← Product Code [IYN](/productcode/IYN) · K172342

# Voluson E6, Voluson E8, Voluson E10 (K172342)

_GE Healthcare · IYN · Sep 29, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172342

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Sep 29, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal

## Device Story

The Voluson E6, E8, and E10 are mobile, general-purpose ultrasound systems for diagnostic imaging and fluid flow analysis. They utilize various linear, curved linear, and matrix phased array transducers to acquire ultrasound signals. The system processes these inputs to provide real-time 2D, 3D, and 4D imaging, measurements, and analysis. Operated by clinicians in clinical settings, the device features a console with a keyboard, touch panel, and display. It supports DICOM networking for image storage and reporting. New features include the IOTA LR2 model (Simple Rules), TomTec Fetal Heart Analysis, and SonoMetrium (3D endometrial cavity analysis). The device aids healthcare providers in diagnostic decision-making by providing high-performance visualization and quantification of anatomical structures and blood flow, potentially improving diagnostic accuracy and patient management.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. The device was evaluated through non-clinical bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in accordance with recognized standards (e.g., IEC 60601-1, ISO 10993-1, NEMA UD 3).

## Technological Characteristics

Mobile console with keyboard, touch panel, and display. Uses linear, curved linear, and matrix phased array transducers. Supports B, M, PW/CW Doppler, Color/Power Doppler, 3D/4D, and harmonic imaging. Connectivity via DICOM. Biocompatible patient-contact materials. Software includes IOTA LR2, Fetal Heart Analysis, and SonoMetrium. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, and NEMA UD 2/3.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Voluson E6_E8_E10 Diagnostic Ultrasound System ([K162269](/device/K162269.md))
- Vivid E80/E90/E95 Diagnostic Ultrasound System ([K150087](/device/K150087.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 29, 2017

GE Healthcare % Mr. Bryan Behn Regulatory Affairs Director 9900 Innovation Drive, Milwaukee County Research Park WAUWATOSA WI 53226

Re: K172342

Trade/Device Name: Voluson E6, Voluson E8, Voluson E10 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: August 17, 2017 Received: August 18, 2017

Dear Mr. Behn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-

Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael D. O'HaraFor

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K172342

#### Device Name

Voluson E10, Voluson E8 and Voluson E6

#### Indications for Use (Describe)

The device is a general-purpose ultrasound system. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatic; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal

| Type of Use (Select one or both, as applicable)                                      |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size: 12px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle is surrounded by a decorative pattern of curved lines and dots, also in blue. The logo is simple, recognizable, and associated with the multinational conglomerate corporation.

### Indications for Use Forms

The following forms represent indications with clinical applications and exam types along with the modes of operation for the Voluson E Series system and for all of its probe/mode combinations. Combinations identified as "P" represents those previously cleared with another GE Ultrasound system. Combinations identified as "E" represents those previously cleared with another GE Ultrasound system and added via Appendix E of the FDA Ultrasound Guidance to this system. This modification does not add to indications to the previously cleared system level or transducer indications or clinical applications.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a circular frame. The logo is blue and white.

#### Diagnostic Ultrasound Indications for Use Form Voluson E6 / Voluson E8 / Voluson E10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                   |                    |                  |                    |                     |                |                  |
|----------------------------------------------------|-------------------|---|---------------|---------------|-------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color#<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic                                         |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Abdominal[1]                                       | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Pediatric                                          | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Small Organ[2]                                     | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [5]              |
| Adult Cephalic                                     | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              |                  |
| Cardiac[3]                                         | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [5]              |
| Peripheral Vascular                                | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Musculo-skeletal Superficial                       | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Other                                              |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access                         |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     | P                 | P | P             |               | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Transvaginal[10]                                   | P                 | P | P             |               | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Transuretheral                                     |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Intraoperative                                     |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                        |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Intravascular                                      |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[1] Abdominal includes renal, GYN/Pelvic, Urology Notes:

[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients

[3] Cardiac is Adult and Pediatric

[5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[9] Elastography imaging- Elasticity

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

[#] 4D color Doppler (only with eM6C probe)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. There are curved lines around the circle, giving the impression of movement or energy. The logo is simple, recognizable, and associated with a well-known multinational corporation.

