K172367 · Stryker Instruments · GEI · Aug 22, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172367
Device Name
Stryker Integrated Bipolar Cord and Tubing Sets
Applicant
Stryker Instruments
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 22, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.
Device Story
Integrated tubing and bipolar cord sets; interface between irrigation unit/electrosurgical generator and irrigating bipolar forceps. Device delivers irrigation fluid and electrical energy simultaneously to surgical site. Available in gravity-type or rotary-type configurations; includes options for flying leads or unitized leads. Used in clinical settings by surgical staff. Provides irrigation and coagulation capabilities; supports surgical workflow by consolidating connections. Benefits patient by enabling precise electrosurgical coagulation with simultaneous irrigation to manage thermal effects.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-5, -10, -11), mechanical/functional testing (durometer, PSI levels), electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-2), and packaging/sterilization validation (ASTM D4169, F1886, F2096, F88; ISO 11135).
Technological Characteristics
Integrated irrigation tubing (silicone/PVC) and bipolar electrical cord. Sterilized via EtO (10-6 SAL). Electrical safety per ANSI/AAMI ES60601-1, IEC 60601-1-2, and IEC 60601-2-2. Packaging per ASTM D4169, F1886, F2096, F88. Biocompatibility per ISO 10993 series.
Indications for Use
Indicated for use with irrigating bipolar forceps to provide simultaneous irrigation and electrosurgical energy during surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Malis Bipolar Cord/Irrigation Tubing Set (K890648)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, featuring three human profiles in a row.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2017
Stryker Instruments Ms. Susanne Galin Principal Regulatory Affairs Specialist 4100 E. Milham Avenue Kalamazoo, Michigan 49001
Re: K172367
Trade/Device Name: Stryker Integrated Bipolar Cord and Tubing Sets Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 2, 2017 Received: August 4, 2017
Dear Ms. Galin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172367
Device Name
Stryker Integrated Bipolar Cord and Tubing Sets
Indications for Use (Describe)
The Stryker Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.
Type of Use (*Select one or both, as applicable*)
| <span></span> Residential Real Estate (CRE) Sales or Financing | <span></span> Other Transaction Not Involving CRE |
|----------------------------------------------------------------|---------------------------------------------------|
|----------------------------------------------------------------|---------------------------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Stryker Integrated Bipolar Cord and Tubing Sets is provided below.
| Device Common Name: | Electrosurgical, Cutting & Coagulation & Accessories |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | Stryker Integrated Bipolar Cord and Tubing Sets |
| Applicant: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001 |
| Contact: | Susanne Galin<br>Principal Regulatory Affairs Specialist<br>Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001<br>Desk Phone +1 269 389 4132<br>Cell Phone +1 812 345 7917<br>Email: susanne.galin@stryker.com |
| Date Prepared: | August 21, 2017 |
| Classification Regulation: | 21 CFR §878.4400- Electrosurgical, Cutting &<br>Coagulation & Accessories, Class II |
| Panel: | General & Plastic Surgery |
| Product Code: | GEI - Electrosurgical, Cutting & Coagulation &<br>Accessories |
| Predicate Device: | K890648, Malis Bipolar Cord/Irrigation Tubing Set |
### Indication for Use:
The Stryker Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.
### Device Description:
The subject devices are the Integrated Tubing and Bipolar Cord Sets. They are integrated irrigation tubing and electrical cord sets to be used with irrigating bipolar forceps, and interface with both the irrigation unit and the electrosurgical generator.
There are two types of sets available from Stryker: one type that is for use with a rotary irrigator unit and one type that is for use with a gravity irrigation unit. Additionally, each type is available with either "flying leads" or a "unitized lead".
