← Product Code [JHI](/productcode/JHI) · K172257

# TrueDX hCG Early Result Pregnancy Test (Midstream Format), TrueDX hCG Early Result Pregnancy Test (Cassette Format), VeriClear Early Result Pregnancy Test (Midstream Format), VeriClear Early Result Pregnancy Test (Cassette Format) (K172257)

_True Diagnostics, Inc. · JHI · Dec 22, 2017 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K172257

## Device Facts

- **Applicant:** True Diagnostics, Inc.
- **Product Code:** [JHI](/productcode/JHI.md)
- **Decision Date:** Dec 22, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1155
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K172257 · Dec 22, 2017 | TrueDX hCG Early Result Pregnancy Test (Midstream Format), TrueDX hCG Early Result Pregnancy Test (Cassette Format), VeriClear Early Result Pregnancy Test (Midstream Format), VeriClear Early Result Pregnancy Test (Cassette Format) | True Diagnostics, Inc. | Urine samples collected from women planning to become pregnant during their natural conception cycles | The study evaluated the device's ability to detect hCG in urine samples collected from women during their natural conception cycles (days -9 to +1 relative to expected period) to support the early pregnancy detection claim. | hCG; Pregnancy test; Early detection; Conception cycle |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Observational cohort study of women planning to become pregnant; Follow-up/Duration: Day -9 to day +1 of expected period | 56 women (25-45 years old) planning to become pregnant; Sample Size: 616 urine samples | Not applicable for this study | Detection rate of hCG relative to expected menstrual period |

## Indications for Use

TrueDX™ hCG Early Result Pregnancy Test (Cassette and Midstream format) is a rapid chromatographic immunoassay for qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. Important note regarding negative result: Some pregnant women will not be able to detect hCG in their urine 5 days before the expected period. If you test negative before your missed period, but think you may still be pregnant, you should test against a few a few days after you missed period. Important note regarding positive results: Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. If you test positive, but think you may not be pregnant, you should check with your doctor. All results should be confirmed by your healthcare provider, especially when making decision about future medical care. This product is intended for prescription use. VeriClear™ Early Result Pregnancy Test (Cassette and Midstream format) is a rapid chromatographic immunoassay for qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period. Important note regarding negative result: Some pregnant women will not be able to detect hCG in their urine 5 days before the expected period. If you test negative before your missed period, but think you may still be pregnant, you should test against a few a few days after you missed period. Important note regarding positive results: Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. If you test positive, but think you may not be pregnant, you should check with your doctor. All results should be confirmed by your healthcare provider, especially when making decision about future medical care. This product is intended for over-the-counter use.

## Device Story

Lateral flow sandwich immunochromatographic assay; detects hCG in urine. Input: urine sample applied to midstream or cassette device. Principle: mouse monoclonal anti-hCG antibodies conjugated to colloidal gold; goat anti-mouse IgG control line. Output: visual lines in test window after 3 minutes; two lines indicate pregnancy, one line indicates non-pregnancy. Used in clinical or home settings; operated by healthcare professionals or lay users. Results aid in early pregnancy detection; positive results require clinical confirmation.

## Clinical Evidence

Clinical study of 56 women (ages 25-45) followed through conception cycles (day -9 to +1). Detection rates: 71% at 5 days before expected period, 100% at 1 day before. Lay user study (n=218) demonstrated 100% agreement with professional users at 10 mIU/ml. Specificity study (n=320) across age groups (pre-, peri-, post-menopausal) showed 0% false-positive rate. Analytical sensitivity confirmed at 10 mIU/ml.

## Technological Characteristics

Lateral flow sandwich immunoassay; nitrocellulose membrane; mouse monoclonal anti-hCG antibodies; colloidal gold conjugate; goat anti-mouse IgG control line. Formats: Cassette and Midstream. Qualitative visual readout. No electronic components or software.