### Diagnostic Ultrasound Indications for Use Form

#### Voluson E6 / Voluson E8 / Voluson E10 with RAB2-5-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |            |            |               |                 |               |                 |                  |             |               |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic                                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Fetal / Obstetrics[7]                              | P                 | P | P          | P          | P             | P               | P             | P               | P                | P           | [ 5,6]        |
| Abdominal[1]                                       | P                 | P | P          | P          | P             | P               | P             | P               | P                | P           | [ 5,6]        |
| Pediatric                                          |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Small Organ[2]                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Neonatal Cephalic                                  |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Adult Cephalic                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Cardiac[3]                                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Peripheral Vascular                                |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Musculo-skeletal Conventional                      | P                 | P | P          | P          | P             | P               | P             | P               | P                | P           | [ 5,6]        |
| Musculo-skeletal Superficial                       |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Other                                              |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Exam Type, Means of Access                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transesophageal                                    |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transrectal                                        |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transvaginal                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transuretheral                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative Neurological                        |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intravascular                                      |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Laparoscopic                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[1] Abdominal includes renal, GYN/Pelvic, Urology Notes:

[5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy (3D/4D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is colored in a shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### Voluson E6 / Voluson E8 / Voluson E10 with RIC5-9-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |            |            |               |                 |               |                 |                  |             |               |  |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes) |  |
| Ophthalmic                                         |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Fetal / Obstetrics[7]                              | P                 | P | P          |            | P             | P               | P             | P               | P                | P           | [ 5,6,9]      |  |
| Abdominal[1]                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Pediatric                                          |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Small Organ[2]                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Neonatal Cephalic                                  |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Adult Cephalic                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Cardiac[3]                                         |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Peripheral Vascular                                |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Musculo-skeletal Conventional                      |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Musculo-skeletal Superficial                       |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Other                                              |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Exam Type, Means of Access                         |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Transesophageal                                    |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Transrectal[8]                                     | P                 | P | P          |            | P             | P               | P             | P               | P                | P           | [ 5,6,9]      |  |
| Transvaginal [10]                                  | P                 | P | P          |            | P             | P               | P             | P               | P                | P           | [ 5,6,9]      |  |
| Transuretheral                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Intraoperative                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Intraoperative Neurological                        |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Intravascular                                      |                   |   |            |            |               |                 |               |                 |                  |             |               |  |
| Laparoscopic                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[5] 3D/4D Imaging Mode Notes:

[6] Includes imaging of guidance of biopsy (3D/4D)

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[9] Elastography Imaging- Elasticity

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular frame. The color of the letters and the frame is a light blue. 

### Diagnostic Ultrasound Indications for Use Form Voluson E6 / Voluson E8 / Voluson E10 with 4C-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br><i>Anatomy/Region of Interest</i> | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|-----------------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Ophthalmic                                                |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                                     | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |
| Abdominal[1]                                              | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |
| Pediatric                                                 |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                            |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                            |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac[3]                                                |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                       | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |
| Musculo-skeletal Conventional                             |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                              |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other                                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access                                |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                           |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal                                               |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transvaginal                                              |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transuretheral                                            |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative                                            |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                               |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                             |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                              |   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology

[6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular frame. The color of the letters and the frame is a light blue. 

### Diagnostic Ultrasound Indications for Use Form Voluson E6 / Voluson E8 / Voluson E10 with IC5-9-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |            |            |               |                 |               |                 |                  |             |               |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic                                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Fetal / Obstetrics[7]                              | P                 | P | P          |            | P             | P               | P             | P               | P                | P           | [ 6,9]        |
| Abdominal[1]                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Pediatric                                          |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Small Organ[2]                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Neonatal Cephalic                                  |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Adult Cephalic                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Cardiac[3]                                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Peripheral Vascular                                |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Musculo-skeletal Conventional                      |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Musculo-skeletal Superficial                       |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Other                                              |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Exam Type, Means of Access                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transesophageal                                    |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transrectal[8]                                     | P                 | P | P          |            | P             | P               | P             | P               | P                | P           | [ 6,9]        |
| Transvaginal[10]                                   | P                 | P | P          |            | P             | P               | P             | P               | P                | P           | [ 6,9]        |
| Transuretheral                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative Neurological                        |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intravascular                                      |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Laparoscopic                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[9] Elastography Imaging- Elasticity

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is colored in a shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### Voluson E6 / Voluson E8 / Voluson E10 with RSP6-16-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Small Organ[2]                                     | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Other                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal[8]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients Notes:

[5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy (3D/4D)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is colored in a shade of blue.