{4}------------------------------------------------
| Part Number | Item Name | Description |
|--------------|-----------------------------------------------------------------------|------------------------------|
| 6790-100-001 | Integrated Tubing and Bipolar Cord Set,<br>Disposable | Gravity-type, flying lead. |
| 6790-100-002 | Integrated Tubing and Bipolar Cord Set,<br>Disposable, Unitized Lead | Gravity-type, unitized lead. |
| 6790-100-003 | Bipolar Irrigator Integrated Tubing Set,<br>Disposable | Rotary-type, flying lead. |
| 6790-100-004 | Bipolar Irrigator Integrated Tubing Set,<br>Disposable, Unitized Plug | Rotary-type, unitized lead. |
## Listing of Subject Devices
# Device Comparison Table:
| Technological<br>Characteristic | Integrated Tubing &<br>Bipolar Cord Set (as<br>cleared in K890648) | Currently Marketed Gravity Tubing & Bipolar<br>Cord Set | |
|-----------------------------------|--------------------------------------------------------------------|---------------------------------------------------------|------------------------------------------------|
| | | Gravity-type | Rotary-type |
| Electrical Cord<br>Length | 12 ft bipolar electrical cord | 10.958 ft ± 0.375 ft | 10.958 ft ± 0.375 ft |
| Lead type | flying leads | flying leads and unitized<br>leads | flying leads and unitized<br>leads |
| Electrical Cord<br>Connector | Female connector which<br>connects to forceps | Female connector which<br>connects to forceps | Female connector which<br>connects to forceps |
| Irrigation<br>Tubing Length | Total Length: 15 ft<br>irrigation tubing | Total length: 15.833 ft<br>(nominal) | Total Length: 15.583 ft<br>(nominal) |
| | 8.5 inch section serves as<br>pump chamber | 8.5 inch serves as pump<br>chamber | 5.5 inch serves as pump<br>chamber |
| Irrigation<br>Tubing<br>Connector | Spike with drip chamber | Spike with no drip<br>chamber | Spike with no drip<br>chamber |
| | Male Luer connects to<br>forceps (tubing) | Male Luer connects to<br>forceps (tubing) | Male Luer connects to<br>forceps (tubing) |
| Materials | Silicone tubing – Pump<br>PVC tubing - bag | Silicone tubing – Pump<br>PVC tubing - Bag | Silicone tubing – Pump<br>PVC tubing - Bag |
| Silicone tubing<br>wall thickness | Not provided in K890648 | OD: 0.167" – 0.173"<br>ID: 0.117" - 0.123" | OD: 0.117 (+0.008/-0.0)"<br>ID: 0.028"- 0.035" |
{5}------------------------------------------------
| Technological<br>Characteristic | Integrated Tubing &<br>Bipolar Cord Set (as<br>cleared in K890648) | Currently Marketed Gravity Tubing & Bipolar<br>Cord Set | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| | | Gravity-type | Rotary-type |
| Packaging | Peel pouch, Mylar to Tyvek | Poly bag inside a peel-<br>pouch (60GA BIAX<br>Nylon/2.0 MIL LDPE,<br>Uncoated Tyvek 1073B) | Poly bag inside a peel-<br>pouch (60GA BIAX<br>Nylon/2.0 MIL LDPE,<br>Uncoated Tyvek 1073B) |
| Sterilization<br>Method | EtO | EtO | EtO |
| Sterilization<br>Parameters | 120° F Temperature<br>50° F Relative Humidity<br>750 Mgs. Per liter of Pure<br>EtO<br>3 hr Gas Exposure<br>2 Air washes | 130° F Temperature<br>Humidity Dwell: 30 min at<br>4.0"HgA<br>790 Mg/L 100% EtO<br>6 hrs 1 min gas exposure<br>2 Air/nitrogen/steam<br>washes | 130° F Temperature<br>Humidity Dwell: 30 min<br>at 4.0"HgA<br>790 Mg/L 100% EtO<br>6 hrs 1 min gas exposure<br>2 Air/nitrogen/steam<br>washes |
| SAL | 10-6 SAL | 10-6 SAL | 10-6 SAL |
| Shelf Life | Not provided in K890648 | 3 years | 3 years |
## Performance Data:
The following summary of V&V Testing is provided to establish substantial equivalence.