## Regulatory Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

## Predicate Devices

- FIRST RESPONSE Early Result Pregnancy Test (k123436)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k172257

B. Purpose for Submission:
New Device

C. Measurand:
Human Chorionic Gonadotropin (hCG)

D. Type of Test:
Qualitative chromatographic immunoassay

E. Applicant:
True Diagnostics, Inc.

F. Proprietary and Established Names:
TrueDX hCG Early Result Pregnancy Test (Midstream Format)
TrueDX hCG Early Result Pregnancy Test (Cassette Format)
VeriClear Early Result Pregnancy Test (Cassette Format)
VeriClear Early Result Pregnancy Test (Midstream Format)

G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1155 Human chorionic gonadotropin (hCG) test System
2. Classification:
Class II
3. Product code:
JHI, LCX

{1}

4. Panel:

Chemistry (75)

H. Intended Use:

1. Intended use(s):

See Indications for use below.

2. Indication(s) for use:

TrueDX hCG Early Result Pregnancy Test

TrueDX hCG Early Result Pregnancy Test (Midstream and Cassette Format) is a rapid chromatographic immunoassay for qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

Important note regarding negative result:

Some pregnant women will not be able to detect hCG in their urine 5 days before the expected period. If you test negative before your missed period, but think you may still be pregnant, you should test against a few a few days after you missed period.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. If you test positive, but think you may not be pregnant, you should check with your doctor.

All results should be confirmed by your healthcare provider, especially when making decision about future medical care.

VeriClear Early Result Pregnancy Test

VeriClear Early Result Pregnancy Test (Midstream and Cassette Format) is a rapid chromatographic immunoassay for qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

Important note regarding negative result:

Some pregnant women will not be able to detect hCG in their urine 5 days before the expected period. If you test negative before your missed period, but think you may still be pregnant, you should test against a few a few days after you missed period.

2

{2}

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. If you test positive, but think you may not be pregnant, you should check with your doctor.

All results should be confirmed by your healthcare provider, especially when making decision about future medical care.

3. Special conditions for use statement(s):

TrueDX hCG Early Result Pregnancy Test (Midstream and Cassette Format) is intended for prescription use

VeriClear Early Result Pregnancy Test (Midstream and Cassette Format) is intended for over-the-counter (OTC) use

4. Special instrument requirements:

None

I. Device Description:

The TrueDX hCG Early Result Pregnancy Test is a qualitative lateral flow immunoassay for the detection of hCG. The device has two formats: Cassette and Midstream. Each device kit includes a pouch with all components to perform the test, an instruction for use and a desiccant package to control the moisture during the storage of the test kit. The cassette and midstream nitrocellulose test strips are mounted in a plastic housing. In addition, the cassette test kit contains a dropper.

The VeriClear Early Result Pregnancy Test and TrueDX hCG Early Result Pregnancy Test are the same devices, except the device names and type of use. the VeriClear Early Result Pregnancy Test is intended for OTC use while the TrueDX hCG Early Result Pregnancy Test is intended for prescription use.

J. Substantial Equivalence Information:

1. Predicate device name(s):

FIRST RESPONSE Early Result Pregnancy Test

2. Predicate 510(k) number(s):

k123436

{3}

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device TrueDX hCG Early Result Pregnancy Test and VeriClear Early Result Pregnancy Test (k172257) | Predicate FIRST RESPONSE Early Result Pregnancy Test (k123436)  |
|  Intended Use | Aid in early detection of pregnancy | Same  |
|  Early Detection claim | Detects pregnancy as early as 5 days before the expected period or as early as 6 days before the day of the missed period. | Same  |
|  Test Principle | Lateral flow sandwich immunochromatographic assay | Same  |
|  Sample Matrix | Urine | Same  |
|  Traceability | WHO 4thInternational Standard |   |
|  Limit of Detection | 10 mIU/mL | Same  |
|  Time to Result | 3 minutes | Same  |
|  Differences  |   |   |
|  Item | Device TrueDX hCG Early Result Pregnancy Test and VeriClear Early Result Pregnancy Test (k172257) | Predicate FIRST RESPONSE Early Result Pregnancy Test (k123436)  |
|  Target User | Prescription use (TrueDX hCG Early Result Pregnancy Test) OTC use (VeriClear Early Result Pregnancy Test) | OTC use only  |
|  Device format | Cassette, Midstream | Midstream  |
|  hCG isoforms detected | Intact hCG Hyperglycosylated hCG hCG β -subunit | Intact hCG Hyperglycosylated hCG hCG β -subunit hCG β -core fragment  |

K. Standard/Guidance Document Referenced (if applicable):

None reference.