#### Diagnostic Ultrasound Indications for Use Form

#### Voluson E6 / Voluson E8 / Voluson E10 with RIC6-12-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                   | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|-----------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application              | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| <i>Anatomy/Region of Interest</i> |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Ophthalmic                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]             | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6]            |  |
| Abdominal[1]                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Small Organ[2]                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Neonatal Cephalic                 |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular               |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Musculo-skeletal Conventional     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Musculo-skeletal Superficial      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Other                             |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                   |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal[8]                    | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6]            |  |
| Transvaginal[10]                  | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6]            |  |
| Transuretheral                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[5] 3D/4D Imaging Mode Notes:

[6] Includes imaging of guidance of biopsy (3D/4D)

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is colored in a shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### Voluson E6 / Voluson E8 / Voluson E10 with 9L-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Small Organ[2]                                     | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Other                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients Notes:

[6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the letters. The logo is presented in a blue color.

### Diagnostic Ultrasound Indications for Use Form Voluson E6 / Voluson E8 / Voluson E10 with 11L-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Small Organ[2]                                     | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Other                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients

[6] Includes imaging of guidance of biopsy (2D)

[9] Elastography Imaging- Elasticity

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the letters. The logo is presented in a blue color.

### Diagnostic Ultrasound Indications for Use Form Voluson E6 / Voluson E8 / Voluson E10 with C1-5-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                               | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|-------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application          | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Anatomy/Region of Interest    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Ophthalmic                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]         | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Abdominal[1]                  | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Pediatric                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Small Organ[2]                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Neonatal Cephalic             |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular           | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Musculo-skeletal Conventional |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Musculo-skeletal Superficial  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Other                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal               |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal                   |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological   |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                 |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology

[6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the letters. The logo is presented in a blue color.

### Diagnostic Ultrasound Indications for Use Form Voluson E6 / Voluson E8 / Voluson E10 with ML6-15-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                               | Mode of Operation |   |         |         |         |         |         |          |          |       |         |  |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|----------|----------|-------|---------|--|
| Clinical Application          | B                 | M | PW      | CW      | Color   | Color M | Power   | Combined | Harmonic | Coded | Other   |  |
| Anatomy/Region of Interest    |                   |   | Doppler | Doppler | Doppler | Doppler | Doppler | Modes*   | Imaging  | Pulse | [Notes] |  |
| Ophthalmic                    |                   |   |         |         |         |         |         |          |          |       |         |  |
| Fetal / Obstetrics[7]         |                   |   |         |         |         |         |         |          |          |       |         |  |
| Abdominal[1]                  |                   |   |         |         |         |         |         |          |          |       |         |  |
| Pediatric                     | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [6,9]   |  |
| Small Organ[2]                | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [ 6,9]  |  |
| Neonatal Cephalic             |                   |   |         |         |         |         |         |          |          |       |         |  |
| Adult Cephalic                |                   |   |         |         |         |         |         |          |          |       |         |  |
| Cardiac[3]                    |                   |   |         |         |         |         |         |          |          |       |         |  |
| Peripheral Vascular           | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [6,9]   |  |
| Musculo-skeletal Conventional | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [ 6,9]  |  |
| Musculo-skeletal Superficial  | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [ 6,9]  |  |
| Other                         |                   |   |         |         |         |         |         |          |          |       |         |  |
| Exam Type, Means of Access    |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transesophageal               |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transrectal                   |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transvaginal                  |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transuretheral                |                   |   |         |         |         |         |         |          |          |       |         |  |
| Intraoperative                |                   |   |         |         |         |         |         |          |          |       |         |  |
| Intraoperative Neurological   |                   |   |         |         |         |         |         |          |          |       |         |  |
| Intravascular                 |                   |   |         |         |         |         |         |          |          |       |         |  |
| Laparoscopic                  |                   |   |         |         |         |         |         |          |          |       |         |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients

[6] Includes imaging of guidance of biopsy (2D)

[9] Elastography Imaging-Elasticity

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo is a blue circle with the letters "GE" in a stylized script inside. T…

---

**Source:** [https://fda.innolitics.com/device/K172342](https://fda.innolitics.com/device/K172342)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