| V&V Testing Summary | |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | |
| Cytotoxicity (MEM<br>Elution) | ISO 10993-5:2009 Biological Evaluation of Medical Devices, Part 5:<br>Tests for In Vitro Cytotoxicity |
| Systemic Toxicity | ISO 10993-11: 2006 Biological Evaluation of Medical Devices Part 11:<br>Tests for Systemic Toxicity |
| Material Mediated<br>Pyrogenicity | ISO 10993-11:2006, Biological Evaluation of Medical Devices, Part 11:<br>Tests for Systemic Toxicity. |
| Irritation/Intracutaneous<br>Reactivity | ISO 10993-10: 2010 Standard, Biological Evaluation of<br>Medical Devices, Part 10: Tests for Irritation and Skin Sensitization. |
| Sensitization (Guinea Pig<br>Maximization) | ISO 10993-10:2010. Biological Evaluation of Medical Devices, Part 10:<br>Tests for Irritation and Skin Sensitization. |
| Cytotoxicity (MEM<br>Elution) after 2X<br>sterilization | ISO 10993-5:2009 Biological Evaluation of Medical Devices, Part 5:<br>Tests for In Vitro Cytotoxicity |
{6}------------------------------------------------
| Mechanical and Performance Testing | |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional Testing of<br>Cord & Tube Set | Functional Testing of Cord & Tube Set after Packaging Testing<br>Testing conducted on: |
| | • Unaged Samples |
| | • Samples Accelerated Aged to 1.5 years |
| | • Samples Accelerated Aged to 3 years |
| | Cord Set Durometer Testing |
| | Verification PSI Level Testing for Disposable Cable with Tubing, Rotary |
| Electrical Safety Testing | |
| Electrical Safety Test<br>Report | ANSI/AAMI ES60601-1:2005, Medical Electrical Equipment – Part 1:<br>General Requirements for Basic Safety and Essential Performance;<br>Consolidated Reprint (2009); Amendment 2 (2010) |
| | ANSI/AAMI IEC 60601-1-2:2007(R)2012, Medical Electrical Equipment –<br>Part 1-2: General Requirements for Safety and Essential Performance –<br>Collateral Standard : Electromagnetic Compatibility – Requirements and<br>Tests (Edition 3) |
| | ANSI/AAMI IEC 60601-2-2:2009, Medical Electrical Equipment – Part<br>2-2: Particular Requirements for the Basic Safety and Essential<br>Performance of High Frequency Surgery Equipment and High Frequency<br>Surgical Accessories |
| Sterilization Validation | |
| Sterilization Validation | ANSI / AAMI / ISO 11135: 2014 Sterilization of health care products–<br>Ethylene Oxide-Requirements for the development, validation and routine<br>control of a sterilization process for medical devices. |
| Packaging Design Verification Testing | |
| T=0 Design Verification | ASTM D4169, Standard Test Method for<br>Performance Testing of Shipping Containers and Systems |
| T=1.5 Year Accelerated<br>Aging 60C | ASTM F1886/F1886M, Standard Test Method for Determining Integrity<br>of Seals for Flexible Packaging by Visual Inspection |
| T=3 Year Accelerated<br>Aging 60C | ASTM F2096, Standard Test Method for Detecting<br>Gross Leaks in Porous Medical Packaging by Internal Pressurization |
| | ASTM F88-15, Standard Test Method for Seal Strength of Flexible<br>Barrier Materials |
{7}------------------------------------------------
### Substantial Equivalence Conclusion:
As shown in the table above and evidenced by the Verification and Validation Activities described in the submission, the differences in technological characteristics between the currently marketed Integrated Tubing and Cord Sets and the versions cleared in K890648 do not raise new questions of safety and effectiveness, and can be found substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.