L. Test Principle:

The TrueDX Early Result Pregnancy Test (Midstream and Cassette Format) and VeriClear Early Result Pregnancy Test (Midstream and Cassette Format) are sandwich immunoassays employing mouse monoclonal antibodies specific for hCG, which are immobilized on the

{4}

membrane as test line, and goat anti-mouse IgG immobilized on the membrane as control line. After the urine specimen is applied to the device, the hCG present in the specimen will react with the mouse anti-hCG monoclonal antibody-colloidal gold conjugate. The complex migrates along the membrane towards the test and control zones. Two lines in the test window 3 minutes after the urine application indicate that hCG has been detected (pregnant); one line (control line) indicates that no hCG has been detected (not pregnant).

## M. Performance Characteristics (if/when applicable):

The VeriClear Early Result Pregnancy Test and TrueDX hCG Early Result Pregnancy Test are identical; therefore, only one set of representative performance data is presented below.

## 1. Analytical performance:

### a. Precision/Reproducibility:

A precision study was performed using samples with hCG concentrations of 3.0, 5.0, 8.5, 10, 25, 50, and 100 mIU/ml, traceable to the WHO 4th International Standard. Each sample was tested in 10 replicates each day over 5 days, using 3 lots of devices (for both formats) by 5 operators. A total of 250 replicates (10 replicates/run x 5 operators x 5 days) per lot of device for each sample were obtained. The positive results observed over total number of tests are summarized in the following tables:

Precision Study for three lots Devices-Midstream Format

|  hCG level(mIU/ml) | Lot 1 | Lot 2 | Lot3 | % Positive  |
| --- | --- | --- | --- | --- |
|  0 | 0/250 | 0/250 | 0/250 | 0 %  |
|  3.0 | 0/250 | 0/250 | 0/250 | 0 %  |
|  5.0 | 0/250 | 0/250 | 0/250 | 0 %  |
|  8.5 | 135/250 | 139/250 | 130/250 | 53.8 %  |
|  10 | 250/250 | 250/250 | 250/250 | 100%  |
|  25 | 250/250 | 250/250 | 250/250 | 100%  |
|  50 | 250/250 | 250/250 | 250/250 | 100%  |
|  100 | 250/250 | 250/250 | 250/250 | 100%  |

Precision Study for three lots devices - Cassette Format

|  hCG level(mIU/ml) | Lot 1 | Lot 2 | Lot3 | % Positive  |
| --- | --- | --- | --- | --- |
|  0 | 0/250 | 0/250 | 0/250 | 0 %  |
|  3.0 | 0/250 | 0/250 | 0/250 | 0 %  |
|  5.0 | 0/250 | 0/250 | 0/250 | 0 %  |
|  8.5 | 115/250 | 110/250 | 135/250 | 48%  |
|  10 | 250/250 | 250/250 | 250/250 | 100%  |
|  25 | 250/250 | 250/250 | 250/250 | 100%  |
|  50 | 250/250 | 250/250 | 250/250 | 100%  |

{5}

6

|  hCG level(mIU/ml) | Lot 1 | Lot 2 | Lot3 | % Positive  |
| --- | --- | --- | --- | --- |
|  100 | 250/250 | 250/250 | 250/250 | 100%  |

Precision Study for five Operators –Midstream Format

|  hCG level (mIU/ml) | Operator 1 | Operator 2 | Operator 3 | Operator 4 | Operator 5 | % Positive  |
| --- | --- | --- | --- | --- | --- | --- |
|  0 | 0/150 | 0/150 | 0/150 | 0/150 | 0/150 | 0 %  |
|  3.0 | 0/150 | 0/150 | 0/150 | 0/150 | 0/150 | 0 %  |
|  5.0 | 0/150 | 0/150 | 0/150 | 0/150 | 0/150 | 0 %  |
|  8.5 | 84/150 | 82/150 | 79/150 | 74/150 | 85/150 | 53.8%  |
|  10 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |
|  25 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |
|  50 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |
|  100 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |

