
| Contraindications | Use of the MicroSTAAR Injector with other intraocular implants<br>and cartridges that have not been tested or approved for use with<br>the MicroSTAAR Injector System and Cartridges | Same |
| Anatomical site | The human eye | Same |
| Injector<br>configuration<br>(single use) | The MicroSTAAR Injector Cartridges are used in conjunction<br>with appropriate injector systems with 1-piece and 3-piece<br>phakic and aphakic Collamer lenses. | Same<br>(As modified for the model<br>CQ cartridge only) |
| How is the device<br>used | The lens is placed into the loading area of the cartridge and<br>folded shut. Then the cartridge is loaded onto the injector. When<br>the injector is advanced, the lens is ejected from the cartridge<br>nozzle into the human eye. | Same |
| Single use? | The MicroSTAAR Injector Cartridges are intended for single<br>use only | Same |
| Is the product<br>sterile? | The MicroSTAAR Injector Cartridges are supplied sterile | Same |
| How sterilized | Ethylene Oxide (EtO) | Same |
| How supplied | The MicroSTAAR Injector Cartridge is placed in a tray, sealed<br>with a Tyvek lid and labeled. The tray is sealed in a pre-printed,<br>labeled Tyvek pouch and terminally sterilized. | Same |
# COMPARISON OF PREDICATE DEVICE TO THE PROPOSED DEVICE
{5}------------------------------------------------
The technological characteristics of the modified CQ Cartridge are substantially equivalent to the predicate CQ Cartridge, cleared in K101782. As demonstrated above, the modified and predicate CQ Cartridges share the same intended use, indications, operating principle, design, packaging and sterilization method.
#### F. Brief Summary of Nonclinical Test and Results:
Results from validation testing of the modified MicroSTAAR Injector CQ Cartridge demonstrate that the cartridge functions as intended. In accordance with applicable tests in ISO 11979-3:2012 Ophthalmic implants -- Intraocular lenses -- Part 3: Mechanical properties and test methods (i.e., Section 5, Recovery of Properties following simulated surgical manipulation), intraocular lenses recovered to specifications after being folded and deformed by the cartridge.
A shelf-life study was initiated and is on-going for the modified cartridge under real time and accelerated conditions in accordance with applicable tests in ISO 11979-6:2014, Ophthalmic implants -- Intraocular lenses -- Part 6: Shelf-life and transport stability testing.
The CQ Cartridge composed of the modified polypropylene formulation has been tested for biocompatibility to ensure that the new material formulation meets the requirements of ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, and the FDA Guidance for Industry and Food and Drug Administration Staff dated June 16, 2016, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
#### Conclusions Drawn from Non-clinical Studies G.
A summary of the results of the non-clinical studies is provided below.
| Study | Result | Is the Subject Device<br>as safe and effective as<br>the Predicate Device? |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Recovery of Properties | The change in material formulation does<br>not impact the performance characteristics<br>of the cartridge or lens. | Yes |
| Shelf-Life | The change in material formulation does<br>not impact the performance characteristic<br>of the cartridge over the stated shelf life. | Yes |
| Biocompatibility | The biocompatibility tests performed on the<br>modified formulation model CQ cartridge<br>met all acceptance criteria. | Yes |
### Non-Clinical Study Results
{6}------------------------------------------------
The data generated from the non-clinical studies of the modified MicroSTAAR Injector cartridge, Model CQ support the subject device substantial equivalence to the predicate device.
The comparison of technical characteristics and data generated from the non-clinical studies demonstrate the substantial equivalence of the subject device and the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.