← Product Code [GEX](/productcode/GEX) · K172363

# DEKA SMARTXIDE ULTRASPEED (K172363)

_El.En Electronic Engineering Spa · GEX · Nov 17, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172363

## Device Facts

- **Applicant:** El.En Electronic Engineering Spa
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Nov 17, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

SmartXide UltraSpeed laser is indicated for the incision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, endoscopic/laproscopic general surgery, gastroenterology, otorhinolaryngology, neurosurgery, oculoplastic, orthopedic, pulmonary/thoracic surgery, and urology for surgical and aesthetic applications. Also, the SmartXide UltraSpeed laser is indicated for periodontal applications such as, but not limited to: · Removal of diseased or Inflamed soft tissue in the periodontal pocket (sulcular debridement) · Biopsies - Frenectomy, Frenum release - Gingivoplasty - Papillectomy - Vestibuloplasty - · Hyperplasia - · Operculectomy - · Drainage (abscess) - · Flap surgery - · Fibroma (nonmalignant tumor, mucosa, tongue) - Epulis - · Aphthous ulcers - · Removal of soft tissue, cysts, and tumors · Laser assisted new attachment procedure (cementum-mediated periodontal ligament newattachment to the root surface in the absence of long junctional epithelium).

## Device Story

10,600 nm CO2 laser system; 40W maximum power. Laser output delivered via interchangeable handpieces; footswitch-activated. Used by clinicians in surgical/aesthetic settings for soft tissue management (incision, ablation, vaporization, coagulation). Device transforms electrical energy into laser energy for tissue interaction. Healthcare providers use visual feedback and handpiece control to perform procedures; output affects tissue state directly, facilitating surgical precision and patient recovery.

## Clinical Evidence

No clinical data provided. Bench testing only.

## Technological Characteristics

10,600 nm CO2 laser; 40W max power. Features 0.2mm/0.4mm spot sizes; 0.02-70ms pulse width; 5-100Hz repetition rate. Includes 4.0mW diode aiming beam. Complies with ISO 10993-1 (biocompatibility), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1 (laser product safety).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- DEKA Smart US 20D/UltraSpeed and Smart Clinic laser ([K081181](/device/K081181.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

November 17, 2017

ELEN Electronic Engineering Spa Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, 50041 Italy

Re: K172363

Trade/Device Name: Deka Smartxide Ultraspeed Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 22, 2017 Received: August 24, 2017

Dear Paolo Peruzzi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@Ida.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Jennifer R. Stevenson -\$3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

K172363

Device Name

SmartXide UltraSpeed

Indications for Use (Describe)

The SmartXide UltraSpeed laser is incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, endoscopic general surgery, gastroenterology, gynecology, otorhinolaryngology, neurosurgery, oculoplastic, orthopedic, pulmonary/thoracic urgery, and urology for surgical and aesthetic applications.

Also, the SmartXide UltraSpeed laser is indicated for periodontal applications such as, but not limited to:

- · Removal of diseased or Inflamed soft tissue in the periodontal pocket (sulcular debridement)
- · Biopsies
- Frenectomy, Frenum release
- Gingivoplasty
- Papillectomy
- · Vestibuloplasty
- · Hyperplasia
- Operculectomy
- · Drainage (abscess)
- · Flap surgerv
- · Fibroma (nonmalignant tumor, mucosa, tongue)
- Epulis
- · Aphthous ulcers
- · Removal of soft tissue, cysts, and tumors

· Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long iunctional enithelium).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

## 510(K) Summary

## DEKA SmartXide UltraSpeed

### Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

# Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

### Date Summary Prepared:

August 02, 2017

Device Trade Name:

DEKA SmartXide UltraSpeed

### Common Name:

Medical Laser system

### Classification Name:

Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology (GEX)

### Classification Number:

21 CFR 878.4810

# Equivalent Devices:

DEKA Smart US 20D/UltraSpeed and Smart Clinic laser (K081181).

### Device Description:

DEKA SmartXide UltraSpeed is a 10,600 nm CO2 laser devices whit a maximum power of 40W. The laser source delivers the laser output through interchangeable laser handpieces. Handpiece activation is by footswitch.

# Intended Use:

SmartXide UltraSpeed laser is indicated for the incision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, endoscopic/laproscopic general surgery, gastroenterology, otorhinolaryngology, neurosurgery, oculoplastic, orthopedic, pulmonary/thoracic surgery, and urology for surgical and aesthetic applications. Also, the SmartXide UltraSpeed laser is indicated for periodontal applications such as, but not limited to:

· Removal of diseased or Inflamed soft tissue in the periodontal pocket (sulcular debridement)

· Biopsies

- Frenectomy, Frenum release

{4}------------------------------------------------

K172363 Page 2 of 3

- Gingivoplasty
- Papillectomy
- Vestibuloplasty
- · Hyperplasia
- · Operculectomy
- · Drainage (abscess)
- · Flap surgery
- · Fibroma (nonmalignant tumor, mucosa, tongue)
- Epulis
- · Aphthous ulcers
- · Removal of soft tissue, cysts, and tumors
- · Laser assisted new attachment procedure (cementum-mediated periodontal ligament newattachment to the root surface in the absence of long junctional epithelium).

# Substantial equivalence discussion:

The DEKA SmartXide UltraSpeed is substantially equivalent to the Smart US 20D/UltraSpeed and Smart Clinic laser (K081181).

| Device Trade Name    | DEKA SmartXide UltraSpeed                | Predicate Device<br>K081181<br>Smart US 20D/UltraSpeed<br>and Smart Clinic lasers |
|----------------------|------------------------------------------|-----------------------------------------------------------------------------------|
| Laser Type           | CO2                                      | CO2                                                                               |
| Wavelength (nm)      | 10600                                    | 10600                                                                             |
| Power (Watts)        | 40                                       | Smart US20D/Ultraspeed: 25<br>Smart Clinic: 50                                    |
| Spot Sizes<br>(mm)   | 0.2; 0.4                                 | 0.2; 0.4                                                                          |
| Exposure             | Continuous, single and repeated<br>pulse | Continuous, single and<br>repeated pulse                                          |
| Pulse width (ms)     | 0.02-70                                  | 0.2-80                                                                            |
| Repetition rate (Hz) | 5-100                                    | 5-100                                                                             |
| Aiming beam          | Diode laser, 4.0 mW                      | Diode laser, 1.0 mW                                                               |

The DEKA SmartXide UltraSpeed has the same indications for use as the abovementioned predicate device, with same principle of operation and essentially the same performances.

# Clinical Performance Data:

None

# Non-Clinical Performance Data:

{5}------------------------------------------------

K172363 Page 3 of 3

Non clinical Performance Tests according to the following standards have been performed:

- ISO 10993-1 Fourth edition 2009-10-15 biological evaluation of medical devices - part 1: evaluation and testing within a risk management process
- IEC 60601-1-2 Edition 3: 2007-03, medical electrical equipment part 1-2: general . requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests.
- IEC 60601-2-22 Edition 3.1 2012-10, medical electrical equipment part 2-22: particular ● requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
- IEC 60825-1 Edition 2.0 2007-03, safety of laser products part 1: equipment classification, ● and requirements.

Tests results demonstrated compliance to related standards.

# Conclusion:

Base on substantially equivalence discussion and on non Clinical Performance data, we can conclude that the DEKA SmartXide UltraSpeed laser system is substantially equivalent to the predicate device .

# Additional Information:

None

---

**Source:** [https://fda.innolitics.com/device/K172363](https://fda.innolitics.com/device/K172363)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com