Precision Study for five Operators –Cassette Format

|  hCG level (mIU/ml) | Operator 1 | Operator 2 | Operator 3 | Operator 4 | Operator 5 | % Positive  |
| --- | --- | --- | --- | --- | --- | --- |
|  0 | 0/150 | 0/150 | 0/150 | 0/150 | 0/150 | 0 %  |
|  3.0 | 0/150 | 0/150 | 0/150 | 0/150 | 0/150 | 0 %  |
|  5.0 | 0/150 | 0/150 | 0/150 | 0/150 | 0/150 | 0 %  |
|  8.5 | 69/150 | 69/150 | 71/150 | 77/150 | 74/150 | 48%  |
|  10 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |
|  25 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |
|  50 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |
|  100 | 150/150 | 150/150 | 150/150 | 150/150 | 150/150 | 100%  |

b. Linearity/assay reportable range:

Linearity is not applicable since this is a qualitative test.

High dose hook effect study:

Negative urine samples were spiked with hCG at concentrations up to 450,000 mIU/ml, and then tested in 2 replicates per lot using two lots of devices for each format. The results demonstrated that no hook effect was observed at hCG concentrations up to 450,000 mIU/ml.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The test is calibrated against the WHO 4th International Standards for hCG.

{6}

The stability testing protocol and acceptance criteria used to support the shelf life were reviewed and found to be acceptable. The sponsor claims a 24-month shelf life for both formats when stored in the sealed foil pouch at 39-86°F (4-30°C).

d. Detection limit:

An analytical sensitivity study was performed using negative human urine sample spiked with hCG traceable to the WHO 4th International Standards for hCG to obtain concentration of 0, 3.0, 5.0, 7.5, 8.5, 9.0, 10, 12.5, 15, and 25 mIU/ml hCG. The samples were measured in 15 replicates, using 3 different lots of each test format. The tests were performed by 3 different operators for 3 consecutive days. A different set of operators tested each format of the device. The positive results observed over the total number of tests are summarized in the following tables.

Midstream Format

|  Concentration (mIU/ml) | Lot 1 | Lot 2 | Lot 3 | % Positive  |
| --- | --- | --- | --- | --- |
|  0 | 0+/15 | 0+/15 | 0+/15 | 0%  |
|  3 | 0+/15 | 0+/15 | 0+/15 | 0%  |
|  5 | 0+/15 | 0+/15 | 0+/15 | 0%  |
|  7.5 | 3+/15 | 2+/15 | 3+/15 | 17.7%  |
|  8.5 | 8+/15 | 9+/15 | 9+/15 | 57.7%  |
|  9.0 | 14+/15 | 14+/15 | 14+/15 | 93.3 %  |
|  10 | 15+/15 | 15+/15 | 15+/15 | 100 %  |
|  12.5 | 15+/15 | 15+/15 | 15+/15 | 100 %  |
|  15 | 15+/15 | 15+/15 | 15+/15 | 100 %  |
|  25 | 15+/15 | 15+/15 | 15+/15 | 100 %  |

Cassette Format

|  hCG Concentration (mIU/ml) | Lot 1 | Lot 2 | Lot 3 | % Positive  |
| --- | --- | --- | --- | --- |
|  0 | 0+/15 | 0+/15 | 0+/15 | 0 %  |
|  3 | 0+/15 | 0+/15 | 0+/15 | 0%  |
|  5 | 0+/15 | 0+/15 | 0+/15 | 0%  |
|  7.5 | 3+/15 | 2+/15 | 4+/15 | 20%  |
|  8.5 | 8+/15 | 9+/15 | 8+/15 | 55 %  |
|  9.0 | 13+/15 | 14+/15 | 13+/15 | 88.8%  |
|  10 | 15+/15 | 15+/15 | 15+/15 | 100 %  |
|  12.5 | 15+/15 | 15+/15 | 15+/15 | 100 %  |
|  15 | 15+/15 | 15+/15 | 15+/15 | 100 %  |
|  25 | 15+/15 | 15+/15 | 15+/15 | 100 %  |

The results demonstrated that the analytical sensitivity of the new device (the lowest concentration that yields 100% positive results) is 10 mIU/ml.

{7}

# e. Analytical specificity:

Negative and positive urine containing  $10\mathrm{mIU / mL}$  hCG were spiked with potentially interfering exogenous or endogenous substances, and then tested in 5 replicates per lot using two lots of devices for each device format. No interference effect was observed at the concentrations tested below:

|  Substance tested | Highest Concentration tested that demonstrated no interference Concentration  |
| --- | --- |
|  Acetaminophen | 20 mg/dL  |
|  Acetylsalicylic acid | 20 mg/dL  |
|  Human serum Albumin | 2000 mg/dL  |
|  Ampicillin | 20 mg/dL  |
|  Ascorbic acid | 20 mg/dL  |
|  Atropine | 20 mg/dL  |
|  Caffeine | 20 mg/dL  |
|  Cortisol | 200 ng/dL  |
|  EDTA | 80 mg/dL  |
|  Phenylpropanolamine | 20 mg/dL  |
|  Ephedrine | 20 mg/dL  |
|  Gentisic acid | 20 mg/dL  |
|  Glucose | 2000 mg/dL  |
|  Tetracycline | 20 mg/dL  |
|  Uric acid | 10 mg/dL  |
|  Bilirubin | 20 mg/dL  |
|  Ethanol | 0.1 %  |
|  Salicylic Acid | 20 mg/dL  |

# Cross reactivity of structurally similar compounds

Negative and positive (10 mIU/ml hCG) urine samples were spiked with various potential cross reactants (hLH, hFSH, and hTSH), then tested in 5 replicates per lot using two lots of devices for each format. No cross reactivity was observed at the concentrations tested below:

|  Reactant | Concentration  |
| --- | --- |
|  hLH | 1000 mIU/mL  |
|  hFSH | 1000 mIU/mL  |
|  hTSH | 1000 μIU/mL  |

{8}

9

# Effect of urine pH and Specific Gravity:

Negative and positive (10 mIU/ml hCG) urine samples were adjusted to have pH values of 4.0, 5.0, 6.0, 7.0, 8.0, and 9.0, and then tested in 2 replicates per lot using two lots of devices for each format. The positive and negative hCG results were not affected by urine pH levels between the ranges of 4.0 and 9.0.

Negative and positive (10 mIU/ml hCG) urine samples were adjusted to have specific gravities of 1.000, 1.005, 1.010, 1.015, 1.020, 1.025, 1.030 and 1.035, and then tested in 2 replicates per lot using two lots of devices for each format. The positive and negative hCG results were not affected by urine specific gravity concentrations between 1.000 and 1.035.

# Effect of hCG β-core fragment:

Negative urine samples (0 and 5 mIU/ml hCG) and positive urine samples (10, 25 and 20,000 mIU/ml hCG) were spiked with hCG β-core fragment at concentrations of 50,400, 102,000, 204,000 and 408,000 pmol/L, and then tested in 5 replicates per lot using two lots of devices for each format. All samples yielded correct results with hCG β-core fragment concentrations up to 408,000 pmol/L.

## f. Assay cut-off:

See detection limit section M.1.d.

## 2. Comparison studies:

### a. Method comparison with predicate device:

Urine samples were collected from 166 women at physician offices for pregnancy testing. Of the 166 women, 65 of them were suspected to be pregnant. Patient samples were randomly collected at various times throughout the day. Ages of these women ranged from 19 to 41 years. Samples were masked and randomized prior to testing by two health care professionals. Two lots of each device format (Midstream and Cassette) were tested with every sample. For the midstream format, one lot of test device was tested in the simulated mid-stream method, while the other lots were tested in dip method. The results are summarized in table below:

|   | Predicate Device |   | Total  |   |
| --- | --- | --- | --- | --- |
|   |   |  hCG + | hCG- |   |
|  Candidate Device | hCG + | 65 | 0 | 65  |
|   |  hCG- | 0 | 101 | 101  |
|  Total |   | 65 | 101 | 166  |

### b. Matrix comparison:

Not applicable. The devices are intended for urine samples only.

{9}

10

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

Detection of hCG in Early Pregnancy Clinical Samples

A total of 616 urine samples were collected from 56 different women (25 - 45 years old) who planned to become pregnant. These women were followed throughout their conception cycles with urine collected from day -9 to day +1 of their expected period. The Candidate device (both formats) detected hCG in 71% of samples from, five days before the expected menstrual period, and 100% of samples from one day before the expected menstrual period. The early pregnancy detection results are summarized below:

|  Day in cycle relative to EMP | Midstream Format | Cassette Format | Overall Pregnancy Detection Rate (%)  |
| --- | --- | --- | --- |
|  -9 days | 0% | 0% | 0%  |
|  -8 days | 0% | 0% | 0%  |
|  -7 days | 12.5% | 12.5% | 12.5%  |
|  -6 days | 41% | 41% | 41%  |
|  -5 days | 71% | 71% | 71%  |
|  -4 days | 94% | 94% | 94%  |
|  -3 days | 98% | 98% | 98%  |
|  -2 days | 98% | 98% | 98%  |
|  -1 days | 100% | 100% | 100%  |
|  0 days | 100% | 100% | 100%  |
|  +1 days | 100% | 100% | 100%  |

{10}

# Lay user study

A lay user study was performed at an intended use site with a total of 218 females with diverse educational and professional backgrounds and ages, 18 years and older. 110 lay users tested with midstream format devices and 108 lay users tested with cassette format devices. Lay users were only provided with the package insert prior to performing the study. Lay user results compared to professional user results are listed below.

|  Pregnancy Result | Lay user vs. professional  |   |   |
| --- | --- | --- | --- |
|   |  Midstream | Midstream-Dip | Cassette  |
|  Pregnant | 9/9 (100%) | 9/9(100%) | 6/6 (100%)  |
|  Non-pregnant | 101/101 (100%) | 101/101 (100%) | 102/102 (100%)  |
|  Total | 110/110 (100%) | 110/110 (100%) | 108/108 (100%)  |

Besides testing their own urine samples, lay users also tested 4 spiked urine samples around the cut-off level, at concentrations of 3.0, 7.0, 8.5 and  $10\mathrm{mIU / ml}$  hCG. A total of 110 samples at each level were tested with the candidate device in both Cassette and Midstream formats. Among the 110 midstream format devices tested at each hCG level, 53 were tested in simulated midstream method while 57 were tested in dip method. An aliquot of each of the urine samples was also tested by a laboratory professional using the candidate device. The results are summarized below.

Midstream format, spiked urine samples, lay user vs professional

|  No. of samples | hCG Concentration (mIU/mL) | Lay person results |   | Professionals results  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Number of Positive | % Positive | Number of Positive | % Positive  |
|  110 | 3.0 | 0/110 | 0% | 0/110 | 0%  |
|  110 | 7.0 | 28/110 | 25.4% | 27/110 | 24.5%  |
|  110 | 8.5 | 66/110 | 60.0% | 70/110 | 63.6%  |
|  110 | 10 | 110/110 | 100% | 110/110 | 100%  |

Cassette format, spiked urine samples, lay user vs professional

|  No. of samples | hCG Concentration (mIU/mL) | Lay person results |   | Professionals results  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Number of Positive | % Positive | Number of Positive | % Positive  |
|  108 | 3.0 | 0/108 | 0% | 0/108 | 0%  |
|  108 | 7.0 | 23/108 | 21.3% | 26/108 | 24.1%  |
|  108 | 8.5 | 81/108 | 75.0% | 84/108 | 77.8%  |
|  108 | 10 | 108/108 | 100% | 108/108 | 100%  |

A Flesch-Kincaid reading analysis was performed on each OTC package insert and the score demonstrated a reading Grade Level of 7. The results of the questionnaire reflected that the consumers found the test easy to use and that they did not have trouble understanding the labeling or interpreting results.

{11}

12

Specificity study to determine false-positive result rate

A study was performed to determine the incidence of positive test results using the candidate device among non-pregnant women in three age groups: 18 - 41 years of age (Pre-menopausal), 42-55 years of age (peri-menopausal) and &gt;55 year of age (post-menopausal). A total of 320 subjects provided urine samples: 100 from pre-menopausal subjects, 111 from peri-menopausal subjects, and 109 from post-menopausal subjects. Three lots of each test format of the candidate devices were used for this study. No positive results were observed in any of the age groups.

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
Not applicable.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K172257](https://fda.innolitics.com/device/K172257)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
